K Number
DEN150011
Device Name
Sonablate 450
Date Cleared
2015-10-09

(200 days)

Product Code
Regulation Number
876.4340
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sonablate® 450 is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue.

Device Description

The Sonablate® 450 is a computer-controlled device designed for transrectal delivery of high intensity focused ultrasound (HIFU) energy to the prostate to ablate regions of unwanted prostatic tissue via thermal ablation. The device makes use of integrated biplanar ultrasound imaging for real-time monitoring, planning, and pre- and post-ablation imaging of the prostate. The system consists of the following main components: Sonasource Console, Sonablate Probe 30/40, Sonachill, Sonablate Multi-Axis Stepper and Probe Arm, and Single Use Components (Probe Tip Kit and Water Path Kit).

AI/ML Overview

The provided text describes the Sonablate® 450, a High Intensity Focused Ultrasound (HIFU) system for prostate tissue ablation. The information covers device description, non-clinical studies (biocompatibility, shelf life, electrical safety, software), animal studies, and a clinical study. The acceptance criteria and the study proving the device meets these criteria are detailed, primarily within the "PERFORMANCE TESTING - BENCH" and "SUMMARY OF CLINICAL INFORMATION" sections, and implicitly derived from the "Special Controls" which outline the requirements for devices of this type.

Here's a breakdown of the requested information:

Acceptance Criteria and Device Performance

The acceptance criteria for the Sonablate® 450 are primarily outlined in the "PERFORMANCE TESTING - BENCH" section and in the "SPECIAL CONTROLS" section. The reported device performance is presented in the "Results" columns of the tables in the "PERFORMANCE TESTING - BENCH" section and summarized in the "SUMMARY OF CLINICAL INFORMATION" and "Clinical Results Summary" sections.

Table of Acceptance Criteria and Reported Device Performance

Test/CharacteristicAcceptance CriteriaReported Device Performance
Transducer Characterization
Schlieren test (Beam profile & alignment)Image: b(4) - Specific numerical criteria are redacted but based on the overall context, it would relate to acceptable beam width and misregistration tolerance.All test samples (n=7) met the acceptance criteria.
Electrostatic test (Focal length)Image: b(4) - Specific numerical criteria are redacted but would define the acceptable focal length range for each transducer (30mm and 40mm).All test samples (n=7) met the acceptance criteria. "These results support the targeting accuracy and reproducibility of the HIFU output."
Force balance test (Total Acoustic Power- TAP)Image: b(4) - Specific output power range redacted.All test samples (n=7) met the acceptance criteria. (No acceptance criteria for focal intensity and acoustic pressure, as they were for modeling).
Imaging Verification
Phantom testing (Resolution, Penetration, SNR)Both transducers: Image: b(4) - Specific numerical criteria for resolution, penetration depth, and signal-to-noise ratio redacted.The test sample (n=1) met the acceptance criteria. "The resolution is sufficient for tissue differentiation of the prostate and surrounding anatomy."
Imaging frame rate testImage: b(4) - Specific numerical criteria for linear and sector frame rate redacted.The test sample (n=1) met the acceptance criteria. "The frame rate is sufficient to provide feedback to the physician during probe positioning."
Image display testUsing each transducer: the system performs transverse and longitudinal imaging using each probe, and displays them simultaneously.The test sample (n=1) met the acceptance criteria.
HIFU Verification/Safety
Ex vivo tissue phantom tests (Ablation volume/depth)b(4) - Specific numerical criteria for max ablation volume, depth, and lesion dimensions redacted. Criteria ensure volumes up to 40 cm³ at 40 mm depth, and creation of contiguous volume lesions of specified dimensions.The test sample (n=1) met the acceptance criteria. "The system is capable of targeting a tissue volume of up to 40 cm³, at a depth of up to 40 mm. The system is capable of creating single and volume lesions in x, y, and z dimensions that meet the acceptance criteria. These results support the targeting accuracy and reproducibility of the HIFU output."
HIFU ON-time/OFF-time VerificationTo confirm the ON/OFF timing of the system's HIFU application cycle.The test sample (n=1) met the acceptance criteria. "The system meets ON time and OFF time requirements."
Clinical Effectiveness (Primary Study Focus)Evidence of prostatic tissue ablation, as demonstrated by reductions in prostate volume and PSA level, and increase in the rate of negative biopsy, between baseline and 12 months post-HIFU (Implicitly, significant, statistically relevant changes are expected).Prostate Volume Reduction: Mean reduction of 11.8 cm³ (95% CI: 9.7, 14.0) in 73 patients, representing a mean reduction of 46% from baseline. PSA Reduction: 83% of 117 patients had a reduction in PSA post-ablation (95% CI: 74.6, 89.0). Negative Biopsy Rate: 61% of 117 patients had a negative post-ablation biopsy (95% CI: 51.2, 69.5). (Note: 30 non-biopsied patients were considered "positive").
Clinical SafetyCharacterization of the adverse event profile, ensuring risks are outweighed by benefits.Adverse Events: 99% of patients experienced some application-related complication. Serious AEs included urinary retention (5%), hematuria (4%), rectal fistula (4%), UTI (4%), osteomyelitis (3%). Non-serious AEs included erectile dysfunction (71%), UTI (50%), urinary incontinence (47%), urinary retention (43%), hematuria (31%), urinary obstruction (20%), and urethral stricture (15%). Duration varied, some resolved, others required intervention. (Detailed table of AE percentages provided in the source text).

2. Sample sizes used for the test set and the data provenance

Test Set (Bench Testing):

  • Transducer Characterization: n=7 for Schlieren, Electrostatic, and Force balance tests.
  • Imaging Verification: n=1 for Phantom testing, Imaging frame rate test, and Image display test.
  • HIFU Verification/Safety: n=1 for Ex vivo tissue phantom tests and HIFU ON-time/OFF-time Verification Testing.

Test Set (Animal Studies):

  • Study 1 ("Safety and Effectiveness of Tissue Ablation"): n=5 dogs.
  • Study 2 ("Efficacy Improvement Software Test Report Localization and Ablation"): n=4 dogs.
  • Study 3 ("Treatment Volume and Accuracy Study Report"): Preliminary dog study (n not specified, but likely small).
  • Study 4 ("Thermocouple Response to Treatment Test"): Preliminary dog study (n not specified, but likely small).
  • Study 5 ("Safety of HIFU Energy Application to Adjacent Tissues"): Preliminary dog study (n not specified, but likely small).

Test Set (Clinical Study):

  • Clinical Dataset: Interim analysis of the initial 117 patients enrolled, with 116 undergoing HIFU application.
  • Data Provenance: Multi-center, prospective, single-arm study involving 20 sites across the U.S. and Canada. The study was structured as a clinical trial.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The provided text does not specify the number or qualifications of experts used to establish ground truth for the test set.

  • For bench and animal studies, "ground truth" is established through physical measurements, ex-vivo tissue analysis, and histological evaluation. These are typically performed by trained technical personnel and pathologists, but specific expert involvement (e.g., number and qualifications of pathologists for dog study histology) is not detailed.
  • For the clinical study, "ablation effectiveness" was assessed by analyzing changes in prostate volume, PSA level, and negative biopsy rate. PSA levels are objective measurements. Prostate volume changes and biopsy results would involve radiologists/urologists for imaging interpretation and pathologists for biopsy analysis. However, the exact number and qualifications of these healthcare professionals serving as ground truth experts are not given in the document.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The text does not describe any explicit adjudication method for observations or interpretations in the test set (bench, animal, or clinical).

  • For the clinical study, "investigator reported adverse events" were collected, implying individual site investigators reported. It does not mention a central adjudication committee for AEs or efficacy endpoints. The endpoint for negative biopsy rate notes that "30 patients who did not have a 12-month biopsy were considered 'positive'," which is a pre-defined rule for handling missing data, not an adjudication method for conflicting interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. The Sonablate® 450 is a therapeutic device for tissue ablation, not an imaging diagnostic device assisted by AI. Therefore, the concept of human readers improving with AI assistance is not applicable in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The Sonablate® 450 is a medical device for therapeutic ablation, not a diagnostic algorithm. Therefore, the concept of "standalone (algorithm only)" performance is not applicable in the typical sense of AI/ML software. The device's performance is inherently linked to its physical operation and the physician's control and monitoring. The "PERFORMANCE TESTING - BENCH" section details the device's functional performance in controlled (standalone power/accuracy) environments, but this is not an "algorithm only" evaluation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Bench Testing: Ground truth established via physical measurements (e.g., beam width, focal length, acoustic power, frame rate) and ex-vivo tissue analysis for lesion characterization.
  • Animal Studies: Ground truth established via histological evaluation of ablated tissue, temperature measurements using thermocouples, and comparison of planned vs. actual ablation zones.
  • Clinical Study: Ground truth for effectiveness was established using objective clinical outcomes data:
    • Prostate volume reduction (measured via imaging, presumably ultrasound, pre and post).
    • PSA level reduction (blood test, an objective biochemical marker).
    • Negative biopsy rate (histopathology, which is a gold standard for malignancy/tissue presence).
    • Ground truth for safety was established by investigator-reported adverse events and patient-completed questionnaires.

8. The sample size for the training set

The provided text does not include information about any training set. This is because the Sonablate® 450 is a hardware device for therapeutic ablation, not an AI/ML diagnostic or therapeutic algorithm that typically undergoes distinct training/test phases. While the system involves software that controls its functions, the documentation focuses on the verification and validation of this software, not its "training" on a dataset in the machine learning sense.

9. How the ground truth for the training set was established

As there is no explicit "training set" described for an AI/ML algorithm, the concept of establishing ground truth for it is not applicable in this context. The software's functionality is validated against design specifications and performance requirements through the various bench and animal tests, rather than being "trained" on a dataset with external ground truth.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR SONABLATE® 450

REGULATORY INFORMATION

FDA identifies this generic type of device as:

High intensity ultrasound system for prostate tissue ablation. A high intensity ultrasound system for prostate tissue ablation is a prescription device that transmits high intensity therapeutic ultrasound (HITU) energy into the prostate to thermally ablate a defined, targeted volume of tissue, performed under imaging guidance. This classification does not include devices that are intended for the treatment of any specific prostate disease and does not include devices that are intended to ablate nonprostatic tissues/organs.

NEW REGULATION NUMBER: 21 CFR 876.4340

CLASSIFICATION: II

PRODUCT CODE: PLP

BACKGROUND

DEVICE NAME: SONABLATE® 450

SUBMISSION NUMBER: DEN150011

DATE OF DE NOVO: MARCH 24, 2015

  • SONACARE MEDICAL, LLC CONTACT: DAWN BURLESON, RN, MBA, CCRA VICE PRESIDENT OF CLINICAL AFFAIRS 10130 PERIMETER PARKWAY, SUITE 250 CHARLOTTE, NC 28216

REQUESTER'S RECOMMENDED CLASSIFICATION: II

INDICATIONS FOR USE

The Sonablate® 450 is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue.

LIMITATIONS

The sale, distribution, and use of the Sonablate® 450 is limited to prescription use only.

{1}------------------------------------------------

Use of the Sonablate® 450 must be prescribed and administered under the direct supervision of a qualified and trained physician, after appropriate urologic patient evaluation.

Limitations on device use are also achieved through the following statements included in the Instructions for Use:

Contraindications:

  • Large (> 10 mm) cysts or calcifications in the planned ablation zone ●
  • Metal implants or stents in the urethra
  • Brachytherapy seeds in the planned ablation zone ●
  • . Pre-existing inflammatory disease of the colon or rectum (such as proctitis or ulcerative colitis)
  • Prior significant rectal surgery
  • Inability to insert or tolerate a transrectal ultrasound probe
  • Active urinary tract infection
  • Urethral stricture
  • Latex allergy ●

Warnings:

The effectiveness of the Sonablate® 450 in treating any specific prostate disease has not been established.

The safety and effectiveness of the Sonablate® 450 have not been established in patients with the following conditions:

  • . interest in future fertility.
  • I bleeding disorders,
  • l clinical or histological evidence of urinary bladder cancer,
  • l renal impairment,
  • I functional bladder problems, including neurological disorders that might affect bladder function,
  • I post-void residual urine > 250 mL、
  • I urinary retention requiring an indwelling catheter,
  • I enlarged medial lobe of the prostate protruding into the urinary bladder,
  • I clinical evidence of prostatitis within 6 months prior to HIFU.
  • bladder calculus, and
  • I prostates > 40 cm3.

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

{2}------------------------------------------------

DEVICE DESCRIPTION

The Sonablate® 450 is a computer-controlled device designed for transrectal delivery of high intensity focused ultrasound (HIFU) energy to the prostate to ablate regions of unwanted prostatic tissue via thermal ablation. The device makes use of integrated biplanar ultrasound imaging for real-time monitoring, planning, and pre- and post-ablation imaging of the prostate. The system consists of the following main components:

Sonasource Console:

Image /page/2/Picture/3 description: The image shows a Sonablate machine, which is a medical device used for non-invasive ablation of tissue. The machine is mounted on wheels and has a monitor and keyboard attached to an arm. The monitor displays medical images, and the keyboard is used to control the machine. The side of the machine has the word "Sonablate" written on it.

The Sonasource Console contains the system computer and electronics for HIFU planning and delivery. The monitor displays all system, patient, imaging, and HIFU delivery information via a graphical user interface. The keyboard and trackball are used for data and user-command entry. The trackball allows the operator to define the dimensions of the target zone and to measure the tissue distance from the rectal wall to the center of the target zone.

Planning is performed under physician control, while HIFU delivery is automatically performed by the system under physician monitoring.

Key system features:

  • . Rectal Wall Distance (RWD) monitor: A feature used by the system to continually measure the distance between the therapy transducer and the rectal wall, to maintain a safe distance.
  • . Reflectivity Index Monitor (RIM): A feature used by the system to detect undesired/excessive cavitation bubble formation at the rectal wall.
  • . Dual Stack: A visual aid tool for the physician to use during planning, which allows rapid view of sequential transverse and linear images.
  • . MR/US Fusion: Fuses the ultrasound image with a prior magnetic resonance (MR) image.

{3}------------------------------------------------

Sonablate Probe 30/40:

Image /page/3/Figure/1 description: The image contains three different pictures. The first picture shows a Sonablate device in a black case. The second picture shows a diagram of the device and how it is used to create transverse and sagittal planes. The third picture shows a diagram of the device being used to treat the prostate with a focal point.

The Sonablate Probe 30/40 (probe) is designed to provide gray-scale ultrasound images of the prostate and to deliver HIFU pulses to targeted tissue while minimizing damage to intervening tissue. Prior to HIFU application, the probe is inserted into the patient's rectum, with the ultrasound transducers positioned at the level of the prostate. The probe has a dual-sided piezoelectric transducer, with different curvatures (i.e., 30 mm and 40 mm focal lengths) on opposite sides. The dual-sided transducer allows focal HIFU application of the entire volume of prostate tissue using only one transrectal probe (30 mm transducer for the posterior region of the gland, 40 mm probe for the anterior region).

The probe is connected to the Sonasource Console via a cable. Prior to insertion in the patient, the tip of the probe is covered with a single use probe sheath (an off the shelf condom), which is secured by O-rings.

The probe incorporates two motors (linear and sector) to guide transducer movement, dedicated, closed loop coolant water path, and a thermistor for temperature control. To reduce the possibility of thermal injury to the rectal wall, HIFU application is automatically terminated if the probe temperature exceeds 30°C.

In the Imaging Mode, the transducer moves in a longitudinal direction to provide linear images of the prostate and oscillates in the transverse plane for sector images. In the Therapy Mode, the transducer sequentially moves to pre-specified positions (determined during HIFU planning), and stays at each position while the energy is delivered to the tissue. Each thermal lesion created in the prostate is 10-12 mm in length and 1-2 mm wide. Multiple lesions must be sequentially positioned within the prostate to fill the intended tissue volume.

{4}------------------------------------------------

Key technical specifications:

  • . Two piezoelectric crystals of different curvatures - 30 mm and 40 mm focal lengths on opposite sides
  • Focal lengths: 30 mm +2.0/-1.0 mm, 40 mm +2.0/-1.0 mm
  • Site intensity in the focal zone 1200 to 2000 W/cm-
  • Pulse ON/OFF durations: 3 s ON/6s OFF ●
  • HIFU application mode frequency 4.0 MHz ●
  • Imaging mode frequency 6.5 MHz ●

Sonachill:

Image /page/4/Picture/8 description: This image shows a medical device on a stand. The device is labeled "Senachtill" on the front. The stand has a five-legged base with wheels, allowing it to be easily moved. A power cord is visible, indicating the device requires electricity to operate.

The Sonachill circulates degassed cold water to the probe tip (flow contained within the probe sheath). The circulating water functions to cool the rectal wall and HIFU transducer surface, and to provide ultrasound coupling between the transducer and tissue. To prevent contaminated coolant water from the patient entering and contaminating the internal fluid pathways of the Sonachill device (which could occur if the probe sheath breaks), there are two independent fluid pathways. The first fluid path is through the probe and the single use components of the Water Path Kit. The second fluid path is isolated from the probe (i.e., within the Sonachill), and circulates between the heat exchanger and the pump.

Sonablate Multi-Axis Stepper and Probe Arm:

Image /page/4/Picture/11 description: The image shows two different medical devices. The device on the left is a metallic surgical instrument with multiple adjustable parts and knobs, set against a blue background. The device on the right is a metallic arm with a black headpiece, also featuring adjustable knobs, and is set against a white background. Both devices appear to be designed for precise positioning and fixation during medical procedures.

After the probe is inserted in the patient's rectum. the Multi-Axis Stepper and Probe Arm are locked to hold the probe in a fixed position for the duration of the procedure. Adjustments are made to the stepper to optimize the position of the probe in relationship to the probe arm.

{5}------------------------------------------------

Single Use Components:

Single use, disposable components used during a Sonablate® 450 HIFU application are provided in (i) a Probe Tip Kit, and (ii) a Water Path Kit. The Probe Tip Kit contains the single use components to cover and prepare the probe tip for insertion (i.e., probe sheaths (legally marketed condoms), packets of ultrasound transmission gel, O-rings, and O-ring applicator). The Water Path Kit contains the tubing, reservoir, syringe, and other disposable components used to set up and establish coolant water flow between the probe and the Sonachill.

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

HIFU application time ranges from 2 to 4 hours, and all patient contact is limited to the rectum. Therefore, components that contact the patient have limited (< 24 hr.) contact with intact mucosa. In accordance with ISO 10993-1: Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing within a Risk Management Process, the following biocompatibility tests should be considered given this type and duration of contact: cytotoxicity, sensitization, and intracutaneous reactivity (or irritation).

ComponentMaterial(s)Direct/indirect contactAdditional information
Probe sheathPolyurethane or latexdirectOff-the-shelf finished device (Regulated under 884.5300)
Ultrasound transmission gelWater-based, off-the shelf geldirectOff-the-shelf finished device Regulated under 892.1570)
O-ringb(4) rubberdirect--
Probe componentsVarious (see diagram and table below)Direct and indirect (if the probe sheath ruptures)--
Water Path Kit componentsVarious (see table below)indirect (if the probe sheath ruptures)--

During normal clinical use, the components/materials that contact the patient are:

{6}------------------------------------------------

Image /page/6/Figure/0 description: This image is a technical drawing of a probe. The drawing labels the different parts of the probe, including the probe tip plastic tubing, probe O-rings, main bushing, U-cup seals, front bushing, probe tip neck sub-assembly, thermistor, transducer shaft, transducer housing, transducer, and probe tip button insert. The drawing provides a detailed view of the probe's construction.

{7}------------------------------------------------

ComponentMaterial(s)PatientContact
Collar Materialb(4)Direct
Probe Tip Neck Materialb(4)Direct
Probe Tip Buttonb(4)Direct
Probe O-ringsb(4)Direct
Transducerb(4)Direct
Transducer shaftb(4)Direct
Transducer housingb(4)Direct
U-Cup SealsIndirect
ThermistorIndirect
Front BushingIndirect
Probe Tip Neck SubAssemblyIndirect
Main BushingNocontact
Shaft to housing assemblyDirect
Procedure Treatment Kit1. reservoir assembly2. 2 Masterflex Tubing3. Tygon flexible plastictubing (5x6 inches, 18inches, 2x24 inches)4. 2 Syringe assembly5. 7 male luer lock to barbconnectors6. 10 Female Luers7. 2 inline Degasser Filters8. 3 port Y infusion9. 4 Male Luers with taper10. tygon double lumen tubing11. blue luer nylon insertassembly12. red male luerIndirect
Reservoir BottleIndirect
LubrisealIndirect

Components/materials that contact the patient in the event of a probe sheath rupture:

The probe sheath and ultrasound transmission gel are legally marketed products classified under separate classification regulations, and obtained by SonaCare in final, finished form. Because the probe sheath and ultrasound transmission gel are legally marketed medical devices, no further biocompatibility testing is needed.

{8}------------------------------------------------

The components of the probe tip, Probe Tip Kit, and Water Path Kit that directly or indirectly contact the patient in instances of probe sheath breakage, including the O-rings that are used to secure the probe sheath to the probe tip, separately underwent cytotoxicity, sensitization, acute systemic toxicity, and irritation testing in accordance with ISO 10993 methods, and the results were found to be acceptable.

SHELF LIFE/STERILITY/REUSE

Sterilization:

The Probe Tip Kit and Water Path Kit are packaged by SonaCare Medical in a polyethylene tray with Tyvek lid, and are sterilized by b(4) to ensure a sterility assurance level of 10°. The sterilization process validation and routine monitoring comply with ISO 11137 (Sterilization of Health Care Products -Requirements for Validation and Routine Control - Radiation Sterilization).

The hardware components andsubassemblies that comprise the Sonablate® 450 (i.e., Sonasource Console, Sonachill, probe, multi-axis stepper, and probe arm) are provided non-sterile for reuse.

Shelf Life:

The components of the Probe Tip Kit have individually-labeled expiration dates (i.e., the probe sheaths and packets of ultrasound transmission gel are legally marketed devices with expiration dates from the original manufacturers, and the O-rings and O-ring applicator are labeled with a 3-year shelf life based on stability testing conducted by SonaCare Medical). Therefore, the labeled shelf life of the Probe Tip Kit is based on the minimum (soonest) expiration date of the internal components. A 3-vear shelf life of the sterile packaging of the Probe Tip Kit components is supported by accelerated shelf life testing of packaging integrity (i.e., seal strength and bubble emission tests).

The Water Path Kit is labeled with a 12-month shelf life. This shelf life is supported by accelerated shelf life testing of packaging integrity (i.e., seal strength and bubble emission tests) and component functional integrity/performance.

Reuse:

The hardware components/subassemblies that comprise the Sonablate® 450 (i.e., Sonasource console, Sonachill, probe arm, multi-axis stepper, and probe) are reusable, and must be either cleaned, or cleaned and either high level disinfected or sterilized, between uses:

  • Sonasource console: cleaning ●
  • Sonachill: cleaning ●
  • Probe arm: cleaning and steam sterilization ●
  • Multi-axis Stepper: cleaning and steam sterilization .

{9}------------------------------------------------

  • . Probe: cleaning, and either high level disinfection (HLD) or ethylene oxide (EO) sterilization
    The Sonasource console and Sonachill have non-critical Spaulding classifications, and require only routine manual cleaning of external surfaces between uses.

The probe arm and multi-axis stepper also have non-critical Spaulding classifications, but have increased risk of contamination by the physician's gloved hand. Therefore, the labeling instructs users to both manually clean and steam sterilize these components between uses.

The probe has a semi-critical Spaulding classification, particularly in the event that the probe sheath breaks. Therefore, the probe must be manually cleaned and either undergo HLD or EO sterilization between uses.

Validation testing was performed for each of these reprocessing steps:

  • The cleaning validation of each reusable component was adequately performed. ● and all samples met the pre-specified acceptance criteria.
  • . The recommended HLD procedure for the probe was validated to achieve a 6-log reduction of mycobacteria. Cytotoxicity testing performed on probes that underwent HLD followed by simulated clinical usage verified that unsafe levels of disinfectant solution do not remain on the device.
  • . The recommended steam sterilization cvcle for the probe arm and multi-axis stepper was validated for a SAL of 10-9.
  • The recommended EO sterilization cvcle for the probe was validated for a SAL of . 10 °. EO residual levels were verified to be below unsafe levels.

ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY

Electromagnetic Compatibility (EMC):

The Sonablate® 450 was evaluated for conformance to IEC 60601-1-2, and was found to comply with all applicable requirements of this EMC testing standard. Emissions were within acceptable levels, and the device exhibited adequate immunity to electromagnetic disturbances. The user's manual was reviewed from the EMC perspective, and was found to contain comprehensive warnings to the user on recognizing and avoiding common situations that could cause electromagnetic interference (EMI), as well as a summary of the EMC testing per the IEC 60601-1-2 standard.

Electrical Safety:

The Sonablate® 450 was evaluated for conformance to IEC 60601-1 (general requirements) and IEC 60601-2-37 (particular requirements for ultrasonic diagnostic and monitoring equipment). Review of the results concluded that the device complies with all of the electrical safety requirements specified in these standards.

{10}------------------------------------------------

SOFTWARE

The Sonablate® 450 software performs/controls the following functions:

  • . HIFU application procedure,
  • Imaging, monitoring, interface with the display, printer, and keyboard, ●
  • Data storage, .
  • Motors controlling transducer placement, ●
  • Watchdog timer for safety, and ●
  • . Alarms.

This information was provided and reviewed in accordance with the FDA documents "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Guidance for Industry. FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices," and found to be complete. Details of the software review are described below:

  • The operating system is Windows XP Professional. .
  • The Sonablate® 450 uses the following off-the-shelf software: Java, Co-routine . for Java, SnagIt, and Embedded board code.
  • The following software documentation, specified in the FDA document . "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," was provided and is adequate:
Version: 6.0
Level of Concern: Major
YesNo
Software descriptionX
Device Hazard AnalysisX
Software Requirements SpecificationsX
Architecture Design ChartX
Design SpecificationsX
Traceability Analysis/MatrixX
DevelopmentX
Verification & Validation TestingX
Revision level historyX
Unresolved anomaliesX

PERFORMANCE TESTING - BENCH

Ultrasound characterization and output performance tests:

Transducer characterization:

Ultrasound transducer characterization and calibration tests were performed on the

{11}------------------------------------------------

Sonablate® 450. Specifically, to demonstrate that the ultrasound output from the transducers is consistent and within expected levels, the technical parameters described in the table below were measured:

TestPurposeAcceptance CriteriaResults
Schlieren test(Uses images of thetherapy and imagingtransducers, along andperpendicular to thedirection of apropagating lightbeam, to measurepatterns of refractionproduced by localchanges in waterdensity.)To measure the beamprofile (width) andalignment(misregistration) ofeach transducer.Image: b(4)All test samples(n=7) met theacceptance criteria.
Electrostatic test(Applies an electricalimpulse to thetransducer, andanalyzes the resultingacoustic wave.)To determine the focallength of eachtransducer.Image: b(4)All test samples(n=7) met theacceptance criteria.These results supportthe targetingaccuracy andreproducibility of theHIFU output.
Force balance test(Measures themechanical energyproduced by thetransducer.)To measure totalacoustic power (TAP)from each transducer,and calculate the focalintensity and acousticpressure.Image: b(4)*There are noacceptance criteria forfocal intensity andacoustic pressure; theywere calculated topermit computermodeling of focal sitetemperature.Image: b(4)All test samples(n=7) met theacceptance criteria.

Transducer Characterization Test Summary:

Imaging verification:

The tests summarized in the table below were performed to verify the imaging functionality of the ultrasound transducers:

{12}------------------------------------------------

TestPurposeAcceptance CriteriaResults
Phantom testing usinga standard ATS LabsphantomTo measure thefollowingcharacteristics of eachtransducer:Linear & sectorresolutionPenetration depthSignal-to-noise ratio(SNR)Both transducers:Image: b(4)The test sample (n=1)met the acceptancecriteria.The resolution issufficient for tissuedifferentiation of theprostate andsurrounding anatomy.
Imaging frame ratetestTo measure the linearand sector frame rateof each transducer.Image: b(4)The test sample (n=1)met the acceptancecriteria.The frame rate issufficient to providefeedback to thephysician duringprobe positioning.
Image display testTo verify that thesystem is capable ofperforming transverseand longitudinalimaging using eachprobe, andsimultaneouslydisplaying themUsing each transducer:the system is performstransverse andlongitudinal imagingusing each probe, anddisplays themsimultaneouslyThe test sample (n=1)met the acceptancecriteria.

Imaging Verification Test Summary:

HIFUverification/safety:

The tests summarized in the table below were performed to verify the key aspects of the Sonablate® 450 that relate to its therapeutic ultrasound (i.e., HIFU) output:

Verification Test Summary:
TestPurposeAcceptance CriteriaResults
Ex vivo tissuephantom tests(HIFU ablation inchicken breast tissue.)To confirm themaximumablationvolume anddepth that can beachieved by the systemb(4)The test sample (n=1)met the acceptancecriteria.The system iscapable of targeting a

{13}------------------------------------------------

(based on therequirements forprostate use).To verify thedimensions of theHIFU lesions createdby each transducer,and that lesions arepositioned to create acontiguous volume.tissue volume of upto 40 cm³, at a depthof up to 40 mm. Thesystem is capable ofcreating single andvolume lesions in x,y, and z dimensionsthat meet theacceptance criteria.These results supportthe targetingaccuracy andreproducibility of theHIFU output.
HIFU ON-time/OFF-time VerificationTesting(Using anoscilloscope tomeasure the durationsof each period of theHIFU applicationcycle.)To confirm theON/OFF timing of thsystem's HIFUapplication cycle.The test sample (n=1)met the acceptancecriteria.The system meetsON time and OFFtime requirements.

General functional/mechanical/performance tests:

SonaCare Medical conducted a variety of tests to verify the proper operation/performance of the system components. These tests included verification of:

  • The ability of Sonablate® 450 subsystems (console, probe, chiller) to . communicate/interface with one another per design specifications, including simulated use testing.
  • Operation of system alarms and automatic application interruption. ●
  • Total acoustic power display (ability of the TAP meter to display this HIFU . application parameter).
  • Probe coupling performance.
  • . User interface control of energy.
  • Inability to access remotely (i.e., unable to connect to outside communication . networks).
  • Sonachill fluid cooling performance (temperature, flow rate). ●
  • Multi-axis stepper movement sensitivity, carrying capacity, reliability, and . durability.
  • Probe arm carrying capacity, resistance to probe motion (when locked), degree of . motion (when unlocked), and clamp attachment.

{14}------------------------------------------------

Each of these tests evaluated the device in an environment simulating clinical use, and confirmed that the device (or device component) successfully met the functional requirement of each test.

PERFORMANCE TESTING - ANIMAL

The following animal testing was performed to demonstrate that the Sonablate® 450 thermally ablates targeted tissue in a controlled manner, without unsafe heating or ablation of adjacent, non-target tissues:

  • Study 1 ("Safety and Effectiveness of Tissue Ablation"): A dog prostate model . study (n=5), which demonstrated that the Sonablate® 450 effectively destroys prostate tissue without rectal wall injury or ablation outside the planned target zones. This study used the 3 seconds ON/6 seconds OFF HIFU application cycle (i.e., the only HIFU application cycle available on the Sonablate® 450). In this study, thermocouples positioned in the prostate and adjacent to the rectal wall documented that the system is capable of raising target tissue temperature to above 60°C for 3 seconds without rectal wall temperatures exceeding 43°C. After HIFU application, histological evaluation of the prostate confirmed the presence of tissue coagulation in the targeted volume.
  • Study 2 ("Efficacy Improvement Software Test Report Localization and . Ablation"): A dog prostate model study (n=4), which demonstrated (i) the safety and effectiveness of a 90 degree application angle, and (ii) the ablation effectiveness of an alternate application cycle (that is not available on the Sonablate® 450).
  • Study 3 ("Treatment Volume and Accuracy Study Report"): This preliminary dog study confirmed accurate spatial localization of a specific target using ultrasound imaging, verifying that temperatures within the focal zone are sufficient to achieve coagulative necrosis (> 60℃), and that tissue outside of the focal zone (including the rectal wall) do not reach temperatures capable of coagulative necrosis.
  • . Study 4 ("Thermocouple Response to Treatment Test"): This preliminary dog study compared the accuracy of the planned HIFU application zones versus actual target tissue volume.
  • . Study 5 ("Safety of HIFU Energy Application to Adjacent Tissues"): This preliminary dog study was performed to verify that energy delivery to prostatic tissue does not damage adjacent structures, and that the use of the Sonachill cooling system during HIFU cools the rectal wall tissue.

Study 1 (primary study) was performed using the Sonablate® 450, whereas Studies 2, 3, 4, and 5 (supporting studies) were performed using a different device version that has several additional features.

SUMMARY OF CLINICAL INFORMATION

Data to support that the Sonablate® 450 safely and effectively ablates prostatic tissue were obtained from the US Salvage Study. In this study, a population of post-radiation patients

{15}------------------------------------------------

received whole gland ablation using the Sonablate® 450 as a treatment for recurrent prostate cancer. While the oncological outcomes from this study are inconclusive, the results provide reasonable assurance of safety and effectiveness of the device in the context of prostate tissue ablation.

The design of the US Salvage Study is briefly summarized below:

  • Study Design: Multi-center, non-randomized, prospective, single arm study involving . 20 sites (U.S. and Canada) and 200 subjects. Sixteen sites actively enrolled subjects.
  • . Patient Population: Men (40-85 years of age) with histologically confirmed, organconfined, recurrent, non-metastatic prostatic adenocarcinoma. All participants had documented disease relapse two or more years following external beam radiation therapy (EBRT). All participants had a prostate specific antigen (PSA) level > 0.5 ng/mL and < 10 ng/mL, and clinical stage T1c-T2 prior to radiation. Participants were candidates for salvage therapy on the basis of an assessment of general health and inter-current comorbidities.
  • . Endpoints: Effectiveness - PSA nadir of ≤ 0.5 ng/mL within 12 months, and negative prostate biopsy at 12 months. Safety - Rate of adverse event occurrence by type and severity.
  • . HIFU Application: Whole gland ablation of the prostate using the Sonablate® 450. All applications were performed under general or regional anesthesia. Prior to application, patients received a suprapubic catheter.
  • . Follow-Up Exams: Telephone follow-up occurred on day 2 (to assess for signs of infection and subject's comprehension of suprapubic catheter care). Another follow-up occurred to between 2-4 weeks post ablation to remove the catheter. Follow-up visits were scheduled for 6 weeks, 3, 6, 9, 12, months post-ablation, and annually thereafter until 5 years post-ablation.

The original objective of this study was to evaluate device use as a treatment for recurrent prostate cancer. However, to evaluate the safety and effectiveness of the device for the functional indication of prostate tissue ablation, changes were made to the endpoints and associated analyses. Specifically, "ablation effectiveness" was assessed by analyzing changes in prostate volume, PSA level, and negative biopsy rate, assessed at baseline and 12 months postablation. Safety evaluations were conducted using patient completed questionnaires and investigator reported adverse events through 12 months post-ablation.

The clinical dataset is based on an interim analysis of the initial 117 patients enrolled (of which 116 underwent HIFU application). Patients ranged in age from 53 to 83 years (mean = 69.8). These men had a mean pre-ablation PSA of 4.9 ng/mL.

Effectiveness:

The following results demonstrate that the Sonablate® 450 effectively ablates targeted prostatic tissue:

  • Among the 73 patients who had prostate volumes determined both pre- and post-ablation. . the mean reduction in volume was 11.8 cm (95% confidence limits = 9.7, 14.0). This change represents a mean reduction of 46% from the baseline volume. Of the 44 patients

{16}------------------------------------------------

who did not have volumes calculated post-ablation, there was only a single case of a positive biopsy.

  • Eighty-three percent (83%) of the total cohort of 117 patients had a reduction in PSA ● post-ablation (95% confidence limits = 74.6, 89.0). For this analysis, missing data were treated using the last value carried forward method.
  • Sixty-one percent (61%) of the total cohort of 117 patients had a negative post-ablation . biopsy (95% confidence limits = 51.2, 69.5). For this analysis, 30 patients who did not have a 12-month biopsy were considered "positive."

Safety:

The following adverse events were observed in this whole gland study of post-radiation patients (n=117 unless otherwise noted):

Adverse EventPercentage ofSubjects
Erectile Dysfunction(n=51 potent at baseline)71%
Urinary Tract Infection50%
Urinary Incontinence(leak)47%
Hematuria43%
Urinary Retention43%
Urinary Frequency35%
Urinary Incontinence(pads)32%
Urinary Urgency29%
Urinary Tract Obstruction20%
Dysuria18%
Urethral Stricture15%
Epididymitis/Orchitis8%
Bladder Neck Contracture5%
Rectal Fistula4%
Urinary Fistula2.6%
Osteomyelitis2.6%

A total of 27 device/procedure related serious adverse events (SAEs) were reported in 21 subjects. These consisted of urinary retention (n=6), hematuria (n=5), rectal/urinary fistula (n=5), UTI/urosepsis (n=5), osteomyelitis (n=3), urinary incontinence (n=1), urethral stricture (n=1), and small intestinal obstruction (n=1). Each of these events resolved.

Two patients died during the course of the study, both for reasons unrelated to HIFU application (i.e., aspiration following esophageal surgery, and motor vehicle accident).

{17}------------------------------------------------

Clinical Results Summary:

The observed changes in prostate volume, PSA level, and negative biopsy rate provide complimentary evidence that the Sonablate® 450 is capable of ablating prostate tissue. The observed 46% reduction in total volume of the prostate demonstrates bulk removal of a large quantity of prostatic tissue when a whole-gland strategy is employed with the Sonablate® 450, and the percentage of patients experiencing a PSA reduction post-ablation (i.e., 83%) demonstrates that the ablation effect was achieved in a large majority of study subjects. The increase in negative biopsy rate associated with HIFU application (i.e., from 0% at baseline to 61% at 12 months post-ablation) provides further evidence from histology that Sonablate® 450 ablates prostate tissue.

With the number of patients treated and the duration of follow-up, the risk proffile associated with Sonablate® 450 ablation is well-characterized in the context of this study population. In this study, 99% of patients experienced some type of application-related complication. Serious adverse events (SAEs) occurring in the population included urinary retention/obstruction (5%), hematuria (4%), rectal fistula (4%), urinary tract infection (4%), and osteomyelitis (3%). Nonserious adverse events primarily consisted of erectile dysfunction (71%), UTI (54%), urinary incontinence (47%), urinary retention (43%), hematuria (31%), urinary obstruction (20%), and urethral stricture (15%). The duration of a harmful event was dependent on the indication for prostate ablation and current disease status. Some of the harmful events such as voiding symptoms resolve on their own over weeks to months, and some such as infections resolve with medical therapies. Of the three patients with osteomyelitis, one patient underwent cystectomy and ileal conduit diversion. Some of the patients with urinary incontinence and erectile dysfunction resolve spontaneously over a period of months while a significant percentage have lifelong symptoms requiring medical and surgical intervention. These include placement of an artificial urinary sphincter for treatment of incontinence and medications or devices for treatment of erectile dysfunction. The adverse event profile associated with use of the Sonablate® 450 may be different in patients who receive ablation to less than the entire prostate, or who have not had a prior treatment to their prostate.

While this HIFU application using Sonablate® 450 has documented morbidity, the benefits outweigh the risks in certain populations of men seeking ablation of unwanted prostate tissue.

LABELING

Labeling has been provided that includes instructions for the physician, and an appropriate prescription statement as required by 21 CFR 801.109. The labeling includes:

  • directions on device set-up and HIFU application planning, delivery, and completion . using the Sonablate® 450 device;
  • . a "Clinical Studies" section that summarizes the clinical trial results (i.e., evidence of prostate tissue ablation and the adverse event data), and states that the safety and effectiveness of the device in other clinical usages and patient populations have not been established;
  • a summary of the physician training program: .

{18}------------------------------------------------

  • a training course providing education on the procedure, transrectal ultrasound for o HIFU, and post-HIFU recovery and care,
  • a didactic presentation, HIFU simulator training, and completion of at least five O cases with a trained technician onsite, and
  • completion of at least five cases that are remotely monitored by a trained o technician;
  • statement that use of the Sonablate® 450 must be prescribed and administered under the ● direct supervision of a qualified and trained physician, after appropriate urologic patient evaluation;
  • . contraindication statements to avoid use of the device in patients in whom the benefits do not outweigh the risks, such as:
    • large (> 10 mm) cysts or calcifications in the planned ablation zone, O
    • metal implants or stents in the urethra. O
    • brachytherapy seeds in the planned ablation zone, о
    • pre-existing inflammatory disease of the colon or rectum (such as proctitis or O ulcerative colitis).
    • prior significant rectal surgery. O
    • inability to insert or tolerate a transrectal ultrasound probe, O
    • active urinary tract infection. o
    • urethral stricture, and O
    • latex allergy; o
  • . warning and precaution statements to mitigate potential risks in the clinical setting, such as:
    • the Sonablate® 450 is not capable of ablating anterior prostate tissue > 4 cm from o the transducer (due to physical limitations of the device);
    • in patients who have previously received radiation to the pelvic region, the o Denonvillers' fascia should not be included within the targeted ablation zone, and the manufacturer's recommended values of total acoustic power should be followed:
    • the effectiveness of the device in treating any specific prostate disease has not о been established:
    • the safety and effectiveness of the Sonablate® 450 have not been established in o patients with the following conditions:
      • interest in future fertility, ■
      • I bleeding disorders.
      • l clinical or histological evidence of urinary bladder cancer,
      • l renal impairment,
      • functional bladder problems, including neurological disorders that might affect bladder function,
      • . post-void residual urine > 250 mL,
      • I urinary retention requiring an indwelling catheter,
      • l enlarged medial lobe of the prostate protruding into the urinary bladder,
      • I clinical evidence of prostatitis within 6 months prior to HIFU,
      • l bladder calculus, and
      • prostates > 40 cm3.

{19}------------------------------------------------

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of the high intensity ultrasound system for prostate tissue ablation and the measures necessary to mitigate these risks.

Identified RiskMitigation Measure
Thermal injury from high intensityultrasound exposure to non-target tissue:-erectile dysfunction-urinary incontinence-rectal fistula-osteomyelitis pubisNon-clinical performance testing Software verification, validation, and hazard analysis In vivo testing Clinical testing Labeling Physician training
Thermal injury from high intensityultrasound exposure to target tissue:-urethral stricture-bladder neck contracture-urinary retention-tissue debris/obstruction-voiding dysfunction-dysuria-hematuria-ejaculation disorderClinical testing Labeling Physician training
Mechanical injury from unintentionalmovement of ultrasound components:-patient rectal injury-operator hand injurySoftware verification, validation, and hazard analysis Clinical testing Labeling Physician training
InfectionSterilization validation Reprocessing validation Shelf life validation Labeling
Electrical shock/ElectromagneticinterferenceElectrical safety testing Electromagnetic compatibility testing Labeling
Adverse tissue reactionBiocompatibility testing

SPECIAL CONTROLS :

In combination with the general controls of the FD&C Act, the high intensity ultrasound system for prostate tissue ablation is subject to the following special controls:

  • (1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Characterization of acoustic pressure and power output at clinically relevant levels.

{20}------------------------------------------------

  • (B) Measurement of targeting accuracy and reproducibility of high intensity ultrasound output.
  • (C) Ultrasound-induced heating verification testing at target and non-target tissues.
  • (D) Electrical safety testing.
  • (E) Electromagnetic compatibility testing.
  • (2) Software verification, validation, and hazard analysis must be performed.
  • (3) The elements of the device that may contact the patient's mucosal tissue must be demonstrated to be biocompatible.
  • (4) Performance data must demonstrate the sterility of the device components that contact the patient's mucosal tissue.
  • (5) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the identified shelf life.
  • (6) Performance data must support the instructions for reprocessing all reusable components.
  • (7) In vivo testing must demonstrate that the device thermally ablates targeted tissue in a controlled manner without thermal injury to adjacent, non-target tissues.
  • (8) Clinical testing must document the adverse event profile, provide evidence of prostatic ablation, and demonstrate that the device performs as intended under anticipated conditions of use.
  • (9) Training must be provided so that upon completion of the training program, the physician can:
    • (A) Use all safety features of the device:
    • (B) Accurately target the high intensity ultrasound energy within the desired region of the prostate;
    • (C) Perform the ablation procedure in a manner that minimizes damage to non-target tissues;
  • (10) Labeling must include:
    • (A) A section that summarizes the clinical testing results, including the adverse event profile and evidence of prostate ablation achieved.
    • (B) An expiration date or shelf life for single use components.

BENEFIT/RISK DETERMINATION

The risks of the device are based on the non-clinical laboratory studies, as well as data collected in a clinical study described above. The pivotal study indicated that nearly all (i.e., 99%) subjects experienced a HIFU application-related complication. Serious adverse events (SAEs) occurring in the population included urinary retention (5%), hematuria (4%), rectal fistula (4%), urinary tract infection (4%), and osteomyelitis (3%). Non-serious adverse events primarily consisted of erectile dysfunction (71%), UTI (50%), urinary incontinence (leak) (47%), hematuria (43%), urinary retention (43%), urinary obstruction (20%), and urethral stricture (15%). The duration of a harmful event was dependent on the indication for prostate ablation and current disease status. Some of the harmful events such as voiding symptoms resolve on their own over weeks to months, and some such as infections resolve with medical therapies. Of the three patients with osteomvelitis, one patient underwent cystectomy and ileal conduit diversion. Some of the patients with urinary incontinence and erectile dysfunction resolve spontaneously over a period of months while a significant percentage have lifelong symptoms

{21}------------------------------------------------

requiring medical and surgical intervention. These include placement of an artificial urinary sphincter for treatment of incontinence and medications or devices for treatment of erectile dysfunction.

The probable benefits of the device are also based on the non-clinical laboratory studies as well as the data collected in a clinical study described above. For this de novo, effectiveness was defined as evidence that Sonablate® 450 ablates prostate tissue within the targeted region, as demonstrated by reductions in prostate volume and PSA level, and increase in the rate of negative biopsy, between baseline and 12 months post-HIFU. Ninety percent (90%) of the perprotocol population had a decrease in prostate size. The average reduction was approximately 50%. Across the entire population, 83% had a reduction in PSA at 12 months. Sixty-one percent (61%) of the intent to treat protocol population had a negative biopsy 12 months post-HIFU.

Additional factors to be considered in determining probable risks and benefits for the high intensity ultrasound system for prostate tissue ablation include:

  • Robustness was enhanced by the use of objective endpoints for safety and documentation . of tissue ablation. However, robustness was challenged by the reliance on a single-arm study and by missing data.
  • It is likely that the device can ablate prostate tissue in a range of clinical settings.
  • Post-radiation patients have limited options for removal or ablation of unwanted prostate . tissue.
  • The submission did not include patient perspective information: however, prior input ● from the advisory panel indicates that there is a low tolerance for risks since the majority of patients with this condition are currently observed.
  • . An extensive physician training program is in place to mitigate risks. Limitations in the labeling also mitigate risks.
  • . The technology is a novel strategy for ablating prostate tissue.
  • . The effectiveness of the device in ablating prostate tissue appears to be similar to that of existing methods, with comparable safety.

In conclusion, given the available information above, the data support that, for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue. the probable benefits outweigh the probable risks for the high intensity ultrasound system for prostate tissue ablation. The device provides substantial benefits and the risks can be mitigated by the use of general and the identified special controls.

{22}------------------------------------------------

CONCLUSION

The de novo for the high intensity ultrasound system for prostate tissue ablation is granted and the device is classified under the following:

Product Code: PLP Device Type: High intensity ultrasound system for prostate tissue ablation Class: II Regulation: 21 CFR 876.4340

§ 876.4340 High intensity ultrasound system for prostate tissue ablation.

(a)
Identification. A high intensity ultrasound system for prostate tissue ablation is a prescription device that transmits high intensity therapeutic ultrasound energy into the prostate to thermally ablate a defined, targeted volume of tissue, performed under imaging guidance. This classification does not include devices that are intended for the treatment of any specific prostate disease and does not include devices that are intended to ablate non-prostatic tissues/organs.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Characterization of acoustic pressure and power output at clinically relevant levels;
(ii) Measurement of targeting accuracy and reproducibility of high intensity ultrasound output;
(iii) Ultrasound-induced heating verification testing at target and non-target tissues;
(iv) Electrical safety testing; and
(v) Electromagnetic compatibility testing.
(2) Software verification, validation, and hazard analysis must be performed.
(3) The elements of the device that may contact the patient's mucosal tissue must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device components that contact the patient's mucosal tissue.
(5) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the identified shelf life.
(6) Performance data must support the instructions for reprocessing all reusable components.
(7)
In vivo testing must demonstrate that the device thermally ablates targeted tissue in a controlled manner without thermal injury to adjacent, non-target tissues.(8) Clinical testing must document the adverse event profile, provide evidence of prostatic ablation, and demonstrate that the device performs as intended under anticipated conditions of use.
(9) Training must be provided so that upon completion of the training program, the physician can:
(i) Use all safety features of the device;
(ii) Accurately target the high intensity ultrasound energy within the desired region of the prostate; and
(iii) Perform the ablation procedure in a manner that minimizes damage to non-target tissues.
(10) Labeling must include:
(i) A section that summarizes the clinical testing results, including the adverse event profile and evidence of prostate ablation achieved; and
(ii) An expiration date or shelf life for single use components.