K Number
DEN130007
Manufacturer
Date Cleared
2013-12-18

(302 days)

Product Code
Regulation Number
876.5015
Panel
GU
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AXIOS Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, with ≥ 70% fluid content that are adherent to the bowel wall. Once placed, the AXIOS Stent functions as a port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution.

Device Description

The AXIOS Stent and Delivery System consists of two major components, the catheter-based delivery system and the implantable stent. The stent is a flexible, fully-covered. self-expanding Nitinol stent, which is preloaded within the catheter-based delivery system. The system is intended to cannulate the transgastric or transduodenal wall into a pancreatic pseudocyst for endoscopic drainage. The stent serves as a conduit for passive drainage of pseudocyst contents directly into the GI tract. The large lumen diameter provides a short path and may allow secure access to the pseudocyst interior for additional diagnostic or therapeutic procedures. The AXIOS Stent is designed with a duplicate "anchor" or "flange" on each end to achieve tissue apposition and prevent migration. The stent is radiopaque and is fully covered with silicone to prevent leakage, minimize tissue in-growth, and facilitate removal. The AXIOS Stent will be provided in two lumen diameters (10mm and 15mm) and one length (10mm). The AXIOS Delivery System consists of a catheter and an integrated handle with manual controls for positioning and deploying the AXIOS stent. The AXIOS Delivery System is sized to fit within commercially available echoendoscopes with a working channel of 3.7 mm diameter or larger. The Delivery System catheter is 138 cm in working length and attached directly to the endoscope handle. The AXIOS Delivery System requires an access site of at least 10 Fr. The AXIOS Stent and Delivery System is provided sterile, disposable and intended for use during a single patient procedure.

AI/ML Overview

Acceptance Criteria and Device Performance for the AXIOS Stent and Delivery System

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the AXIOS Stent and Delivery System were primarily drawn from the "Safety Endpoints" and "Effectiveness Endpoints" sections of the pivotal clinical study.

Acceptance Criteria CategorySpecific Acceptance CriterionReported Device Performance (Pivotal Study, Per-Protocol population unless otherwise specified)
EffectivenessTechnical success (placement of AXIOS Stent using AXIOS Delivery System and removal with standard endoscopic snare)Placement: 90.9% (30/33) in ITT population. Removal: 96.6% (28/29) using standard endoscopic tools; one was inadvertently dislodged.
Stent lumen patency at 30 days and/or 60 days (protocol defined drainage, including partial drainage, as patent)93.1% (27/29) had stent lumens that were patent at stent removal. (Note: 9 required debridement, 3 required supplemental stenting, 2 of which were non-patent at removal and considered failures for patency, though partial drainage considered success for the third).
Stent removability at 30 days and/or 60 days96.6% (28/29) successfully removed using standard endoscopic tools.
Clinical success (at least a 50% decrease in pseudocyst size, based on radiographic analysis, at 30 days and/or 60 days)86.2% (25/29) achieved at least a 50% decrease in pseudocyst size at time of stent removal. (Note: This includes 75.9% at 30 days, and 77.8% for the 9 subjects remaining at 60 days. The failures included 2 with

§ 876.5015 Pancreatic drainage stent and delivery system.

(a)
Identification. A pancreatic drainage stent is a prescription device that consists of a self-expanding, covered, metallic stent, intended for placement to facilitate transmural endoscopic drainage of pancreatic pseudocysts. This stent is intended to be removed upon confirmation of pseudocyst resolution. This device may also include a delivery system.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device and elements of the delivery device that may contact the patient must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of patient-contacting components of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the requested shelf life.
(4) Non-clinical testing data must demonstrate that the stent and delivery system perform as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing of the stent and delivery system must be conducted under simulated use conditions.
(ii) Removal force testing must be conducted. The removal force testing must demonstrate that the stent can be safely removed, and that the stent will remain in place when subjected to forces encountered during use.
(iii) Expansion force testing must be conducted. The expansion force must demonstrate that the forces exerted by the stent will not damage the tissue surrounding the stent.
(iv) Compression force testing must be conducted. The compression force must demonstrate that the stent will withstand the forces encountered during use.
(v) Dimensional verification testing must be conducted.
(vi) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(vii) Fatigue testing must be conducted. Material strength must demonstrate that the stent will withstand forces encountered during use.
(viii) Corrosion testing must be conducted. Corrosion resistance must demonstrate that the stent will withstand conditions encountered during use.
(5) Non-clinical testing must evaluate the compatibility of the stent in a magnetic resonance (MR) environment.
(6) Well-documented clinical experience must demonstrate safe and effective use, and capture any adverse events observed during clinical use.
(7) Labeling must include the following:
(i) Appropriate instructions, warnings, cautions, limitations, and information related to the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device.
(ii) A warning that the safety and patency of the stent has not been established beyond the duration of the documented clinical experience.
(iii) Specific instructions and the qualifications and clinical training needed for the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device.
(iv) Information on the patient population for which the device has been demonstrated to be effective.
(v) A detailed summary of the clinical experience pertinent to use of the device.
(vi) A detailed summary of the device technical parameters.
(vii) A detailed summary of the device- and procedure-related complications pertinent to use of the device.
(viii) An expiration date/shelf life.