K Number
DEN120008
Date Cleared
2012-01-20

(14 days)

Product Code
Regulation Number
866.3336
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STRATIFY JCVTM Antibody ELISA testing service provided by Focus Diagnostics is intended for the qualitative detection of antibodies to John Cunningham virus in human serum or plasma. The assay is intended for use in conjunction with other clinical data, in multiple sclerosis and Crohn's disease patients receiving natalizumab therapy, as an aid in risk stratification for progressive multifocal leukoencephalopathy development. The assay is for professional use only and is to be performed only at Focus Diagnostics' Reference Laboratory.

The assay is not intended for donor screening. The performance of this assay has not been established for use in other immunocompromised patient populations or in neonates and pediatrics patient populations.

Device Description

The STRATIFY JCV™ Antibody ELISA consists of two devices. The first is an initial detection ELISA and the second device is a confirmatory (inhibition) ELISA. Both tests utilize the same recombinant antigen which is used in two different formats as described in section L. The devices include the following reagents; Recombinant JC virus like particles (VLP), a JCV high and low positive controls, consisting of human sera that is positive for JCV antibodies, and JCV negative control consisting of human sera that is negative for JCV antibodies. The conjugate, substrate, wash buffers and blocking buffers needed for the test are not supplied with the device and are listed together with other reagents and consumables in the device labeling.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies that support the performance of the STRATIFY JCV™ Antibody ELISA, based on the provided 510(k) summary:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance (Summary)
Precision (Detection Assay)Total % CV should be acceptable for intended clinical use.Total % CV ranged from 7.0% to 18.3% for plasma, and 9.1% to 12.9% for serum.
Precision (Detection Assay - Indeterminate)Indeterminate samples should consistently yield indeterminate results.Serum indeterminate pool returned indeterminate 96.7% (87/90) of the time; plasma indeterminate pool returned indeterminate 100% of the time.
Precision (Confirmation Assay)Indeterminate samples, when confirmed, should consistently yield positive results.Serum indeterminate pool confirmed positive 100% (86/86) of the time; plasma indeterminate pool confirmed positive 100% (90/90) of the time.
Analytical Specificity (Cross-Reactivity - Spiked Antibodies)No observed reactivity with common cross-reacting antibodies.No observed reactivity (0/3 detected for each) with antibodies to E. coli, M. tuberculosis, and Pneumocystis jiroveci.
Analytical Specificity (Cross-Reactivity - Seroprevalence)JCV seroprevalence in groups with other antibodies should not be significantly higher than expected background.Three groups (C. pneumoniae, HSV1, T. pallidum) exhibited slightly higher positivity than expected the normal population, but no direct cross-reactivity was explicitly proven in this study design. M. pneumoniae showed lower.
Analytical Specificity (Interferences - Endogenous Substances)Signal change due to interferents should not alter result interpretation.Most interferents (Albumin, Ascorbic Acid, Cholesterol) showed acceptable shifts. Triglyceride, Hemoglobin, and Bilirubin in plasma, and Hemoglobin and Bilirubin in serum, demonstrated >20% shift from baseline, indicating potential interference.
Analytical Specificity (Interferences - Concomitant Medications)Concomitant medications should not interfere with JCV antibody detection.No potential interfering effect was observed for commonly used medications (Paracetamol, Ibuprofen, Methylprednisolone, Amoxicillin, Acetylsalicylic Acid, Multi-Vitamin, Amantadine, Diclofenac) based on longitudinal analysis in patient cohorts.
False Negative Rate (JCV Viruric Patients)False negative rate should be low and acceptable.Initial assessment from STRATA study: 2.5% (95% CI: 0.5 to 4.9%) based on 5/204 viruric patients. Confirmed in STRATIFY-1 study: 2.7% (95% CI: 0.9 to 6.2%) based on 5/186 viruric patients.
Method Comparison (with LDT)Acceptable agreement with a laboratory-developed test.Positive Percent Agreement (PPA) of 91.0% (122/134) with 95% CI: 85.0% to 94.8%. Negative Percent Agreement (NPA) of 89.4% (59/66) with 95% CI: 79.7% to 94.8%.
Matrix Comparison (Serum vs. Plasma)Qualitative results from serum and plasma should show equivalence.100% (50/50) qualitative agreement between paired serum and plasma samples. Mean % difference for nOD values was 4.9%. Regression slope was 0.99 (95% CI: 0.9695 to 1.0103) and intercept 0.0 (95% CI: -0.0034 to 0.0043).
Clinical Performance (PML Risk Stratification)High sensitivity for JCV antibodies in PML patients prior to diagnosis. Ability to stratify PML risk.100% sensitivity (37/37) with 95% CI: 90.6% to 100% in pre-PML samples from natalizumab-treated patients. JCV antibody positive status results in a 1.82-fold increased risk of PML.

Study Details

2. Sample size used for the test set and the data provenance:

  • Precision Studies (Detection & Confirmation):
    • Sample Size: 90 aliquots for each plasma and serum level (High Positive, Indeterminate, Low Positive, Negative) in the detection assay. The indeterminate pools then subjected to confirmation (90 plasma indeterminate, 86 serum indeterminate).
    • Data Provenance: Not explicitly stated, but likely from laboratory internal samples/controls, potentially spiked to achieve different concentrations.
  • Analytical Specificity (Cross-Reactivity - Spiked Antibodies):
    • Sample Size: 3 replicates for each of 3 spiked antibodies (total 9 tests for serum, 9 for plasma).
    • Data Provenance: Laboratory, using commercially available human antibodies spiked into JCV negative serum/plasma.
  • Analytical Specificity (Cross-Reactivity - Seroprevalence):
    • Sample Size: Panels of up to twenty remnant specimens (e.g., 20 for C. pneumoniae, HSV6, VZV, Candida, HSV1, HSV2; 4 for T. pallidum; 9 for M. pneumoniae; 15 for CMV; 20 for HIV1). Total 188 samples across various sero-positive groups.
    • Data Provenance: Remnant clinical specimens, likely retrospective from various patient populations.
  • Analytical Specificity (Interferences - Endogenous Substances):
    • Sample Size: Not explicitly stated as a number of samples, but involved spiking a single low-concentration JCV antibody native serum and plasma panel with various interferents and comparing to baseline.
    • Data Provenance: Laboratory, using native serum/plasma samples and spiked interferents.
  • Analytical Specificity (Interferences - Concomitant Medications):
    • Sample Size: Subset of 585 MS patients from the AFFIRM study.
    • Data Provenance: Prospective clinical trial (AFFIRM study) data, with serial samples collected over 30 months. Patients were treated for Multiple Sclerosis.
  • False Negative Rate (JCV Viruric Patients):
    • STRATA Study: 204 viruric patients.
    • STRATIFY-1 Study: 186 urinary JCV DNA positive subjects (out of 1073 total subjects).
    • Data Provenance: Prospective clinical studies (STRATA and STRATIFY-1), involving MS patients.
  • Method Comparison with LDT:
    • Sample Size: 200 samples (62 negative, 71 low positives near cut-point, 67 positive).
    • Data Provenance: Not explicitly stated, but presumably clinical samples, likely retrospective.
  • Matrix Comparison:
    • Sample Size: 50 paired plasma and serum samples.
    • Data Provenance: Clinical samples (presumably from patients).
  • Clinical Performance (PML Risk Stratification - Pre-PML Samples):
    • Sample Size: 37 available serum samples from confirmed PML patients.
    • Data Provenance: Retrospective collection of serum samples from patients who developed PML, collected specifically from clinical trials and post-marketing reports prior to PML diagnosis.
  • Clinical Performance (PML Risk Stratification - MS Population JCV Positivity):
    • Sample Size: 5,896 MS patients.
    • Data Provenance: Clinical trials (AFFIRM, TYGRIS-US, STRATIFY-1) and a national MS registry (Sweden). This is a large, geographically diverse, retrospective and potentially prospective (from ongoing studies) cohort of MS patients.
  • Clinical Performance (JCV Antibody Prevalence - CD Patients):
    • Sample Size: 313 CD patients.
    • Data Provenance: Two completed Phase 3 clinical trials (CD301 and CD303) of natalizumab in Crohn's disease patients. Retrospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not mention the use of experts to establish ground truth for the test sets (patient samples, viruric samples, etc.). The ground truth for these studies relies on other laboratory methods (e.g., urinary JCV DNA positivity for viruric patients) or clinical diagnoses (e.g., confirmed PML diagnosis for the pre-PML samples).

4. Adjudication method for the test set:

  • The document does not describe an adjudication method involving experts for any of the test sets. Ground truth is established by laboratory results (e.g., confirmed urinary JCV DNA positivity) or clinical diagnosis of PML, not by expert consensus on the samples themselves.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is an ELISA (Enzyme-Linked Immunosorbent Assay) for detecting antibodies, not an imaging device or AI-assisted diagnostic tool that would involve human readers. Therefore, an MRMC study is not applicable and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Yes, effectively. The STRATIFY JCV™ Antibody ELISA is a laboratory assay. Its performance characteristics (precision, specificity, false negative rates, and clinical performance) are evaluated as a standalone test, meaning the results are generated directly by the assay without real-time human interpretation or modification of the assay output. Human intervention occurs in performing the assay and interpreting the final quantitative (nOD, % Inhibition) or qualitative (Positive/Negative/Indeterminate) results based on predefined cut-offs.

7. The type of ground truth used:

  • JCV Viruric Patients: JCV DNA positivity in urine (confirmed infection).
  • PML Risk Stratification: Clinical diagnosis of Progressive Multifocal Leukoencephalopathy (PML), along with historical records of natalizumab treatment and prior immunosuppressant use.
  • Method Comparison: The Laboratory Developed Test (LDT) served as the comparative method, acting as a de facto reference in that specific study.
  • For other analytical studies (precision, cross-reactivity, interference), ground truth was established by known concentrations/statuses of analytes or known presence/absence of cross-reactants/interferents.

8. The sample size for the training set:

  • The document does not explicitly describe a "training set" in the context of a machine learning algorithm. This is a traditional IVD ELISA kit.
  • However, the cut-off for the assay (nOD > 0.25 positive, < 0.1 negative, 0.1-0.25 indeterminate) was established based on the distribution of serological responses from JCV viruric patients. Specifically, "None of the JC viruric subjects had a reactivity below 0.1, which was therefore selected as the lower cut point..."
  • The "training" data for establishing the initial cut-off therefore involved an unspecified number of "JCV viruric patients" as a positive reference. For the false negative rate, 204 viruric patients from the STRATA study and 186 viruric patients from the STRATIFY-1 study were used to validate the cut-off.

9. How the ground truth for the training set was established:

  • As mentioned above, the primary ground truth for setting the initial cut-off was based on samples from JCV viruric patients. These patients were confirmed as viruric through detection of JCV DNA in their urine. The rationale is that JCV infection is often asymptomatic and hard to confirm by other means, but urinary shedding of JCV DNA is a clear indicator of infection.

{0}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:

K112394

B. Purpose for Submission:

New Device

C. Measurand:

Anti- John Cunningham Virus (JCV) antibodies

D. Type of Test:

Enzyme Linked Immunosorbent Assay (ELISA)

E. Applicant:

Focus Diagnostics Inc.

F. Proprietary and Established Names:

STRATIFY JCV™ Antibody ELISA

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.3336: John Cunningham Virus serological reagents
    1. Classification:
      Class II (de novo)
    1. Product code:
      OYP
    1. Panel:
      Microbiology (83)

{1}------------------------------------------------

H. Intended Use:

1. Intended use(s):

The STRATIFY JCVTM Antibody ELISA testing service provided by Focus Diagnostics is intended for the qualitative detection of antibodies to John Cunningham virus in human serum or plasma. The assay is intended for use in conjunction with other clinical data, in multiple sclerosis and Crohn's disease patients receiving natalizumab therapy, as an aid in risk stratification for progressive multifocal leukoencephalopathy development. The assay is for professional use only and is to be performed only at Focus Diagnostics' Reference Laboratory.

The assay is not intended for donor screening. The performance of this assay has not been established for use in other immunocompromised patient populations or in neonates and pediatrics patient populations.

    1. Indication(s) for use:
      Same as intended use
    1. Special conditions for use statement(s):
      N/A
    1. Special instrument requirements:
      N/A

I. Device Description:

The STRATIFY JCV™ Antibody ELISA consists of two devices. The first is an initial detection ELISA and the second device is a confirmatory (inhibition) ELISA. Both tests utilize the same recombinant antigen which is used in two different formats as described in section L. The devices include the following reagents; Recombinant JC virus like particles (VLP), a JCV high and low positive controls, consisting of human sera that is positive for JCV antibodies, and JCV negative control consisting of human sera that is negative for JCV antibodies. The conjugate, substrate, wash buffers and blocking buffers needed for the test are not supplied with the device and are listed together with other reagents and consumables in the device labeling.

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      None

{2}------------------------------------------------

    1. Predicate 510(k) number(s): None
    1. Comparison with predicate: N/A

K. Standard/Guidance Document Referenced (if applicable):

A Special Controls Guidance Document will be promulgated.

L. Test Principle:

The STRATIFY JCV™ Antibody ELISA, consists of two assays performed in 2 steps. The first test detects anti-JCV antibodies and the second test is a confirmation (inhibition) assay to confirm the indeterminate results obtained by the first assay. In the first step (or assay) JCV-VLP are coated onto 96-well microtiter plates. Samples and controls are incubated in the wells to allow the binding of JCV specific antibodies in the samples to the JCV-VLP in the solid phase. After washing away the unbound reactants the plates are incubated with horseradish peroxidase labeled donkey antihuman IgG (H+L). Then, the excess conjugate is washed and the reaction developed using tetramethylbenzidine peroxidase substrate and stopped by the addition of a diluted acid stop solution. The results are read as optical density (OD) values by spectrophotometry. Results are recorded as a normalized value (nOD). Samples with nOD values greater than 0.25 are reported as positive for detectable anti-JCV antibodies, samples with nOD values less than 0.1 are reported as negative for detectable anti-JCV antibodies. Samples with nOD values between 0.25 and 0.1 are reported as indeterminate and requiring evaluation in the confirmation (inhibition) assay.

The second step is intended to identify samples containing antibodies which cross react with denatured capture antigen or with other polyomaviruses such as BK virus (BK V). In the second step, the confirmation (inhibition) assay, samples are preincubated with the JCV-VLP in solution prior to the incubation in the JCV-VLPcoated plates, competing with the plate bound JCV-VLP. JCV specific antibodies will bind to the free VLP. The ELISA is then completed as described for the first step. False positive results are characterized by the minimum competition. The percent inhibition is calculated to confirm the presence of anti-JCV antibodies. Samples with a percent inhibition > 40% are reported as confirmed positive, and samples with percent inhibition ≤ 40% are reported as confirmed negative.

M. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision/Reproducibility:

The precision of the detection assay was characterized in one laboratory. The

{3}------------------------------------------------

study design included testing a panel of serum and plasma samples by three operators, two runs per day over a period of 5 non-consecutive days spread across two weeks. Each sample pool was diluted in three aliquots; each individual aliquot was assayed in duplicate providing one nOD result per aliquot. The STRATIFY JCVTM Antibody ELISA detection assay demonstrates a total % CV ranging from 7.0% to 18.3% for the plasma pools and 9.1% to 12.9% for the serum pools. The serum indeterminate pool returned an indeterminate result 96.7% (87/90) of the time, the plasma indeterminate pool returned an indeterminate result 100% of the time. The confirmation (inhibition) assay was performed in batches by random operators on the indeterminate pools. The indeterminate pool was confirmed as positive in the inhibition assay 100% (86/86) of the time for the serum indeterminate pool and 100% (90/90) for the plasma indeterminate pool.

Precision - Detection Assay (nOD values)

SampleMatrixSampleLevel*QualitativeResult usingnODNMeanBetweenOperatorsBetweenDaysBetweenRunsWithin-Run(repeatability)Total
DINDSD%CVSD%CVSD%CVSD%CVSD%CV
PlasmaHigh Positive90900.9930.0363.6050.0292.9600.0414.0830.0323.1780.0696.97
Indeterminate90900.1500.0032.0450.0117.4020.0064.0100.0063.9530.0149.52
Low Positive873900.3370.03811.290.03710.820.0277.8610.0175.1720.06218.25
Negative684900.0820.0010.7750.00910.790.0066.8420.0033.8110.01113.36
SerumHigh Positive90900.9780.0616.2180.0919.2530.0393.9940.0464.7120.12512.75
Indeterminate873900.1160.0000.0000.0086.5700.0043.5180.0065.1360.0119.05
Low Positive864900.2960.0000.0000.03311.060.0134.3460.0144.8760.03812.85
Negative90900.0580.0000.0000.0046.5860.0034.5130.0034.6160.0059.22
  • D = Detected, I = Indeterminate, ND = Not Detected

{4}------------------------------------------------

Precision - Confirmation Assay

Sample*QualitativeResult using%InhibitionNMin5thPercentile95thPercentileMaxMeanSD%CV
DND
Indeterminate(Plasma)909061.1162.9380.1081.5071.574.976.95
Indeterminate(Serum)86**8644.7147.4065.7168.5455.285.269.52
  • D = Detected, ND = Not Detected

** Three replicates of this sample were negative in the detection assay, per the laboratory procedure the confirmation assay is not performed on negative samples. One dilution was inadvertently omitted by the operator.

  • b. Linearity/assay reportable range:
    N/A

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    N/A

  • d. Detection limit:
    N/A

  • e. Analytical specificity:

Cross Reactivity

Cross reactivity was evaluated in a two part study. The first part of the study evaluated cross reactivity with commercially available human antibodies spiked into JCV negative serum and plasma as determined by the STRATIFY JCVTM Antibody ELISA. The potentially cross reacting antibodies were spiked at high concentrations. The concentrations tested and the test results are presented in the table below. Each antibody tested was not detected with the STRATIFY JCVTM Antibody ELISA. There was no observed reactivity with the three potentially cross-reacting antibodies tested.

Cross Reactivity - Part One - Spiked Antibodies

Cross ReactantConcentrationQuantity Detected
--------------------------------------------------

{5}------------------------------------------------

SerumPlasma
Antibody to Escherichia coli0.4 µg/mL0/30/3
Antibody to Mycobacterium tuberculosis0.4 µg/mL0/30/3
Antibody to Pneumocystis jiroveciunquantified0/30/3

In the second part of the study a panel of up to twenty remnant specimens that previously tested positive for each of the potential cross reacting antibodies was evaluated. Each member of the panel was evaluated using the STRATIFY JCV™ Antibody ELISA along with appropriate controls. The seroprevalence of JCV for each group of potential cross reactants in the panel was compared to the expected seroprevalence of JCV in the normal population. If a group demonstrated a higher than expected seroprevalence of JCV it was considered as an indicator of potential cross reactivity.

Three groups of patients (C. pneumoniae, HSV1 and T. pallidum) exhibited a positivity rate that was slightly above that observed in previously reported studies: one group of patients (M. pneumoniae) exhibited a positivity rate that was lower than the observed rate in previous studies. The results suggest that these groups demonstrate potential cross reactivity.

Sero-positive SamplesAntibodies to JCV DetectedQuantity Tested% Positive
C. pneumoniae152075.0%
C. trachomatis112055.0%
HSV 6112055.0%
Varicellea Zoster Virus122060.0%
Candida132065.0%
HSV 1172085.0%
HSV 272035.0%
Treponema pallidum44100.0%
Mycoplasma pneumoniae1911.1%
CMV101566.7%
HIV 1142070.0%
Overall11518861.2%

Cross Reactivity - Part Two - Sero-prevalence Comparison

Interferences

A two-fold approach was used to evaluate potential interference due to endogenous substances and concomitant medications. Endogenous substances were studied in two studies as described in the following paragraphs. The potential interference caused by concomitant medications was evaluated by reviewing clinical information provided as part of the drug trials. Potential interference due to endogenous substances were evaluated using a sample panel of native sera and plasma samples that contain a low concentration of

{6}------------------------------------------------

JCV antibody which are spiked with the potential interferent as compared to baseline of the same serum and plasma samples that did not contain the interferent. The interferents were spiked at the highest possible endogenous level. For all of the potential interferents with the exception of y globulin; the observed differences in signal did not cause any changes in interpretation of the final result. Plasma spiked with triglyceride, hemogolobin, and bilirubin and sera spiked with hemoglobin and bilirubin demonstrated a >20% shift from baseline, indicating potential interference. Commercially available y globulin is produced using normal human serum containing IgG antibodies: since the seroprevalence of antibodies to JCV virus is approximately 55% in the normal population y globulin is expected to react with this assay.

PotentialInterferent PanelPlasmaSerum
Mean nODStd Dev%ChangefromBaselineMean nODStd Dev%ChangefromBaseline
Baseline (Part 1)0.3850.031N/A0.4400.025N/A
Triglyceride (10mg/mL)0.2950.003-23.3%0.4110.017-6.4%
Hemoglobin (200mg/mL)0.2830.008-26.3%0.3480.013-20.9%
Bilirubin (0.2mg/mL)0.2890.008-24.9%0.3370.005-23.3%
Baseline (Part 2)0.3150.0074N/A0.2960.0139N/A
Albumin (12 g/dL)0.2760.0083-12.60.2670.0172-9.6
Ascorbic Acid (3mg/dL)0.2840.0451-9.90.2730.0199-7.9
Cholesterol (500mg/dL)0.3020.0175-4.40.2890.0120-2.2
y globulin (6 g/dL)1.2060.0136282.41.3080.0616342.1

Interference Summary - Signal Comparison to Baseline

1 The observed decrease in signal may be in part due to a decrease in JCV antibody concentration caused by the additional volume of the spiked interferent

Interference Due To Concomitant Medications

In order to evaluate the potential effect of concomitant medications on the ability to detect JCV antibodies, results were evaluated for serum samples collected serially over a period of 30 months from a subset of 585 MS patients in the AFFIRM study using the STRATIFY JCV™ Antibody ELISA. AFFIRM is a completed clinical trial where patients with multiple sclerosis were randomized to receive placebo or natalizumab, and blood samples were collected every 6 months for 30 months. The AFFIRM study utilized an

{7}------------------------------------------------

ongoing Concomitant Medications Log for each patient to capture all concomitant medications used by the patient during the course of the study. A total of 557 of the 585 patient subset (95%) used concomitant medications during the 30 month AFFIRM study

In order to assess whether the presence of concomitant medications caused interference in the STRATIFY JCV™ Antibody ELISA, each of the concomitant medications used by ≥10% and ≥20% of the patients in AFFIRM study were evaluated separately. The proportion of patients who tested: JCV antibody positive at any time point, JCV antibody positive at all time points, or JCV antibody negative at all time points over the 30 month study period were evaluated in those patients who received or did not receive each individual concomitant medication. If interference occurred at any time point over the 30 month sampling period due to a particular concomitant medication, the proportion of patients who tested JCV antibody positive or negative in these categories would be expected to be different in those who received the medication compared to those patients who did not. The following table illustrates the results of the study.

Concomitant Medications UsedAFFIRM Who Also Provided Longitudinal Samples
Tested for JCV Antibody Status
MEDICATIONTotal Numberof PatientsWho Receiveda ConcomitantMedication(%)*JCVANTIBODYSTATUS*PATIENTSREPORTINGMEDICATION USEPATIENTS NOTREPORTINGMEDICATION USE
SampleSizePercentSampleSizePercent
Paracetamol238 (41%)Positive atANY timepoint15163(57.0to69.6)20960 (54.9to65.4)
Positive atALL timepoints12251 (44.7 to57.8)16748 (42.8 to 53.5)
Negative atALL timepoints8737 (30.4-43.0)13840 (34.6 to 45.1)
Ibuprofen211 (36%)Positive atANY timepoint12559 (52.3 to65.9)23563 (57.7 to 67.7)
Positive atALL timepoints9645 (38.6 to52.5)19352 (46.4 to 56.8)
Negative atALL timepoints8641 (34.1 to47.7)13937 (32.3 to 42.3)
MEDICATIONTotal Numberof PatientsWho Receiveda ConcomitantMedication(%)*JCVANTIBODYSTATUS*PATIENTSREPORTINGMEDICATION USEPATIENTS NOTREPORTINGMEDICATION USE
Methylprednisolone167 (29%)Positive atANY timepoint10764 (56.3 to71.3)25361 (55.7 to 65.2)
Positive atALL timepoints9154 (46.6 to62.2)19847 (42.5 to 52.3)
Negative atALL timepoints6036 (28.7 to43.7)16539 (34.8 to 44.3)
Amoxicillin115 (20%)Positive atANY timepoint7162 (52.2 to70.6)28961 (56.9 to 65.9)
Positive atALL timepoints5346 (36.8 to55.6)23650 (45.6 to 54.8)
Negative atALL timepoints4438 (29.4 to47.8)18139 (34.1 to 43.1)
Acetylsalicylic Acid78 (13%)Positive atANY timepoint5165 (53.8 to75.8)30961 (56.5 to 65.2)
Positive atALL timepoints4558 (46.0 to68.8)24448 (43.7 to 52.6)
Negative atALL timepoints2735 (24.2 to46.2)19839 (34.8 to 43.5)
Multi -Vitamin69 (12%)Positive atANY timepoint3957 (44.0 to68.4)32162 (57.9 to 66.4)
Positive atALL timepoints2739 (27.6 to51.6)26251 (46.4 to 55.2)
Negative atALL timepoints3043 (31.6 to56.0)19538 (33.6 to 42.1)
Amantadine73 (12%)Positive atANY timepoint4663 (50.9 to74.0)31461 (57.0 to 65.6)
MEDICATIONTotal Numberof PatientsWho Receiveda ConcomitantMedication(%)*JCVANTIBODYSTATUS*PATIENTSREPORTINGMEDICATION USEPATIENTS NOTREPORTINGMEDICATION USE
SampleSizePercentSampleSizePercent
Diclofenac66 (11%)Positive atALL timepoints3649 (37.4 to61.3)25349 (45.0 to 53.8)
Negative atALL timepoints2737 (26.0 to49.1)19839 (34.4 to 43.0)
Positive atANY timepoint3452 (38.9 to64.0)32663 (58.5 to 67.0)
Positive atALL timepoints2639 (27.6 to52.2)26351 (46.3 to 55.1)
Negative atALL timepoints3248 (36.0 to61.1)19337 (33.0 to 41.5)

{8}------------------------------------------------

{9}------------------------------------------------

The proportion of patients in each category was consistent between those patients who received the medications and those who did not. The results did not show any potential interfering effect of these commonly used medications on the performance of the assay.

f. Assay cut-off:

To characterize antibody responses against infectious agents in humans, it is critical to have reference sera from both infected and non-infected individuals However, JCV infection is clinically asymptomatic, thereby making it difficult to generally distinguish JCV infected from non-infected individuals. Therefore it is difficult to obtain samples from confirmed infected and noninfected individuals. Even though about 20-30% of JCV infected individuals shed viral DNA in the urine, JCV DNA is often not detected in the blood or urine of infected individuals, even when the infection results in the development of PML. There is no evidence that testing for JCV DNA in blood or urine can identify all JCV infected individuals, however, those who shed virus in the urine are confirmed to be infected with JCV. Therefore, sera from viruric patients were used to establish the positive reference sera for the ELISA. The assay cut point was established from the distribution of the serological responses of samples collected from JCV viruric patients in the JCV antibody ELISA. None of the JC viruric subjects had a reactivity below 0.1, which was therefore selected as the lower cut point in the JCV antibody

{10}------------------------------------------------

ELISA.

The false negative rate of the assay was calculated from the number of subjects who were urinary JCV DNA positive (viruric), but negative for serum JCV antibodies in the STRATIFY JCV™ Antibody ELISA. An initial assessment of the STRATIFY JCV antibody ELISA false negative rate of 2.5% (95%CI: 0.5 to 4.9%) resulted from 5 of 204 viruric patients from the STRATA study whom tested seronegative. The false negative rate of the JCV antibody assay was confirmed using serum and urine samples collected at enrollment from 1073 subjects in the STRATIFY-1 study. Of the 1073 subjects, 186 (17%) were found to be urinary JCV DNA positive. The false negative rate of the assay in STRATIFY-1 was determined to be 2.7% (95%CI: 0.9 to 6.2%) as 5 out of the 186 urinary JCV DNA positive subjects did not have detectable serum JCV antibodies. The seropositivity rate for CD patients was assessed using the same cut-off points and found to be similar to that of MS patients. No matrix differences were found between MS patients, CD patients or healthy volunteers. The false negative rate for CD population was not determined; however, the data supports the applicability of the cut-off and the false negative rate established for MS population to the CD population.

2. Comparison studies:

a. Method comparison with predicate device:

There is no predicate device for this assay. However, the sponsor performed a method comparison study with a laboratory developed test for the detection of anti-JCV antibodies using a panel of 100 samples that cover the detection range of their assay. The purpose of this study was to evaluated the comparative performance with a different anti-JCV antibody detection test particularly in samples close to the cut-off of the STRATIFY Antibody ELISA. The sample panel consisted of 62 negative, 71 low positives near the cut-point and 67 positive samples. Low positives include samples between two cut points (0.1 and 0.25) and a small group of samples just above the upper cut point (the range is from nOD ≤ 0.1 to ≤0.35). The Positive percent agreement (PPA) was 91.0% (122/134) with 95% CI: 85.0% to 94.8%, the Negative percent agreement (NPA) was 89.4% (59/66) with 95% CI: 79.7% to 94.8%. The results of the analytical comparison are shown below.

Comparative Assay
DetectedNot Detected
Sample Category based on initial nODSample Category based on initial nOD
Positive (nOD>0.35)Low Positive* (0.1 ≤ nOD ≤ 0.35)Negative ( nOD < 0.1)Positive (nOD> 0.35)Low Positive* (0.1 ≤ nOD ≤ 0.35)Negative ( nOD < 0.1)

{11}------------------------------------------------

STRATIFYJCVAntibodyELISAtestPositive
6755007**0
Negative02**1007**52
e
67571001452
Total13466
200
* 63 out of 71 samples fell in screening "Indeterminate" zone ((0.1 < nOD < 0.25) which were followed

| 人 ((0.1 by confirmation (%Inhibition) in Focus assay.

**2+7+7=16 samples were part of 63 samples which required confirmation assay in Focus assay.

b. Matrix comparison:

A total of fifty paired plasma and serum samples were assayed to evaluate the equivalence between both sample types. The comparison of the qualitative results from the serum and plasma samples is presented in the table below. A graphical representation of the scatter plot of the data and the Passing-Bablok regression analysis of nOD values are shown in the following figure. The slope was 0.99 with 95% CI: 0.9695 to 1.0103 and an intercept was 0.0 with 95% CI: -0.0034 to 0.0043. The mean % difference for nOD values obtained for serum and plasma samples was 4.9% . Additionally, 50 out of 50 (100%) qualitative results obtained for plasma samples matched the qualitative results obtained for matched serum samples. The results demonstrate the equivalency of both matrices when used on the STRATIFY JCV™ Antibody ELISA.

Result based on Detection Assay nOD
Serum
PlasmaNot DetectedIndeterminateDetectedTotal
Not Detected15116
Indeterminate1616
Detected1818
Total15171850

Sample Matrix Comparison - Qualitative Results

{12}------------------------------------------------

Image /page/12/Figure/0 description: The image is a scatter plot titled "STRATIFY JCVTM Antibody ELISA Assay Matrix Comparison Study Passing-Bablok Regression Fit with Confidence Bound". The x-axis is labeled "Serum nOD", and the y-axis is labeled "Plasma nOD". The plot shows a linear relationship between Serum nOD and Plasma nOD, with a regression line of P-B Reg Line = 0 + 0.99 * X. The 95% confidence interval for the intercept is (-0.0034,0.0043), and the 95% confidence interval for the slope is (0.9695,1.0103), with n=50.

    1. Clinical studies:
    • a. Clinical Sensitivity:

N/A

  • b. Clinical specificity:
    N/A

  • Other clinical supportive data (when a. and b. are not applicable): C.
    Because PML is an infrequent event in natalizumab-treated patients, data collected from both clinical trial and post-marketing reports of confirmed cases of PML were used to assess the clinical performance of the STRATIFY JCV™ Antibody ELISA test for PML risk stratification. A clinical plan was developed for collection of serum samples obtained from natalizumab-treated patients prior to the onset of PML for JCV antibody testing. As of September 2011, 37 available serum samples from confirmed PML patients collected at least 6 months prior to clinical diagnosis of PML were tested for JCV antibody status using the STRATIFY JCV™ Antibody ELISA. A total of 36 of these patients had received at least 18 infusions of natalizumab at the time of PML

{13}------------------------------------------------

diagnosis, while one patient had received 17 infusions of natalizumab. All 37 pre-PML samples tested positive for JCV antibodies with an estimated sensitivity of 100% (37/37) with 95% CI: 90.6% to 100%. In addition, there has only been one natalizumab treated CD patient with an available pre-PML sample who has developed PML. This sample also tested positive for JCV antibodies.

Samples from 5,896 MS patients were tested by the STRATIFY JCV Antibody ELISA. A total of 3,239 patients tested positive for antibodies to JCV using the test, and the positivity rate was estimated to be 54.9% (3.239/5.896) with 95% Cl: 53.7% to 56.2%. An assay is statistically informative if the percentage of positive results in patients with the disease of interest is higher than the percentage of positive results in the population at risk. The 100% JCV antibody positivity demonstrated in the 37 natalizumab-treated PML patients prior to PML diagnosis was significantly different than the 54.9% JCV antibody positivity in the MS population, and represents a 1.82-fold (95% CI: 1.65 to 1.86) increased risk of PML compared to the PML incidence in the overall natalizumabtreated population.

Estimated incidence of PML by STRATIFY JCV Antibody ELISA test results based on statistical modeling with an assumption that there is one hypothetical PML case with a negative test result (38 PML cases: 37 positive and 1 hypothetical negative) and assumption that study has 13,227 patients and all they were tested by STRATIFY JCV Antibody ELISA test.

Number withPMLNumberwithout PMLTotal numberpatients treated
STRATIFYJCV AntibodyELISA testPositive377,2297,266
STRATIFYJCV AntibodyELISA testNegative1*5,9605,961
Total3813,18913,227
Risk of PML (per 1,000) treated with≥ 18 months for Positive result5.0995% CI: 3.70 to 7.01
Risk of PML (per 1,000) treated with≥ 18 months for Negative result0.1795% CI: 0.03 to 0.95
Relative risk30.495% CI: 5.3 to 437.4

*For the negative result, a hypothetical case was assumed in order to allow for calculation of the point estimate.

Data from three studies performed in MS patients were analyzed to determine whether use of prior immunosuppressant (IS) therapy affects JCV antibody status. The data showed that the percentage of positive results obtained by the STRATIFY JCV Antibody ELISA test was not dependent on prior IS therapy. Data from the STRATIFY-1 study

{14}------------------------------------------------

was analyzed to determine whether duration of natalizumab treatment affects JCV antibody status. The data showed that the percentage of positive results obtained by the STRATIFY JCV Antibody ELISA test analyzed by dosing interval of natalizumab was not dependent on treatment duration.

The STRATIFY JCV antibody ELISA testing service results can therefore be used along with the other established PML risk factors of prior immunosuppressant use and natalizumab treatment duration, to stratify an individual patient's risk of PML. JCV antibody positive status results in a 1.82-fold increased risk of PML. The risk of PML based on prior IS use and treatment duration is therefore multiplied by 1.82 to estimate the PML risk based on the three risk factors combined

Risk of PML in JCV Antibody Positive Patients Based on Prior IS Use and Treatment Duration (per 1,000)

NatalizumabExposure†STRATIFY JCV Antibody ELISA testPositive Result*
No Prior ImmunosuppressantUsePrior Immunosuppressant Use
1-24 months<1/1,0002/1,000
25-48 months4/1,00011/1,000

Notes: Based on postmarketing PML data and natalizumab use data as of September 1, 2011. "Data beyond 4 years of treatment are limited.

*Risk in anti-JCV antibody positive patients was estimated based on the assumptions that 18% of Natalizumab-treated MS patients have a history of prior immunosuppressant treatment and that 55% of natalizumab-treated MS patients are anti-JCV antibody positive.

    1. Clinical cut-off:
      N/A

5. Expected values/Reference range:

The STRATIFY JCV™ Antibody ELISA has been used to evaluate JCV antibody positivity rate in serum and plasma samples from a geographically diverse cohort of 5,896 MS patients. The sample cohort was comprised from MS patients from clinical trials including a completed Phase 3 clinical study of natalizumab in MS (AFFIRM C-1801), an ongoing study to evaluate seroprevalence in the MS population (STRATIFY-1 [101JC401]), an observational study in natalizumab treated MS patients (TYGRIS-US [101MS402]) and a national MS registry from Sweden. The clinical characteristics for the MS patients within each study are shown in Table 2. The age and gender distribution of the MS cohort tested with

{15}------------------------------------------------

the STRATIFY JCVTM Antibody ELISA are similar to the age and gender distribution of MS patients treated with natalizumab in the post-marketing setting. JCV antibody positivity rate in the MS cohort was 47-59% which is consistent with what has been reported in the literature. JCV antibody prevalence was shown to increase with age and was lower in women compared to men which is also consistent with what has been reported in the literature in healthy adults using similar assay methodologies

AFFIRM(N=823)TYGRIS-US(N=1480)STRATIFY-1(N=1096)Swedish MSRegistry(N=2497)
Age (years)
● Range18-5018-7512-7512-75
● Mean35.944.344.437.5
● Median36444537
Gender (%)*
● Male252/823(30.6%)350/1451(24.1%)266/1096(24.3%)700/2494(28.1%)
● Female571/823(69.4%)1101/1451(75.9%)830/1096(75.7%)1794/2494(71.9%)
GeographyNorth Americaand EU/Rest ofWorldUS and CanadaUSSweden
JCV Antibody Positivity Rate(95% CI)
● % JCV AntibodyPositive449/823(54.6%)(51.1 to 58.0)704/1480(47.6%)(45.0 to 50.1)614/1096(56.0%)(53.0 to 59.0)1472/2497(59.0%)(57.0 to 60.9)
● % JCV AntibodyNegative374/823(45.4%)(42.0 to 48.9)776/1480(52.4%)(49.9 to 55.0)482/1096(44.0%)(41.0 to 47.0)1025/2497(41.0%)(39.1 to 43.0)

Demographic Data and JCV Antibody Prevalence for MS Patients

  • In the TYGRIS-US dataset, 1451 of the 1480 patients had age and gender information available. In the Swedish MS dataset, 2464 of 2497 patients had age information available and 2494 of 2497 patients had gender information available.

The STRATIFY JCVTM Antibody ELISA has also been used to evaluate JCV antibody prevalence in serum samples from a geographically diverse cohort in CD patients. The sample cohort was comprised of patients from two completed Phase

{16}------------------------------------------------

3 clinical trials of natalizumab in CD (CD301 and CD303). The clinical characteristics for the CD patients within these studies are shown in Table 3. The age and gender distribution of the CD cohort tested with the STRATIFY JCVTM Antibody ELISA are similar to the age and gender distribution of CD patients treated with natalizumab in the post-marketing setting. JCV antibody prevalence in the CD cohort was 55.6% and was shown to increase with age, consistent with what has been reported in the literature 9, 12. In general there was no difference in the prevalence of JCV antibodies between genders.

Demographic and JCV Antibody Positivity Rate for CD Patients

Clinical Studies CD301 andCD303(N=313)
Age (years)
• Range18-74
• Mean38.4
• Median37
Gender (%)
• Male123/313 (39.3%)
• Female190/313 (60.7%)
GeographyNorth America and EU/Rest ofWorld
JCV Antibody Positivity Rate(95% CI)
• % JCV Antibody Positive174/313 (55.6%)(49.9 to 61.2)
• % JCV Antibody Negative139/313 (44.4%)(38.8 to 50.1)

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The petition for Evaluation of Automatic Class III Designation for this device is accepted. The device is classified as Class II under regulation 21 CFR 866.3336 with special controls. The special control guidance document "Class II Special Controls Guidance Document: John Cunningham Virus Serological Reagents," will be available shortly.

§ 866.3336 John Cunningham Virus serological reagents.

(a)
Identification. John Cunningham Virus serological reagents are devices that consist of antigens and antisera used in serological assays to identify antibodies to John Cunningham Virus in serum and plasma. The identification aids in the risk stratification for the development of progressive multifocal leukoencephalopathy in multiple sclerosis and Crohn's disease patients undergoing natalizumab therapy. These devices are for adjunctive use, in the context of other clinical risk factors for the development of progressive multifocal leukoencephalopathy.(b)
Classification. Class II (special controls). The special control for this device is the FDA guideline document entitled “Class II Special Controls Guideline: John Cunningham Virus Serological Reagents.” For availability of the guideline document, see § 866.1(e).