K Number
DEN090004
Device Name
OVA1 TEST
Manufacturer
Date Cleared
2009-09-11

(51 days)

Product Code
Regulation Number
866.6050
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OVA1™ Test is a qualitative serum test that combines the results of five immunoassays into a single numerical score. It is indicated for women who meet the following criteria: over age 18; ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. The OVA 1 Test is an aid to further assess the likelihood that malignancy is present when the physician's independent clinical and radiological evaluation does not indicate malignancy. The test is not intended as a screening or stand-alone diagnostic assay.

PRECAUTION: The OVA1™ Test should not be used without an independent clinical/radiological evaluation and is not intended to be a screening test or to determine whether a patient should proceed to surgery. Incorrect use of the OVA1™ Test carries the risk of unnecessary testing, surgery, and/or delayed diagnosis.

Device Description

The OVA1™ Test uses OvaCalc Software to incorporate the values for 5 analytes from separately run immunoassays (described below) into a single numerical score between 0.0 and 10.0.

The cleared test system consists of the software, instruments, assays and reagents used to obtain the OVA1™ Test result. The immunoassays and reagents are sold separately from the OvaCalc Software. Users are instructed to use only those lots identified by Vermillion. The immunoassays are performed according to the manufacturers' directions detailed in each product insert. The analytes and corresponding tests and calibrators used in the OVA1™ Test are:

AnalyteDevice (Assay and Calibrator)Instrument
CA 125Elecsys CA 125 IICA125 II CalSetRoche Elecsys 2010
PrealbuminN Antisera to Human Prealbumin andRetinal-binding ProteinN Protein Standard SL (human)Siemens BN II
ApolipoproteinA-1N-Antisera to Human Apolipoprotein A-1and Apolipoprotein BN Apolipoprotein Standard Serum (human)Siemens BN II
β2-microglobulinHuman Beta-2 Microglobulin LatexEnhanced Nephelometric Kit (Binding Site)Siemens BN II
TransferrinN Antisera to Human Transferrin andHaptoglobinN Protein Standard SL (human)Siemens BN II

The user enters results of the five analytes manually into an Excel spreadsheet together with the headers needed by OvaCalc Software. There is no physical or electronic connection between the immunoassay devices and the OvaCalc Software. Using an algorithm and the values of these 5 analytes, the OvaCalc Software generates a single unit-less numerical score from 0.0 to 10.0.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study detailed in the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of target performance metrics (e.g., minimum sensitivity or specificity values). Instead, it presents the performance characteristics observed in the clinical validation study. The closest thing to acceptance criteria for the clinical performance seems to be the demonstration that the "True Positive Rate (TPR) exceeded the False Positive Rate (FPR)" with statistical significance. The other criteria relate to analytical performance, such as precision and stability, which are met through demonstrated performance.

Criterion TypeAcceptance Criteria (Implicit/Explicit)Reported Device Performance
Analytical Performance
Precision (Total %CV)Acceptable within-run, between-run, between-day, between-operator, and between-site %CV. (Explicitly stated < 8.6% for lot-to-lot and < 8.9% for reproducibility).Overall within-run %CV: ≤ 7.3%. Between-run %CV: ≤ 2.4%. Between-day %CV: ≤ 3.9%. Between-operator %CV: ≤ 5.3%. Between-site %CV: < 4.4%. Total Reproducibility %CV: < 8.9%. Lot-to-Lot Imprecision %CV: < 8.6%.
LinearityClaimed in package inserts for individual analytes; measurement intervals used in OVA1™ Test within demonstrated linear range.Demonstrated for measurement intervals corresponding to those used in the OVA1™ Test.
TraceabilityCalibrators and controls traceable to recognized standards (WHO 1st International Preparation for β2M; protein reference preparation CRM 470 for transferrin and prealbumin).β2M calibrators/controls traceable to WHO 1st International Preparation. Transferrin and prealbumin calibrators/controls traceable to protein reference preparation CRM 470.
Stability (Reagents)Reagent stability acceptable with specified storage conditions and timeframes.Open Vial Stability: Apo A-1, Prealbumin, Transferrin (Siemens BNII): 4 weeks at 2-8°C. CA 125 II (Roche Elecsys® 2010): 6 weeks. β2M reagents (reconstituted): up to 1 week. Calibrators and controls: up to 4 weeks.
Stability (Specimen)Specimen stability acceptable under specified storage conditions and timeframes.Fresh serum: +2 to +8°C up to 8 days. Frozen serum (within 24 hr) at -20°C: up to 9 weeks. Frozen serum (within 24 hours) at -65 to -85°C: up to 12 weeks.
Limit of Detection (LoD)LoD/LoQ from individual assay package inserts confirmed and incorporated into algorithm.Confirmed and incorporated; results outside measuring interval do not yield OVA1™ Test score.
Analytical Specificity (Interference)< 10% difference between sample with interferent and control, with the exception of rheumatoid factor at high concentrations.No significant interference observed for hemoglobin, bilirubin, triglycerides. Rheumatoid factor > 250 RU/mL caused significant interference; specimens with RF > 250 RU/mL are not appropriate for the test.
Clinical Performance
Statistical InformativenessTrue Positive Rate (TPR) must exceed False Positive Rate (FPR) with statistical significance for combined data, pre-menopausal subjects, and post-menopausal subjects.All combined data: TPR (87.5%) > FPR (49.2%); difference of 38.3% (95% CI: 26.5% to 47.8%) statistically significant. Pre-menopausal: TPR (80.8%) > FPR (43.2%); difference of 37.6% (95% CI: 16.7% to 52.2%) statistically significant. Post-menopausal: TPR (91.3%) > FPR (58.2%); difference of 33.1% (95% CI: 17.3% to 46.1%) statistically significant.
Adjunctive Information Value (Dual Assessment for Non-GO)Dual assessment (Physician's pre-surgical assessment + OVA1™ Test) should provide additional information compared to physician's assessment alone, specifically by increasing sensitivity for malignancy and maintaining (or improving) NPV. The benefit of detecting additional true positive cases should outweigh the additional false positives for the intended use population.Sensitivity: Increased from 72.2% (single assessment) to 91.7% (dual assessment). Specificity: Decreased from 82.7% to 41.6%. PPV: Decreased from 60.5% to 36.5%. NPV: Increased from 89.1% to 93.2%. (95% CI for the 4.1% increase in NPV was -0.5% to 8.7%, borderline statistical significance). Conclusion notes "sufficient benefit".
Adjunctive Information Value (Dual Assessment for GO)Corroborative results to non-GO analysis regarding additional information provided by dual assessment.Sensitivity: Increased from 77.5% (single assessment) to 98.9% (dual assessment). Specificity: Decreased from 74.7% to 25.9%. PPV: Decreased from 63.3% to 42.9%. NPV: Increased from 85.5% to 97.6% (95% CI for the 12.1% increase in NPV was 5.7% to 18.6%, statistically significant). Conclusion notes "corroborative, but not dispositive" for intended use.

2. Sample Sizes and Data Provenance

Test Set (Clinical Validation Study):

  • Total Enrolled: 743 patients.
  • Training Set (from enrollment): 146 subjects were set aside for training, with 21 not evaluable, leaving 125 for training.
  • Final Evaluable Test Set: 516 subjects/samples (after excluding training set and those with missing info/lack of sample).
    • Non-GO Physician Evaluated Subset: 269 patients.
    • GO Physician Evaluated Subset: 247 patients.
  • Data Provenance: Prospective, multicenter, double-blind clinical study. Samples collected from 27 demographically mixed subject enrollment sites in the US (implied by typical FDA submission context and "demographically mixed" implies US diversity).

Training Set:

  • Training Set 1: 284 pre-operative serum samples from the University of Kentucky.
    • Complete laboratory data for 274 samples (109 malignant, 175 benign).
  • Training Set 2: A randomly selected subset of 146 pre-operative serum samples collected under a clinical trial specimen repository.
    • 21 not evaluable, leaving 125 samples (89 benign, 10 LMPs, 19 EOCs, 1 primary, 3 non-primary ovarian cancers, 3 other malignancies).

3. Number of Experts and Qualifications for Test Set Ground Truth

The document does not specify a "number of experts" used to establish ground truth in the sense of independent review of imaging or clinical data. Instead, the ground truth for malignancy status in the clinical validation study was established through histopathology results from tissue samples obtained during surgical intervention.

The clinical assessments made by physicians (non-GO and GO) were used for comparative purposes against the device and as part of the "dual assessment" scenario, but these were not explicitly designated as "expert ground truth" for the device's diagnostic accuracy. The ultimate ground truth for classification of benign vs. malignant was pathology.

4. Adjudication Method for the Test Set

The document does not describe an "adjudication method" in the typical sense of multiple expert readers reviewing cases and resolving disagreements.

  • The OVA1™ Test results were generated by an algorithm from five immunoassay values.
  • The clinical pre-surgical assessments were made by individual physicians (non-GO or GO).
  • The ground truth outcome was histopathology.

The comparison was between these individual components or combinations, not between adjudicated expert readings.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study, as typically understood for imaging devices where human readers interpret cases with and without AI assistance, was not performed.

This study compared:

  1. The OVA1™ Test alone.
  2. The physician's (non-GO or GO) pre-surgical assessment alone.
  3. A "Dual Assessment" (physician's assessment OR OVA1™ Test positive).

The "effect size of how much human readers improve with AI vs without AI assistance" is not reported in the traditional MRMC sense. Instead, the document quantifies the change in clinical performance metrics (Sensitivity, Specificity, PPV, NPV) when the OVA1™ Test is combined with the physician's pre-surgical assessment, rather than directly measuring physician improvement while using AI.

For non-GO physicians, when using dual assessment:

  • Sensitivity for malignancy increased from 72.2% (single assessment) to 91.7% (dual assessment).
  • NPV increased from 89.1% to 93.2%.

For GO physicians, when using dual assessment:

  • Sensitivity for malignancy increased from 77.5% (single assessment) to 98.9% (dual assessment).
  • NPV increased from 85.5% to 97.6%.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance assessment (algorithm only without human-in-the-loop performance) was done for the OVA1™ Test.

The "Performance Characteristics of the OVA1™ Test Alone" section directly presents its sensitivity, specificity, NPV, and PPV compared to histopathology for patients evaluated by non-GO physicians (and similarly for GO physicians, though these were deemed less relevant for the intended use population).

Standalone Performance (Non-GO Physician Population):

  • Sensitivity: 87.5% (63/72)
  • Specificity: 50.8% (100/197)
  • NPV: 91.7% (100/109)
  • PPV: 39.4% (63/160)

7. Type of Ground Truth Used (for Clinical Studies)

The primary ground truth used for establishing clinical performance was histopathology results from tissue samples obtained during surgical intervention. Malignancy status was determined based on these reports.

8. Sample Size for the Training Set

The algorithm was derived using two independent training datasets:

  • Training Set 1: 284 pre-operative serum samples (274 evaluable: 109 malignant, 175 benign).
  • Training Set 2: 146 pre-operative serum samples (125 evaluable: 89 benign, 10 LMPs, 19 EOCs, 1 primary and 3 non-primary ovarian cancers, 3 other malignancies).

9. How the Ground Truth for the Training Set Was Established

The document states that the training sets consisted of "pre-operative serum samples" which were classified into categories like "benign diseases," "ovarian tumors of low malignant potential (LMP)," "epithelial ovarian cancers," etc. This classification would universally be based on histopathology obtained from the surgical specimens after mass removal.

{0}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

  • A. 510(k) Number: K081754
  • B. Purpose for Submission: New Device
  • C. Measurand: Score based on 5 serum analytes
  • D. Type of Test: Software algorithm and 5 immunoassays
  • E. Applicant: Vermillion, Inc.
  • F. Proprietary and Established Names: OVA1™ Test

G. Regulatory Information:

    1. Regulation section:
    • 21 CFR 866.6050 Ovarian adnexal mass assessment score test system
    1. Classification: Class II
    1. Product code: ONX Serum, algorithm, ovarian cancer assessment test
    1. Panel:

Immunology (82)

H. Intended Use: 1. Intended use:

The OVA1™ Test is a qualitative serum test that combines the results of five immunoassays into a single numerical score. It is indicated for women who meet the following criteria: over age 18; ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. The OVA 1 Test is an aid to further assess the likelihood that malignancy is present when the physician's independent clinical and radiological evaluation does not indicate malignancy. The test is not intended as a screening or stand-alone diagnostic assay.

PRECAUTION: The OVA1™ Test should not be used without an independent clinical/radiological evaluation and is not intended to be a screening test or to determine whether a patient should proceed to surgery. Incorrect use of the OVA1™ Test carries the risk of unnecessary testing, surgery, and/or delayed diagnosis.

    1. Indications for Use: Same as Intended Use.
    1. Special conditions for use statement(s): Prescription Use only.
    1. Special instrument requirements: The Siemens BNTM II System for the measurement of Prealbumin (also known as

{1}------------------------------------------------

Transythretin; TT), Apolipoprotein A-1 (Apo A-1), B2-microglobulin (B2M), and Transferrin (Tfr). The Roche Elecsys® 2010 for the measurement of CA 125. Both instrument systems are FDA cleared.

I. Device Description:

The OVA1™ Test uses OvaCalc Software to incorporate the values for 5 analytes from separately run immunoassays (described below) into a single numerical score between 0.0 and 10.0.

The cleared test system consists of the software, instruments, assays and reagents used to obtain the OVA1™ Test result. The immunoassays and reagents are sold separately from the OvaCalc Software. Users are instructed to use only those lots identified by Vermillion. The immunoassays are performed according to the manufacturers' directions detailed in each product insert. The analytes and corresponding tests and calibrators used in the OVA1™ Test are:

AnalyteDevice (Assay and Calibrator)Instrument
CA 125Elecsys CA 125 IICA125 II CalSetRoche Elecsys 2010
PrealbuminN Antisera to Human Prealbumin andRetinal-binding ProteinN Protein Standard SL (human)Siemens BN II
ApolipoproteinA-1N-Antisera to Human Apolipoprotein A-1and Apolipoprotein BN Apolipoprotein Standard Serum (human)Siemens BN II
β2-microglobulinHuman Beta-2 Microglobulin LatexEnhanced Nephelometric Kit (Binding Site)Siemens BN II
TransferrinN Antisera to Human Transferrin andHaptoglobinN Protein Standard SL (human)Siemens BN II

J. Substantial Equivalence Information:

    1. Predicate device name(s): Not applicable
    1. Predicate K number(s): Not applicable
    1. Comparison with predicate: Not applicable

K. Standard/Guidance Document Referenced (if applicable):

ISO 14971:2007 Medical Devices-Application of Risk Management to Medical Devices, International Organization for Standardization, 200 Edition. CLSI guideline EP5-A2 "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline."

L. Test Principle:

The individual assays for prealbumin, apolipoprotein A1 and transferrin each contain a biomarker specific rabbit polyclonal antibody which forms an immune complex with the target when reacted with a serum specimen. The immune complexes are

{2}------------------------------------------------

proportional to the concentration of biomarker in the serum specimen for each specific assay. The assay for ß2-microglobulin consists of polystyrene particles coated with a monospecific antiserum to ß2-microglobulin which aggregate when mixed with serum specimen ß2-microglobulin. These aggregates are proportional to the concentration of B2-microglobulin in the serum specimen. The Siemens BNTM II System is an automated immunonephelometer.

The CA 125 II assay uses 2 mouse monoclonal antibodies to CA 125. The quantity of CA 125 present is then measured by chemiluminescence emission. The Roche Elecsys® 2010 is an automated analyzer with electrochemiluminescence detection. The amount of analyte in each assay is determined against the calibration curve. Each assay uses its own specific calibrator and controls.

The user enters results of the five analytes manually into an Excel spreadsheet together with the headers needed by OvaCalc Software. There is no physical or electronic connection between the immunoassay devices and the OvaCalc Software. Using an algorithm and the values of these 5 analytes, the OvaCalc Software generates a single unit-less numerical score from 0.0 to 10.0.

M. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision: Precision performance of the OVA1™ Test was evaluated in accordance with CLSI guideline EP5-A2 "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline." Five serum specimens spanning the OVA1™ Test score range (range of numerical results for OVA1™ Test: 0.0 to 10.0) were tested over 20 days, two runs per day, and two replicates per run. There were no unevaluable results. Total percent coefficient of variation (%CV) ranged from 1.0 to 7.4%.
SamplenOVA1(Mean)Within-runBetween-runTotal
SD%CVSD%CVSD%CV
1802.740.0652.40.0110.40.0913.3
2803.390.1013.00.0992.90.1594.7
3803.740.1494.00.0972.60.1925.1
4804.690.2906.20.0000.00.3497.4
5809.940.0610.60.0510.50.0981.0

Lot-to-Lot precision: Five serum specimens and a minimum of two control sera from the same lot that vield two different OVA1™ Test scores (e.g., high and low, or low and near cutoff score) were analyzed at one site with three different reagent kit lots and calibrators over three different days by one operator. Reagents and calibrators were analyzed as shown:

{3}------------------------------------------------

DayCalibratorReagent Lots
11ABC
22BCB
33CAA

Note: ß2M calibrators and reagents are kit specific and are not interchangeable between kits. Therefore the three lots were analyzed on 3 separate days, and calibrators and reagents were not mixed between the lots (as shown) for this assay. For each of the five specimens and the two controls, the mean and standard deviation (SD) for the between-lot component of variance for each sample were calculated. The %CV for imprecision was < 8.6% for any of the 5 samples.

Reproducibility: Five serum specimens spanning the OVA1™ Test score range and a minimum of two controls per day were tested in duplicate, two runs each day, over 6 days, by two operators at each of three sites. Each operator performed the test on three nonconsecutive days. i.e., operator 1 ran the test on days 1, 3, and 5; operator 2 ran the test on days 2, 4, and 6. At each site, the test was run with the same lots of calibrators, kit reagents, and controls for the duration of the study. Each operator performed the complete analysis of the OVA1™ Test for the day on the Siemens BN™ II System and on the Roche Elecsys® 2010 and imported the five biomarkers into Vermillion's OvaCalc Software for generation of the OVA1™ Test score. For each of the five specimens, the mean, SD, %CV, median, and range of OVA1™ Test scores were identified for an assessment of components of imprecision in reproducibility study. For OVA1™ Test scores, the %CV for within-run (repeatability) was ≤ 7.3, the %CV for between-run ≤2.4%, %CV between-day was ≤3.9%, the %CV between-operator ≤5.3% and the %CV between-site <4.4% . The total imprecision (reproducibility) was <8.9%. The results for the OVA1™ Test are shown below:

ParametersSpecimen
12345
OVA1 result
Mean2.673.213.755.009.71
Repeatability (within run)SD0.0690.0940.1570.3640.157
%CV2.62.94.27.31.6
Between runSD0.0340.0870.0910.0000.129
%CV1.32.72.40.01.3
Between daySD0.0000.0000.1460.0320.045
%CV0.00.03.90.60.5
Between operatorSD0.0420.0390.1050.2650.000
ParametersSpecimen
12345
%CV1.61.22.85.30.0
Between sitesSD0.0570.1410.0000.1030.212
%CV2.14.40.02.12.2
Reproducibility (total)SD0.0980.1760.2500.4470.271
%CV3.75.56.78.92.8
Individual Protein Biomarkers
Apolipoprotein A1, mg/dL
Mean148.6156.5174.5192.8155.4
Repeatability (within run)SD4.4894.7023.4385.8404.091
%CV3.03.02.03.02.6
Between runSD2.0580.0004.2210.0003.456
%CV1.40.02.40.02.2
Between daySD3.1053.0753.2274.9691.622
%CV2.12.01.82.61.0
Between operatorSD3.4463.7444.2643.7683.569
%CV2.32.42.42.02.3
Between sitesSD6.5356.3386.7335.5706.408
%CV4.44.13.92.94.1
Reproducibility (total)SD8.5138.3779.2409.5258.340
%CV5.75.45.34.95.4
Beta-2 microglobulin, mg/L
Mean1.661.641.791.942.11
Repeatability (within run)SD0.0460.0390.0450.0400.053
%CV2.72.42.52.12.5
Between runSD0.0550.0280.0360.0250.011
%CV3.31.72.01.30.5
Between daySD0.0000.0150.0380.0250.034
%CV0.00.92.11.31.6
Between operatorSD0.0300.0170.0000.0140.038
%CV1.81.10.00.71.8
Between sitesSD0.0290.0240.0270.0260.018
%CV1.71.41.51.30.9
Reproducibility (total)SD0.0800.0560.0720.0590.074
%CV4.83.54.03.03.5
Specimen
Parameters12345
CA125 II, U/mL
Mean9.0214.0417.0220.92352.1
Repeatability (within run)SD0.3540.2100.8390.5255.131
%CV3.91.54.92.51.5
Between runSD0.1760.5900.6791.05420.53
%CV2.04.24.05.05.8
Between daySD0.1400.1760.3860.0005.054
%CV1.61.32.30.01.4
Between operatorSD0.4530.2940.0000.0003.306
%CV5.02.10.00.00.9
Between sitesSD0.4760.3800.1380.2365.182
%CV5.32.70.81.11.5
Reproducibility (total)SD0.7080.7661.1461.18722.22
%CV7.95.56.75.76.3
Transferrin, mg/dL
Mean270.9263.3290.2308.4280.4
Repeatability (within run)SD12.027.98010.738.62711.02
%CV4.43.03.72.83.9
Between runSD7.51410.7710.7310.806.947
%CV2.84.13.73.52.5
Between daySD9.5091.7252.0581.3350.000
%CV3.50.70.70.40.0
Between operatorSD0.0000.9924.7994.5871.581
%CV0.00.41.71.50.6
Between sitesSD16.7819.9920.4620.0618.23
%CV6.27.67.16.56.5
Reproducibility (total)SD21.8421.2623.1321.9319.89
%CV8.18.18.07.17.1
Prealbumin (transthyretin), mg/dL
Mean21.4825.0429.3334.7824.78
Repeatability (within run)SD0.6190.9360.9511.6770.694
%CV2.93.73.24.82.8
Between runSD0.5340.4710.9380.5870.417
%CV2.51.93.21.71.7
Between daySD0.3020.0000.1610.0000.556
Specimen
Parameters12345
Between operator%CV1.40.00.50.02.2
Between operatorSD0.5740.4320.5100.9790.848
Between sites%CV2.71.71.72.83.4
Between sitesSD0.5380.6500.1240.0000.164
Reproducibility (total)%CV2.52.60.40.00.7
Reproducibility (total)SD1.1071.2401.4231.9901.255
Reproducibility (total)%CV5.25.04.95.75.1

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

  • b. Linearity/assay reportable range: For each analyte, measurement linearity (as claimed in the package inserts for the individual analytes) was demonstrated for measurement intervals corresponding to those used in the OVA1™ Test.
  • Traceability, Stability (controls, calibrators, or methods): C. Each assay uses its own calibrator and controls. The calibrator and control for ß2-microglobulin are traceable to WHO 1st International Preparation. For transferrin and prealbumin, the calibrators and controls are traceable to protein reference preparation CRM 470.

Stability - Closed Vial: For each assay, users are instructed to refer to the individual stability information in the package insert.

Stability - Open Vial: Studies were performed to investigate the stability of the reagents used in the OVA1™ Test and impact on the OVA1™ Test score when they are opened for use and stored at 2-8°C. Three (3) serum specimens spanning the OVA1™ Test score range (one low, one high and one at the cutoff) were tested in duplicate after opening reagents and storing at various time points. The data demonstrate open vial reagent stability is acceptable when stored after opening as follows: Four (4) weeks at 2-8°C for Apo A-1, Prealbumin, Transferrin, on the Siemens BNII, and six (6) weeks for CA 125 II on the Roche Elecsys® 2010. Stability of the ß2M reagents once reconstituted is up to 1 week at 2-8°C. Calibrators and controls are stable up to 4 weeks at 2-8°C.

Specimen Stability: Serum samples from 10 healthy pre- and post-menopausal women were used to evaluate stability (consistency) of the OVA1™ Test scores. The mean, SD, median and range were used to describe the OVA1TM Test score and each of the 5 analyte values for each patient sample stored under each condition, at each time point in duplicate when compared to time point 0. Storage conditions consisted of room temperature (15-25°C) up to 48 hours, refrigerated (2-8°C) up to 9 days, and frozen (-15 to -35°C and -60 to -85°C) up to 13 weeks. The summary table below shows the mean change

{7}------------------------------------------------

from the initial result, the standard error (SE) and the 95% CI for the change, for the latest time interval. The results support the following specimen stability claims: fresh serum stored at +2 to +8°C can be used up to 8 days. Serum frozen within 24 hr at -20°C is stable up to 9 weeks. Serum frozen within 24 hours and stored at -65 to -85° is stable up to 12 weeks.

N=2215 to 25°Cat 48 hours2 to 8°Cat 9 days-15 to -35°Cat 13 weeks-60 to -85°Cat 13 weeks
OVA1™ Test
Mean Change0.030.050.01-0.10
SE0.0380.0390.0560.037
95% CI-0.04 to 0.10-0.03 to 0.13-0.10 to 0.12-0.17 to -0.03
Apo A-1 (mg/dL)
Mean Change-0.051.051.183.82
SE1.1241.1581.9751.266
95% CI-2.25 to 2.15-1.22 to 3.32-2.69 to 5.051.34 to 6.30
β2-microglobulin (mg/L)
Mean Change-04-0.03-0.01*0.02*
SE0.0110.0140.0160.021
95% CI-0.06 to 0.02-0.05 to 0.000.04 to 0.03-0.03 to 0.06
CA-125 II (U/mL)
Mean Change0.441.303.293.45
SE0.0970.1730.0940.08
95% CI0.25 to 0.630.96 to 1.643.11 to 3.473.29 to 3.61
Transferrin (mg/dL)
Mean Change-9.09-17.59-2.5-17.41
SE2.0532.4252.8272.245
95% CI-13.11 to -5.07-22.34 to -12.84-8.04 to 3.04-21.81 to -13.01
Prealbumin (mg/dL)
Mean Change0.390.08-4.62-2.94
SE0.2130.1830.2240.192
95% CI-0.03 to 0.81-0.28 to 0.44-5.06 to -4.18-3.32 to -2.56

*Data obtained at 4 weeks.

  • d. Limit of Detection
    The limits of detection and limits of quantitation reported in each assay's package insert were confirmed and they are incorporated into the algorithm such that results outside of the measuring interval are not imported and do not yield an OVA1™ Test score.

  • e. Analytical specificity:
    Interference: Three pooled serum samples with low (control score ~3.0). medium (control score ~4.5) and high (control score ~10.00) OVA1™ Test scores were evaluated for interference by hemoglobin, bilirubin (conjugated and unconjugated), triglycerides and rheumatoid factor. The effect of each interfering substance on the OVA1™ Test score was assessed using a mean of 4 repeated measurements on each sample and compared to control measurements. No significant interference at the concentrations evaluated

{8}------------------------------------------------

was observed on the OVA1™ Test score for any of the interferents evaluated except for rheumatoid factor. Specimens with rheumatoid factor greater than 250 RU/mL are not appropriate for the OVA1™ Test. Acceptance criteria for interference were < 10% difference between the sample with interferent and control.

InterferentSubstanceConc.OVA1(low)OVA1(med)OVA1(high)
% difference fromcontrol
Hemoglobin (g/L)9.00.00-1.060.00
Bilirubin (Conjugated) (g/L)0.90.000.000.00
Bilirubin (Unconjugated) (g/L)0.90.00-0.540.00
Triglycerides (g/L)5.0-0.800.540.00
Rheumatoid Factor (IU/mL)2509.7613.7-0.75
Rheumatoid Factor (IU/mL)50032.229.62.76
Rheumatoid Factor (IU/mL)101194.260.6-15.3
Rheumatoid Factor (IU/mL)250044.535.7-3.50
  • f. Assay cut-off: See clinical cut-off
    1. Comparison studies:
    • Method comparison with predicate device: a. Not applicable
    • b. Matrix comparison: Serum is the only claimed matrix for each of the 5 analytes evaluated.

3. Clinical studies:

  • Clinical Sensitivity and Specificity: a.
    The OVA 1™ Test score is based on 5 biomarkers (Prealbumin, Apo A-1, B2M, Transferrin, and CA 125) measured individually as 5 immunoassays and incorporated into a single numerical score using an algorithm. The algorithm was derived using two independent training data sets from preoperative serum samples. The output of the algorithm is a numeric index between 0.0 and 10.0. Two cutoffs, 5.0 and 4.4 for pre- and post-menopausal patients respectively, were identified based on the training data. The cutoff score classifies a patient based on her OVA1™ Test score as low probability or high probability for presence of ovarian malignancy.

Training set 1 consisted of 284 pre-operative serum samples from women with adnexal mass, obtained from the University of Kentucky: 175 benign diseases, 29 ovarian tumors of low malignant potential (LMP), 64 epithelial ovarian cancers, 3 other primary ovarian malignancies and 13 other

{9}------------------------------------------------

malignancies. Complete laboratory data was available for 274 samples of which 109 were malignant and 175 were benign controls. Training set 2 consisted of a randomly selected subset of 146 pre-operative serum samples collected under a collection/enrollment protocol from the clinical trial serum specimen repository. Twenty-one (21) of these samples were not evaluable. The remaining set of 125 consisted of 89 benign diseases, 10 LMPs, 19 epithelial ovarian cancers, 1 primary and 3 non-primary ovarian cancers and 3 other malignancies.

Clinical Validation Study:

The clinical validation study was a prospective, multicenter, double-blind clinical study. Study samples were collected from 27 demographically mixed subject enrollment sites that are representative of institutions where patients with ovarian masses typically undergo a complete clinical evaluation prior to surgical intervention. These sites included large and small medical centers (universities/community hospitals), small gynecology/obstetrics groups, gynecology/oncology practices, and subjects from HMO groups. Subjects were women over 18 with a documented pelvic mass following physical examination and clinical examination. Enrollment in the study was limited to those patients with planned surgical intervention. Patients who had a diagnosis of malignancy within the last 5 years with the exception of melanoma were excluded from the study. Pre-surgical assessments identifying the mass as benign or malignant were made based on a variety of clinical assessments. One imaging test was required and had to be performed within 12 weeks of surgery. Blinded sample testing was conducted at 3 laboratories using bar-coded serum aliquots. The OVA1™ Test results, in conjunction with other clinicopathologic variables (e.g., patient's symptoms, physical findings, imaging, CA-125 value), were compared to histopathology results for detecting the presence of ovarian malignancy.

A total of 743 patients were enrolled in the study. A total of 146 subjects were set aside as a training set (as described above). Seventy-four (74) subjects/specimens were eliminated due to missing information or lack of sample resulting in a final total of 516 evaluable subjects/samples. Menopausal status was self-reported. In cases where menopausal status was not identified. a cut off of 50 years of age was applied. A pre-surgical clinical assessment based on radiological findings and other clinical data was obtained by the physician for each patient.

OVA1™ Test score cut points for negative versus positive test result were as follows:

Pre-menopausal:

low probability for malignancyOVA1TM Test score < 5.0
high probability for malignancyOVA1TM Test score ≥ 5.0

{10}------------------------------------------------

Post-menopausal:

low probability for malignancyOVA1TM Test score < 4.4
high probability for malignancyOVA1TM Test score ≥ 4.4

Summary statistics for enrolled subjects (age, menopausal status, and pathology) for all evaluable patients are in the following Table. For 269 patients, the pre-surgical evaluations were made by physicians who are not gynecologic oncologists (non-GO, i.e., gynecologists and primary care physicians). For 247 patients, the pre-surgical evaluations were made by gynecologic oncologists (GO). The demographic and outcome data are presented for all patients and for the two pre-surgical evaluation subsets.

Demographic Characteristics and Pathology Results for Evaluable Subjects with aPresurgical Clinical Assessment
All EvaluableSubjects withPresurgicalAssessment(N= 516)Subjects withEvaluation byNon-GOPhysicians(N= 269)Subjects withEvaluation by GOPhysicians(N= 247)
Age, years
N516269247
Mean (SD)52.0 (13.9)49.7 (13.6)54.6 (13.8)
Range (min, max)18 to 9219 to 9018 to 92
Menopausal Status, n (%)
Pre235 (45.5%)144 (53.5%)91 (36.8%)
Post281 (54.5%)125 (46.5%)156 (63.2%)
Pathology Diagnosis, n (%)
Benign ovarian conditions355 (68.8%)197 (73.2%)158 (64.0%)
Epithelial ovarian cancer96 (18.6%)45 (16.7%)51 (20.6%)
Other primary ovarianmalignancies (not EOC)9 (1.7%)5 (1.9%)4 (1.6%)
Low malignant potential(Borderline)28 (5.4%)12 (4.5%)16 (6.5%)
Non-primary ovarianmalignancies withinvolvement of the ovaries18 (3.5%)5 (1.9%)13 (5.3%)
Non-primary ovarianmalignancies with noinvolvement of ovaries10 (1.9%)5 (1.9%)5 (2.0%)

All major racial groups were represented (21.7% (128/590) of subjects from ethnic groups).

{11}------------------------------------------------

For the 105 surgically confirmed cases of primary ovarian malignancy, 99 had tumor grade information (11 grade 1, 25 grade II, and 63 grade III), and 103 had stage information (31 stage I, 18 stage II, 51 stage III, and 3 stage IV). Summary statistics for OVA1™ Test scores in these patients are given by stage in the table below.

Stage IStage IIStage IIIStage IVNot Given
No. ofSubjects31185132
OVA1 ScoreMean (SD)6.48(1.786)8.04(1.596)8.26(1.357)8.70(1.054)6.05(1.626)
Median6.308.608.808.606.05
Range(min, max)3.6 to 10.05.0 to 10.05.0 to 10.07.7 to 9.84.9 and 7.2

OVA1™ Test scores for the 105 EOC and non-EOC primary ovarian malignancies by stage:

The distribution of histological classifications for the 105 patients with primary malignant ovarian tumors included fifty-five (55) serous, eight (8) mucinous, ten (10) endometroid, eight(8) clear cell, and twenty-four (24) other.

Performance Characteristics of the OVAI™ Test as Used by Physicians Who Are Not Gynecologic Oncologists (non-GO):

From a total of 516 evaluable subjects, 269 patients (age range 19-90) were evaluated by physicians who are not gynecologic oncologists (non-GO, i.e., gynecologists and primary care physicians). Of the 269 patients, 144 subjects were identified as pre-menopausal and 125 as post-menopausal.

For each patient, the OVA1™ Test result was compared to the pathology report from tissue. Malignant masses were those that included epithelial ovarian cancer (EOC), ovarian tumor of low malignant potential (LMP), other ovarian primary malignancy, malignancy involving but not arising in the ovary, and malignancy neither involving nor arising in the ovary, The table below shows performance characteristics of the OVA1™ Test alone for all subjects evaluated by non-GO, and separately for pre-menopausal and postmenopausal subjects. NPV denotes negative predictive value, and PPV denotes positive predictive value.

{12}------------------------------------------------

Performance of OVA1™ Test Alone Compared to Histopathology in
Patients Evaluated by Non-GO
AllPre-menopausalPost-menopausal
N269144125
Sensitivity87.5% (63/72)80.8% (21/26)91.3% (42/46)
95% CI77.9% - 93.3%62.1% - 91.5%79.7% - 96.6%
Specificity50.8% (100/197)56.8% (67/118)41.8% (33/79)
95% CI43.8% - 57.7%47.8% - 65.4%31.5% - 52.8%
NPV91.7% (100/109)93.1% (67/72)89.2% (33/37)
PPV39.4% (63/160)29.2% (21/72)47.7% (42/88)
Prevalence26.8% (72/269)18.1% (26/144)36.8% (46/125)

In order to conclude that the test is statistically informative, it is necessary to demonstrate that the TPR (True Positive Rate, same as Sensitivity) differs from the FPR (False Positive Rate, same as 1 - Specificity). For a non-informative test (a random test, no better than the toss of a coin), the percent of positive test results does not depend on whether a subject has a target condition or not (TPR = FPR).

The OVA1™ Test was statistically informative for the combined data and for premenopausal and post-menopausal subjects separately.

  • . The True Positive Rate (TPR) exceeded the False Positive Rate (FPR) for all combined data: the estimate of TPR was 87.5% (63/72) and the estimate of FPR was 49.2% (97/197); the difference of TPR and FPR was statistically significant (38.3% with 95% CI: 26.5% to 47.8%).
  • The TPR exceeded the FPR for the pre-menopausal subjects: the estimate . of TPR was 80.8% (21/26) and the estimate of FPR was 43.2% (51/118); the difference of TPR and FPR was statistically significant (37.6% with 95% CI: 16.7% to 52.2%).
  • The TPR exceeded FPR for the post-menopausal subjects: the estimate of ● TPR was 91.3% (42/46) and the estimate of FPR was 58.2% (46/79); the difference of TPR and FPR was statistically significant (33.1% with 95% CI: 17.3% to 46.1%).

The information provided by the OVA1™ Test should be used by the physician only as an adjunctive test to complement, not replace, other diagnostic and clinical procedures. To examine whether the OVA1™ Test provides additional information when used in combination with the physician's pre-surgical assessment, the ability to contribute to the physician's pre-surgical assessment was analyzed.

The analysis examined whether patient referral to a gynecological oncologist is supported when positive OVA1™ Test results occur in the setting of negative clinical evaluations by non-GO physicians. The mass was declared potentially malignant if the pre-surgical clinical assessment, the OVA1™ Test score, or both were positive.

{13}------------------------------------------------

In the study, there were 269 subjects:

  • 24.1% (65/269) were subjects with "Positive" Pre-surgical assessment ● and positive OVA1™ Test results,
  • 7.8% (21/269) were subjects with "Positive" Pre-surgical assessment and negative OVA1™ Test results,
  • 35.3% (95/269) were subjects with "Negative" Pre-surgical assessment ● and positive OVA1™ Test results, and
  • 32.7% (88/269) were subjects with "Negative" Pre-surgical assessment . and negative OVA1™ Test results.
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive6595160
Negative2188109
Total86183269

Among 269 subjects, there were 72 subjects with malignancy by pathology and 197 subjects with no malignancy by pathology.

Malignancy by Pathology
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive491463
Negative369
Total522072
No Malignancy by Pathology
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive168197
Negative1882100
Total34163197

The following table presents the observed frequencies of malignancy tabulated according to pre-surgical evaluation and OVA1™ Test results from the 269 patients evaluated by non-GO physicians:

{14}------------------------------------------------

Frequency ofMalignancy95% CI
Prevalence of malignancy among patients with adnexal mass assessed bynon-GO physicians: 26.8% (72/269)
Pre-surgical assessment alone"Positive"60.5%(52/86)49.9% to 70.1%
Pre-surgical assessment alone"Negative"10.9%(20/183)7.2% to 16.3%
OVA1TM Test Score alone"Positive"39.4%(63/160)32.1% to 47.1%
OVA1TM Test Score alone"Negative"8.3%(9/109)4.4% to 15.0%
Pre-surgical assessment "Positive"andOVA1TM Test Score "Positive"75.4%(49/65)63.7% to 84.2%
Pre-surgical assessment “Positive”andOVA1TM Test Score “Negative”14.3%(3/21)5.0% to 34.6%
Pre-surgical assessment “Negative”andOVA1TM Test Score "Positive"14.7%(14/95)9.0% to 23.2%
Pre-surgical assessment “Negative”andOVA1TM Test Score “Negative”6.8%(6/88)3.2% to 14.1%

The same information about the frequencies of malignancy can be presented also by the likelihood ratios: Likelihood ratio (Result) = Pr(Result|Malignancy) / Pr(Result|No Malignancy). Likelihood ratio is a way of quantifying how much a given test result changes the pre-test probability of malignancy in a patient.

Results for PatientLikelihoodRatio
Pre-surgical assessment alone “Positive”4.18
Pre-surgical assessment alone “Negative”0.34
OVA1TM Test Score alone “Positive”1.78
OVA1TM Test Score alone “Negative”0.25
Pre-surgical assessment “Positive” andOVA1TM Test Score “Positive”8.37
Pre-surgical assessment “Positive” andOVA1TM Test Score “Negative”0.46
Pre-surgical assessment “Negative” andOVA1TM Test Score “Positive”0.47
Pre-surgical assessment “Negative” andOVA1TM Test Score “Negative”0.20

{15}------------------------------------------------

The Table below shows performance characteristics for the test applied to all subjects evaluated by non-GO physicians. For Single Assessment, only the pre-surgical assessment was used, without reference to an OVA1™ Test result. For Dual Assessment, the adnexal mass was declared potentially malignant if the pre-surgical clinical assessment, the OVA1™ Test score, or both were positive.

PerformanceSingle Assessment (Pre-surgical Assessment)Dual Assessment (Pre-surgical Assessment and OVA1TM Test)
Sensitivity72.2% (52/72)91.7% (66/72)
Specificity82.7% (163/197)41.6% (82/197)
PPV60.5% (52/86)36.5% (66/181)
NPV89.1% (163/183)93.2% (82/88)
Prevalence26.8% (72/269)

With dual assessments, sensitivity for malignancy increased from 72% to 92%. That is, approximately two-thirds of the malignancies missed by pre-surgical assessment alone were called positive when a dual assessment was used. Specificity for malignancy decreased from 83% to 42% with dual assessment. The ratio of false positive results to true positive results for the study population increased from 34:52 (0.65:1) with single assessment to 115:66 (1.74:1) with dual assessment (PPV with dual assessment decreased from 60% to 37%). However, NPV with dual assessment increased from 89% to 93%. supporting improved performance with dual assessment. The confidence interval for the observed 4.1% increase was -0.5% to 8.7% (calculated by bootstrap). The statistical significance of the observed increase in NPV was borderline.

For the already surgery-bound intended use population, the additional true positive cases detected by dual assessment present a sufficient benefit (in terms of opportunity for optimal treatment) compared to the additional false positive cases (for which an unneeded referral poses no medical risk to the patients).

Pre-menopausal subjects

Among 144 pre-menopausal subjects, with pre-surgical assessment by a non-GO, there were 26 subjects with malignancy by pathology and 118 subjects with no malignancy by pathology.

Malignancy by Pathology
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive16521
Negative145
Total17926

{16}------------------------------------------------

{17}------------------------------------------------

No Malignancy by Pathology
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive74451
Negative135467
Total2098118
PerformanceSingle Assessment(Pre-surgical Assessment)Dual Assessment(Pre-surgical Assessmentand OVA1™ Test)
Sensitivity65.4% (17/26)84.6% (22/26)
Specificity83.1% (98/118)45.8% (54/118)
PPV45.9% (17/37)25.6% (22/86)
NPV91.6% (98/107)93.1% (54/58)
Prevalence18.1% (26/144)

PPV of the dual assessment decreased from 46% to 26% and NPV of the dual assessment increased from 92% to 93%. The confidence interval for the observed 1.5% increase was -3.3% to 6.4% (calculated by bootstrap).

Post-menopausal subjects

Among 125 post-menopausal subjects, there were 46 subjects with malignancy by pathology and 79 subjects with no malignancy by pathology.

Malignancy by Pathology
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive33942
Negative224
Total351146
No Malignancy by Pathology
Non-GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive93746
Negative52833
Total146579

{18}------------------------------------------------

PerformanceSingle Assessment(Pre-surgical Assessment)Dual Assessment(Pre-surgical Assessmentand OVA1™ Test)
Sensitivity76.1% (35/46)95.7% (44/46)
Specificity82.3% (65/79)35.4% (28/79)
PPV71.4% (35/49)46.3% (44/95)
NPV85.5% (65/76)93.3% (28/30)
Prevalence36.8% (46/125)

PPV of the dual assessment decreased from 72% to 46% and NPV of the dual assessment increased from 86% to 93%. The confidence interval for the observed 7.8% increase was -1.6% to 17.2% (calculated by bootstrap).

Summary statistics for OVA1™ Test scores, for subjects who were both evaluated by a non-GO physician and had a primarv ovarian malignancy (EOC and non-EOC), are given by cancer stage in the table below. There were no Stage IV cancers among these patients.

Stage IStage IIStage III
No. of Subjects141125
OVA1 Mean6.898.218.36
(SD)(2.313)(1.600)(1.289)
Median6.558.608.70
Range(min, max)3.6 to 10.05.1 to 10.05.5 to 10.0
OVA1 Positive111125
OVA1 Negative300
OVA1 Sensitivity78.6 %100%100%

Table # OVA1™ Test Scores by Cancer Stage for primary ovarian malignancies in All Evaluable Subjects with a Pre-surgical Clinical Assessment from Non-GO

Performance Characteristics of the OVAT™ Test as Used by Gynecologic-Oncologists (GO):

The characteristics of patients evaluated by GO differ from those of patients evaluated by non-GO physicians. Due to selective referral, the prevalence of malignancy is typically higher for adnexal masses in patients who are evaluated by GO, and malignancy found at surgery for these patients may be higher stage disease which can impact test performance characteristics. In addition, the spectrum of disease (i.e., the kind and frequency of benign adnexal masses) may vary between GO and non-GO settings and impact the

{19}------------------------------------------------

performance characteristics of the test. As a result, performance characteristics within already-referred patients may corroborate, but are insufficient to establish, performance characteristics in a not-yet-referred patient population.

There were 247 patients evaluated by gynecologic oncologists in the OVA1™ Test clinical study, with 91 premenopausal and 156 postmenopausal subjects. The Table below presents results for all subjects evaluated by GO and separately by menopausal status for pre-menopausal and post-menopausal subjects.

Performance of OVA1™ Test compared to Histology in patients Evaluatedby GO
AllPre-menopausalPost-menopausal
N24791156
Sensitivity96.6% (86/89)94.7% (18/19)97.1% (68/70)
(95% CI)90.6% - 98.9%75.4%-99.1%90.2% -99.2%
Specificity32.9% (52/158)43.1% (31/72)24.4% (21/86)
(95% CI)26.1% - 40.6%32.2% - 54.6%16.6% - 34.5%
NPV94.5% (52/55)96.9% (31/32)91.3% (21/23)
PPV44.8% (86/192)30.5% (18/59)51.1% (68/133)
Prevalence36.0% (89/247)20.9% (19/91)44.9% (70/156)

For the patients evaluated by gynecologic oncologists, the observed positive likelihood ratio (PLR) was 1.44 and the observed negative likelihood ratio (NLR) was 0.10; for the patients evaluated by non-GO physicians, PLR was 1.78 and NLR was 0.25. Sensitivity of the OVA1™ Test was notably higher, and specificity was notably lower, for GO-evaluated patients compared to non-GO-evaluated patients. Nevertheless, the NPV and PPV figures were slightly higher in the GO-evaluated patients (predictive values depend on the corresponding likelihood ratios and prevalence).

Analyses combining pre-surgical assessment with the OVA1™ Test result were completed for GO-evaluated patients, echoing the analyses performed for patients evaluated by non-GO physicians. The results are corroborative, but are not dispositive, concerning safety and effectiveness of the test for the intended use population.

GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive9696192
Negative134255
Total109138247

Among 247 subjects, there were 89 subjects with malignancy by pathology and 158 subjects with no malignancy by pathology.

{20}------------------------------------------------

Malignancy by Pathology
GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive671986
Negative213
Total692089
No Malignancy by Pathology
GO Pre-surgical Assessment
PositiveNegativeTotal
OVA1Positive2977106
Negative114152
Total40118158

The following table presents the observed frequencies of malignancy tabulated according to pre-surgical evaluation and OVA1™ Test results from the 247 patients evaluated by gynecologic oncologists:

Frequency ofMalignancy95% CI
Prevalence of malignancy among patients with adnexal mass assessed by gynecologic oncologists: 36.0% (89/247)
Pre-surgical assessment alone"Positive"63.3%(69/109)53.9% to 71.8%
Pre-surgical assessment alone"Negative"14.5%(20/138)9.6% to 21.3%
OVA1™ Test Score alone"Positive"44.8%(86/192)37.9% to 51.9%
OVA1™ Test Score alone"Negative"5.5%(3/55)1.9% to 14.9%
Pre-surgical assessment “Positive”andOVA1™ Test Score “Positive”69.8%(67/96)60.0% to 78.1%
Pre-surgical assessment “Positive”andOVA1™ Test Score “Negative”15.4%(2/13)4.3% to 42.2%
Pre-surgical assessment “Negative”andOVA1™ Test Score "Positive"19.8%(19/96)13.1% to 28.9%
Pre-surgical assessment “Negative”andOVA1™ Test Score "Negative"2.4%(1/42)0.4% to 12.3%

The same information about the frequencies of malignancy is presented also by the

{21}------------------------------------------------

observed likelihood ratios:

Results for PatientLikelihoodRatio
Pre-surgical assessment alone “Positive”3.06
Pre-surgical assessment alone “Negative”0.30
OVA1™ Test Score alone “Positive”1.44
OVA1™ Test Score alone “Negative”0.10
Pre-surgical assessment “Positive” andOVA1™ Test Score “Positive”4.10
Pre-surgical assessment “Positive” andOVA1™ Test Score “Negative”0.32
Pre-surgical assessment “Negative” andOVA1™ Test Score “Positive”0.44
Pre-surgical assessment “Negative” andOVA1™ Test Score “Negative”0.04

The Table below shows results for all subjects evaluated by gynecological oncologists, with conclusions for malignancy either based on a positive pre-surgical assessment alone (single assessment), or based on a positive result from the pre-surgical assessment or from the OVA1™ Test result or from both (dual assessment).

PerformanceSingle Assessment(Pre-surgical Assessment)Dual Assessment(Pre-surgical Assessmentand OVA1TM Test)
Sensitivity77.5% (69/89)98.9% (88/89)
Specificity74.7% (118/158)25.9% (41/158)
PPV63.3% (69/109)42.9% (88/205)
NPV85.5% (118/138)97.6% (41/42)
Prevalence36.0% (89/247)

With dual assessments, sensitivity for malignancy increased from 77% to 99% and specificity for malignant calls decreased from 75% to 26% with dual assessment. Also with dual assessment, the PPV decreased from 63% to 43%, and the NPV increased from 86% to 98%. The confidence interval for the observed 12.1% increase in NPV was 5.7% to 18.6% (calculated by bootstrap). The observed increase in NPV was statistically significant.

Summary statistics for OVA1™ Test scores, for subjects who were both evaluated by a gynecologic oncologist and had a primary ovarian malignancy (EOC and non-EOC) are given by cancer stage in the table below:

Table # OVA1™ Test Scores by Cancer Stage for Evaluable Subjects with a Pre-surgical Clinical Assessment from Gynecologic Oncologists

{22}------------------------------------------------

Stage IStage IIStage IIIStage IVNot Given
No. ofSubjects1772632
OVA1Mean(SD)6.14(1.167)7.79(1.681)8.17(1.439)8.70(1.054)6.05(1.626)
Median6.308.608.858.606.05
Range(min,max)4.4 to 7.95.0 to 9.85.0 to 10.07.7 to 9.84.9 and 7.2
OVA1Positive1772632
OVA1Negative00000
Sensitivity(%)100100100100100
  • b. Other clinical supportive data (when a is not applicable): Not applicable.
    1. Clinical cut-off:

The results are interpreted as follows:

Pre-menopausal:

low probability for malignancyhigh probability for malignancyOVA1TM Test score < 5.0OVA1TM Test score ≥ 5.0
Post-menopausal:
low probability for malignancyOVA1TM Test score < 4.4
high probability for malignancyOVA1TM Test score ≥ 4.4
  • Expected values/Reference interval: 5.
    To determine the reference interval of OVA1™ Test in healthy women, 69 pre-menopausal patients and 78 post-menopausal patients were tested (total = 147 evaluable subjects). Ages ranged from 18 to 85 and represented whites (81.3%), Hispanic/Latino (8.7%) and African American (7.3%) subjects. Using a cut-off of 5.0 for OVA1™ Test scores, the results from the pre- and post-menopausal populations are presented below:

{23}------------------------------------------------

All EvaluableSubjects(N= 147)Pre-menopausalSubjects(N= 69)Post-menopausalSubjects(N= 78)
N1476978
OVA1™ Test Score
Mean (SD)4.18 (0.858)4.34 (0.871)4.04 (0.827)
Median4.104.303.95
Range (min, max)2.7, 6.72.9, 6.72.7, 6.1
Reference interval(5th percentile, 95th percentile)(3.0, 5.8)(3.0, 5.8)(2.9, 5.9)
OVA1™ Test Score' n (%)
Positive46 ( 31.3% )20 ( 29.0% )26 ( 33.3% )
Negative101 ( 68.7% )49 ( 71.0% )52 ( 66.7% )

Expected Values in Non-Ovarian Malignancy Condition: To evaluate the performance of the OVA1™ Test in subjects with other benign and malignant conditions, the OVA1™ Test was evaluated in subjects with (breast cancer, Stage 1 to Stage IV; endometrial cancer, cervical cancer, bladder cancer, lung cancer, colon cancer, leukemia and lymphoma) and in women with benign conditions (anemia of chronic disease, iron deficiency anemia, acute and chronic pelvic inflammatory disease, malnutrition, cardiac disease, hepatitis, kidney diseases and autoimmune disease). A total of 360 evaluable specimens were analyzed.

Evaluable specimens
All Specimens360
Bladder cancer16
Breast cancer45
Cervical cancer12
Colon cancer40
Endometrial cancer44
Leukemia10
Lung cancer13
Lymphoma13
Autoimmune disease10
Cardiac disease12
Diabetes40
Endometriosis40

{24}------------------------------------------------

Evaluable specimens
Hepatitis13
Kidney disease12
Pregnant women10
Anemia11
Pelvic inflammatory disease9
Malnutrition10

The mean, standard deviation, 5th to 95th percentiles as observed in the data are shown for each condition group. Using a cut-off of 5.0 for OVA1™ Test scores, the number of positive and negative cases is presented below:

BladderCancer(N= 16)BreastCancer(N= 45)CervicalCancer(N= 12)ColonCancer(N= 40)EndometrialCancer(N= 44)Leukemia(N= 10)LungCancer(N= 13)Lymphoma(N= 13)
OVA1TM Test Score, statistics
N1645124044101313
Mean(SD)4.57(1.109)4.34(1.151)6.40(1.979)4.86(1.259)5.00(1.467)6.04(0.651)4.17(0.884)5.26(1.669)
Median4.44.16.64.84.66.14.04.6
5th to 95thpercentiles2.9 to6.52.9 to6.03.4 to 9.83.0 to7.03.5 to 8.34.6 to 7.03.0 to5.53.0 to 8.6
OVA1TM Test Score, n
Positive61181815936
Negative1034422291107
% negativeresults62.575.633.355.065.910.076.953.8

Summary of OVA1 Scores for Specimens from Subjects with Non-cancer Conditions

AutoimmuneDisease(N= 10)CardiacDisease(N= 12)Diabetes(N= 40)Endo-metriosis(N= 40)Hepatitis(N= 13)
OVA1TM Test Score, statistics
N1012404013
Mean (SD)5.23 (1.083)5.28 (1.345)4.39 (1.069)5.01 (1.189)4.46 (0.832)
Median5.35.64.04.74.2
5th to 95th percentiles3.4 to 6.63.3 to 8.03.2 to 6.43.4 to 7.13.5 to 6.0

{25}------------------------------------------------

AutoimmuneDisease(N= 10)CardiacDisease(N= 12)Diabetes(N= 40)Endo-metriosis(N= 40)Hepatitis(N= 13)
OVA1TM Test Score, n
Positive5710173
Negative55302310
%negative results50.041.775.057.576.9
KidneyDisease(N= 12)PregnantWomen(N= 10)Anemia(N= 11)Pelvicinflammatorydisease(N= 9)Malnutrition(N= 10)
OVA1™ Test Score, statistics
N121011910
Mean (SD)6.13 (0.094)4.86 (0.686)3.73 (0.673)4.52 (1.072)4.30 (0.948)
Median6.24.84.04.44.4
5th to 95th percentiles5.9 to 6.24.0 to 6.22.7 to 4.73.2 to 6.32.8 to 5.7
OVA1™ Test Score, n
Positive123022
Negative071178
%negative results0.070.010077.880.0

The number of cases is small within each of the examined diseases and conditions, but the results suggest that caution is warranted when interpreting OVA 1™ Test results for patients with cervical cancer, leukemia, kidney disease and anemia.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The petition for Evaluation of Automatic Class III Designation for this device is accepted. The device is classified as Class II under regulation 21 CFR 866.6050 with special controls. The special control guidance document, "Class II Special Controls Guidance Document: Ovarian Adnexal Mass Assessment Score Test System" accompanies this device.

§ 866.6050 Ovarian adnexal mass assessment score test system.

(a)
Identification. An ovarian/adnexal mass assessment test system is a device that measures one or more proteins in serum or plasma. It yields a single result for the likelihood that an adnexal pelvic mass in a woman, for whom surgery is planned, is malignant. The test is for adjunctive use, in the context of a negative primary clinical and radiological evaluation, to augment the identification of patients whose gynecologic surgery requires oncology expertise and resources.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Ovarian Adnexal Mass Assessment Score Test System.” For the availability of this guidance document,see § 866.1(e).(c)
Black box warning. Under section 520(e) of the Federal Food, Drug, and Cosmetic Act these devices are subject to the following restriction: A warning statement must be placed in a black box and must appear in all advertising, labeling, and promotional material for these devices. That warning statement must read: