K Number
DEN070006
Date Cleared
2007-06-13

(82 days)

Product Code
Regulation Number
866.3402
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Binax NOW® Malaria Test is an in vitro immunochromatographic assay for the qualitative detection of Plasmodium antigens circulating in human venous and capillary EDTA whole blood of individuals with signs and symptoms of malarial infection. The test targets the histidine-rich protein II (HRPII) antigen specific to Plasmodium falciparum (P.f.) and a pan-malarial antigen, common to all four malaria species capable of infecting humans - P. falciparum, P. vivax (P.v.), P. ovale (P.o.), and P. malariae (P.m.). It is intended to aid in the rapid diagnosis of human malaria infections and to aid in the differential diagnosis of Plasmodium falciparum (P.f.) infections from other less virulent malarial infections. Negative results must be confirmed by thin / thick smear microscopy.

The Binax NOW® Malaria Test is for the laboratory diagnosis of malaria in individuals with signs and symptoms consistent with malaria infection.

Device Description

In vitro immunochromatographic immunoassay

The Binax NOW® Malaria Test is an immunochromatographic membrane assay that uses monoclonal antibodies to detect Plasmodium falciparum antigen and pan-malarial antigen (an antigen shared by all Plasmodium species causing human malaria) in venous and capillary whole blood specimens. These antibodies, and a control antibody, are immobilized on a membrane support as three distinct lines and are combined with a sample pad, which is impregnated with visualizing particles conjugated to control and anti-malarial antibodies, to create a test strip. This test strip is mounted in a book-shaped. hinged test device, along with wash and absorbent pads, intended to aid in the clearing of the membrane when the device is closed.

To perform the test, whole blood is applied to the sample pad. Malarial antigen present in the sample reacts to bind the anti-malaria conjugated antibody. Reagent A is added to the bottom of the test strip and allows the antigen-conjugate complexes to migrate along the test strip, where they are captured by the immobilized antibodies, forming the Test Line(s). Immobilized control antibody captures control conjugate, forming the Control Line. Once the blood sample has migrated the length of the test strip, the device is closed, allowing Reagent A that has been added to the wash pad to clear the test strip of excess blood.

Test results are interpreted by the presence or absence of visually detectable pink-to-purple colored lines. A positive test result, read in 15 minutes, will include the detection of both a Test Line (or Test Lines) and a Control Line. A negative test result, read in 15 minutes, will produce only a Control Line, indicating that malarial antigens were not detected in the sample. Failure of the Control Line to appear, whether the Test Line(s) is present or not, indicates an invalid result.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the supporting study for the Binax NOW® Malaria Test, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied)Reported Device Performance
Clinical Sensitivity for P. falciparum (P.f.)Overall SensitivityHigh, especially at higher parasitemia levels95.3% (531/557) Overall
Sensitivity at > 5000 parasites/μLVery High99.7% (326/327)
Sensitivity at 1000-5000 parasites/μLHigh99.2% (126/127)
Sensitivity at 500-1000 parasites/μLModerate to High92.6% (25/27)
Sensitivity at 100-500 parasites/μLModerate89.2% (33/37)
Sensitivity at 0-100 parasites/μLLower, due to low parasite burden53.9% (21/39)
Clinical Specificity for P. falciparum (P.f.)SpecificityHigh94.2% (3297/3500) in endemic population; 100% in low incidence population
Clinical Sensitivity for P. vivax (P.v.)Overall SensitivityModerate to High68.9% (818/1187) (increases to 74.6% with two lines)
Sensitivity at > 5000 parasites/μLHigh93.5% (462/494)
Sensitivity at 1000-5000 parasites/μLModerate to High81.0% (277/342)
Sensitivity at 500-1000 parasites/μLLower47.4% (37/78)
Sensitivity at 100-500 parasites/μLLow23.6% (34/144)
Sensitivity at 0-100 parasites/μLVery Low6.2% (8/129)
Clinical Specificity for P. vivax (P.v.)SpecificityVery High99.8% (2863/2870)
Clinical Sensitivity for P. malariae (P.m.)Overall SensitivityLower (User must establish)43.8% (7/16) (increases to 75.0% with two lines)
Clinical Sensitivity for P. ovale (P.o.)Overall SensitivityLower (User must establish)50% (1/2)
Clinical Sensitivity for Mixed P.f./P.v. InfectionOverall SensitivityHigh94.1% (32/34)
Analytical SpecificityAbsence of interference from common pathogensNo cross-reactivity28 pathogenic microorganisms (bacteria, protists, viruses) tested negative
Absence of interference from anti-malarial drugsNo interference at specified concentrations8 anti-malarial drugs tested, no interference
Absence of interference from antibioticsNo interference at specified concentrations4 antibiotics tested, no interference
Absence of interference from anti-inflammatory drugsNo interference at specified concentrations3 anti-inflammatory drugs tested, no interference
Absence of interference from endogenous blood componentsNo interferenceHemoglobin, protein, bilirubin, triglycerides tested, no interference
Absence of interference from unrelated medical conditionsHigh specificity5 false positives out of 116 specimens (from 4 rheumatoid factor, 1 HAMA)
Precision/ReproducibilityAgreement with expected resultsHigh97% (140/144) agreement
Limit of Detection (LOD)P.f. parasitemia levelClear threshold1001-1500 parasites per μL
P.v. parasitemia levelClear threshold5001-5500 parasites per μL

Study Information

  1. Sample size used for the test set and the data provenance:

    • Clinical Efficacy (Endemic Population): 4,122 whole blood specimens collected from patients in regions considered endemic for malaria. Prospective study conducted in 2001 outside the U.S.
    • Clinical Specificity (Low Incidence Population): 100 whole blood specimens from febrile patients. Prospective study conducted in the eastern US in 2006-2007.
    • Precision/Reproducibility: Not explicitly stated as "test set," but involved a panel of blind coded specimens (negative, LOD, low positive for P.f. and P.v.).
    • Analytical Specificity (Microorganisms/Drugs/Components): Number of samples varied per test (e.g., 28 microorganisms, various drug concentrations, 116 specimens for unrelated medical conditions). Provenance not specified but likely lab-generated or commercial controls.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The primary ground truth for clinical studies was Giemsa malaria microscopy.
    • The document does not specify the number of individual experts who performed the microscopy or their specific qualifications (e.g., "microscopist with X years of experience"). It implies standard laboratory practices for identifying malaria parasites via microscopy.
  3. Adjudication method for the test set:

    • The document does not explicitly state an adjudication method for conflicting microscopy results. It simply states "Microscopy was considered positive only when asexual malaria forms were detected." For the Binax NOW® results, interpretation was based on visually detectable pink-to-purple lines, and invalid results were noted if the control line didn't appear.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test, not an AI-powered diagnostic for image interpretation involving human readers. Therefore, there's no mention of "AI assistance" or "human reader improvement." The comparison was between the device (Binax NOW®) and traditional microscopy (human interpretation).
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance study was done. The Binax NOW® test is a lateral flow immunochromatographic assay. Its performance against the ground truth (microscopy) represents its standalone capability as an "algorithm only" in the sense that the test strip itself produces the result, interpreted visually ("visually detectable pink-to-purple colored lines"). There's no human intervention in processing or algorithm output, only in the visual interpretation of the final lines.
  6. The type of ground truth used:

    • For the clinical performance studies, the ground truth was expert microscopy (Giemsa malaria microscopy). Specifically, "Microscopy was considered positive only when asexual malaria forms were detected."
    • For analytical specificity regarding microorganisms, the ground truth was the known presence/concentration of those microorganisms.
  7. The sample size for the training set:

    • The document does not mention a training set in the context of machine learning or AI. This is an immunoassay, which does not typically involve training sets in the same way an AI model would. Assay development for immunoassays involves reagent optimization and validation, not a "training set" of patient data for an algorithm.
  8. How the ground truth for the training set was established:

    • As there is no mention of a training set for an AI/ML algorithm, this question is not applicable.

{0}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number: K061542

  • B. Purpose for Submission: De Novo clearance
  • C. Measurand: Plasmodium antigens

D. Type of Test: Qualitative, in vitro immunochromatographic assay E. Applicant:

Binax, Inc., d/b/a Inverness Medical Professional Diagnostics

  • F. Proprietary and Established Names: Binax NOW® Malaria

G. Regulatory Information:

a) Regulation section: Plasmodium species antigen detection assay 21 CFR 866.3402. b) Classification: Class II Product Code: OAX c) Panel: 83 Microbiology

H. Intended Use:

  • a) Intended use(s):
    The Binax NOW® Malaria Test is an in vitro immunochromatographic assay for the qualitative detection of Plasmodium antigens circulating in human venous and capillary EDTA whole blood of individuals with signs and symptoms of malarial infection. The test targets the histidine-rich protein II (HRPII) antigen specific to Plasmodium falciparum (P.f.) and a pan-malarial antigen, common to all four malaria species capable of infecting humans - P. falciparum, P. vivax (P.v.), P. ovale (P.o.), and P. malariae (P.m.). It is intended to aid in the rapid diagnosis of human malaria infections and to aid in the differential diagnosis of Plasmodium falciparum (P.f.) infections from other less virulent malarial infections. Negative results must be confirmed by thin / thick smear microscopy.

b) Indication(s) for use:

The Binax NOW® Malaria Test is for the laboratory diagnosis of malaria in individuals with signs and symptoms consistent with malaria infection.

  • c) Special condition for use statement(s): The device is for prescription use only

{1}------------------------------------------------

  • d) Special instrument Requirements: NA
  • I. Device Description:
    • In vitro immunochromatographic immunoassay
  • J. Substantial Equivalence Information:
    • N/A De Novo

K. Standard/Guidance Document Referenced (if applicable):

L. Test Principle:

The Binax NOW® Malaria Test is an immunochromatographic membrane assay that uses monoclonal antibodies to detect Plasmodium falciparum antigen and pan-malarial antigen (an antigen shared by all Plasmodium species causing human malaria) in venous and capillary whole blood specimens. These antibodies, and a control antibody, are immobilized on a membrane support as three distinct lines and are combined with a sample pad, which is impregnated with visualizing particles conjugated to control and anti-malaria antibodies, to create a test strip. This test strip is mounted in a book-shaped. hinged test device, along with wash and absorbent pads, intended to aid in the clearing of the membrane when the device is closed.

To perform the test, whole blood is applied to the sample pad. Malarial antigen present in the sample reacts to bind the anti-malaria conjugated antibody. Reagent A is added to the bottom of the test strip and allows the antigen-conjugate complexes to migrate along the test strip, where they are captured by the immobilized antibodies, forming the Test Line(s). Immobilized control antibody captures control conjugate, forming the Control Line. Once the blood sample has migrated the length of the test strip, the device is closed, allowing Reagent A that has been added to the wash pad to clear the test strip of excess blood.

Test results are interpreted by the presence or absence of visually detectable pink-topurple colored lines. A positive test result, read in 15 minutes, will include the detection of both a Test Line (or Test Lines) and a Control Line. A negative test result, read in 15 minutes, will produce only a Control Line, indicating that malarial antigens were not detected in the sample. Failure of the Control Line to appear, whether the Test Line(s) is present or not, indicates an invalid result.

M. Performance Characteristics (if/when applicable):

Analytical performance:

a) Precision/Reproducibility:

A blind study of the Binax NOW® Malaria Test was conducted at 3 separate sites using panels of blind coded specimens containing negative, limit of detection, and low positive P.f. and P.v. samples. Participants tested each sample multiple times on 3 different days. There was 97% (140/144) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (3 different days), between sites (3 sites), or between

{2}------------------------------------------------

operators (6 operators). The overall percent detection of each sample type is summarized below.

Sample Type% Detection
P.f. Low Positive94% (17/18)
P.f. LOD97% (35/36)
P.v. Low Positive94% (17/18)
P.v. LOD100% (36/36)
Negative3% (1/36)*

Overall Percent Detection of P.f. and P.v. Samples

  • One operator called a negative sample a P.f. positive.
  • a. Linearity/assay reportable range:
    NA

  • Traceability, Stability, Expected values (controls, calibrators, or method): b. NA

  • c. Detection limit:

P.f. and P.v. Limits of Detection:

In the study described above. Binax NOW® test clinical limit of detection (LOD) for P.f., defined as the parasitemia level in infected blood that produces positive Binax NOW® test results approximately 95% of the time, was determined to be 1001-1500 parasites per ul and the clinical LOD for P.v. was determined to be 5001-5500 parasites per ul.

d. Analytical specificity:

To determine the analytical specificity of the Binax NOW® Malaria Test, 28 pathogenic microorganisms (7 bacteria, 5 protists and 16 viruses) that may be present in whole blood were tested. All were negative when tested at the concentrations listed below.

{3}------------------------------------------------

TYPEPATHOGEN TESTEDCONCENTRATIONTESTED
BacteriaBorrelia burgdorferi(N40 strain)$2.3 x 10^6$ organisms/ml
Leptospira interrogans(icterohaemorrhagiae)$1.0 x 10^7$ organisms/ml
Leptospira biflexa(andamana)$1.0 x 10^7$ organisms/ml
Treponema pallidum$1.0 x 10^5$ organisms/ml
Rickettsia conorii(Malish 7)$1.0 x 10^7$ organisms/ml
Rickettsia typhi(Wilmington)$1.0 x 10^7$ organisms/ml
Orientia tsutsugamushi -Rickettsia (Karp)$1.0 x 10^7$ organisms/ml
Babesia microti(RMNS strain)$4.4 x 10^7$ parasites/ml
ProtistsTrypanosoma cruzi(Y strain)$1.3 x 10^6$ parasites/ml
Leishmania donovani$1.0 x 10^6$ parasites/ml
Leishmania infantum$1.0 x 10^6$ parasites/ml
Leishmania chagasi$1.0 x 10^6$ parasites/ml
Cytomegalovirus (CMV)(AD169)$1.2 x 10^5$ PFU/ml
Epstein-Barr virus (EBV)$1.1 x 10^4$ copies/ml
Dengue virus - West Pac 74$1.2 x 10^5$ PFU/ml
Dengue virus - S16803$3.9 x 10^4$ PFU/ml
Dengue virus - CH53489$1.3 x 10^4$ PFU/ml
Dengue virus - TVP360$1.4 x 10^5$ PFU/ml
Yellow Fever virus$7.9 x 10^6$ PFU/ml
VirusesWest Nile virus$1.6 x 10^5$ PFU/ml
Chikungunya virus$4.0 x 10^5$ PFU/ml
Ross-River virus$1.0 x 10^6$ PFU/ml
Influenza A - Bayern/7/95$2.5 x 10^7$ TCID50/ml
Influenza B - Victoria/2/87$1.0 x 10^7$ TCID50/ml
HIV-1 (Subtype B)$1.4 x 10^5$ copies/ml
Hepatitis B$2.0 x 10^5$ IU/ml
Hepatitis C$1.9 x 10^5$ IU/ml
Rubella virus> $2.0 x 10^2$ TCID50/ml

{4}------------------------------------------------

Interference from Exogenous Blood Components:

The following substances that may be artificially introduced into whole blood were evaluated in the Binax NOW® Malaria Test at the concentrations listed and were found not to affect test performance. Note: The analytical effects of these drugs on the Binax NOW® test were studied by taking whole blood and spiking it with quantities at high therapeutic concentrations and then testing these samples.

Substance TypeSubstanceConcentration
Anti-malarial drugs(prevention)Mefloquine (Lariam®)1 mg/ml
Doxycycline* (Vibramycin®)1 mg/ml
Chloroquine1 mg/ml
Hydroxychloroquine sulfate1 mg/ml
Paludrine (Proguanil®)1 mg/ml
Primaquine1 mg/ml
Quinine1 mg/ml
Sulfadoxine andPyrimethamine (Fansidar®)1 mg/ml
Antibiotic(treatment)Amoxicillin (Trimox®)0.1 mg/ml
Cephalexin0.1 mg/ml
Ciprofloxacin0.1 mg/ml
Erythromycin0.1 mg/ml
Anti-InflammatoryDrugs(treatment)Aspirin1 mg/ml
Acetaminophen1 mg/ml
Ibuprofen (NSAID)1 mg/ml
  • Doxcycline is also used as an antibiotic, typically at a lower dose than that tested in this study.

Interference from Endogenous Blood Components:

The Binax NOW® Malaria test was evaluated for possible interference from high levels of endogenous blood components, based on guidelines described in CLSI EP7. EDTA whole blood samples were tested that contained hemoglobin, protein, bilirubin (conjugated and unconjugated) or triglycerides at concentrations above physiological levels. None of the endogenous blood components affected test performance.

Interference from Unrelated Medical Conditions:

To assess the impact of unrelated medical conditions on the specificity of the Binax NOW® Malaria Test, 116 specimens from subjects with a variety of medical conditions unrelated to malaria were tested. Only five (5) of the 116 specimens tested produced a false positive result on the Binax NOW® Test, four (4) from subjects known to be positive for rheumatoid factor and one (1) from a subject with a positive human antimouse antibody (HAMA) titer.

Comparison studies:

a. Method comparison with predicate device:

N/A

  • b. Matrix comparison:

{5}------------------------------------------------

Clinical studies:

a. Clinical sensitivity: Clinical Sample Performance – Binax NOW® Malaria Test Sensitivity & Specificity - Endemic Population:

The performance of the Binax NOW® test was compared to Giemsa malaria microscopy in a multi-center prospective study conducted in 2001 outside the U.S., in regions considered endemic for malaria. A total of 4,122 whole blood specimens collected from patients presenting with malaria-like symptoms were evaluated on the Binax NOW® test. Microscopy was considered positive only when asexual malaria forms were detected, since asexual forms (not gametocytes) are indicative of active infection.

Forty-four percent (1.796/4.122) of the tested population was microscopy positive for malaria, including 557 patients with P.f., 1,187 with P.v., 16 with P.m., 2 with P.o., and 34 with mixed P.f./P.v. infections. Fifty-nine percent of patients were male, 41% female, 19% pediatric (<18 years) and 81% adult (≥18 years). Binax NOW® test performance for detection of the individual malaria species and for mixed P.f./P.v. infections is summarized below.

No differences in Binax NOW® Malaria Test performance were observed based on patient age or gender. Binax NOW® test specificity for P.f. trends slightly lower (89.4%) in the 5% of patients who were on anti-malarial drug therapy, than in patients not receiving therapy (94.4%), but does not achieve statistical significance.

Detection of P.f. Infection

Binax NOW® test sensitivity and specificity for detection of P.f. vs. microscopy is presented below. Sensitivity was evaluated based on the levels of parasitemia (parasites per ul) observed in microscopy.

SENSITIVITY FOR P.f.
Parasitemia Level% Sensitivity95%CI
> 500099.7% (326 /327)98 - 100%
1000 -500099.2% (126 /127)96 - 100%
500 -100092.6% (25 / 27)76 - 99%
100 - 50089.2% (33 / 37)75 - 97%
0-10053.9% (21 / 39)37 - 70%
Overall95.3% (531 /557)93 - 97%

Binax NOW® Malaria Test Sensitivity and Specificity for P.f. vs. Microscopy

{6}------------------------------------------------

SPECIFICITY FOR P.f.
% Specificity95% CI
94.2% (3297 / 3500)93-95%

Detection of P.v. Infection

Binax NOW® test sensitivity and specificity for detection of P.v. vs. microscopy is presented below. Sensitivity was evaluated based on the levels of parasitemia (parasites per ul) observed in microscopy. There were 68 samples generating two Binax NOW® test lines that were microscopy positive for P.v. only. When these samples are included in the true positive calculation, Binax NOW® test sensitivity for overall detection of P.v. increases from 68.9% to 74.6% (886/1,187).

Binax NOW® Malaria Test Sensitivity and Specificity for P.v. vs. Microscopy

SENSITIVITY FOR P.v.
Parasitemia Level% Sensitivity95%CI
> 500093.5% (462 / 494)91 -96%
1000 -500081.0% (277 / 342)76 -85%
500 -100047.4% (37 / 78)36 -59%
100 - 50023.6% (34 / 144)17 -31%
0 - 1006.2% (8 / 129)3 -12%
Overall68.9% (818 / 1187)66 -72%
SPECIFICITY FOR P.v.
% Specificityતે જેન્જી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો
CI
99.8% (2863 /99-
2870)100%

Detection of P.m. and P.o. Infection

Binax NOW® test sensitivity was 43.8% (7/16) for detection of P.m. and 50% (1/2) for detection of P.o. When five P.m. microscopy positive samples that generated two test

{7}------------------------------------------------

lines in the Binax NOW® test are included in the true positive calculation, Binax NOW® test sensitivity for P.m. increases from 43.8% to 75.0% (12/16).

Detection of Mixed P.f./P.v. Infection

Thirty four samples were both P.f. and P.v. positive by microscopy, based on the detection of asexual forms of both species. The Binax NOW® test detected 32 of these samples by generating both test lines, for a sensitivity of 94.1% (95% CI of 81-98%).

b. Clinical specificity:

The performance of the Binax NOW® test was compared to Giemsa malaria microscopy in a prospective study conducted in the eastern US in 2006-2007. One hundred (100) whole blood specimens collected from febrile patients were evaluated on the Binax NOW® test and on microscopy. All 100 samples were negative for malaria on microscopy, and 99 of these samples generated negative Binax NOW® test results, yielding a specificity of 99% (99/100) in this low incidence population. Binax NOW® test specificity versus microscopy is presented below.

Binax NOW® Malaria Test Specificity vs. Microscopy

- / -+ / -% Spec95% CI
P.f.1000100%96-100%
P.v., P.o., P.m.99199%95-100%

N. Proposed labeling:

The labeling is sufficient and it satisfies the requirement of 21 CFR Part 809.10.

WARNING

This test should only be used by laboratories that have or can acquire blood samples containing Plasmodium falciparum for use as a positive control. It is recommended that the level of the positive control used challenge the assay cutoff.

Clinical performance has not been adequately established for P. ovale (P.o.) and P. malariae (P.m.). The user must establish performance characteristics of this test with these Plasmodium species.

The test is not intended for use in screening asymptomatic populations.

O. Conclusion:

The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3402 Plasmodium species antigen detection assays.

(a)
Identification. APlasmodium species antigen detection assay is a device that employs antibodies for the detection of specific malaria parasite antigens, including histidine-rich protein-2 (HRP2) specific antigens, and pan malarial antigens in human whole blood. These devices are used for testing specimens from individuals who have signs and symptoms consistent with malaria infection. The detection of these antigens aids in the clinical laboratory diagnosis of malaria caused by the four malaria species capable of infecting humans:Plasmodium falciparum ,Plasmodium vivax ,Plasmodium ovale , andPlasmodium malariae , and aids in the differential diagnosis ofPlasmodium falciparum infections from other less virulentPlasmodium species. The device is intended for use in conjunction with other clinical laboratory findings.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document:Plasmodium species Antigen Detection Assays.” See § 866.1(e) for the availability of this guidance document.