K Number
DEN060001
Device Name
PHICAL TEST
Manufacturer
Date Cleared
2006-04-26

(34 days)

Product Code
Regulation Number
866.5180
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PhiCal™ test is a quantitative ELISA for measuring, in human stool, concentrations of fecal calprotectin, a neutrophilic protein that is a marker of mucosal inflammation. The PhiCal™ test can be used as an in vitro diagnostic to aid in the diagnosis of inflammatory bowel diseases (IBD): Crohn's disease and ulcerative colitis, and to differentiate IBD from irritable bowel syndrome; when used in conjunction with other diagnostic testing and the total clinical picture.

Device Description

The PhiCal test assay consists of: microtiter plate coated with polyclonal rabbit antibodies for calprotectin (12 strips, 8 wells per strip); alkaline phosphatase labeled rabbit anti-calprotectin IgG in buffer with Proclin 300; substrate in buffer; 20X washing solution; 10X dilution solution with Proclin 300; 5X extraction solution with Proclin 300; 5 vials of calprotectin solution at concentrations of 6.25, 12.5, 25, 50 and 100 ng/mL; a low and a high control.

AI/ML Overview

Acceptance Criteria and Study for the PhiCal™ Test

1. Table of Acceptance Criteria and Reported Device Performance

The submission document primarily focuses on demonstrating the analytical and clinical performance of the PhiCal™ Test. The acceptance criteria are implicitly derived from the established performance metrics and the intended use of the device.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Analytical Performance
Intra-Assay Precision (%CV)Low CV, indicating high repeatability within a single assay run.Single extraction, 20x assay reads: Low matrix: 4.07%, High matrix: 6.17%.12 samples, 6 replicates: Range 2.9% - 14.3%.
Inter-Assay Precision (%CV)Low CV, indicating high reproducibility across different runs.6 stool pools, 20 separate extraction & assay runs: Range 8.9% - 18.1%.10 samples (5+, 5-), 5 extractions, 4 replicates each, 5 different days: Range 5.8% - 20.1%.
Extraction Repeat ReproducibilityLow CV, demonstrating consistent extraction process.Low range sample: 12.60% CV, High range sample: 12.13% CV.
Split Sample Comparison (correlation)High correlation (r²) with a reference laboratory.y = 0.9603x + 9.7691, r² = 0.9618 (comparison with Fagerhol laboratory in Norway).
Aqueous Linearity (R²)High R² and acceptable linearity.y = 0.8771x + 2.8302, R² = 0.9962 (from 6.25 - 100 ng/ml).
Matrix Linearity (R²)High R² and acceptable linearity.y = 1.0569x - 0.1658, R² = 0.9866 (from 7.73 - 114.90 ng/ml).
Accuracy/Recovery (% Recovery)% Recovery within an acceptable range for spiked samples.Recoveries ranged from 99.1% to 118.6%.
Analyte StabilityStable under recommended storage and transport conditions.Calprotectin levels remained relatively consistent for up to 11 days under variable temperature conditions simulating specimen transport (2-8°C, 37°C, room temperature). Extract stability also demonstrated at -20°C for 11 days.
Minimum Detection Limit (MDL)Result below the functional sensitivity of the assay.Determined by 20 replicates of assay buffer, resulted in MDT below functional sensitivity.
Functional Sensitivity (%CV)%CV < 20% at the low concentration range.<20% CV for 12.5 ng/mL, 6.25 ng/mL, and 3.125 ng/mL across 6 days. Limited to 6.25 ng/mL (15.6 mg calprotectin/kg stool) for reporting.
Analytical SpecificityNo interference from common bacteria or oral pharmaceuticals.No interference noted from E. coli, C. freundii, K. pneumoniae, Salmonella, Shigella, Yersinia, or various oral pharmaceuticals (Prednisone, Sulfamethoxazole, Pentasa, Prevacid, Vancomycin, Asacol, Azathioprine, Ciprofloxacin HCL, Ferrous Sulfate, Multiple vitamin, Vitamin E, Zelnorm). Minimal interference from gastrointestinal bleeding (increase of no more than 6 mg/L for 100 mL/day bleeding).
Clinical Performance
Overall Clinical Sensitivity (at >120 mcg/gm cut-off)High sensitivity for detecting IBD/Inflamed.86% (254/296) when borderline cases are excluded (n=132).
Overall Clinical Specificity (at >120 mcg/gm cut-off)High specificity for differentiating IBS/Normal.94% (451/480) when borderline cases are excluded (n=132).
ROC Analysis (at 120 µg/g cut-off)Acceptable sensitivity and specificity.95% specificity and 70% sensitivity at 120 µg/g.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Performance (Test Set): 908 subjects (total cohort used for clinical sensitivity and specificity).
    • IBD: 255
    • IBS: 410
    • Other bowel diseases: 82
    • Normal subjects: 161
  • Data Provenance: Not explicitly stated regarding country of origin or specific study sites. The Split Sample Comparison Study mentions the "Fagerhol laboratory in Norway" as a reference for comparison, suggesting at least some interaction with data or expertise from Norway. The clinical study itself is presented as a single comprehensive cohort without geographical breakdown. Given the nature of a 510(k) submission for a new device, it is likely a retrospective analysis of collected samples for this initial submission, or a prospective collection for this specific study. The document does not specify "retrospective" or "prospective."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts used to establish the ground truth for the clinical test set. The clinical study categorizes subjects into "IBD/Inflamed 'Organic'" and "IBS/Normal 'Non-Organic'". The diagnosis of these conditions (IBD, IBS, other bowel diseases, normal) would typically be established by Gastroenterologists based on a combination of clinical history, endoscopic findings, histopathology, imaging, and other diagnostic tests. However, the exact process or number/qualifications of experts involved in the classification of these 908 subjects (which serves as the "ground truth" for the device's clinical performance) are not detailed in this summary.

Regarding the calibrators reference standard, it mentions comparison to the "Gold Standard" preparation provided by Dr. Magne Fagerhol, Norway. Dr. Fagerhol is a prominent researcher in the field of calprotectin, implying his expertise in establishing this "gold standard."

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for establishing the diagnostic classifications of the 908 subjects. The diagnoses used as ground truth are presented as given, presumably established via standard diagnostic clinical practice.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The device is a quantitative ELISA test, not an imaging-based AI system that would typically involve human readers. Therefore, the concept of human readers improving with AI vs. without AI assistance is not applicable in this context.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance study was done. The entire submission describes the performance of the PhiCal™ test as a "quantitative ELISA for measuring... fecal calprotectin," which is an in vitro diagnostic device. The clinical sensitivity and specificity are reported based on the device's measurements compared to the established clinical diagnoses (ground truth) of the 908 subjects. This is the definition of standalone performance for this type of diagnostic test – the assay provides a result without direct human interpretation in the loop that influences the result itself.

7. Type of Ground Truth Used

The ground truth for the clinical performance evaluation (sensitivity and specificity) was based on clinical diagnoses. Subjects were classified as "IBD/Inflamed 'Organic'," "IBS/Normal 'Non-Organic'," or "other bowel diseases" and "normal subjects." These classifications would typically involve a combination of:

  • Clinical outcomes data: Patient symptoms, disease course.
  • Pathology: Biopsy results (especially for IBD).
  • Endoscopy: Visual assessment of the bowel lining.
  • Other diagnostic testing: Blood tests, imaging, etc.
  • Expert consensus: The overall clinical picture, as defined by medical professionals.

8. Sample Size for the Training Set

The document does not explicitly delineate a "training set" in the context of traditional machine learning where an algorithm is explicitly trained on data. This is an ELISA assay, not an AI/ML model. The "training" for such an assay primarily involves:

  • Assay Development and Optimization: Iterative laboratory work to establish optimal reagent concentrations, incubation times, standard curves, etc. The document mentions "Establishing the cut-off for the assay was done by testing 124 normal sera." This group of 124 normal sera was likely used to help define the initial cut-off, which could be considered a form of "training" or optimization data for the clinical interpretation.
  • Calibrators: The 5 calibrators used in each assay run (6.25, 12.5, 25, 50, and 100 ng/mL) are integral to "training" the curve from which sample concentrations are derived.

9. How the Ground Truth for the Training Set Was Established

As noted above, for an ELISA assay, the concept of a "training set" and associated ground truth is different from AI/ML.

  • For the 124 normal sera used to establish the cut-off: The "ground truth" would be that these individuals were clinically healthy or considered "normal," likely confirmed through absence of symptoms and standard clinical assessment.
  • For the Calibrators: The ground truth for the concentrations of the calibrators (6.25, 12.5, 25, 50, and 100 ng/mL) is established by careful gravimetric and volumetric preparation and validated against the internal primary reference, which itself is compared to the "Gold Standard" preparation provided by Dr. Magne Fagerhol, Norway. This "Gold Standard" essentially defines the true concentration of calprotectin.

{0}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

  • A. 510(k) Number: K050007
  • B. Purpose for Submission: New device.
  • C. Measurand: Fecal calprotectin
  • D. Type of Test: Quantitative, ELISA
  • E. Applicant: Genova Diagnostics
  • F. Proprietary and Established Names: PhiCal™ Test

G. Regulatory Information:

    1. Regulation section: None
    1. Classification: De novo
    1. Product Code: NXO, Calprotectin, fecal
    1. Panel: Immunology (82)

H. Intended Use:

    1. Intended use(s):
    • The PhiCal™ test is a quantitative ELISA for measuring, in human stool, concentrations of fecal calprotectin, a neutrophilic protein that is a marker of mucosal inflammation. The PhiCal™ test can be used as an in vitro diagnostic to aid in the diagnosis of inflammatory bowel diseases (IBD): Crohn's disease and ulcerative colitis, and to differentiate IBD from irritable bowel syndrome.
    1. Indication(s) for use:

The PhiCal™ test is a quantitative ELISA for measuring, in human stool, concentrations of fecal calprotectin, a neutrophilic protein that is a marker of mucosal inflammation. The PhiCal™ test can be used as an in vitro diagnostic to aid in the diagnosis of inflammatory bowel diseases (IBD): Crohn's disease and ulcerative colitis, and to differentiate IBD from irritable bowel syndrome; when used in conjunction with other diagnostic testing and the total clinical picture.

    1. Special condition for use statement(s): The device is for prescription use only.
    1. Special instrument requirements: ELISA reader (405 nm filter), digital scale (40-150 mg), vortex mixer, shaker, micro-centrifuge
  • I. Device Description:

{1}------------------------------------------------

The PhiCal test assay consists of: microtiter plate coated with polyclonal rabbit antibodies for calprotectin (12 strips, 8 wells per strip); alkaline phosphatase labeled rabbit anti-calprotectin IgG in buffer with Proclin 300; substrate in buffer; 20X washing solution; 10X dilution solution with Proclin 300; 5X extraction solution with Proclin 300; 5 vials of calprotectin solution at concentrations of 6.25, 12.5, 25, 50 and 100 ng/mL; a low and a high control.

J. Substantial Equivalence Information:

    1. Predicate device name(s): Not applicable
    1. Predicate K number(s): Not applicable
    1. Comparison with predicate:
Similarities
ItemDevicePredicate
Differences
ItemDevicePredicate

K. Standard/Guidance Document Referenced (if applicable):

None referenced

L. Test Principle:

The test is performed on random stool samples, collected without preservatives. The samples should be tested within 7 days, or frozen at -20° C until tested. An extract is prepared by combining approximately 0.1 gm of stool with 5 mL of extraction buffer and mixing for 30 minutes. Following centrifugation, 20uL of the supernatant is diluted 1:50 with dilution buffer (final dilution 1:2500).

The assay uses a polyclonal rabbit antibody against calprotectin as the capture antibody in an enzyme linked immunosorbent assay system. Calprotectin present in the diluted sample is bound by the antibody adsorbed onto the surface of the microtiter plate. The enzyme conjugated antibodies (rabbit IgG antibodies against calprotectin) bind to the captured antigen. A substrate (alkaline phosphatase) is added and subsequently the enzyme catalyses the conversion of the substrate to a colored product. The intensity of the color is proportional to the amount of conjugate bound, and thus to the amount of captured calprotectin. The concentration of calprotectin in the samples is interpreted from a standard curve using 5 calibrators.

M. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • i. Precision/Reproducibility:

Intra-Assay precision was determined by extracting one low and one high matrix sample, and assaying each extract twenty times within a single assay run. Precision is calculated as the % CV obtained for each level. Respective CV's were 4.07% and 6.17%. Additionally 12 samples were

{2}------------------------------------------------

assayed in 6 replicates within a singly assay run. Range of %CV was
2.9% to 14.3%. The resultant %CVs demonstrates that the PhiCal Test is
precise in the reportable range.
Calprotectin Intra-Assay Precision ( g/g)
123456789101112
AnalysisMean177.555.3229.431.196.9106.817.518.720.8182.855.660.2
StdDev5.76.66.72.311.59.40.82.71.910.83.92.5
%CV3.212.02.97.411.98.84.714.39.25.97.04.1

Inter-assay precision was assessed by evaluating results for six different pools of stool. Points are from 20 separate extraction and assay runs. The %CV range was 8.9% to 18.1%. Ten different samples (5 positive, 5 negative) were each extracted 5 separate times from individual stool aliquots. Each extract was then assayed in 4 replicates on 5 separate EIA runs performed on 5 different days. The %CV range was 5.8% to 20.1%. These results demonstrate the assav is reproducible within acceptable limits along the reportable range of the assay.

Calprotectin Inter-Assay Precision ( g/g)
AnalysisReplicateS1S2S3S4S5S6S7S8S9S10
Mean36.912.916.620.226.9131.7188.289.8101.163.3
Std Dev2.12.12.50.92.010.236.910.413.312.7
%CV5.816.714.94.47.67.719.611.613.220.1

Extraction Repeat Reproducibility: Two samples that represent the low and high ends of the reportable range were tested. Each was extracted 24 times, and each extract was tested. Precision was evaluated according to the %CV obtained for each level. Respective CV's of 12.60% and 12.13% for the low and high levels demonstrate that results are reproducible along the reportable range.

Split Sample Comparison Study:

Forty stool samples that spanned the reportable range of the test were assayed both by GENOVA and by the Fagerhol laboratory in Norway. The forty samples were assayed in small batches over 5 days to introduce day-to-day variability and to minimize any systematic error that might occur in a single run. The comparison plot showed y = 0.9603x + 9.7691, r- = 0.9618.

ii. Linearity/assay reportable range:

To confirm the reportable range of the PhiCal™ assay, standard curves were generated for both aqueous and matrix linearity. The results with serially diluted aqueous calprotectin in triplicate showed acceptable

{3}------------------------------------------------

linearity from 6.25 - 100 ng/ml. The following equation was obtained: Standard curve Aqueous, y = 0.8771x + 2.8302, R2=0.9962.

Image /page/3/Figure/2 description: The image is a graph titled "Calprotectin Aqueous Linearity 2-26-03". The graph plots "Observed ng/ml" on the y-axis versus "Theoretical ng/ml" on the x-axis. A black line is plotted through the data points, and the equation of the line is "y = 0.8771x + 2.8302" with an R-squared value of 0.9962.

The results with serially diluted fecal calprotectin in triplicate showed acceptable linearity and accuracy from 7.73 - 114.90 ng/ml. The following equation was obtained: Standard curve Matrix, y = 1.0569x -0.1658, R2=0.9866.

Image /page/3/Figure/4 description: The image is a scatter plot titled "Matrix Linearity Calprotectin". The x-axis is labeled "Theoretical Value ng/ml", and the y-axis is labeled "Observed Value ng/ml". The plot shows a linear relationship between the theoretical and observed values, with the equation of the line being y = 1.0569x - 0.1658 and R^2 = 0.9866.

Accuracy/ Recovery

Extracts from five different stool samples were each spiked with calprotectin. The calprotectin used for the spike was obtained from established serum pools. The baseline extract for each sample was "spiked" with assay buffer to compensate for volume adjustments made to

{4}------------------------------------------------

Calprotectin Recovery Data
# 1a# 2a# 3a# 4a# 5a
Baseline (ug/g)49.173.466.46.473.6
Spike Value (ug/g)58.557.655.2રરું રે59.0
Theoretical (Base + Spike) (ug/g)107.6131.0121.761.7132.7
Observed (Base + Spike) (ug/g)106.7129.9144.366.1140.0
% Recovery99.299.1118.6107.2105.5

the calprotectin-spiked extracts. Each of the spiked extracts was then assayed per kit protocol. Data is shown in table below.

iii. Traceability, Stability, Expected values (controls, calibrators, or methods):

Calibrators:

The primary internal reference is calibrated by comparison to the "Gold Standard" preparation provided by Dr. Magne Fagerhol, Norway.

Analyte Stability:

Studies were conducted to determine the stability of calprotectin in neat stool and in stool extracts. To determine stability in neat stool, several temperature conditions were evaluated: stool collected fresh and subsequently stored at 2-8°C: stool collected fresh and then stored at -20°C; and stool collected fresh and subjected to variable temperatures. Stability in stool extract was evaluated by extracting fresh samples and then storing aliquots of the extract at -20°Cm where a different aliquot was thawed for each analysis over 11 days.

For the variable temperature stability study the collected samples were stored at 2-8°C for 24 hours, after which they were stored in an incubator at 37°C for approximately 21 hours. Once removed from the incubator they were left at room temperature for approximately 5½ hours then stored at 2-8°C for the remainder of the study. This study concluded that calprotectin levels stay relatively consistent (no trends observed) for up to 11 days after collection, under conditions that may be experienced during specimen transport. The studies support the specimen collection, transport and storage recommendations in the package insert.

  • iv. Detection limit (functional sensitivity):

Minimum Detection Limit:

The minimum detection limit (MDT) was determined by running 20 replicates of the assay buffer (treated as patient) on one assay plate. Two standard deviations were added to the mean OD reading, and this value was plugged into the calibration curve to arrive at the MDT. The result was below the functional sensitivity of the assay.

Functional sensitivity:

The functional sensitivity was estimated by determining a precision profile at the low concentration range and then selecting the concentration at

{5}------------------------------------------------

which a 20% CV is obtained. To establish the functional sensitivity for the assay, 3 concentrations of kit standards, 12.5 ng/mL, 6.25 ng/mL and 3.125 ng/mL (6.25 ng/mL standard diluted 1:2 with dilution buffer), were assayed in either triplicate or quadruplicate over 6 days. %CV for all three concentrations were <20%. For purposes of result reporting, the low end of the reportable range will be limited to the lowest calibrator, 6.25ng/mL. This corresponds to 15.6 mg calprotectin/kg stool after conversion at the typical sample dilution of 1:2500.

  • Analytical specificity: V.
    Microorganisms: The following bacteria, which occur frequently in the stool, or are common to infectious diarrhea were evaluated as potential interfering factors – Escherichia coli, Citrobacter freundii, and Klebsiella pneumoniae, Salmonella, Shigella, Yersinia. The presence of these bacteria in stool samples does not interfere with the PhiCal Test.

Oral Pharmaceuticals: The following oral pharmaceuticals and nutritional supplements were tested for potential interference: Prednisone; Sulfamethoxazole, Pentasa; Prevacid; Vancomycin; Asacol; Azathioprine; Ciprofloxacin HCL; Ferrous Sulfate; Multiple vitamin; Vitamin E; Zelnorm. The concentration of the drug in the suspension was sufficient to approximate the concentration to be expected in a patient's stool based on the appropriate dosages and average stool volume per twenty four hours. No interference was noted.

Gastrointestinal Bleeding: Bleeding of as much as 100 mL per day would increase the fecal calprotectin concentration by no more than 6 mg/L (15 ug/g).23

  • vi. Assay cut-off: See clinical cut-off.
    1. Comparison studies:
    • Method comparison with predicate device: i. Not applicable.
    • ii. Matrix comparison: Not applicable.
    1. Clinical studies:
    • Clinical sensitivity and specificity: a.

Fecal calprotectin results from the complete cohort of 908 subjects were used to calculate the overall clinical sensitivity and specificity of the PhiCal test. The 908 subjects included IBD n=255, IBS n=410, other bowel diseases n=82, IBD, and normal subjects n=161. The results and calculations are shown below:

N = 908IBD/ Inflamed"Organic"IBS/ Normal"Non-Organic"Totals
>120 mcg/gm25429283
50-120 mcg/gm4389132

{6}------------------------------------------------

<50 mcg/gm42451483
Totals296+43480+89$776+132 = 908$

When borderline cases (n=132) are excluded from the calculations: Clinical sensitivity: 86% (254/296) Clinical specificity: 94% (451/480)

  • b. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
    1. Clinical cut-off:

Establishing the cut-off for the assay was done by testing 124 normal sera. The mean value was 40μg/g, and with a standard deviation of 40μg/g, the upper cut point was 120μg/g. In addition, ROC analysis using results from the total 908 patients showed that 120 µg/g yielded 95% specificity and 70% sensitivity.

Image /page/6/Figure/6 description: The image is a Receiver Operator Curve (ROC) for Calprotectin. The y-axis represents sensitivity, ranging from 0 to 100, while the x-axis represents 100-Specificity, also ranging from 0 to 100. The ROC curve starts near the top of the y-axis and gradually approaches the top right corner, indicating the trade-off between sensitivity and specificity for the Calprotectin test. A dashed diagonal line represents the line of no discrimination.

Calprotectin ConcentrationInterpretationFollow-Up
<15.625 – 50 µg/gNormalNone
50 – 120 µg/gBorderlineRe-evaluate @ 4-6 weeks
>120 µg/gAbnormalRepeat as clinically indicated

{7}------------------------------------------------

    1. Expected values/Reference range:
      The expected value in the normal population is <50 µg/g. The highest levels are found in Crohn's disease and ulcerative colitis with levels as high as five to several thousand times the upper limit for healthy individuals. Calprotectin is generally not elevated in healthy subjects or in irritable bowel syndrome but levels can overlap with those of organic disease.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The petition for Evaluation of Automatic Class III Designation for this device is accepted. The device is classified as Class II under regulation 21 CFR 866.5180 with special controls. The special control guidance document, "Class II Special Controls Guidance Document: Fecal Calprotectin Immunological Test Systems" will be available shortly.

§ 866.5180 Fecal calprotectin immunological test system.

(a)
Identification. A fecal calprotectin immunological test system is anin vitro diagnostic device that consists of reagents used to quantitatively measure, by immunochemical techniques, fecal calprotectin in human stool specimens. The device is intended forin vitro diagnostic use as an aid in the diagnosis of inflammatory bowel diseases (IBD), specifically Crohn's disease and ulcerative colitis, and as an aid in differentiation of IBD from irritable bowel syndrome.(b)
Classification. Class II (special controls). The special control for these devices is FDA's guidance document entitled “Class II Special Controls Guidance Document: Fecal Calprotectin Immunological Test Systems.” For the availability of this guidance document, see § 866.1(e).