K Number
DEN050003
Manufacturer
Date Cleared
2005-04-18

(49 days)

Product Code
Regulation Number
866.4070
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PAXgene™ Blood RNA System consists of a blood collection tube (PAXgene™ Blood RNA Tube) and nucleic acid purification kit (PAXgene™ Blood RNA Kit). It is intended for the collection, storage, and transport of blood and stabilization of intracellular RNA in a closed tube and subsequent isolation and purification of host RNA from whole blood for RT-PCR used in molecular diagnostic testing.

Performance characteristics for the PAXgene™ Blood RNA System have only been established with "cfos and IL1B." The user is responsible for establishing appropriate PAXgene™ Blood RNA System performance characteristics for other target transcripts.

Device Description

The PAXgene™ Blood RNA System consists of:

  • the PAXgene™ Blood RNA tubes and .
  • the PAXgene™ Blood RNA kit. .

The PAXgene™ Blood RNA tube is of a sterile, plastic, evacuated blood collection tube containing stabilization solution (tetradecyl trimethyl-ammonium oxalate and tartaric acid. These components serve to lyse cells, protect RNA molecules from degradation by ribonucleases (RNases) and prevent induction of gene expression.

The kit consists of 5 aqueous buffer solutions for resuspending, binding, washing, and eluting RNA, RNase-free water, proteinase K, an RNase-Free DNase set, spin columns, microcentrifuge tubes, processing tubes, and secondary blood collection tube closures.

AI/ML Overview

The provided text details the performance characteristics and studies for the PAXgene™ Blood RNA System, intended for the collection, storage, and purification of intracellular RNA from whole blood for RT-PCR.

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance
RNA Yield≥ 3 µg/tube (A260 nm)Achieved (specific values not always given, but "all samples... fulfilled the acceptance criteria" and "in the expected ranges")
RNA PurityA260/A280 ratio between 1.8 and 2.2Achieved ("all samples... fulfilled the acceptance criteria" and "in the expected ranges")
Integrity (CFOS/18S rRNA)CFOS within 2.34 CTAchieved ("within the acceptance range")
Integrity (IL1B/18S rRNA)IL1B within 1.94 CTAchieved ("within the acceptance range")
Genomic DNA (gDNA) ContaminationPercentage of gDNA in total nucleic acid preparationAll samples "fulfilled the acceptance criteria"
RT-PCR InhibitionKit components do not introduce inhibition for an RT-PCR assay"All samples demonstrated that the kit components do not introduce any inhibition"
Reproducibility (% CV)Within-run, inter-user, inter-lotVaries: - Repeatability (quadruplicate per donor pool, per lot/user): Min/max % CV 3.4-28.8, overall 11.9% - Reproducibility (per user, between lots): Min/max % CV 8.7-20.1, overall 14.9% - Reproducibility (between lots and users): Min/max % CV 8.7-23.1, overall 16.4%
PAXgene Tube Stability (25°C)Draw volume, liquid additive volume, closure performance, pH, conductivity, density, RNA yield, purity, CFOS/18S rRNA & IL1B/18S rRNA stabilityAll physical/chemical attributes and functional performance were within expected ranges for 6 months (sponsor claims 19 months based on accelerated studies)
PAXgene Kit Component StabilityRecovery of input RNA, variability of recovery, degree of inhibition, pH, conductivity, density, Proteinase K & DNase I activity, bioburdenAll parameters in expected ranges for 6 months real-time storage; no bioburden; acceptable performance after extreme temperature simulation
RNA In Situ Stability (various temperatures/times/freeze-thaw)RNA yield, purity, integrity (CFOS/18S rRNA, IL1B/18S rRNA) within acceptance rangeAll samples at all time points, temperatures, and freeze/thaw cycles were within the acceptance range (with one unspecified exception)
Linearity/Reportable Range (CFOS)0.2 to 7.8 ng total input nucleic acid; up to 25% DNA without affecting performanceMet
Linearity/Reportable Range (IL1B)0.1 to 7.8 ng total input nucleic acid; up to 5% DNA without affecting performanceMet

2. Sample Sizes and Data Provenance

  • Test Set (Precision/Reproducibility of RT-PCR Assays):
    • K2EDTA samples: 5 donors, 5 tubes/donor (25 tubes total), 10 ml blood/tube. RNA isolated via OIAzol and RNeasy, pooled, concentrated.
    • PAXgene samples: 48 donors, 8 tubes/donor (384 PAXgene tubes total), 2.5 ml blood/tube. RNA purified, DNase treated, concentrated, pooled.
    • Reproducibility Design: 3 different component lots, 3 different experimenters, 3 different laboratories (equipment), on 3 different days.
      • Each "run" (9 runs total) involved: 2 RNA samples (K=calibrator, T=test), each with 10 replicates. This generated 180 ΔCT values and 90 ΔΔCT values.
  • Test Set (Precision/Reproducibility of PAXgene Blood RNA System - Experiment 1):
    • 14 donors (WBC: 4.8-11.0 x 10^6 cells/ml blood)
    • 12 tubes/donor (168 PAXgene tubes total)
  • Test Set (Precision/Reproducibility of PAXgene Blood RNA System - Experiment 2):
    • 30 donors (WBC: 4.8-11.0 x 10^6 cells/ml blood)
    • 12 tubes/donor (360 PAXgene tubes total)
    • Blood from 3 donors pooled and aliquoted into empty tubes to create 10 donor pools with 36 tubes each.
  • Test Set (RNA Purity, DNA Contamination, RNA Yield):
    • 10 donors
    • 2 PAXgene tubes/donor (20 tubes total) + 1 EDTA tube/donor (10 EDTA tubes) for WBC count.
  • Test Set (RT-PCR Inhibition):
    • 22 blank eluates from columns.
    • HeLa cell RNA used as template.
  • Test Set (PAXgene Tube Stability):
    • Functional tests: 60 tubes/time point (TTP) for accelerated studies (120 tubes total), 60 tubes/TTP for real-time study (810 tubes scheduled, 6 months data presented).
    • Functional performance: 10 donors per time point for RNA yield, purity, CFOS/IL1B relative levels (3 time points for blood-filled tubes storage at 18-25°C x 2 preparations = 60 measurements).
  • Test Set (PAXgene Kit Component Stability):
    • Two reagent sets per time point for functional stability (4 blank eluates, 8 RNA eluates).
  • Test Set (RNA In Situ Stability):
    • Multiple experiments with 10 donors each. Number of PAXgene tubes per donor varied (6, 10, 12, 14 tubes/donor), resulting in totals of 60, 100, 120, and 140 tubes for different conditions.
  • Data Provenance: The studies appear to be prospective, performed by the applicant (PreAnalytiX GmbH) to demonstrate the device's performance. The applicant is based in Switzerland (GmbH). The data is generated in a lab setting, not clinical real-world data from multiple countries.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of experts to establish a ground truth for the "test set" in the context of diagnostic interpretation (e.g., radiologists evaluating images). This device is a sample collection and purification system, not an interpretative diagnostic device. The ground truth for its performance is instead based on quantitative measurements of RNA yield, purity, integrity, and the absence of inhibitors. These are objective measures determined by laboratory instrumentation (e.g., spectrophotometers, Q-RT-PCR machines, BioAnalyzer).

However, "experimenters" and "technicians" are mentioned:

  • Precision/Reproducibility of O-RT-PCR Duplex Assays: "3 different experimenters"
  • Precision/Reproducibility of PAXgene Blood RNA System (both Experiments 1 & 2): "3 technicians"

Their specific qualifications (e.g., years of experience) are not provided, but they are implied to be trained laboratory personnel.

4. Adjudication Method

Not applicable. The "ground truth" for this device is based on objective, quantifiable laboratory measurements (e.g., RNA yield, purity ratios, Ct values), not on subjective expert interpretation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. An MRMC comparative effectiveness study is not applicable for this type of device (RNA collection and purification system). Such studies are typically used for diagnostic imaging or interpreted assays where human readers are involved in making diagnostic decisions, and the AI aims to assist or replace them. This device is a sample preparation tool, and its "performance" is measured analytically, not by human reader accuracy.

6. Standalone (Algorithm Only) Performance

Yes, the studies described are essentially "standalone" performance evaluations of the device and its components. The PAXgene Blood RNA System's analytical performance (e.g., RNA yield, purity, integrity) is assessed directly through laboratory assays, without human intervention in the "decision-making" loop. Its function is to prepare RNA for downstream molecular diagnostic testing (RT-PCR), meaning its performance is evaluated on its ability to produce high-quality RNA suitable for such assays. The Q-RT-PCR assays run on the RNA extracted by the system are designed to evaluate the RNA quality produced by the device, not to perform a diagnosis.

7. Type of Ground Truth Used

The ground truth used is primarily based on analytical measurements derived from standard laboratory techniques:

  • Spectrophotometric measurements: Absorbance at 260 nm (for RNA yield), A260/A280 ratio (for RNA purity).
  • Quantitative Reverse Transcription-Polymerase Chain Reaction (Q-RT-PCR) assays: Using CFOS and IL1B transcripts normalized to 18S rRNA (for RNA integrity and relative transcription levels). This involves comparing threshold cycle (Ct) values.
  • Electropherograms (BioAnalyzer): For RNA purity and integrity.
  • Beta-actin PCR: To determine genomic DNA contamination.
  • Biological assays: Proteinase K and DNase I activity tests.
  • Physical and chemical property tests: pH, conductivity, density, draw volume, liquid additive volume, closure performance.
  • Microbiological assays: Bioburden analysis.

8. Sample Size for the Training Set

The document does not explicitly describe a separate "training set" in the context of machine learning or AI algorithms, as this is not an AI-based device. Instead, the studies detail the performance validation of the system. The various studies use different cohorts of human donors for the collection of blood samples to assess different aspects of the device's performance (e.g., 5 donors for K2EDTA comparator, 48 donors for PAXgene in initial RT-PCR assay validation, 14 and 30 donors for system precision/reproducibility, 10 donors for purity/DNA contamination, 10 donors for each in situ stability condition).

9. How the Ground Truth for the Training Set Was Established

As mentioned above, there is no "training set" in the AI sense for this device. The establishment of "ground truth" for the various performance metrics (RNA yield, purity, integrity, etc.) is through well-established and standardized laboratory analytical methods as described in point 7. These methods themselves are considered the "gold standard" for measuring these specific biochemical and molecular parameters.

{0}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

  • A. 510(k) Number: K042613
  • B. Purpose for Submission: Clearance of new device
  • C. Measurand: Ribonucleic acid (RNA)

D. Type of Test:

Collection, stabilization, and purification of intercellular RNA from whole blood for RT-PCR use in molecular diagnostic testing.

E. Applicant:

PreAnalytiX GmbH

F. Proprietary and Established Names:

PAXgene™ Blood RNA System

G. Regulatory Information:

    1. Regulation section: 21 CFR §866.4070, RNA Preanalytical Systems
    1. Classification: Class II (de novo)
    1. Product code: NTW
    1. Panel: Immunology (82)

H. Intended Use:

    1. Intended use(s):
      The PAXgene™ Blood RNA System consists of a blood collection tube (PAXgene™ Blood RNA Tube) and nucleic acid purification kit (PAXgene™ Blood RNA Kit). It is intended for the collection, storage, and transport of blood and stabilization of intracellular RNA in a closed tube and subsequent isolation and purification of host RNA from whole blood for RT-PCR used in molecular diagnostic testing.

Performance characteristics for the PAXgene™ Blood RNA System have only been established with "cfos and IL1B." The user is responsible for establishing appropriate PAXgene™ Blood RNA System performance characteristics for other target transcripts.

{1}------------------------------------------------

    1. Indication(s) for use: See intended use above
    1. Special conditions for use statement(s): For professional use only
    1. Special instrument requirements: General laboratory equipment (centrifuges, etc.)

I. Device Description:

The PAXgene™ Blood RNA System consists of:

  • the PAXgene™ Blood RNA tubes and .
  • the PAXgene™ Blood RNA kit. .

The PAXgene™ Blood RNA tube is of a sterile, plastic, evacuated blood collection tube containing stabilization solution (tetradecyl trimethyl-ammonium oxalate and tartaric acid. These components serve to lyse cells, protect RNA molecules from degradation by ribonucleases (RNases) and prevent induction of gene expression.

The kit consists of 5 aqueous buffer solutions for resuspending, binding, washing, and eluting RNA, RNase-free water, proteinase K, an RNase-Free DNase set, spin columns, microcentrifuge tubes, processing tubes, and secondary blood collection tube closures.

J. Substantial Equivalence Information:

    1. Predicate device name(s): None
    1. Predicate 510(k) number(s): None
    1. Comparison with predicate: Not applicable

K. Standard/Guidance Document Referenced (if applicable);

Not applicable

L. Test Principle:

Blood (2.5 mL) is collected into the PAXgene™ Blood RNA tube by venipuncture. It can be either stored or immediately processed (according to specifications in the labeling). To isolate RNA from the sample, the blood is centrifuged to sediment the nucleic acids and the pellet is resuspended in and washed with RNase-free water. The resulting pellet is dissolved in buffer BR1 and buffer BR2, and proteinase K is added for the enzymatic digestion of proteins. Genomic DNA is sheared by running the sample through a shredder spin column, and ethanol is added to adjust the binding conditions. RNA is selectively bound to the membrane in the PAXgene™ Blood RNA spin column by a short centrifugation. The bound RNA is washed, and residual genomic DNA is removed from the membrane column by treatment of the bound nucleic acid with DNase I. RNA is eluted from the column using the elution buffer

{2}------------------------------------------------

provided.

M. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision/Reproducibility:

CFOS/18S rRNA and IL1B/18S rRNA Q-RT-PCR Duplex assays.

Objective: The sponsor submitted validation and performance data for two uncleared O-RT-PCR assays for the assessment of performance of the PAXgene Blood RNA System in comparison to a K2EDTA blood collection tube and acid phenol RNA extraction. The assays determine quantitativly the relative level of the c-fos and IL-1ß transcripts compared to 18S rRNA in RNA samples derived from human whole blood. Since CFOS and IL1B offer no diagnostic information, these assays are not intended for diagnostic use. They are ONLY intended for demonstrating that the PAXgene Blood RNA System can be used for RT-PCR.

The validation study uses two types of samples:

K2EDTA (TOTAL: 250 ml blood):

  • 5 donors with 5 tubes per donor ●
  • 10 ml blood per EDTA tube ●
  • RNA was isolated via OIAzol and RNeasy columns, pooled and concentrated via MinElute spin column
  • Purified, concentrated RNA was analyzed for yield and purity, aliquoted, and stored at -20 °C.

PAXgene (TOTAL: 960 ml blood):

  • 48 donors, 8 tubes per donor (total of 384 PAXgene tubes) ●
  • 2.5 ml blood per PAXgene tube
  • tubes were incubated at RT for 2 hrs
  • RNA was purified, DNase treated, concentrated via MinElute spin ● column and pooled.
  • Purified, concentrated RNA was analyzed for yield and purity, aliquoted, and stored at -20 °C.

Experimental design of validation experiments:

The sponsor addressed repeatability and reproducibility and defines these terms within the document as follows:

The repeatability of the assay (within-run precision or intra-assay variability), is the variability of the results from repeated measurements:

  • using the same two source samples (n = 10 for each sample) .
  • . in the same laboratory (same equipment and same experimenter),
  • on the same day, ●
  • with the same reagents (identical lots) and ●

{3}------------------------------------------------

  • . in the same run.
    The repeatability of individual runs is expressed as the weighted mean repeatability used for the calculation of overall precision. The weighted mean repeatability is determined and is included in the overall precision calculation.

The reproducibility of the assay, (robustness, run-to-run precision, or inter-assay variability), is the variability of results from repeat measurements

  • using the same two source samples .
  • in different laboratories (different lots, different equipment and . different experimenters)
  • . on different days
  • . and in different runs.

Therefore, reproducibility investigations involve repeats of the complete experiment in different laboratories. To establish the suitability of the precision for day-to-day work, different lots of reagents are investigated as additional parameters that may have a negative impact on precision. The reproducibility between laboratories is determined, but also not specified in detail, because it is included in the overall precision calculation and therefore will be assessed as its part.

Combi-nation1)ReagentCombinationRun No. (k)Run2)RNAsample3)Replicatesper RNAsampleNumber ofΔCTvalues per RNAsampleNumber ofΔΔCT values per run
A1A1K101010
T101010
2A2K101010
T101010
3A3K101010
T101010
B4B1K101010
T101010
5B2K101010
T101010
6B3K101010
T101010
C7C1K101010
T101010
8C2K101010
T101010
9C3K101010
T101010
Total18090

Table 6: Summary of measurement runs for the determination of overall precision

Combination of experimenter, laboratory (equipment), day, TagMan plate

21 Measurement series (run), Established by reagent lots within the combination (A. B. C)
determined by reagent lots 3 K = calibrator RNA, T = test RNA

{4}------------------------------------------------

Note: Under 1, technician-to-technician, lab-to-lab, and day-to-day variability has been combined in this study.

Results: The sponsor performed an appropriate validation of their submitted assay by using 3 different component lots, in the hands of 3 different experimenters, in 3 different laboratories (equipment), on 3 different days. The sponsor provided all raw CT data and demonstrated the performance of their assays within the expected limits.

Precision/Reproducibility of the PAXgene Blood RNA System.

Objective: Determination of the repeatability, reproducibility and reliability of the PAXgene Blood RNA System.

Two experiments were performed: Experiment 1:

  • 14 donors (WBC counts within 4.8 to 11.0 x 10° cells/ml blood) ●
  • 12 tubes per donor (total 168 PAXgene tubes)
  • PAXgene Blood RNA system Lot No. VL1 (ONLY one lot!)
  • 3 technicians. Each technician processed 4 tubes per subject (total of 56 ● tubes per technician) in 3 runs (Run #1: donor 1-3 (12 tubes); Run #2: donor 4-9 (24 tubes); Run #3: donor 10-14 (20 tubes)

RNA vield (absorbance at 260 nm) and purity (260nm/280nm absorbance ratio) were determined for each sample.

Repeatibility was determined by evaluating the vield and purity of four samples from one donor processed by one technician.

Reproducibility was determined by evaluating the yield and purity of 12 samples from one donor processed by three technicians. The repeatability and reproducibility was determined separately for each donor because of the differences between individuals.

Experiment 2:

  • 30 donors (WBC counts within 4.8 to 11.0 x 10° cells/ml blood) ●
  • 12 tubes per donor (total 360 PAXgene tubes) ●
  • Blood from filled PAXgene tubes from 3 donors were pooled and realiquoted into empty tubes, generating 10 donor pools with 36 tubes each.
  • PAXgene Blood RNA system Lot No. VL1, VL2, VL3
  • 3 technicians. Each technician processed 4 tubes per donor pool per ● day. This was repeated with 2 additional kit lots.

RNA yield (absorbance at 260 nm) and purity (260nm/280nm absorbance ratio) were determined for each sample. For this experiment the RNA purity was also evaluated by electropherograms generated by BioAnalyzer. The isolated total RNA was further tested for degradation with the CFOS and IL1B Q-RT-PCR assay.

{5}------------------------------------------------

Results: the sponsor was able to show that the RNA vield and the RNA purity and quality were within the specification ranges. The min/max % CV repeatability for the quadruplicate preparation per donor pool within each lot and user were 3.4 and 28.8, respectively, with an overall % CV of 11.9. The min/max % CVs reproducibility within each user and between all lots were 8.7 and 20.1, respectively, with an overall % CV of 14.9. The min/max % CV reproducibility within all lots and between all users were 8.7 and 23.1, respectively, with an overall % CV of 16.4. The electropherograms demonstrated that the device repeatedly generates high quality total cellular RNA from whole blood. The Q-RT-PCR assay demonstrated that the transcripts CFOS and IL1B can be repeatedly measured within the assays limits utilizing different device lots, at different days, with different users in different laboratories.

RNA Purity, DNA Contamination, RNA Yield.

Objective:

Determination of the purity of RNA in samples, the genomic DNA (gDNA) in RNA samples and the nucleic acid yield from human whole blood specimens taken from normal subject population resulting from collection and processing of human whole blood in the PAXgene RNA Blood System.

Experimental design:

  • . 10 donors
  • 2 tubes per donor (total 20 PAXgene tubes) + 1 EDTA tube per . donor for WBC count (10 EDTA tubes)
  • storage for 24 hrs at 18-22 ℃, tubes were processed, RNA isolated ●
  • . RNA vield and purity were determined via absorbance measurement at 260 and 280 nm
  • . Percent gDNA in total nucleic acid preparation was determined using beta-actin PCR

Results: All samples for all parameter fulfilled the acceptance criteria.

RT-PCR Inhibition.

Objective:

Determination of RT-PCR inhibition introduced into RNA isolates by components of the PAXgene Blood RNA Kit

Experimental Design:

  • 22 blank eluates form columns
  • template RNA for RT-PCR assay was isolated from confluent layer of HeLa cells

{6}------------------------------------------------

  • Test samples were formulated each containing 40 ng of HeLa RNA template plus 0, 10, 20, 30, 40% volume of volume of blank eluate
  • Template RNA with eluate was tested in duplicate in a one-step ACTB RT-PCR assay

Results: All samples demonstrated that the kit components do not introduce any inhibition for an RT-PCR assay.

  • b. Linearity/assay reportable range:
    The linearity and range was determined within 0.2 to 7.8 ng for CFOS and 0.1 to 7.8 ng for IL1B total input nucleic acid per reaction. The total input mass of nucleic acid per reaction can contain up to 25% DNA for CFOS and up to 5% DNA for IL1B without affecting the performance characteristics of the assay.

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C. Stability of PAXgene Tubes:
    Objective: Determine the shelf life of the PAXgene Blood RNA Tube. Determination of physical attributes of the tubes:

  • draw volume

  • liquid additive volume

  • closure performance .

Chemical attributes of the liquid reagent:

  • . pH
  • conductivity ●
  • density
  • . chemical component concentration

Analysis of tube function:

  • RNA yield (A260 nm, 95% ≥3 µg/tube) ●
  • Purity (ratio of A260 and A280, between 1.8 2.2) .
  • Relative levels of CFOS and IL1B (CFOS/18S rRNA and IL1B/18S rRNA . duplex Q-RT-PCR assay, CFOS within 2.34 CT and IL1B within 1.94 CT).

Blood was collected in PAXgene tubes that had been stored at designated test time points (TTP's). RNA stabilizing properties of the tubes were investigated by preparing purified RNA from blood-filled tubes stored at 18-25°C for 0, 1 and 3 days with the PAXgene blood RNA kit (10 donors x 3 times of storage of blood filled tubes x 2 preparations). All PAXgene Blood RNA kits used in these studies were stored prior to use at 2-8°C to minimize kit aging. IMPORTANT: All claims of stability that are based on data from accelerated stability studies will be confirmed in parallel in real time stability studies. Table 1 shows the number of tubes tested in functional tests at each TTP.

{7}------------------------------------------------

Stability ConditionsTTPs at thistemperaturenumber of tubes forthistemperature studytotal
40°C study (accelerated)19w, 28w2 TTP x 60tubes/TTP120
50°C study (accelerated)10w, 15w2 TTP x 60 tubes/TTP120
25°C study (real time)0, 6m, 9m, 12m,13m, 15m, 16m,18m, 19m9 TTP x 60 tubes/TTP810

Table 1: Functional Test TTPs and number of tubes scheduled for testing

Results: Accelerated stability studies have been completed at the time of this report. Real time stability studies are ongoing, and only results of 6 months storage at 25°C are presented. The sponsor interprets that the data support a shelf live of 19 months storage at 25 °C.

The physical attributes were evaluated for 3 different lots of tubes. The draw volume, liquid additive volume, and closure performance meet the acceptance criteria. For the chemical attributes as well as the functional performance it is not clear whether 1 or 3 lots of tubes were used. For the presented data the pH. conductivity, and density as well as the RNA yield, purity and CFOS and IL1B transcript stability were in the expected ranges for the 6 months real time storage.

Stability of PAXgene Blood RNA Kit component:

Objective: Determine the shelf life of the PAXgene Blood RNA Kit component. Evaluation of different storage conditions via:

  • Real time and accelerated stability study
  • Open bottle study
  • . Simulated transportation study (simulated extreme temperatures, -20°C-40°C)

Determination of functional stability done with purified HeLa cell RNA as starting material. Two reagent sets were used per tested time point (4 blank eluates (RNase-free water as starting material), and 8 RNA eluates. The following parameters were addressed:

  • Determination of recovery of input RNA (HeLa cell RNA) ●
  • Determination of variability of recovery of input RNA ●
  • . Determination of the degree of inhibition of an RT-PCR assay (via ACTB RT-PCR monoplex assay.

Evaluation of physical and chemical parameters:

  • pH
  • conductivity ●
  • . density

Bioburden Analysis:

● Test of buffers for total bacterial and yeast counts per ml

Results: The RNA recovery, CV of recovery, degree on inhibition, pH, conductivity, and density (for all buffers) were in the expected ranges for the 6 months real time storage. Furthermore, the Proteinase K and the DNase I activity performed to the sponsors claims for the 6 month real time storage.

{8}------------------------------------------------

The bioburden analysis did not show any bacteria, yeast of fungi growth in the same time period. The transportation simulation also revealed acceptable performance of all kit components after extreme temperature simulation.

RNA In Situ Stability:

To determine the stability of RNA in samples stored at different temperatures and times, the sponsor performed several RNA In Situ stability experiments that are summarize in Table 2.

Objective: Evaluation of the quantity, quality and integrity of whole blood intracellular RNA after storage in the PAXgene tubes.

Storage Temperature2 – 8 °C18, 22, 25 °C- 20 °C- 70 °CFreeze/Thaw
# of donors1010101010
Eligibility1.8 to1.9 to1.10 to1.11 to1.12 to
WBC per ml11.0 x 106 cells11.0 x 106 cell11.0 x 106 cell11.0 x 106 cell11.0 x 106 cells
PAXgene tubes/donor121414106
Total # of tubes12014014010060
Used in experiment
Storage before startof experiment2 hrs RT2 hrs RT2 hrs RT2 hrs RT24 hrs RT
Time points(# of tubes)t0 (2 tubes)t1D (2 tubes)t2D (2 tubes)t3D (2 tubes)t4D (2 tubes)t5D (2 tubes)t0 (2 tubes)t1D (2 tubes)t3D (2 tubes)t0 (2 tubes)t1m (2 tubes)t2m (2 tubes)t3m (2 tubes)t4m (2 tubes)t5m (2 tubes)t6m (2 tubes)t0 (2 tubes)t3m (2 tubes)t6m (2 tubes))t0 (2 tubes)1st freeze/thaw(2 tubes)2nd freeze/thaw(2 tubes)
Yield (> 3 µg/tube)260 nm260 nm260 nm260 nm260 nm
Purity (ratio 1.8-2.2)260nm/280nm260nm/280nm260nm/280nm260nm/280nm260nm/280nm
CFOS/18S rRNAXXXXNA
Il1b/18S rRNAXXXXX

Table 2: Experimental Design of RNA In Situ Stability at different conditions

The RNA yield, purity, integrity based on the sponsors developed Q-RT-PCR assays showed that all samples at all the time points, temperatures and freeze/thaw cycles were within the acceptance range for the individual parameters (one exception). The sponsor limited the samples to blood donors with White Blood Cell (WBC) counts between 4.8 – 11.0 x 10° cell/ml of blood. The performance is only evaluated for this range of WBC counts, and is specified in the package insert.

d. Detection limit:

Performance evaluation demonstrated that the PAXgene™ Blood RNA system could purify at least 3 µg of RNA with an A260/A280 ratio between 1.8 and 2.2.

{9}------------------------------------------------

  • e. Analytical specificity: See above
  • Assay cut-off: f. Not applicable.
    1. Comparison studies:
    • a. Method comparison with predicate device: Not applicable.
    • b. Matrix comparison: Not applicable.
    1. Clinical studies:
    • a. Clinical Sensitivity: Not applicable.
    • b. Clinical specificity: Not applicable.
    • c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
    1. Clinical cut-off: Not applicable
    1. Expected values/Reference range: Not applicable

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The petition for Evaluation of Automatic Class III Designation for this device is accepted. The device is classified as Class II under regulation 21 CFR 866.4060 with special controls. The special control guidance document "RNA Preanalytical Systems (RNA Collection, Stabilization and Purification Systems for RT-PCR used in Molecular Diagnostic Testing)" will be available shortly.

§ 866.4070 RNA Preanalytical Systems.

(a)
Identification. RNA Preanalytical Systems are devices intended to collect, store, and transport patient specimens, and stabilize intracellular RNA from the specimens, for subsequent isolation and purification of the intracellular RNA for RT-PCR used in in vitro molecular diagnostic testing.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: RNA Preanalytical Systems (RNA Collection, Stabilization and Purification System for RT-PCR Used in Molecular Diagnostic Testing).” See § 866.1(e) for the availability of this guidance document.