Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182503

    Validate with FDA (Live)

    Date Cleared
    2019-06-03

    (264 days)

    Product Code
    Regulation Number
    870.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilizable Internal Defibrillation Paddles are intended for use with LIFEPAK defibrillators to internally detect ECG rhythm and provide defibrillation or synchronized cardioversion directly to the surgically exposed heart within a sterile use environment.

    Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia.

    Synchronized cardioversion is indicated for the treatment of atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, supraventricular tachycardia, and in relatively stable patients, ventricular tachycardia.

    Device Description

    The Sterilizable Internal Defibrillation Paddles is an accessory designed to be used with biphasic LIFEPAK defibrillator/monitors during open heart cardiac surgery. The Sterilizable Internal Defibrillation Paddles are intended for use by highly-trained medical professionals to internally detect electrocardiogram (ECG) rhythm and deliver internal defibrillation and synchronized cardioversion therapy directly to the surgically exposed heart. This accessory consists of metal paddle electrodes, and molded plastic handles with discharge control (located on the right handle) which connects directly to the defibrillator.

    The Sterilizable Internal Defibrillation Paddles accessory requires sterilization before initial use and after each use, per the Sterilizable Internal Defibrillation Paddles Instructions for Use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Sterilizable Internal Defibrillation Paddles. It's important to clarify that this document is for a medical device (hardware), not an AI/ML algorithm. Therefore, many of the typical questions for AI/ML device acceptance criteria and study design (like ground truth, expert adjudication, MRMC studies, training/test set sizes) are not directly applicable in the same way.

    Instead, the acceptance criteria for such a device typically revolve around demonstrating substantial equivalence to a predicate device, as well as meeting established safety and performance standards through design verification and validation.

    However, I will extract the information that is comparable or relevant to your questions, and specify where the requested information is not applicable for this type of device.


    Acceptance Criteria and Device Performance

    The acceptance criteria for this device are implicitly tied to demonstrating substantial equivalence to the predicate device and meeting design specifications related to safety and performance.

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
    Intended Use EquivalenceThe device's intended use (internally detect ECG rhythm, defibrillation/synchronized cardioversion directly to surgically exposed heart in a sterile environment) must be equivalent to the predicate device.The "Indications for Use" and "Intended Use" are stated to be equivalent to the predicate device. The device is used with biphasic LIFEPAK defibrillator/monitors during open heart cardiac surgery.
    Functional EquivalenceFeatures and functional characteristics must be equivalent to the predicate device, despite material/design changes (integrated handle-electrode assembly, improved durability/resistance to cleaning/sterilization).The "features, and functional characteristics of the proposed Sterilizable Internal Defibrillation Paddles are equivalent to the predicate device." The changes (integrated assembly, material for durability/sterilization) "have not raised any new issues when compared to the existing predicate devices."
    Design VerificationAll established design specifications (including those related to electrical performance, mechanical integrity, material properties, and sterilization efficacy) must be met."Design Verification Testing" was completed to "demonstrate that the proposed Sterilizable Internal Defibrillation Paddles meet the safety and performance requirements established in the design specifications."
    BiocompatibilityThe materials in contact with human tissue must be biocompatible. --------------------"Biocompatibility Testing" was completed to demonstrate safety.
    Electrical Safety & EMCThe device must meet electrical safety standards and electromagnetic compatibility requirements. --------------------"Electrical Safety and Electromagnetic Compatibility Testing" was completed.
    Design ValidationThe device must perform effectively and safely under simulated use conditions and, where applicable, in animal models, consistent with its intended clinical application. --------------------"Design Validation Testing" was completed, described as including "Animal Testing" and "Simulated Use Testing."
    No New Safety/Efficacy IssuesThere must be no new questions of safety or effectiveness raised by the device compared to the predicate. --------------------The submission concludes that the information "demonstrates that the Sterilizable Internal Defibrillation Paddles are substantially equivalent to the predicate Internal Defibrillation Handles and Electrodes with respect to performance," implying no new safety or efficacy issues.

    Non-Applicable / General Information for Hardware Devices

    The following questions are primarily relevant to AI/ML software performance studies and are not directly applicable to this hardware device's 510(k) submission as described.

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for a hardware device in the context of AI/ML test sets. Performance testing for hardware involves laboratory, animal, and simulated use testing, not "test sets" of data in the AI/ML sense. Data sources for such tests are internal (e.g., test benches, animal studies).
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/ML test set is not a concept used for this type of hardware device's testing.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a hardware device, not an AI assistance tool for human readers.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is an accessory to a defibrillator and requires a highly-trained medical professional for use. It is a human-operated device.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for a hardware device means meeting its engineering and safety specifications, and performing its intended function.
    • 8. The sample size for the training set: Not applicable for hardware.
    • 9. How the ground truth for the training set was established: Not applicable for hardware.

    Summary of Device-Specific Study Information (Performance Data)

    The submission highlights that no human clinical studies were submitted as part of this 510(k) Premarket Notification. This is typical for devices seeking substantial equivalence where predicate device experience or non-clinical data is deemed sufficient.

    The performance data that supports the acceptance criteria for this hardware device includes:

    • Design Verification Testing: To ensure the device meets established design specifications.
    • Biocompatibility Testing: To ensure materials are safe for biological contact.
    • Electrical Safety and Electromagnetic Compatibility Testing: To ensure the device operates safely within electrical and electromagnetic parameters.
    • Design Validation Testing: Which encompassed:
      • Animal Testing: To evaluate performance in a living biological system.
      • Simulated Use Testing: To assess performance under conditions mimicking clinical use.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1