P970031 · Medtronic, Inc. · LWR · Nov 26, 1997 · Cardiovascular
Device Facts
Record ID
P970031
Device Name
MEDTRONIC FREESTYLE AORTIC ROOT BIOPROSTHESIS
Applicant
Medtronic, Inc.
Product Code
LWR · Cardiovascular
Decision Date
Nov 26, 1997
Decision
APPR
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The Medtronic FREESTYLE® Aortic Root Bioprosthesis (hereafter called the Medtronic FREESTYLE® Bioprosthesis), is indicated for the replacement of malfunctioning native or prosthetic aortic valves with the option of aortic root replacement. No clinical data are available which evaluate the long-term impact of the AOA treatment in patients.
Device Story
The Medtronic FREESTYLE® Bioprosthesis is a stentless porcine aortic root valve preserved in glutaraldehyde and treated with alpha amino oleic acid (AOA) to inhibit calcification. It is designed for surgical implantation in the aortic position using one of three techniques: subcoronary (replacing leaflets), root-inclusion (replacing valve mechanism), or full-root (replacing valve and aortic root). The device is implanted by cardiac surgeons in an operating room. It functions as a replacement heart valve, providing hemodynamic performance comparable to allografts. The device is supplied in sizes 19mm to 27mm. Clinical benefits include improved NYHA functional class and hemodynamic parameters (mean pressure gradient, effective orifice area). Potential risks include thromboembolism, endocarditis, structural deterioration, and reoperation. The device is for single use only and must not be resterilized.
Clinical Evidence
Prospective non-randomized international study of 882 patients (640 subcoronary, 159 full-root, 83 root-inclusion) with mean follow-up of 17 months. Primary endpoints: NYHA classification, hemodynamic performance (mean pressure gradient, EOA, regurgitation), and safety (adverse events). Results showed significant improvement in NYHA class (e.g., 75% of subcoronary patients were Class III/IV at baseline vs 2% at one year). Hemodynamic performance showed low mean pressure gradients and trivial/mild regurgitation across all techniques. Adverse events included thromboembolism, permanent neurological events, and reoperation. Bench testing included biocompatibility, hydrodynamic performance, and accelerated wear (200 million cycles). Animal studies (sheep) confirmed handling characteristics and AOA efficacy in reducing leaflet calcification.
Technological Characteristics
Porcine aortic root tissue preserved in 0.2% glutaraldehyde; AOA (alpha amino oleic acid) antimineralization treatment. Features polyester fabric covering on myocardium and inflow circumference; polyester sutures. Available in 19, 21, 23, 25, 27 mm diameters. Stentless design. Sterilized in 0.2% buffered glutaraldehyde solution at 38°C-42°C. Shelf life: 3 years.
Indications for Use
Indicated for patients requiring replacement of malfunctioning native or prosthetic aortic valves, with the option of aortic root replacement. Contraindications: None known. Warnings: Accelerated calcific degeneration may occur in children, adolescents, young adults, and patients with altered calcium metabolism (e.g., chronic renal failure, hyperparathyroidism).
Reference Devices
Hancock Standard valve (Model 242)
Hancock Model 105
Related Devices
P970030 — TORONTO SPV VALVE · St. Jude Medical, Inc. · Nov 4, 1997
{0}
# SUMMARY of SAFETY and EFFECTIVENESS DATA
## MEDTRONIC FREESTYLE® AORTIC ROOT BIOPROSTHESIS
### Table of Contents
1. General Information ... 1
2. Indications For Use ... 1
3. Contraindications ... 1
4. Warnings and Precautions ... 1
4.1 Warnings ... 1
4.2 Precautions ... 1
Precautions Prior to Use ... 2
Precautions During Use ... 2
5. Device Description ... 2
6. Alternative Practices and Procedures ... 2
7. Marketing History ... 3
8. Adverse Effects of the Device on Health ... 3
8.1 Observed Adverse Events - Subcoronary Technique ... 3
Table 1. Observed Adverse Events for the Subcoronary Technique ... 3
8.2 Observed Adverse Events - Full-Root Technique ... 4
Table 2. Observed Adverse Events for the Full-Root Technique ... 4
8.3 Observed Adverse Events - Root-Inclusion Technique ... 5
Table 3. Observed Adverse Events for the Root-Inclusion Technique ... 5
8.4 Potential Adverse Events ... 6
9. Summaries of Pre-clinical Studies ... 6
9.1 Bench Testing ... 6
9.1.1 Biocompatibility, Immunology and Toxicology Studies ... 6
9.1.2 Hydrodynamic Performance ... 7
9.1.2.1 Test Chamber Development ... 7
Table 4. Summary of Hydrodynamic Performance Testing ... 8
9.1.2.2 Steady State Flow Pressure Drop ... 9
FULL ROOT CONFIGURATION ... 9
ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS ... 9
9.1.2.3 Pulsatile Flow Pressure Drop ... 9
FULL ROOT CONFIGURATION ... 9
ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS ... 9
9.1.2.4 Pulsatile Flow Regurgitation ... 9
FULL ROOT CONFIGURATION ... 9
ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS ... 10
9.1.2.5 Dynamic Regurgitation ... 10
9.1.2.6 In Vitro Doppler Ultrasound ... 10
9.1.2.7 Flow Visualization ... 10
9.1.3 Structural Performance (Accelerated Wear) ... 11
9.1.3.1 Accelerated Wear ... 11
FULL ROOT CONFIGURATION ... 11
ROOT INCLUSION & SUBCORONARY CONFIGURATIONS ... 11
9.1.3.2 Accelerated Wear - Hydrodynamic Assessment ... 11
FULL ROOT CONFIGURATION ... 12
ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS ... 12
9.2 Animal Studies ... 12
9.3 Sterilization ... 13
9.4 Shelf Life ... 13
9.4.1 Package Integrity ... 13
9.4.2 Product Integrity ... 13
10.0 Summary of Clinical Studies ... 13
10.1 Objectives ... 13
10.2 Study Design ... 14
10.2.1 Inclusion Criteria ... 14
10.2.2 Exclusion Criteria ... 14
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page i
{1}
10.2.3 Data Collection...14
Table 5. Number of Patients by Duration of Follow-up and Implant Technique...14
10.3 Analysis for Gender Bias...15
10.4 Description of Patients...15
Table 6. Patient Characteristics: Subcoronary Technique; All patients analyzed, N = 631...15
Table 7. Patient Characteristics: Full-Root Technique; All patients analyzed, N = 157...15
Table 8. Patient Characteristics: Root-Inclusion Technique; All patients analyzed, N = 80...16
10.4 Data Analyses...16
10.4.1 Functional NYHA...16
Table 9. Effectiveness Outcomes, Functional NYHA: Subcoronary Technique...16
Table 10. Effectiveness Outcomes, Functional NYHA: Full-Root Technique...16
Table 11. Effectiveness Outcomes, Functional NYHA: Root-Inclusion Technique...17
10.4.2 Hemodynamics, Valvular Regurgitation...17
Table 12. Effectiveness Outcomes, Hemodynamics, Valvular Regurgitation: Subcoronary Technique...17
Table 13. Effectiveness Outcomes, Hemodynamics, Valvular Regurgitation: Full-Root Technique...17
Table 14. Effectiveness Outcomes, Hemodynamics, Valvular Regurgitation: Root-Inclusion Technique...18
10.4.3 Hemodynamics, Mean Pressure Gradient...18
Table 15. Effectiveness Outcomes, Hemodynamics, Mean Pressure Gradient: Subcoronary Technique...18
Table 16. Effectiveness Outcomes, Hemodynamics, Mean Pressure Gradient: Full-Root Technique...18
Table 17. Effectiveness Outcomes, Hemodynamics, Mean Pressure Gradient: Root-Inclusion Technique...19
10.4.4 Hemodynamics, Effective Orifice Area...19
Table 18. Effectiveness Outcomes, Hemodynamics, Effective Orifice Area: Subcoronary Technique...19
Table 19. Effectiveness Outcomes, Hemodynamics, Effective Orifice Area: Full-Root Technique...19
Table 20. Effectiveness Outcomes, Hemodynamics, Effective Orifice Area: Root-Inclusion Technique...20
11. Conclusions Drawn from Studies...20
13. Panel Recommendations...20
14. FDA Decision...21
15. Approval Specifications...21
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page ii
10
{2}
P970031
# SUMMARY of SAFETY and EFFECTIVENESS DATA
## MEDTRONIC FREESTYLE® AORTIC ROOT BIOPROSTHESIS
## 1. General Information
Device Generic Name: Replacement Heart Valve
Device Trade Name: Medtronic FREESTYLE® Aortic Root Bioprosthesis
Applicant's Name and Address: Medtronic Heart Valves, Inc.
7000 Central Avenue NE
Minneapolis, MN 55432
PMA Application Number: P970031
Date of Panel Recommendation: September 15, 1997
Date of Notice of Approval to the Applicant: November 26, 1997
## 2. Indications For Use
The Medtronic FREESTYLE® Aortic Root Bioprosthesis (hereafter called the Medtronic FREESTYLE® Bioprosthesis), is indicated for the replacement of malfunctioning native or prosthetic aortic valves with the option of aortic root replacement.
No clinical data are available which evaluate the long-term impact of the AOA treatment in patients.
## 3. Contraindications
No contraindications for use of this device are known.
## 4. Warnings and Precautions
### 4.1 Warnings
- FOR SINGLE USE ONLY.
- DO NOT RESTERILIZE the valve by any method. Exposure of the bioprosthesis and container to irradiation, steam, ethylene oxide or other chemical sterilants will render the bioprosthesis unfit for use.
- Accelerated deterioration due to calcific degeneration of bioprosthesis may occur in:
- children, adolescents, or young adults;
- patients with altered calcium metabolism (e.g., chronic renal failure, hyperparathyroidism).
### 4.2 Precautions
- Implanting physicians must be familiar with the techniques for implanting an unstented bioprosthesis. These techniques are similar to those required for allograft implantation.
- In vitro testing of the Medtronic FREESTYLE® Bioprosthesis has only been performed in less compliant simulated aorta comparable to the aorta of middle aged or older patients. Data from clinical or in vitro testing are not available from more compliant simulated aorta comparable to the aorta of a young adult.
- Limited implant experience is available for the 19mm bioprosthesis implanted with the root-inclusion technique (none), and with the full-root technique (nine patients); and for the 21mm bioprosthesis implanted with the root-inclusion technique (four patients) (see Section 10, Summary of Clinical Studies).
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 1
{3}
# Precautions Prior to Use
Do not use the Medtronic FREESTYLE® Bioprosthesis:
- if the tamper evident seal is broken;
- if the glutaraldehyde storage solution does not completely cover the bioprosthesis;
- if the bioprosthesis has been exposed to freezing or has had prolonged exposure to heat as indicated by the temperature indicators provided in the packaging;
- if the bioprosthesis is damaged.
# Precautions During Use
- Do not expose the bioprosthesis to storage solutions other than the storage solution in which it was shipped, the sterile isotonic saline solution used during the rinsing procedure, or the sterile isotonic saline used to irrigate the bioprosthesis.
- Do not allow the tissue of the bioprosthesis to dry. Continuous submersion or irrigation is required.
- Do not add antibiotics to either the storage or the rinse solution. Do not apply antibiotics to the bioprosthesis.
- Do not lacerate the leaflet tissue. If a valve is damaged, the valve must be explanted and replaced.
- Do not evert the valve. Eversion will damage valve tissue.
- Passage of a catheter through any bioprosthesis may damage the valve and is, therefore, not recommended.
- Trim suture end close to the knot to prevent abrasion of leaflet tissue.
# 5. Device Description
The Medtronic FREESTYLE® Bioprosthesis, Model 995, is comprised of a porcine aortic root preserved by collagen crosslinking in a buffered 0.2% glutaraldehyde solution. During fixation, a 40 mmHg hydrostatic pressure is applied to the root with a zero pressure differential across the valve leaflets. The preserved aortic root has a thin, single layer of polyester fabric covering the myocardium and the inflow circumference of the bioprosthesis. All stitching is done with polyester suture. The Medtronic FREESTYLE® Bioprosthesis is supplied sterile in a buffered 0.2% glutaraldehyde solution. The Medtronic FREESTYLE® Bioprosthesis is designed for the aortic position only and is available in the following implantation diameters: 19 mm, 21 mm, 23 mm, 25 mm and 27 mm. The function of a stentless bioprosthetic valve is sensitive to surgical implant technique. The Medtronic FREESTYLE® Bioprosthesis is packaged as a single device that is modified at operation for insertion with one of three techniques to replace the native valve leaflets with the subcoronary technique, to replace the entire valve mechanism with the root-inclusion technique, or to replace the entire valve mechanism as well as the aortic root with the full-root technique.
# 6. Alternative Practices and Procedures
The alternative to the Medtronic FREESTYLE® Bioprosthesis is surgical replacement of the malfunctioning aortic valve with a homograft, mechanical prosthetic valve, or a stented bioprosthetic valve for which there is an approved premarket approval application (PMA). The choice of replacement valve depends on an assessment of patient factors which include age, preoperative condition, anatomy and the patient’s ability to tolerate long-term anticoagulant therapy.
Other forms of treatment may include the use of cardiac drug therapy.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 2
12
{4}
# 7. Marketing History
Commercial distribution of the Medtronic FREESTYLE® Bioprosthesis outside the U.S. began in January 1996. Currently the device is distributed in the following countries: Argentina, Austria, Australia, Belgium, Canada, Denmark, Finland, France, Germany, Greece, Israel, Italy, The Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.
The Medtronic FREESTYLE® Bioprosthesis has not been withdrawn from marketing for any reason relating to the safety and/or the effectiveness of the device.
# 8. Adverse Effects of the Device on Health
A prospective non-randomized multicenter international study evaluated the Medtronic FREESTYLE® Bioprosthesis with patient follow-up out to three years. A total of 882 patients received the bioprosthesis. Patients were monitored throughout the entire postoperative period for possible adverse events. The cumulative follow-up was 1246 patient-years with a mean follow-up of 17 months (SD = 12 months, range = 0 to 42 months).
# 8.1 Observed Adverse Events - Subcoronary Technique
A total of 640 Medtronic FREESTYLE® Bioprostheses were implanted with the subcoronary technique in 640 patients at 15 centers. Nine of the 640 patients were excluded from the data summary of adverse events for the following reasons: five patients had their bioprosthesis removed and replaced with another prosthesis during the initial surgery due to difficulty sizing a small aortic root, high mean gradient, or patient prosthesis mismatch; and four patients had either a pre-existing or concomitant implant of a mitral valve prosthesis. The adverse event rates were based on 631 bioprostheses implanted in 631 patients. The cumulative follow-up was 913 patient-years with a mean follow-up of 17 months (SD = 11 months, range = 0 to 42 months).
All patients analyzed, N = 631, Cumulative follow-up = 913 patient-years
Table 1. Observed Adverse Events for the Subcoronary Technique
| | Early Events^{1} | | Late Events | | Actuarial Freedom by Kaplan-Meier (%) | |
| --- | --- | --- | --- | --- | --- | --- |
| | N | % of Patients | N | % / Patient-Year^{2} | 1 Year (95% CI) | 3 Years (95% CI) |
| All Deaths | 31 | 4.9% | 31 | 3.6% | 91.6% [89.0% - 94.2%] | 85.9% [77.5% - 94.3%] |
| Bioprosthesis-Related or Unexplained | 4 | 0.6% | 9 | 1.0% | 97.9% [96.5% - 99.3%] | 97.5% [93.5% - 100.0%] |
| Study Bioprosthesis-Related AEs | | | | | | |
| Thromboembolism^{3} | 14 | 2.2% | 13 | 1.5% | 95.6% [93.6% - 97.6%] | 94.6% [88.8% - 100.0%] |
| Permanent Neurological Event | 11 | 1.7% | 6 | 0.7% | 97.3% [95.8% - 98.8%] | 96.6% [91.9% - 100.0%] |
| Transient Neurological Event | 3 | 0.5% | 6 | 0.7% | 98.5% [97.3% - 99.7%] | 98.2% [94.7% - 100.0%] |
| Bioprosthetic Thrombosis | 0 | 0.0% | 1 | 0.1% | 99.8% [99.4% - 100.0%] | 99.8% [98.6% - 100.0%] |
| Structural Deterioration^{4} | 0 | 0.0% | 0 | 0.0% | 100.0% [99.1% - 100.0%] | 100.0% [93.7% - 100.0%] |
| Nonstructural Dysfunction^{4} | 0 | 0.0% | 0 | 0.0% | 100.0% [99.1% - 100.0%] | 100.0% [93.7% - 100.0%] |
| Major Antithromboembolic-Related Hemorrhage | 10 | 1.6% | 7 | 0.8% | 97.0% [95.4% - 98.6%] | 97.0% [92.6% - 100.0%] |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 3
{5}
| | Early Events1 | | Late Events | | Actuarial Freedom by Kaplan-Meier (%) | |
| --- | --- | --- | --- | --- | --- | --- |
| | N | % of Patients | N | % / Patient-Year2 | 1 Year (95% CI) | 3 Years (95% CI) |
| Primary Paravalvular Leak | 3 | 0.5% | 8 | 0.9% | 98.6% | 97.6% |
| | | | | | [97.5% - 99.7%] | [93.6% - 100.0%] |
| Endocarditis | 2 | 0.3% | 8 | 0.9% | 98.5% | 97.9% |
| | | | | | [97.3% - 99.7%] | [94.2% - 100.0%] |
| Primary Hemolysis4 | 0 | 0.0% | 0 | 0.0% | 100.0% | 100.0% |
| | | | | | [99.1% - 100.0%] | [93.7% - 100.0%] |
| Reoperation | 0 | 0.0% | 7 | 0.8% | 98.9% | 98.6% |
| | | | | | [97.9% - 99.9%] | [95.6% - 100.0%] |
| Explant | 0 | 0.0% | 6 | 0.7% | 99.0% | 98.7% |
| | | | | | [98.0% - 100.0%] | [95.8% - 100.0%] |
NOTES:
AEs = Adverse Events
1. Hospital or 30 day event for death or 30 day event for adverse events.
2. Calculations were based on 864 late patient-years.
3. One late event was peripheral arterial embolus.
4. The number of patients remaining at risk at one year (N = 415) and three years (N = 57) was used for N in the calculations of the lower confidence limits for the actuarial estimates. The calculation methods are those described in the following note.
Note: Adverse event rates were calculated as the percentage of patients for early events. For late adverse events, the linearized rates (%/patient-year) were calculated. For time to first event (early or late), actuarial rates using Kaplan-Meier method and confidence intervals were calculated. For adverse events with no occurrences, the lower two-sided 95% confidence limits for the Kaplan-Meier estimates were calculated as (1 - maximum risk), where (1 - maximum risk) is equal to $(0.025)^{1/N}$. If there was no censoring, N would be the total sample size. Since there was censoring, the number of patients remaining at risk at one and three years was used for N. Using the number of patients remaining at risk ignores the experience of all the patients who were censored before the relevant time points and, therefore, overestimates the maximum risk.
## 8.2 Observed Adverse Events - Full-Root Technique
A total of 159 Medtronic FREESTYLE® Bioprostheses were implanted with the full-root technique in 159 patients at 14 centers. Two of the 159 patients were excluded from the data summary of adverse events for the following reasons: two patients had either a pre-existing or concomitant implant of a mitral valve prosthesis.
The adverse event rates were based on 157 bioprostheses implanted in 157 patients. The cumulative follow-up was 189 patient-years with a mean follow-up of 14 months (SD = 13 months, range = 0 to 38 months).
Table 2. Observed Adverse Events for the Full-Root Technique
All patients analyzed, N = 157, Cumulative follow-up = 189 patient-years
| | Early Events1 | | Late Events | | Actuarial Freedom by Kaplan-Meier (%) | |
| --- | --- | --- | --- | --- | --- | --- |
| | N | % of Patients | N | % / Patient-Year2 | 1 Year (95% CI) | 3 Years (95% CI) |
| All Deaths | 22 | 14.0% | 8 | 4.5% | 79.8% [72.0% - 87.6%] | 78.5% [62.5% - 94.5%] |
| Bioprosthesis-Related or Unexplained | 1 | 0.6% | 4 | 2.2% | 96.3% [92.3% - 100.0%] | 94.7% [85.1% - 100.0%] |
| Study Bioprosthesis-Related AEs | | | | | | |
| Thromboembolism3 | 5 | 3.2% | 6 | 3.4% | 93.3% [87.9% - 98.7%] | 90.1% [76.2% - 100.0%] |
| Permanent Neurological Event | 2 | 1.3% | 3 | 1.75 | 96.6% [92.7% - 100.0%] | 95.0% [85.4% - 100.0%] |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 4
{6}
| | Early Events1 | | Late Events | | Actuarial Freedom by Kaplan-Meier (%) | |
| --- | --- | --- | --- | --- | --- | --- |
| | N | % of Patients | N | % / Patient-Year2 | 1 Year (95% CI) | 3 Years (95% CI) |
| Transient Neurological Ebent | 2 | 1.3% | 3 | 1.7% | 96.3% [92.2% - 100.0%] | 94.7% [84.3% - 100.0%] |
| Bioprosthetic Thrombosis | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
| Structural Deterioration4 | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
| Nonstructural Dysfunction4 | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
| Major Antithromboembolic-Related Hemorrhage | 2 | 1.3% | 1 | 0.6% | 97.7% [94.4% - 100.0%] | 97.7% [91.0% - 100.0%] |
| Primary Paravalvular Leak | 1 | 0.6% | 0 | 0.0% | 99.2% [97.3% - 100.0%] | 99.2% [95.3% - 100.0%] |
| Endocarditis | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
| Primary Hemolysis4 | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
| Reoperation | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
| Explant | 0 | 0.0% | 0 | 0.0% | 100.0% [95.5% - 100.0%] | 100.0% [83.2% - 100.0%] |
NOTES:
AEs = Adverse Events
1. Hospital or 30 day event for death or 30 day event for adverse events.
2. Calculations were based on 179 late patient-years.
3. One late event was peripheral arterial embolus.
4. The number of patients remaining at risk at one year (N = 81) and three years (N = 20) was used for N in the calculations of the lower confidence limits for the actuarial estimates. The calculation methods are those described in the note following Table 1.
## 8.3 Observed Adverse Events - Root-Inclusion Technique
A total of 83 Medtronic FREESTYLE® Bioprostheses were implanted with the root-inclusion technique in 83 patients at 14 centers. Three of the 83 patients were excluded from the data summary of adverse events for the following reasons: three patients had their bioprosthesis removed and replaced with another prosthesis during the initial surgery due to a leaflet tear during suturing, difficulty seating and orienting the coronaries, or a paravalvular leak.
The adverse event rates were based on 80 bioprostheses implanted in 80 patients. The cumulative follow-up was 139 patient-years with a mean follow-up of 21 months (SD = 13 months, range = 0 to 39 months).
All patients analyzed, N = 80, Cumulative follow-up = 139 patient-years
Table 3. Observed Adverse Events for the Root-Inclusion Technique
| | Early Events1 | | Late Events | | Actuarial Freedom by Kaplan-Meier (%) | |
| --- | --- | --- | --- | --- | --- | --- |
| | N | % of Patients | N | % / Patient-Year2 | 1 Year (95% CI) | 3 Years (95% CI) |
| All Deaths | 4 | 5.0% | 5 | 3.8% | 88.8% [80.9% - 96.7%] | 87.0% [72.5% - 100.0%] |
| Bioprosthesis-Related or Unexplained | 3 | 3.8% | 3 | 2.3% | 91.5% [84.4% - 98.6%] | 91.5% [79.2% - 100.0%] |
| Study Bioprosthesis-Related AEs | | | | | | |
| Thromboembolism3 | 3 | 3.8% | 7 | 5.3% | 91.3% [83.8% - 98.8%] | 87.2% [70.2% - 100.0%] |
| Permanent Neurological Event | 0 | 0.0% | 2 | 1.5% | 98.3% [95.0% - 100.0%] | 96.3% [87.3% - 100.0%] |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 5
{7}
| | Early Events^{1} | | Late Events | | Actuarial Freedom by Kaplan-Meier (%) | |
| --- | --- | --- | --- | --- | --- | --- |
| | N | % of Patients | N | % / Patient-Year^{2} | 1 Year (95% CI) | 3 Years (95% CI) |
| Transient Neurological Event | 3 | 3.8% | 5 | 3.8% | 93.0%
[86.1% - 99.9%] | 87.6%
[71.5% - 100.0%] |
| Bioprosthetic Thrombosis | 0 | 0.0% | 0 | 0.0% | 100.0%
[93.5% - 100.0%] | 100.0%
[81.5% - 100.0%] |
| Structural Deterioration^{4} | 0 | 0.0% | 0 | 0.0% | 100.0%
[93.5% - 100.0%] | 100.0%
[81.5% - 100.0%] |
| Nonstructural Dysfunction^{4} | 0 | 0.0% | 0 | 0.0% | 100.0%
[93.5% - 100.0%] | 100.0%
[81.5% - 100.0%] |
| Major Antithromboembolic-Related Hemorrhage | 2 | 2.5% | 0 | 0.0% | 97.4%
[93.2% - 100.0%] | 97.4%
[90.1% - 100.0%] |
| Primary Paravalvular Leak | 2 | 2.5% | 0 | 0.0% | 97.4%
[93.2% - 100.0%] | 97.4%
[90.1% - 100.0%] |
| Endocarditis | 1 | 1.3% | 1 | 0.8% | 96.9%
[92.4% - 100.0%] | 96.9%
[89.0% - 100.0%] |
| Primary Hemolysis^{4} | 0 | 0.0% | 0 | 0.0% | 100.0%
[93.5% - 100.0%] | 100.0%
[81.5% - 100.0%] |
| Reoperation | 2 | 2.5% | 2 | 1.5% | 94.4%
[88.5% - 100.0%] | 94.4%
[84.1% - 100.0%] |
| Explant | 2 | 2.5% | 2 | 1.5% | 94.4%
[88.5% - 100.0%] | 94.4%
[84.1% - 100.0%] |
NOTES:
AEs = Adverse Events
1. Hospital or 30 day event for death or 30 day event for adverse events.
2. Calculations were based on 179 late patient-years.
3. One late event was peripheral arterial embolus.
4. The number of patients remaining at risk at one year (N = 81) and three years (N = 20) was used for N in the calculations of the lower confidence limits for the actuarial estimates. The calculation methods are those described in the note following Table 1.
## 8.4 Potential Adverse Events
Adverse events potentially associated with the use of bioprosthetic heart valves (in alphabetical order) include:
- cardiac dysrhythmias
- endocarditis
- hemorrhage, anticoagulant/antiplatelet-related
- nonstructural dysfunction (pannus, suture, inappropriate sizing, or other)
- thromboembolism
- death
- hemolysis
- leak, transvalvular or paravalvular
- structural deterioration (calcification, leaflet tear, or other)
- valve thrombosis
## 9. Summaries of Pre-clinical Studies
### 9.1 Bench Testing
#### 9.1.1 Biocompatibility, Immunology and Toxicology Studies
Biocompatibility tests were performed on alpha amino oleic acid (AOA) and glutaraldehyde tanned porcine tissue treated with AOA. All samples used in these tests were subjected to the maximum number of resterilization cycles (2X) with the exception of the testing of the AOA compound.
The biocompatibility of the AOA compound and the AOA treated Medtronic FREESTYLE® Bioprosthesis were assessed by determining its influence on: the inflammatory response, cytotoxicity, hemolysis, mutagenicity (Ames genotoxicity), USP rabbit pyrogenicity, systemic toxicity, intracutaneous toxicity, sensitization (Kligman guinea pig), and thrombogenicity. The approach used for the biocompatibility assessment of the Medtronic FREESTYLE® Bioprosthesis is consistent with the intent of the Biological
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
14
{8}
Evaluation of Medical Devices Memorandum (G95-1) issued by FDA on May 1, 1995. The results of the biocompatibility studies performed support the biocompatibility of the Medtronic FREESTYLE® Bioprosthesis.
A health risk assessment was done to assess the amount of the glutaraldehyde that leaches out over time from the tissue. All tests were negative, and the amount of glutaraldehyde leaching out was 0.069 mg/l, which peaked at 4 hours, and was less than the 20.5 mg/l used in the acute toxicity test. The applicant did not conduct carcinogenicity, subchronic, chronic or reproductive toxicity testing because glutaraldehyde tanned porcine tissue has a long history of use, the device does not contact soft tissue, and no leachable components were detected during acute testing.
## 9.1.2 Hydrodynamic Performance
### 9.1.2.1 Test Chamber Development
Test chamber development for the *in vitro* evaluation of the Medtronic FREESTYLE® Bioprosthesis took into consideration that the aortic valve in it’s natural state does not have a fixed shape and can only be described at a given time in the cardiac cycle such as mid-systole or mid-diastole. Another consideration was that the aorta expands quite easily at low internal pressure but stiffens at higher pressures to prevent ballooning.
In order to produce simulated aortas of consistent, repeatable geometry and dimensions, steel compression molds were produced. Model aortas were constructed from silicone rubber formulated to produce the desired compliance for the *in vitro* tests, and cured until the molded part was stabilized. Dimensions were patterned from data available through published clinical literature. The geometry of the model aorta was based on the normal human adult aorta rather than a diseased aorta, since the geometry of the diseased aorta varies with the type and extent of the disease.
Hydrodynamic performance testing was conducted on Medtronic FREESTYLE® Bioprostheses mounted in these compliant test chambers. Table 4 is a summary of the hydrodynamic performance testing.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 7
{9}
Table 4. Summary of Hydrodynamic Performance Testing
| Test | Number of Samples: Freestyle | Number of Samples Control: Model 242 | Pass/Fail Criteria | Results |
| --- | --- | --- | --- | --- |
| Steady State Flow Pressure Drop | 15 (3 of each size) for each implant technique (full-root, root inclusion, subcoronary) | 4 (3 size 31mm, 1 size 19mm) | Pressure Drop to be ≤ the Control valve | Pass |
| Pulsatile Flow Pressure Drop | 15 (3 of each size) for each implant technique (full-root, root inclusion, subcoronary) | 4 (3 size 31mm, 1 size 19mm) | Pressure Drop to be ≤ the Control valve. | Pass |
| Pulsatile Flow Regurgitation (Regurgitant Volume) | 15 (3 of each size) for each implant technique (full-root, root inclusion, subcoronary) | 4 (3 size 31mm, 1 size 19mm) | Regurgitant Volume to be ≤ to Control valve of equivalent flow area. | Pass |
| Dynamic Regurgitation (Leakage Volume vs Back Pressure) | 15 (3 of each size) for each implant technique (full-root, root inclusion, subcoronary) | 4 (3 size 31mm, 1 size 19mm) | Leakage Volume to be ≤ to Control valve of equivalent flow area. | Pass |
| In-Vitro Doppler Ultrasound | 3 for each implant technique (full-root, root inclusion, subcoronary) size 19mm, 23mm, and 27mm | N/A | Assessment: Determine if the Bernoulli equation ΔP = 4(V₂² - V₁²) may be used to assess peak and mean pressure gradients | Bernoulli equation works well within the experimental error (5%). |
| Flow Visualization (Color Doppler Flow Mapping) | 1 (27mm) | N/A | Qualitatively visualize leaflet motion and downstream flow fields. | Flow accelerates uniformly through the valve with a uniform flat profile downstream. No appearance of a jet with steep velocity gradients. |
| Accelerated Wear tested at room temperature, in sterile/filtered saline, at 1450 RPM, peak closed valve pressure drop = 110±5mmHg | 52 (3 of each size 19mm through 25mm) (6 27mm)-full root, (3 of each size) subcoronary, root inclusion (3 Non-AOA treated, two 19mm; one 23mm)- | 4 (size 31mm) | Wear (as defined by abrasion, tears, holes, coaptation, Delamination and other characteristics) to be ≤ the Control valve. | Pass. Observed Wear- No valve failures for either Freestyle or control. Wear patterns were similar to control wear. Most Frequent wear- 1.wear abrasion caused by contact to fraying aortic sutures; 2. surface abrasion; 3. holes, resulting from abrasion; 4. tears resulting from abrasion. |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 8
{10}
## 9.1.2.2 Steady State Flow Pressure Drop
Steady state flow testing to examine pressure drop as a function of flow rate was performed on each size of the Medtronic FREESTYLE® Bioprosthesis (19 mm, 21 mm, 23 mm, 25 mm & 27 mm). The aortic Hancock Standard valve (31 mm & 19 mm) was used as a control.
### FULL ROOT CONFIGURATION
The results demonstrated that the steady state flow mean pressure drop across the Medtronic FREESTYLE® Bioprosthesis in all sizes was lower than the pressure drop across the 19 mm Hancock Standard control valve. In addition, the mean steady state flow pressure drop across the 25 mm and 27 mm Medtronic FREESTYLE® Bioprostheses was lower than the mean pressure drop across the 31 mm Hancock Standard control valve.
### ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS
The results demonstrated that the steady state flow mean pressure drop across the Medtronic FREESTYLE® Bioprosthesis, in sizes 25 mm and 27 mm, was lower than the mean pressure drop across the 31 mm Hancock Standard control valve.
## 9.1.2.3 Pulsatile Flow Pressure Drop
Pulsatile flow pressure drop testing was completed on each size of the Medtronic FREESTYLE® Bioprosthesis (19 mm, 21 mm, 23 mm 25 mm & 27 mm). The aortic Hancock Standard valve (31 mm & 19 mm) was used as a control. Testing was performed at a cardiac output range of 2.5 to 7.5 L/min, and a pulse rate of 70 beats per minute (bpm), with systole accounting for approximately 35% of the simulated cardiac cycle.
### FULL ROOT CONFIGURATION
The results demonstrated that the pulsatile flow mean pressure drop across the Medtronic FREESTYLE® Bioprosthesis in all sizes was lower than the pressure drop across the 19 mm Hancock standard control valve. In addition, the mean pressure drop across the Medtronic FREESTYLE® Bioprosthesis, in sizes 25 mm and 27 mm, was lower than the mean pressure drop across the 31 mm Hancock Standard control valve.
### ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS
The results demonstrated that the pulsatile flow mean pressure drop across the 25 mm and 27 mm sized Medtronic FREESTYLE® Bioprosthesis was lower than the mean pressure drop across the 31 mm Hancock Standard control valve.
## 9.1.2.4 Pulsatile Flow Regurgitation
Pulsatile flow regurgitation was completed on each size of the Medtronic FREESTYLE® Bioprosthesis (19 mm, 21 mm, 23 mm 25 mm & 27 mm). The aortic Hancock Standard valve (31 mm & 19 mm) was used as a control. Testing was performed at 5.0 L/min and a pulse rate of 70 beats per minute (bpm), with systole accounting for approximately 35% of the simulated cardiac cycle. Regurgitant volume versus back pressure (closing pressures applied to the valve) data were obtained at back pressures (mean arterial pressure) of 90, 110, 130 and 150 mmHg.
### FULL ROOT CONFIGURATION
- **Total Regurgitant Volume**: The pulsatile flow regurgitation data demonstrates that the total regurgitant volume for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve was less than 7 ml/beat.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 9
{11}
- Closing Volume: The pulsatile flow regurgitation data demonstrates that the closing volume was less than 4 ml/beat for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve.
- Leakage Volume: The pulsatile flow regurgitation data demonstrates that the leakage volume was less than 4 ml/beat for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve.
## ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS
- Total Regurgitant Volume: The pulsatile flow regurgitation data demonstrates that the total regurgitant volume for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve was less than 8 ml/beat.
- Closing Volume: The pulsatile flow regurgitation data demonstrates that the closing volume was less than 4 ml/beat for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve.
- Leakage Volume: The pulsatile flow regurgitation data demonstrates that the leakage volume was less than 4 ml/beat for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve.
## 9.1.2.5 Dynamic Regurgitation
Pulsatile flow regurgitant volume versus beat rate (dynamic regurgitation) testing was completed on each size of the Medtronic FREESTYLE® Bioprosthesis (19 mm, 21 mm, 23 mm 25 mm & 27 mm) configured for the full root technique. The aortic Hancock Standard valve (31 mm & 19 mm) was used as a control. Testing was performed using a simulated heart rate of 50, 70 and 100 beats per minute (bpm), with systole accounting for approximately 35% of the simulated cardiac cycle, and a cardiac output of 5.0 L/min. The valves were tested at a back pressure (mean arterial pressure) of 90 mmHg.
The results demonstrated that the mean total Regurgitant volume was less than 7 ml/beat for the Medtronic FREESTYLE® Bioprosthesis and the Hancock Standard control valve.
## 9.1.2.6 In Vitro Doppler Ultrasound
In Vitro Doppler Ultrasound was used to determine if the Bernoulli relationship, $\Delta P = 4(V_2^2 - V_1^2)$, may be used to noninvasively assess pressure gradients in patients implanted with a Medtronic FREESTYLE® Bioprosthesis using the full root, root inclusion and subcoronary technique. Three Medtronic FREESTYLE® Bioprostheses, 19mm, 23mm, and 27mm were used for each technique. The study was conducted at cardiac outputs in the range of 2.5 to 8.0 L/min., at a pulse rate of 70 bpm, with systolic duration of 280 to 320 msec (35% of the simulated cardiac cycle). Mean aortic pressure was 90 to 100 mmHg and mean left atrial pressure was 5 to 10 mmHg. An aqueous solution of saline with a viscosity of 1.0 cP was used as a blood analog fluid at a temperature of 22° C, with 1-2 % cornstarch particles added to the fluid to act as acoustic scatterers.
For the ultrasound Doppler, the root mean square velocities and the mean systolic pressure drops at 30mm and 100mm distal locations indicate that the Bernoulli equation works well within experimental error (within 5%) with the Medtronic FREESTYLE® Bioprosthesis using the full root, root inclusion and subcoronary technique.
## 9.1.2.7 Flow Visualization
Color Doppler Flow Mapping (CDFM) was utilized to qualitatively visualize the downstream flow field of a 27 mm Medtronic FREESTYLE® Bioprosthesis in all configurations. The 27 mm sized valve was chosen since it was the largest annulus size. The study was conducted at a heart rate of 70 bpm, systolic
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
20
{12}
duration of approximately 280 msec (35% of the simulated cardiac cycle), mean aortic pressure of 100 mmHg, and cardiac outputs of approximately 3.0 and 5.0 L/min.
The study showed good opening of the valve leaflets by CDFM, M-mode and 2D echo. The flow accelerated uniformly through the valve and appeared to have a uniform flat profile downstream of the prosthesis. There was no appearance of a jet with steep velocity gradients, as has been observed with stented porcine bioprostheses. The flow visualization study demonstrated that the Medtronic FREESTYLE® Bioprosthesis utilizing the full root technique has a more uniform and less turbulent flow field when compared to stented porcine bioprostheses.
## 9.1.3 Structural Performance (Accelerated Wear)
### 9.1.3.1 Accelerated Wear
Accelerated wear testing was performed on Shelhigh accelerator testers. Wear was defined as abrasion, tears, holes, coaptation, delamination, and other characteristics, and should be less than that of the control valve. A peak closed-valve pressure drop of 110 ± 5 mmHg was maintained. Each valve was visually examined and photographed prior to initiation of the test. All valves were tested for a minimum of 200 x 10⁶ cycles with visual examination and photographs taken at every 20 X 10⁶ cycles. Test parameters were monitored on a routine basis.
All of the Medtronic FREESTYLE® Bioprostheses (AOA and nonAOA treated) demonstrated normal functioning as exhibited by proper opening and closing along with the maintenance of a closed valve pressure drop throughout the 200 X 10⁶ cycles. All observed tissue wear anomalies were determined to be artifactual of the test fixture and environment.
### FULL ROOT CONFIGURATION
Accelerated wear testing was completed on 18 Medtronic FREESTYLE® Bioprostheses (3 per size: 19 mm, 21 mm 23 mm & 25 mm; 6 per size: 27 mm) treated with AOA and 3 non-AOA treated FREESTYLE® Bioprostheses (2 per size: 19 mm and 1 per size: 23 mm). Three (3) aortic Standard Hancock valves (size 31 mm) were used as controls. Tester speed and stroke volume were adjusted to allow typical leaflet excursion (full opening and closing) during cardiac cycling with the tester speed eventually being established at 1310 rpm.
### ROOT INCLUSION & SUBCORONARY CONFIGURATIONS
Accelerated wear testing was completed on three (3) Medtronic FREESTYLE® Bioprostheses of each size (19 mm, 21 mm 23 mm, 25 mm & 27 mm) and configuration for a total of 30 valves. One (1) aortic Standard Hancock (size 31 mm) valve was used as a control. Tester speed and stroke volume were adjusted to allow typical leaflet excursion (full opening and closing) during cardiac cycling with the tester speed being established at 1450 rpm.
### 9.1.3.2 Accelerated Wear - Hydrodynamic Assessment
Pulsatile flow pressure drop and valvular regurgitation were assessed on the accelerated wear test valves and control valves. The test and control valves were the same as those described for the accelerated wear configurations.
Pulsatile flow pressure drop was performed at a cardiac output range of 2.5 to 7.5 L/min, and a pulse rate of 70 beats per minute (bpm), with systole accounting for approximately 35% of the simulated cardiac cycle. Pulsatile flow regurgitant volume tests were performed at 5.0 L/min and a pulse rate of 70 bpm. Regurgitant volume versus back pressure data were obtained at back pressures of 80, 100, 120, 140 and 160 mmHg. Hydrodynamic assessment occurred at the pretest, 60 X 10⁶, 120 X 10⁶ and 200 X 10⁶ cycle intervals of the accelerated wear testing.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
21
{13}
# FULL ROOT CONFIGURATION
The pulsatile flow pressure drop and total regurgitant volume in some valves were significantly lower after durability cycling (accelerated wear) than before cycling. The difference for all valves, except one outlier (19 mm Medtronic FREESTYLE® Bioprosthesis; pre to 2000 X $10^{6}$ difference equals 2.2 mmHg), was found to be within the test system accuracy (±1 mmHg and ±1ml/beat).
# ROOT INCLUSION AND SUBCORONARY CONFIGURATIONS
The pulsatile flow pressure drop and total regurgitant volume in most valves were found to be unaffected by accelerated wear cycling which means that the observed differences were found to be within the accuracy of the test system (±1 mmHg and 1± ml/beat). Some valves, during the course of the accelerated wear testing, did exhibit changes in hydrodynamic performance (i.e., increase in total regurgitant volume). Artifactual related damage to the valves along with the rigidity of the test apparatus were found to be contributing causes to the altered hydrodynamic performance of the device.
## 9.2 Animal Studies
Two *in-vivo* animal studies (acute and chronic) were performed on the Medtronic FREESTYLE® Bioprosthesis. The first was a short-term implant (acute) study in the adult sheep model which was designed to evaluate the hemodynamic efficacy and surgical handling characteristics of the device. The short term study involved implantation of 6 bioprostheses (size 19 mm) in mature sheep. Five of the 6 animals were sacrificed between 6 to 10 weeks post-implant and the sixth animal was sacrificed 16 months post implant. The results of the short term (acute) sheep implantation study demonstrated the following surgical handling characteristics of the Medtronic FREESTYLE® Bioprosthesis:
- comparable to aortic root replacement using an allograft valve;
- accurate sizing was critical;
- coronary artery implantation was without difficulty; and
- cloth reinforcement of the proximal orifice prevented dehiscence.
Hemodynamically, the transvalvular pressure gradient for the 19 mm Medtronic FREESTYLE® Bioprosthesis was found to be comparable to the small mean differential pressures noted clinically with 19 mm aortic homografts. Explant pathology findings described for valves implanted in the short-term study are similar to those reported following studies of implantation in juvenile sheep; however, no leaflet calcification was noted. Hematocrit and hemoglobin values were noted to be normal at the time of explant.
The second study was a long-term implant (chronic) study which was performed in juvenile sheep in an effort to determine the effect of AOA as a potential antimineralization treatment. A total of 16 Medtronic FREESTYLE® Bioprostheses (7 AOA Treated and 9 NonAOA Treated) were implanted as an apico-aortic bypass graft for a 20 week period. Three Hancock, Model 105, size 18mm valved conduits, were used as controls to assess the potential of the animal model to mineralize a commercially available standard high pressure-fixed porcine bioprosthesis. The results of the study showed a significant difference (p<0.0002) in leaflet calcium for AOA-treated valves when compared to the non-AOA treated valves. Although little to no mineralization of treated leaflets was observed, the aortic wall segment of the treated and control valves was found to be focally calcified.
In addition to the sheep studies, a series of subdermal rat implant studies were performed, utilizing treated (AOA) and non-treated samples. The implant duration was 1, 2, 3, and 4 months. For the period of time studied, the AOA treated samples contained significantly less calcium (mg of calcium/mg dry tissue) than the untreated.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 12
{14}
# 9.3 Sterilization
The Medtronic FREESTYLE® Bioprosthesis is sterilized in a 0.2% buffered glutaraldehyde solution with placement of the packaged valve assembly into an incubator for terminal sterilization at 38°C - 42°C for 20-22 hours. After completion of terminal sterilization the product is held in quarantine until sterility is verified in accordance with process specifications. Annual requalification of the sterilization process is performed.
# 9.4 Shelf Life
The shelf life for the Medtronic FREESTYLE® Bioprosthesis underwent qualification testing to ensure that both package and product integrity had been maintained after aging to three years. The package integrity samples were exposed to accelerated aging to three years; whereas, the product integrity samples underwent real time aging to a minimum of three years.
## 9.4.1 Package Integrity
Package integrity was assessed by testing the sterile barrier of the Medtronic FREESTYLE® Bioprosthesis. The testing included a vacuum leak test (60 aged samples), lid removal torque test (60 aged samples/30 non-aged samples), solution volume check test (60 aged samples), and a microbial challenge (37 test samples, 2 positive controls, 1 negative control). All test samples were subjected to a worst-case processing condition for sterilization of 3 cycles (maximum manufacturing specification of two cycles plus one additional cycle) and environmental stress conditioning as well as having undergone a shipping and handling test prior to performance of the sterile barrier tests.
## 9.4.2 Product Integrity
The product integrity portion of the shelf life assessment consisted of a battery of tests which were designed to affirm the functionality of the valve through the examination of multiple aspects of valve performance and structure. The qualification included the following tests: shrink temperature (10 aged samples, 10 non-aged samples), collagen content (enzyme susceptibility) (10 aged samples, 10 non-aged samples, 10 fresh controls), moisture content (10 aged samples, 10 non-aged samples), hydrodynamic performance (3 aged test valves 19/23/27 mm, 1 non-aged control), biaxial mechanical (7 aged valves, 7 non-aged valves), histological evaluation (3 aged valves 19/23/27 mm, 1 non-aged control), storage solution pH (15 samples), and glutaraldehyde percentage (15 samples). All 3 year real time aged samples had been sterilized two times (maximum allowed per manufacturing specification) and underwent environmental stress conditioning before the product integrity assessment was performed.
The pre-established criteria for the package and product integrity testing were found to have been met. Thus, the Medtronic FREESTYLE® Bioprosthesis is considered to be qualified for a three year shelf life.
# 10.0 Summary of Clinical Studies
## 10.1 Objectives
The primary objectives of the study were to evaluate the effectiveness, safety, and clinical performance of the Medtronic FREESTYLE® Bioprosthesis in patients requiring isolated aortic heart valve replacement. Specifically, improvement in coronary function as measured by NYHA classification at baseline and follow-up; and, improvement in hemodynamic performance, in terms of mean pressure gradient, effective orifice area (EOA), cardiac output and prosthesis regurgitation. The study compared the implantation and short- and mid-term performance of the Medtronic FREESTYLE® Bioprosthesis with the published literature on tissue valves, which serves as a source of benchmarks for evaluating efficacy in terms of mean pressure gradient and regurgitation, and EOA. FDA reviewed the published literature and approved three
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 13
{15}
articles which contained results directly comparable to the Medtronic FREESTYLE® Bioprosthesis for mean pressure gradient and regurgitation.
Objective performance criteria were used to evaluate safety in terms of adverse events: thromboembolism, valve thrombosis, hemorrhage, paravalvular leak, endocarditis, hemolysis, valve deterioration, and valve dysfunction. In addition to mortality data, reoperation and explant data were also collected.
## 10.2 Study Design
From 1992 to 1997, a prospective non-randomized international study evaluated the Medtronic FREESTYLE® Bioprosthesis at 21 centers with patient follow-up out to three years. A total of 882 patients received the bioprosthesis using three implant techniques: the subcoronary cohort included 640 patients; the full-root cohort included 159 patients; and, the root-inclusion cohort included 83 patients.
### 10.2.1 Inclusion Criteria
Patients who were diagnosed as having sufficient symptoms to warrant isolated replacement of their natural aortic valve or root, or previously implanted prosthetic or bioprosthetic aortic valve were eligible to enter the study. An isolated procedure was defined as one in which only the aortic valve (or root) was replaced, with the three remaining valves being the native valves. Concomitant procedures (other than valve replacement) were permitted.
### 10.2.2 Exclusion Criteria
Patients who required concomitant valve replacement or who already had a pre-existing prosthetic valve in another position were excluded from enrollment in the study.
### 10.2.3 Data Collection
Patients were evaluated preoperatively, within 30 days post-operatively, at 3 and 6 months, and annually. Patients were monitored throughout the postoperative period for possible adverse events. The cumulative follow-up was 1246 patient-years with a mean follow-up of 17 months (SD = 12 months, range = 0 to 42 months). Table 5 describes patient follow-up by implant technique and valve size.
All patients implanted, approved sizes, N = 882
Table 5. Number of Patients by Duration of Follow-up and Implant Technique
| Implant Technique | Duration of follow-up (years) | | | | |
| --- | --- | --- | --- | --- | --- |
| | 0 | 1 | 2 | 3 | 3.5 |
| | Number of patients by technique by year | | | | |
| Total | 882 | 552 | 348 | 96 | 1 |
| Subcoronary | 640 | 416 | 255 | 58 | 1 |
| Full-Root | 159 | 81 | 48 | 20 | 0 |
| Root-Inclusion | 83 | 55 | 45 | 18 | 0 |
| | | | | | |
| Size | Number of patients by size by year | | | | |
| 19mm | 36 | 22 | 12 | 4 | 0 |
| 21mm | 149 | 91 | 57 | 13 | 0 |
| 23mm | 250 | 156 | 99 | 26 | 1 |
| 25mm | 236 | 153 | 99 | 31 | 0 |
| 27mm | 212 | 130 | 81 | 22 | 0 |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
2H
{16}
# 10.3 Analysis for Gender Bias
Study inclusion and exclusion criteria were designed and the study carried out to avoid gender bias in patient enrollment. Of all patients enrolled, 335 of 631 (53%) were male. This proportion of males (335/296 = 1.1) is consistent with the male to female incidence of patients presenting for valve replacement in the US and Canada.
Based on risk factors analyses, the underlying distribution of complications did not vary by gender. Comparing EOA, mean and peak systolic gradient, and cardiac index between females and males, there was no statistically significant difference in hemodynamic performance based on gender. Hence, the results presented in the following analyses are representative for both men and women.
# 10.4 Description of Patients
The subcoronary cohort included 640 patients implanted with the bioprosthesis. Five patients had perioperative explants and 4 patients had multiple prostheses (mitral valve), and were not included in the analysis. Of the remaining 631 patients, 335 were men and 296 were women. Cumulative follow-up was 913 patient-years with mean follow-up of 1.4 years per patient (SD = 0.9 years, range = 0 to 3.5 years). Table 6 presents patient characteristics for the subcoronary technique.
Table 6. Patient Characteristics: Subcoronary Technique; All patients analyzed, N = 631
| Age at implant in years (mean ± SD, N [min, max] | 71 ± 8, 631 [32,91] |
| --- | --- |
| Gender (% male / % female) | 53% / 47% |
| Etiology | |
| Stenosis - percent of patients with stenosis alone [% (number in subgroup / N)] | 43% (272 / 631) |
| Insufficiency - percent of patients with insufficiency alone [% (number in subgroup / N)] | 6% (39 / 631) |
| Mixed - percent of patients with stenosis and insufficiency [% (number in subgroup / N)] | 50% (318 / 631) |
| Other¹ - percent of patients with etiology other than stenosis or insufficiency [% (number in subgroup / N)] | 0% (2 / 631) |
¹ One patient had incidental replacement of a previously implanted prosthesis and one patient had endocarditis without lesion.
The full-root cohort included 159 patients implanted with the bioprosthesis. Two patients had multiple implants (mitral valve) and were not included in the analysis. Of the remaining 157 patients, 80 were male and 77 were female. Cumulative follow-up was 189 patient-years with mean follow-up of 1.2 years per patient (SD = 1.1 years, range = 0 to 3.2 years). Table 7 presents patient characteristics for the full-root technique.
Table 7. Patient Characteristics: Full-Root Technique; All patients analyzed, N = 157
| Age at implant in years (mean ± SD, N [min, max] | 70 ± 8, 157 [38,87] |
| --- | --- |
| Gender (% male / % female) | 51% / 49% |
| Etiology | |
| Stenosis - percent of patients with stenosis alone [% (number in subgroup / N)] | 31% (49 / 157) |
| Insufficiency - percent of patients with insufficiency alone [% (number in subgroup / N)] | 21% (33 / 157) |
| Mixed - percent of patients with stenosis and insufficiency [% (number in subgroup / N)] | 47% (74 / 157) |
| Other¹ - percent of patients with etiology other than stenosis or insufficiency [% (number in subgroup / N)] | 1% (1 / 157) |
¹ One patient had a dissecting ascending aortic aneurysm.
The root-inclusion cohort included 83 patients implanted with the bioprosthesis. Three patients had perioperative explants and were excluded from the analysis. Of the remaining 80 patients, 58 were male and 22 were female. Cumulative follow-up was 139 patient-years with mean follow-up of 1.7 years per patient (SD = 1.1 years, range = 0 to 3.3 years). Table 8 presents patient characteristics for the root-inclusion technique.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 15
{17}
Table 8. Patient Characteristics: Root-Inclusion Technique; All patients analyzed, $N = 80$
| Age at implant in years (mean ± SD, N [min, max] | 69 ± 9, 80 [40, 90] |
| --- | --- |
| Gender (% male / % female) | 73% / 28% |
| Etiology | |
| Stenosis - percent of patients with stenosis alone [% (number in subgroup / N)] | 41% (33 / 80) |
| Insufficiency - percent of patients with insufficiency alone [% (number in subgroup / N)] | 11% (9 / 80) |
| Mixed - percent of patients with stenosis and insufficiency [% (number in subgroup / N)] | 48% (38 / 80) |
## 10.4 Data Analyses
### 10.4.1 Functional NYHA
At baseline, 75% of the 631 Subcoronary patients were NYHA Classes III and IV. At one year follow-up, 8 patients were in Classes III and IV, or 2% of the 454 patients evaluated at one year. For the 157 Full-Root Technique patients, 78% were in Classes III and IV at baseline; none of the 87 patients evaluated at one year were in Classes III and IV. For the 80 Root-Inclusion Technique patients, 69% were in Classes III and IV at baseline; None of the 54 patients evaluated at one year were in Classes III and IV. These results by implant technique are presented in tables 9, 10 and 11.
Table 9. Effectiveness Outcomes, Functional NYHA: Subcoronary Technique
All patients analyzed: $N = 631$, mean ± SD (number), percent (numerator / N)
| Endpoint | Preoperative | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Functional NYHA | 2.9 ± 0.6 (631) | 1.2 ± 0.5 (525) | 1.2 ± 0.5 (454) |
| I - % of patients in NYHA class I | 2% (14 / 631) | 80% (419 / 525) | 84% (380 / 454) |
| II - % of patients in NYHA class II | 20% (129 / 631) | 18% (96 / 525) | 14% (64 / 454) |
| III - % of patients in NYHA class III | 61% (400 / 631) | 2% (10 / 525) | 2% (8 / 454) |
| IV - % of patients in NYHA class IV | 14% (88 / 631) | 0% (0 / 525) | 0% (2 / 454) |
Table 10. Effectiveness Outcomes, Functional NYHA: Full-Root Technique
All patients analyzed: $N = 157$, mean ± SD (number), percent (numerator / N)
| Endpoint | Preoperative | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Functional NYHA | 2.9 ± 0.7 (155) | 1.1 ± 0.2 (114) | 1.1 ± 0.3 (87) |
| I - % of patients in NYHA class I | 1% (2 / 155) | 94% (107 / 114) | 91% (79 / 87) |
| II - % of patients in NYHA class II | 20% (31 / 155) | 6% (7 / 114) | 9% (8 / 87) |
| III - % of patients in NYHA class III | 61% (95 / 155) | 0% (0 / 114) | 0% (0 / 87) |
| IV - % of patients in NYHA class IV | 17% (27 / 155) | 0% (0 / 114) | 0% (0 / 87) |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 16
{18}
Table 11. Effectiveness Outcomes, Functional NYHA: Root-Inclusion Technique
All patients analyzed: N = 80, mean ± SD (number), percent (numerator / N)
| Endpoint | Preoperative | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Functional NYHA | 2.7 ± 0.6 (80) | 1.2 ± 0.6 (60) | 1.1 ± 0.3 (54) |
| I - % of patients in NYHA class I | 5% (4 / 80) | 87% (52 / 60) | 91% (49 / 54) |
| II - % of patients in NYHA class II | 26% (21 / 80) | 10% (6 / 60) | 9% (5 / 54) |
| III - % of patients in NYHA class III | 68% (54 / 80) | 3% (2 / 60) | 0% (0 / 54) |
| IV - % of patients in NYHA class IV | 1% (1 / 80) | 0% (0 / 60) | 0% (0 / 54) |
## 10.4.2 Hemodynamics, Valvular Regurgitation
The published literature reports 63.4% of patients (2 articles) with no regurgitation postoperatively during the first month or “in-hospital”; 38% of patients (1 article) with no regurgitation at 1 year; and, 7.3% of patients (1 article) with moderate to severe regurgitation at 1 year (or the last evaluation, or within 1 to 81 months). The results for the Medtronic FREESTYLE® Bioprosthesis by implant technique are presented in tables 12, 13 and 14, and indicate that for all implant techniques, valve leakage was trivial or mild and did not increase in severity over time.
Table 12. Effectiveness Outcomes, Hemodynamics, Valvular Regurgitation: Subcoronary Technique
All patients analyzed: N = 631, mean ± SD (number), percent (numerator / N)
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Valvular Regurgitation^{1} | 0.3 ± 0.4 (552) | 0.3 ± 0.5 (523) | 0.3 ± 0.4 (456) |
| 0% of patients with no Rg. | 65% (358/ 552) | 62% (326 / 523) | 65% (296 / 456) |
| < 1+ % of patients with < mild Rg. | 15% (82 / 552) | 17% (87 / 523) | 20% (92 / 456) |
| 1+ % of patients with mild Rg. | 20% (108 / 552) | 19% (101 / 523) | 13% (61 / 456) |
| 2+ % of patients with mod Rg. | 1% (3 / 552) | 2% (8 / 523) | 1% (5 / 456) |
| 3+ / 4+% of patients with mod/severe Rg. | 0% (1 / 552) | 0% (1 / 523) | 0% (2 / 456) |
Rg. = Regurgitation
1 The data reflect regurgitation noted at all locations combined. Data coded as “trivial/mild” were included in the category “< 1+ , < mild regurgitation.” Data in the category “< 1+ ” were coded as “0.5” for the calculation of mean ± SD.
Table 13. Effectiveness Outcomes, Hemodynamics, Valvular Regurgitation: Full-Root Technique
All patients analyzed: N = 157, mean ± SD (number), percent (numerator / N)
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Valvular Regurgitation^{1} | 0.1 ± 0.3 (131) | 0.1 ± 0.3 (113) | 0.1 ± 0.2 (90) |
| 0% of patients with no Rg. | 86% (113/ 131) | 87% (98 / 113) | 87% (78 / 90) |
| < 1+ % of patients with < mild Rg. | 8% (10 / 131) | 8% (9 / 113) | 11% (10 / 90) |
| 1+ % of patients with mild Rg. | 5% (7 / 131) | 4% (5 / 113) | 2% (2 / 90) |
| 2+ % of patients with mod Rg. | 1% (1 / 131) | 1% (1 / 113) | 0% (0 / 90) |
| 3+ / 4+% of patients with mod/severe Rg. | 0% (0 / 131) | 0% (0 / 113) | 0% (0 / 90) |
Rg. = Regurgitation
1 See footnote for Table 12.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 17
{19}
Table 14. Effectiveness Outcomes, Hemodynamics, Valvular Regurgitation: Root-Inclusion Technique
All patients analyzed: N = 80, mean ± SD (number), percent (numerator / N)
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Valvular Regurgitation^{†} | 0.2 ± 0.4 (75) | 0.1 ± 0.3 (63) | 0.2 ± 0.3 (51) |
| 0% of patients with no Rg. | 77% (58/ 75) | 79% (50 / 63) | 76% (39 / 51) |
| < 1+ % of patients with < mild Rg. | 15% (11 / 75) | 14% (9 / 63) | 16% (8 / 51) |
| 1+ % of patients with mild Rg. | 5% (4 / 75) | 6% (4 / 63) | 8% (4 / 51) |
| 2+ % of patients with mod Rg. | 3% (2 / 75) | 0% (0 / 63) | 0% (0 / 51) |
| 3+ / 4+% of patients with mod/severe Rg. | 0% (0 / 75) | 0% (0 / 63) | 0% (0 / 51) |
Rg. = Regurgitation
See footnote for Table 12.
## 10.4.3 Hemodynamics, Mean Pressure Gradient
The FDA selected literature articles report 77% of patients (1 article) with a mean gradient < 10 mm Hg postoperatively for stented valves. The results for the Medtronic FREESTYLE® Bioprosthesis by implant technique are presented in tables 15, 16, and 17, and indicate improved gradients over time for all implant techniques and most valve sizes within a technique, nearing physiologic values at one year.
Table 15. Effectiveness Outcomes, Hemodynamics, Mean Pressure Gradient: Subcoronary Technique
All patients analyzed: N = 631, number in subgroup/N, mean ± SD [min, max]
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Mean Pressure Gradient (mmHg) | | | |
| 19mm | 19/27, 17.6 ± 7.6 [8.0, 42.0] | 19/27, 12.0 ± 5.7 [2.0, 23.0] | 18/27, 11.7 ± 4.7 [5.0, 19.0] |
| 21mm | 100/117, 14.6 ± 8.1 [1.0, 48.0] | 91/117, 9.0 ± 6.2 [1.0, 47.0] | 83/117, 9.8 ± 7.4 [0.8, 51.0] |
| 23mm | 165/191, 12.9 ± 6.9 [2.0, 39.0] | 156/191, 8.9 ± 5.9 [1.0, 35.0] | 138/191, 8.8 ± 6.8 [0.0, 57.0] |
| 25mm | 145/167, 9.1 ± 4.6 [1.0, 28.0] | 146/167, 5.5 ± 3.3 [1.0, 19.0] | 119/167, 5.1 ± 3.3 [0.0, 18.0] |
| 27mm | 112/129, 7.3 ± 4.1 [1.0, 22.0] | 110/129, 5.0 ± 3.7 [1.0, 30.0] | 92/129, 4.4 ± 2.9 [0.7, 13.0] |
Table 16. Effectiveness Outcomes, Hemodynamics, Mean Pressure Gradient: Full-Root Technique
All patients analyzed: N = 157, number in subgroup/N, mean ± SD [min, max]
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Mean Pressure Gradient (mmHg) | | | |
| 19mm | 7/7, 11.9 ± 6.0 [6.0, 23.0] | 7/7, 13.4 ± 6.7 [7.0, 27.0] | 5/7, 16.8 ± 7.4 [9.0, 27.1] |
| 21mm | 14/21, 7.8 ± 3.4 [2.0, 14.0] | 12/21, 7.6 ± 3.5 [2.0, 14.0] | 9/21, 7.2 ± 4.0 [3.0, 14.0] |
| 23mm | 20/32, 6.8 ± 3.8 [0.9, 13.0] | 20/32, 5.4 ± 3.4 [1.0, 12.0] | 13/32, 7.1 ± 3.7 [1.6, 14.0] |
| 25mm | 31/35, 4.9 ± 2.9 [1.0, 12.0] | 29/35, 4.2 ± 2.9 [1.0, 12.0] | 23/35, 4.5 ± 3.4 [1.0, 17.0] |
| 27mm | 59/62, 4.4 ± 3.4 [1.0, 20.0] | 46/62, 3.8 ± 2.9 [0.7, 15.0] | 38/62, 3.2 ± 1.8 [1.0, 8.0] |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 18
{20}
Table 17. Effectiveness Outcomes, Hemodynamics, Mean Pressure Gradient: Root-Inclusion Technique
All patients analyzed: N = 80, number in subgroup/N, mean ± SD [min, max]
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Mean Pressure Gradient (mmHg) | | | |
| 21mm | 4/5, 12.3 ± 7.0 [3.0, 20.0] | 3/5, 8.0 ± 2.6 [6.0, 11.0] | 2/5, 9.0 ± 1.4 [8.0, 10.0] |
| 23mm | 21/25, 13.8 ± 10.7 [2.0, 52.0] | 17/25, 10.3 ± 12.1 [2.0, 54.0] | 14/25, 8.7 ± 8.8 [2.0, 36.0] |
| 25mm | 29/30, 9.6 ± 6.7 [2.0, 28.0] | 25/30, 8.2 ± 6.7 [2.0, 29.0] | 17/30, 7.6 ± 6.8 [2.0, 29.0] |
| 27mm | 19/20, 6.4 ± 3.4 [2.0, 14.0] | 18/20, 4.4 ± 2.4 [1.0, 8.0] | 15/20, 3.7 ± 2.2 [1.0, 8.0] |
## 10.4.4 Hemodynamics, Effective Orifice Area
The results for the Medtronic FREESTYLE® Bioprosthesis by implant technique are presented in tables 18, 19, and 20, and indicate improved effective orifice area over time for all implant techniques and most valve sizes within a technique, when compared to stented valves in FDA selected literature articles, and emphasize a low flow impediment.
Table 18. Effectiveness Outcomes, Hemodynamics, Effective Orifice Area: Subcoronary Technique
All patients analyzed: N = 631, number in subgroup/N, mean ± SD [min, max]
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Mean Pressure Gradient (mmHg) | | | |
| 19mm | 19/27, 0.9 ± 0.2 [0.5, 1.4] | 19/27, 1.1 ± 0.3 [0.6, 1.7] | 18/27, 1.1 ± 0.3 [0.7, 1.7] |
| 21mm | 97/117, 1.3 ± 0.4 [0.5, 2.4] | 91/117, 1.5 ± 0.5 [0.8, 4.3] | 82/117, 1.4 ± 0.4 [0.4, 3.1] |
| 23mm | 160/191, 1.4 ± 0.5 [0.5, 3.7] | 154/191, 1.7 ± 0.5 [0.6, 3.6] | 137/191, 1.7 ± 0.5 [0.8, 3.9] |
| 25mm | 143/167, 1.8 ± 0.6 [0.4, 3.9] | 146/167, 2.0 ± 0.5 [0.9, 3.5] | 119/167, 2.0 ± 0.5 [0.8, 3.5] |
| 27mm | 110/129, 2.2 ± 0.7 [0.8, 5.1] | 109/129, 2.4 ± 0.6 [1.2, 4.1] | 92/129, 2.5 ± 0.7 [1.3, 4.4] |
Table 19. Effectiveness Outcomes, Hemodynamics, Effective Orifice Area: Full-Root Technique
All patients analyzed: N = 157, number in subgroup/N, mean ± SD [min, max]
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Mean Pressure Gradient (mmHg) | | | |
| 19mm | 7/7, 1.3 ± 0.4 [0.5, 1.7] | 7/7, 1.1 ± 0.3 [0.6, 1.5] | 5/7, 1.0 ± 0.1 [0.9, 1.1] |
| 21mm | 14/21, 1.5 ± 0.6 [0.9, 3.3] | 12/21, 1.7 ± 0.4 [1.2, 2.3] | 9/21, 1.4 ± 0.3 [1.1, 1.9] |
| 23mm | 19/32, 1.7 ± 0.4 [0.7, 2.4] | 19/32, 1.8 ± 0.5 [1.2, 3.0] | 13/32, 2.0 ± 0.5 [1.5, 3.3] |
| 25mm | 31/35, 2.1 ± 0.6 [1.1, 3.8] | 29/35, 2.1 ± 0.5 [1.1, 3.8] | 24/35, 2.2 ± 0.6 [1.3, 3.5] |
| 27mm | 59/62, 2.1 ± 0.6 [1.1, 4.0] | 46/62, 2.3 ± 0.5 [1.3, 3.3] | 37/62, 2.4 ± 0.7 [1.3, 4.3] |
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 19
{21}
Table 20. Effectiveness Outcomes, Hemodynamics, Effective Orifice Area: Root-Inclusion Technique
All patients analyzed: N = 80, number in subgroup/N, mean ± SD [min, max]
| Endpoint | < 30 Days | 3 - 6 Months | One Year |
| --- | --- | --- | --- |
| Mean Pressure Gradient (mmHg) | | | |
| 21mm | 4/5, 1.6 ± 0.8 [1.0, 2.8] | 2/5, 1.4 ± 0.2 [1.3, 1.5] | 2/5, 1.0 ± 0.1 [0.9, 1.2] |
| 23mm | 21/25, 1.6 ± 0.6 [0.8, 3.0] | 16/25, 1.8 ± 0.5 [0.9, 3.0] | 14/25, 1.9 ± 0.4 [1.1, 3.1] |
| 25mm | 29/30, 1.9 ± 0.8 [0.9, 3.4] | 22/30, 2.2 ± 0.8 [1.2, 3.8] | 17/30, 2.2 ± 0.8 [1.0, 3.6] |
| 27mm | 19/20, 2.4 ± 0.7 [1.4, 4.0] | 18/20, 2.7 ± 0.7 [1.8, 4.7] | 15/20, 3.0 ± 0.9 [1.7, 5.0] |
## 11. Conclusions Drawn from Studies
The *in vivo* and *in vitro* performance testing for biocompatibility, immunogenicity, toxicity, steady and pulsatile flow pressure drop, pulsatile and dynamic regurgitation, flow visualization, and accelerated wear/hydrodynamic assessment demonstrates that the Medtronic FREESTYLE® Bioprosthesis meets specifications, performs comparably to the aortic Hancock Standard control valve, and suggests that this device is suitable for long-term implant.
The acute and chronic animal studies in sheep demonstrated that hemodynamic performance and handling were similar to allograft valves, and that the AOA treatment reduced the extent of leaflet calcification when compared to non-AOA treated valves, and commercially available porcine valves. However, inhibition of aortic wall mineralization was not observed. A short-term subdermal implant study (0 - 4 months) in rats using AOA-treated and non-treated valves demonstrated that the AOA-treated samples contained less calcium than the untreated. Calcification of the aortic wall is a concern for the bioprostheses implanted by the root-inclusion and full-root techniques. Since the animal studies did not determine the likelihood and effect of aortic wall calcification, further clinical evaluation is necessary.
The results of the clinical studies submitted in the PMA suggest that the performance of the Medtronic FREESTYLE® Bioprosthesis using the Subcoronary Technique (n = 640 patients) is comparable to stented valve prostheses. A similar assessment cannot be made for the Full-Root Technique (n = 159 patients) or the Root-Inclusion Technique (n = 83 patients) as there were fewer patients implanted using these techniques. However, the clinical experience with the Full-Root and Root-Inclusion Techniques is considered sufficient because the bioprosthesis is provided as a single configuration that the implant team modifies to the desired configuration, and to match the patient’s anatomy, prior to insertion by one of the three implant techniques. The labeling will state that there is limited data for the Full-Root Technique in the 19mm, 21mm and 23mm sizes and the Root-Inclusion Technique in the 19mm and 21mm sizes.
This short term study did not reveal calcification of the aortic wall, however, this becomes a concern after longer implant duration. Further clinical evaluation will be necessary to address this issue. The labeling will state that no clinical data are available which evaluate the long-term impact of the AOA treatment in patients.
Therefore, it is reasonable to conclude that the benefits of use of the device for the target population outweigh the risk of illness or injury when used as indicated in accordance with the directions for use.
## 13. Panel Recommendations
The panel recommended to approve all sizes and configurations of the device with the following conditions: a post approval study shall be conducted to assess hemodynamic performance to at least 8 years, with
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 20
{22}
annual clinical assessment which should include NYHA classification, anticoagulant and antiplatelet drugs used, and murmur assessment, with a focus on thromboembolic complications (correlated with age), reoperation, and explant analysis. Physician training was also stipulated, as were labeling refinements to the warnings.
## 14. FDA Decision
FDA agreed with the recommendations of the Panel, and recommended approval of the 19, 21, 23, 25, and 27 mm sizes for the subcoronary, full-root and root inclusion implant techniques. The conditions of approval include two post-approval studies: one to assess long-term bioprosthesis performance for a period of 10 years in 100 patients currently implanted with the bioprosthesis, with aggressive follow-up for safety and effectiveness outcomes using the IDE protocol or an alternate protocol agreed to by FDA; and, a study involving new patients to assess valve-related death, explant, and re-operation, using the tracking database or other appropriate method of determining the implant population (denominator). The focus of these studies is failure mode and calcification to provide the data necessary for future labeling updates. The third condition of approval asked for additional bench testing of the device in a compliant chamber in an attempt to predict performance in younger patients. The applicant agreed to these three conditions and further agreed to implement a physician training program and to modify the labeling as suggested by FDA. FDA issued an approval order on November 26, 1997. The applicant’s manufacturing facility was inspected and was found to be in compliance with the device Good Manufacturing Practice regulations.
## 15. Approval Specifications
Directions for use: See the labeling.
Hazards to health from use of the device: See indications, contraindications, warnings, precautions and adverse events in the labeling.
Postapproval requirements and restrictions: See approval order.
The Approval Order, Summary of Safety and Effectiveness Data, and labeling can be found on the Internet at http://www.fda.gov/cdrh/pmapage.html.
Summary of Safety and Effectiveness, Medtronic FREESTYLE® Bioprosthesis P970031
Page 21
30
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.