MEDTRONIC COREVALVE SYSTEM

P130021 · Medtronic, Inc. · NPT · Jan 17, 2014 · Cardiovascular

Device Facts

Record IDP130021
Device NameMEDTRONIC COREVALVE SYSTEM
ApplicantMedtronic, Inc.
Product CodeNPT · Cardiovascular
Decision DateJan 17, 2014
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The Medtronic CoreValve™ System is indicated for relief of aortic stenosis in patients with symptomatic heart disease due to severe native calcific aortic stenosis (aortic valve area ≤ 0.8 cm², a mean aortic valve gradient of > 40 mmHg, or a peak aortic-jet velocity of > 4.0 m/s) and with native aortic annulus diameters between 18 and 29 mm who are judged by a heart team, including a cardiac surgeon, to be at extreme risk or inoperable for open surgical therapy (predicted risk of operative mortality and/or serious irreversible morbidity ≥50% at 30 days).

Device Story

The Medtronic CoreValve System is a transcatheter aortic valve replacement (TAVR) device used to treat severe aortic stenosis without open-heart surgery. The system comprises a self-expanding Nitinol frame with porcine pericardial leaflets, a delivery catheter system (DCS) with an AccuTrak stability layer, and a compression loading system (CLS). The device is delivered percutaneously (iliofemoral or non-iliofemoral access) by an interventional cardiologist and cardiac surgeon team. The TAV is compressed into the DCS, tracked to the native aortic annulus, and deployed. The device functions by replacing the native valve's function, improving hemodynamic performance (EOA, mean gradient) and patient quality of life. Clinical output is monitored via echocardiography. The device benefits patients deemed inoperable for traditional surgery by reducing mortality and improving functional status (NYHA class).

Clinical Evidence

Prospective, non-randomized, multi-center pivotal trial (IDE G100012) of 656 extreme-risk patients. Primary endpoint: all-cause mortality or major stroke at 12 months. Iliofemoral cohort (n=489) met performance goal (26.0% vs 43% goal, p<0.0001). Secondary endpoints showed significant improvements in EOA, mean gradient, NYHA class, and KCCQ scores. Adverse events included conduction disturbances requiring permanent pacemaker (26.2% at 12 months in iliofemoral cohort), major bleeding, and vascular complications.

Technological Characteristics

Components: TAV (porcine pericardium leaflets, Nitinol frame, AOA antimineralization treatment), DCS (AccuTrak stability layer), CLS (reduction cones). Materials: Nitinol (ASTM F2063), porcine pericardium. Sterilization: TAV (liquid chemical/glutaraldehyde), DCS/CLS (EtO per ANSI/AAMI/ISO 11135). MR Conditional (1.5T/3T).

Indications for Use

Indicated for patients with symptomatic severe native calcific aortic stenosis, aged 18+ (implied by adult heart disease), who are at extreme risk or inoperable for open surgical aortic valve replacement (predicted mortality/morbidity ≥50% at 30 days). Contraindicated in patients with hypersensitivity to aspirin, heparin, bivalirudin, ticlopidine, clopidogrel, Nitinol (Ti/Ni), or contrast media; ongoing sepsis/endocarditis; or preexisting mechanical aortic valve.

Regulatory Classification

Identification

To replace a patient's aortic heart valve. They are different from the classified device (heart valves) in that they are placed percutaneously and do not require open chest surgery or a cardiotomy for placement.

Related Devices

Submission Summary (Full Text)

{0} SUMMARY OF SAFETY AND EFFECTIVENESS DATA I. GENERAL INFORMATION Device Generic Name: Replacement Heart Valve Device Trade Name: Medtronic CoreValve™ System (MCS): Transcatheter Aortic Valve (TAV), Models MCS-P4-23-AOA (23 mm; CoreValve™ Evolut™), MCS-P3-26-AOA (26 mm), MCS-P3-29-AOA (29 mm), and MCS-P3-31-AOA (31 mm); Delivery Catheter System (DCS), Models DCS-C4-18FR and DCS-C4-18FR-23; and Compression Loading System (CLS), Model CLS-3000-18FR Device Procode: NPT Applicant Name and Address: Medtronic CoreValve LLC 3576 Unocal Place Santa Rosa, CA 95403 Date of Panel Recommendation: None Premarket Approval Application (PMA) Number: P130021 Date of FDA Notice of Approval: January 17, 2014 II. INDICATIONS FOR USE The Medtronic CoreValve™ System is indicated for relief of aortic stenosis in patients with symptomatic heart disease due to severe native calcific aortic stenosis (aortic valve area ≤ 0.8 cm², a mean aortic valve gradient of &gt; 40 mmHg, or a peak aortic-jet velocity of &gt; 4.0 m/s) and with native aortic annulus diameters between 18 and 29 mm who are judged by a heart team, including a cardiac surgeon, to be at extreme risk or inoperable for open surgical therapy (predicted risk of operative mortality and/or serious irreversible morbidity ≥50% at 30 days). III. CONTRAINDICATIONS The Medtronic CoreValve System is contraindicated for patients presenting with any of the following conditions: - known hypersensitivity or contraindication to aspirin, heparin (HIT/HITTS) and bivalirudin, ticlopidine, clopidogrel, Nitinol (Titanium or Nickel), or sensitivity to contrast media, which cannot be adequately premedicated PMA P130021: FDA Summary of Safety and Effectiveness Data Page 1 {1} - ongoing sepsis, including active endocarditis - preexisting mechanical heart valve in aortic position # IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the Medtronic CoreValve System labeling. # V. DEVICE DESCRIPTION The Medtronic CoreValve System (MCS) is designed to replace the native aortic heart valve without open heart surgery and without concomitant surgical removal of the failed native valve. It consists of 3 components: the Transcatheter Aortic Valve (TAV), the Delivery Catheter System (DCS), and the Compression Loading System (CLS). # V.1. Transcatheter Aortic Valve (TAV) The TAV (Figure 1) is manufactured by suturing three valve leaflets and skirt, made from a single layer of porcine pericardium, onto a self-expanding, multi-level, radiopaque frame made of Nitinol. The bioprosthesis is processed with alpha-amino oleic acid $(\mathrm{AOA}^{\text{®}})$ , which is an antimineralization treatment derived from oleic acid, a naturally occurring long-chain fatty acid. ![img-0.jpeg](img-0.jpeg) Figure 1: CoreValve Transcatheter Aortic Valve The TAV is available for a range of aortic annulus and ascending aorta diameters as shown in Table 1. Note that the $23\mathrm{mm}$ TAV has its own device name, called CoreValve™ Evolut™. PMA P130021: FDA Summary of Safety and Effectiveness Data {2} Table 1: Patient Anatomical Diameters | Bioprosthesis Model | Size | Aortic Annulus Diameter | Ascending Aorta Diameter | | --- | --- | --- | --- | | CoreValve™ Evolut™ Bioprosthesis | | | | | MCS-P4-23-AOA | 23 mm | 18 mm–20 mm | ≤34 mm | | CoreValve™ Bioprosthesis | | | | | MCS-P3-26-AOA | 26 mm | 20 mm–23 mm | ≤40 mm | | MCS-P3-29-AOA | 29 mm | 23 mm–26 mm | ≤43 mm | | MCS-P3-31-AOA | 31 mm | 26 mm–29 mm | ≤43 mm | # V.2. Delivery Catheter System with AccuTrak Stability Layer (AccuTrak DCS) The DCS (Figure 2) is used to deploy the TAV. The TAV is loaded within the capsule which features an atraumatic, radiopaque tip and protective sheath. The AccuTrak stability layer is fixed at the handle and extends down the outside of the catheter shaft to provide a barrier between the catheter and vessel walls. The handle features macro and micro adjustment control of the retractable capsule sheath. There are two models of the DCS: model DCS-C4-18FR-23 for the $23\mathrm{mm}$ TAV only and model DCS-C4-18FR for the 26, 29, and $31\mathrm{mm}$ TAVs. ![img-1.jpeg](img-1.jpeg) Figure 2: CoreValve Delivery Catheter System # V.3. Compression Loading System (CLS) The CLS (Figure 3) is a system of reduction cones and tubing designed to compress the TAV to an optimal diameter for manual loading into the DCS. Only one model of the CLS is available, i.e., model CLS-3000-18FR. PMA P130021: FDA Summary of Safety and Effectiveness Data {3} ![img-2.jpeg](img-2.jpeg) Figure 3: CoreValve Compression Loading System The CLS comprises the following elements: 1. Inflow tube (straight tube) 2. Outflow cone 3. Outflow cap 4. Outflow tube (tube with flared ends) 5. Inflow cone # VI. ALTERNATIVE PRACTICES AND PROCEDURES Alternatives for patients deemed to be at extreme risk, or non-operable (non-surgical), for surgical aortic valve replacement include: treatment with other approved transcatheter aortic valve implantation therapy, temporary relief using a percutaneous technique called balloon aortic valvuloplasty (BAV), or medical therapy (no obstruction-relieving intervention). Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. # VII. MARKETING HISTORY The current Medtronic CoreValve System is commercially available in over 50 countries, as listed in Table 2. It has not been withdrawn from marketing for any reason related to its safety or effectiveness. Table 2: Countries where Medtronic CoreValve System is commercialized | Commercially Available | | | | | --- | --- | --- | --- | | Afghanistan | Finland | Moldova | Tajikistan | | Albania | France | Netherlands | Thailand | | Argentina | Georgia | New Zealand | Turkmenistan | | Armenia | Germany | Panama | Turkey | | Austria | Greece | Peru | United Kingdom | | Azerbaijan | Guatemala | Philippines | Croatia | | Belgium | Hong Kong | Poland | Israel | | Belarus | Hungary | Portugal | Ukraine | | Bosnia & Herzegovina | Ireland | Romania | Uruguay | | Brazil | Israel | Russia | Uzbekistan | | Canada | Italy | Saudi Arabia | Venezuela | | Chile | Kazakhstan | Serbia | | PMA P130021: FDA Summary of Safety and Effectiveness Data {4} | Commercially Available | | | | | --- | --- | --- | --- | | Colombia | Kyrgyzstan | Slovakia | | | Croatia | Latvia | Slovenia | | | Cyprus | Lithuania | South Africa | | | Czech Republic | Luxembourg | South Korea | | | Denmark | Malaysia | Spain | | | Dominican Republic | Malta | Sweden | | | Ecuador | Mexico | Switzerland | | | Estonia | Montenegro | Taiwan | | VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Potential risks associated with the implantation of the Medtronic CoreValve System may include, but are not limited to, the following: - death - cardiac arrest - coronary occlusion, obstruction, or vessel spasm (including acute coronary closure) - emergent surgery (e.g., coronary artery bypass, heart valve replacement, valve explant) - multi-organ failure - heart failure - myocardial infarction (MI) - cardiogenic shock - respiratory insufficiency or respiratory failure - cardiovascular injury (including rupture, perforation, or dissection of vessels, ventricle, myocardium, or valvular structures that may require intervention) - perforation of the myocardium or a vessel - ascending aorta trauma - cardiac tamponade - cardiac failure or low cardiac output - prosthetic valve dysfunction including, but not limited to, fracture; bending (out-of-round configuration) of the valve frame; under-expansion of the valve frame; calcification; pannus; leaflet wear, tear, prolapse, or retraction; poor valve coaptation; suture breaks or disruption; leaks; mal-sizing (prosthesis-patient mismatch); malposition (either too high or too low)/malplacement; regurgitation; stenosis - thrombosis/embolus (including valve thrombosis) - valve migration/valve embolization - ancillary device embolization - emergent percutaneous coronary intervention (PCI) - emergent balloon valvuloplasty - major or minor bleeding that may or may not require transfusion or intervention (including life-threatening or disabling bleeding) PMA P130021: FDA Summary of Safety and Effectiveness Data {5} - allergic reaction to antiplatelet agents, contrast medium, or anesthesia - infection (including septicemia and endocarditis) - stroke, transient ischemic attack (TIA), or other neurological deficits - permanent disability - renal insufficiency or renal failure (including acute kidney injury) - mitral valve regurgitation or injury - tissue erosion - vascular access related complications (e.g., dissection, perforation, pain, bleeding, hematoma, pseudoaneurysm, irreversible nerve injury, compartment syndrome, arteriovenous fistula, stenosis) - conduction system disturbances (e.g., atrioventricular node block, left-bundle branch block, asystole), which may require a permanent pacemaker - cardiac arrhythmias - encephalopathy - pulmonary edema - pericardial effusion - pleural effusion - myocardial ischemia - peripheral ischemia - bowel ischemia - heart murmur - hemolysis - cerebral infarction-asymptomatic - non-emergent reoperation - inflammation - fever - hypotension or hypertension - syncope - dyspnea - anemia - angina - abnormal lab values (including electrolyte imbalance) For the specific adverse events that occurred in the clinical study, please see Section 10. IX. SUMMARY OF PRECLINICAL STUDIES A. Laboratory Testing A series of non-clinical laboratory studies were performed on the Medtronic CoreValve System as recommended per ISO 5840: 2005, Cardiovascular implants – Cardiac valve prostheses and relevant FDA Guidance Documents. PMA P130021: FDA Summary of Safety and Effectiveness Data {6} # Biocompatibility Biocompatibility evaluations were completed on the components (TAV, DCS, and CLS) of the Medtronic CoreValve System in accordance with ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and FDA's General Program Memorandum No. G95-1, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." The required testing for each component was determined based on the nature and duration of body contact in accordance with ISO 10993-1:2009. Summaries of the test results for the TAV, DCS, and CLS are provided in Table 3-Table 5, respectively. Table 3: Summary of Medtronic CoreValve Transcatheter Aortic Valve Biocompatibility Testing | Biological Effect per ISO 10993-1 | Test Method | Test Result | | --- | --- | --- | | Cytotoxicity | ISO MEM Elution | Pass | | | ISO Agarose Overlay – Direct contact method | Pass | | Sensitization | ISO Guinea Pig Maximization Sensitization Test | Pass | | Irritation | Intracutaneous Irritation Study in Rabbits | Pass | | (Acute) Systemic Toxicity | Systemic Toxicity in Mice | Pass | | | USP Pyrogen Study, Material Mediated | Pass | | Hemocompatibility | ASTM Hemolysis | Pass | | | Partial Thromboplastin Time (PTT) | Pass | | | Complement Activation (C3a, SC5b-9) | Pass | | | In vivo Thrombogenicity in Porcine Model | Pass | | Genotoxicity | Bacterial Reverse Mutation Study | Pass | | | Chromosomal Aberration study in Mammalian Cells | Pass | | | Mouse Peripheral Blood Micronucleus Study | Pass | | Subacute/Subchronic Toxicity | 4-week Systemic Toxicity Study in Rats following Subcutaneous Implantation | Pass | | | 13-week Systemic Toxicity Study in Rats following Subcutaneous Implantation | Pass | | Chronic Toxicity | Chronic toxicity was evaluated as part of the in vivo animal studies | Pass | | Carcinogenicity | As the TAV is made of well-characterized materials and the results from the aforementioned genotoxicity studies demonstrated no mutagenic response, carcinogenicity testing was not conducted. | Not Required | | Biodegradation | The materials used in MCS have no known absorption, distribution, biotransformation, or leachable elimination properties that make them a candidate for this test procedure. Therefore, biodegradation testing was not deemed necessary. | Not Required | PMA P130021: FDA Summary of Safety and Effectiveness Data {7} | Biological Effect per ISO 10993-1 | Test Method | Test Result | | --- | --- | --- | | Reproductive/Developmental Toxicity | The MCS does not have any potential impact on the reproductive potential of the patient, hence this test was not deemed necessary. | Not Required | Table 4: Summary of Medtronic CoreValve Delivery Catheter System Biocompatibility Testing | Biological Effect per ISO 10993-1 | Test Method | Test Result | | --- | --- | --- | | Cytotoxicity | ISO MEM Elution | Pass | | Sensitization | ISO Guinea Pig Maximization Sensitization Test | Pass | | Irritation | Intracutaneous Irritation Study in Rabbits | Pass | | (Acute) Systemic Toxicity | Systemic Toxicity in Mice | Pass | | | USP Pyrogen Study, Material Mediated | Pass | | Hemocompatibility | ASTM Hemolysis | Pass | | | Partial Thromboplastin Time (PTT) | Pass | | | Complement Activation (C3a, SC5b-9) | Pass | | | In vivo Thrombogenicity in Porcine Model | Pass | Table 5: Summary of Medtronic CoreValve Compression Loading System Biocompatibility Testing | Biological Effect per ISO 10993-1 | Test Method | Test Result | | --- | --- | --- | | Cytotoxicity | ISO MEM Elution | Pass | | Hemocompatibility | Modified ASTM Hemolysis (direct contact and extract method) | Pass | | (Acute) Systemic Toxicity | USP Pyrogen Study, Material Mediated | Pass | ## Bench Testing Medtronic conducted comprehensive preclinical bench testing and computational analysis on the Medtronic CoreValve System, including the TAV, the DCS, and the CLS. All testing was conducted in accordance with national and international standards and FDA guidance documents. Testing verified that all components of the Medtronic CoreValve System met its product performance and design specifications. The tests are summarized in Table 6. PMA P130021: FDA Summary of Safety and Effectiveness Data {8} Table 6: Summary of In Vitro Studies for Medtronic CoreValve System (MCS) | Test | Applicable Standards | Test Description | Results | | --- | --- | --- | --- | | Transcatheter Aortic Valve (TAV) | | | | | Frame Raw Material Analysis | ASTM F2063-05, ASTM F2633, ASTM F2516-07, ASTM E8 | This test verified that the incoming raw materials conform to chemical and mechanical property requirements of the MCS TAV frame. | Pass | | Frame Mechanical Property Characterization of Post-Processed Material | ASTM F2516-07, ASTM E8 | This test characterized the mechanical properties of the Nitinol tubing of the MCS TAV frames. | NA – Characterization Testing | | Corrosion Testing | ISO 5840: 2005, ASTM F2129-08 | This test evaluated the corrosion resistance of the MCS TAV in accordance with ASTMF2129 | Pass | | Mechanical Characterization of Porcine Pericardium | ASTM 2063 | This test characterized the mechanical properties of the MCS TAV porcine pericardium. | NA – Characterization Testing | | Dimensional Verification | FDA Guidance Document for Intravascular Stents | This test verified that the dimensions of the MCS TAV frame are within specified requirements. | Pass | | Transformation Temperature, \(A_t\) | ASTM 2082-02 | This test verified that the MCS TAV frames conform to the required Af temperature specification. | Pass | | Frame Radial Force Characterization | EN ISO 14299: 2004, ISO 5480: 2005, FDA Guidance Document for Intravascular Stents | This test characterized the frame radial force of the MCS TAV frame. | NA – Characterization Testing | | Magnetic Resonance Imaging | ASTM F2052-06, ASTM F2503-08, ASTM F2213-06, ASTM F2119-07, ASTM F2182-11a | This test characterized the performance of the MCS TAV in an MR field and determined the compatibility. The following is in the IFU: Nonclinical testing and modeling has demonstrated that the Medtronic CoreValve bioprosthesis is MR Conditional. It can be scanned safely under the following conditions: Static magnetic field of 1.5 tesla and 3 tesla Spatial gradient field of 2500 gauss/cm Normal operating mode only with a maximum whole body SAR of 2.0 W/kg for 15 minutes as read from equipment monitor | Pass | | Radiopacity | ISO 5840: 2005, ISO 25539-1: 2003, FDA Guidance Document for Intravascular Stents | This test evaluated the ability to visualize the MCS TAV and DCS under standard imaging. | Pass | | Finite Element Analysis (FEA) | None | FEA was used to characterize the structural behavior of the MCS TAV frame under in vivo operational conditions. | NA – Characterization Testing | PMA P130021: FDA Summary of Safety and Effectiveness Data {9} PMA P130021: FDA Summary of Safety and Effectiveness Data Page 10 | Test | Applicable Standards | Test Description | Results | | --- | --- | --- | --- | | Device Level Fatigue Testing of TAV Frames (600M) | ISO 5840: 2005, FDA Guidance Document for Heart Valves | This test evaluated the MCS TAV frame fatigue resistance to 600 Million cycles. | NA – Characterization Testing | | Material Fatigue Testing (600M) | ISO 5840: 2005, FDA Guidance Document for Heart Valves | This test determined the Nitinol material fatigue limit using representative material test coupons. | NA – Characterization Testing | | Hydrodynamic Testing | ISO 5840: 2005, FDA Guidance Document for Heart Valves | This test evaluated the hydrodynamic performance of the MCS TAV in round and out of round conditions compared against a commercially approved surgical valve. Pulsatile Flow Test Flow Visualization Test Verification of Bernoulli Relationship | Pass | | Accelerated Wear Testing | ISO 5840: 2005, FDA Guidance Document for Heart Valves | This test evaluated the structural durability of the MCS TAV in round and out of round conditions compared against commercially approved surgical valve to 200 Million cycles. | Pass | | Dynamic Failure Mode | ISO 5840: 2005, FDA Guidance Document for Heart Valves | This test induced valve failure to determine the primary mode and location of failure of the MCS TAV. | NA – Characterization Testing | | Migration | ISO 5840: 2005, FDA Guidance Document for Heart Valves | This test evaluated the migration resistance of the MCS TAV. | Pass | | Delivery Catheter System (DCS) | | | | | Surface Finish Examination/ Dimensional Conformations | ISO 10555-1 (Amd 2, 2004), ISO 25539-1: 2009, FDA Guidance Document for Intravascular Stents | This test verified that the surfaces & dimensions of the MCS DCS meet specification. | Pass. | | Bond/Tubing Tensile Strengths | ISO 10555-1 (Amd 2, 2004), ISO 25539-1: 2009, FDA Guidance Document for Intravascular Stents | This test verified that the bonds and tubing of the MCS DCS meet the strength specifications. | Pass | | Catheter Compressive Strength | ISO 25539-1: 2003(E) | This test verified that the MCS DCS can withstand the forces necessary to deliver the TAV to the treatment site. | Pass | | Kink Resistance | ISO 25539-1: 2003(E) | This test verified the ability of the MCS DCS to accommodate the curvature encountered during clinical use. | Pass | | Flushability | ISO 25539-1: 2009 | This test verified the ability of the MCS DCS to be purged. | Pass | {10} | Test | Applicable Standards | Test Description | Results | | --- | --- | --- | --- | | Corrosion Resistance | ISO 10555-1 (Amd 2, 2004) | This test verified the corrosion resistance of the metallic components of the MCS DCS. | Pass | | Macro and Micro Controls | ISO 25539-1: 2009 | This test verified the macro and micro controls of the MCS DCS handle function as intended. | Pass | | Guidewire Verification / Introducer compatibility | ISO 25539-1: 2009 | This test verified the compatibility with a 0.035" guidewire and 18Fr introducer sheath. | Pass | | Hemostasis | ISO 25539-1: 2009, ISO 11070: 1998 | This test determined the ability of the MCS DCS components to maintain hemostasis. | Pass | | Cather Loading System (CLS) | | | | | Dimensional Verification | None | This test verified that the components of the MCS CLS meet dimensional specifications. | Pass | | MCS System Testing | | | | | Deployment Accuracy | ISO 25539-1: 2009 | This test verified the deployment accuracy of the MCS DCS when used with the TAV. | Pass | | Systems Deployment Force Testing | ISO 25539-1: 2009 | This test evaluated the system's ability to load and characterize the deployment force. | Pass | | Torque Characterization | ISO 25539-1: 2009 | This test characterized the maximum torque that may be applied to the MCS DCS. | NA – Characterization Testing | | TAV Device Foreshortening | ASTM F2081-06, ISO 25539-1:2009, FDA Guidance Document for Intravascular Stents | This test determined the relationship between the MCS TAV frame length and diameter when crimped and deployed. | NA – Characterization Testing | | Frame & Valve Integrity post-Tracking and Deployment | ISO 25539-1: 2009, FDA Guidance Document for Heart Valves, FDA Guidance Document for Intravascular Stents | This test evaluated the effects of crimping, tracking, and deployment on MCS TAV frame and valve integrity. | Pass | | System Usability | ISO 25539-1:2003(E), ANSI/AAMI HE74:2001, BS EN 62366:2008 | This test assessed the user's ability to use the MCS DCS with TAV and CLS. | Pass | PMA P130021: FDA Summary of Safety and Effectiveness Data Page 11 {11} B. Animal Studies Four animal studies were performed in support of the safety and performance of the current Medtronic CoreValve System (MCS). Two of those four studies were conducted to evaluate the chronic in vivo safety and performance of the MCS TAV in an ovine and a porcine model, respectively. The other two studies were simulated use evaluation of the performance of models DCS-C4-18FR and DCS-C4-18FR-23 of the AccuTrak DCS using an in vivo porcine model. These studies are summarized in Table 7. PMA P130021: FDA Summary of Safety and Effectiveness Data Page 12 {12} Table 7: Summary of In Vivo Studies for Medtronic CoreValve System | Study Information | Chronic Orthotopic Study | Chronic Descending Aorta Study | Simulated use study for AccuTrak DCS (DCS-C4-18FR) | Simulated use study for AccuTrak Short Capsule DCS (DCS-C4-18FR-23) | | --- | --- | --- | --- | --- | | Device evaluated | 26mm TAV | 26mm TAV | AccuTrak DCS (DCS-C4-18FR) | AccuTrak Short Capsule DCS (DCS-C4-18FR-23) | | Animal Model | Micro-Yucatan pig | Sheep | Yorkshire pigs | Yorkshire pigs | | Methods | Percutaneous delivery of the MCS in the pig's native aortic valve. | Percutaneous delivery of the MCS in the proximal descending aorta (Hufnagel) after creation of sufficient aortic insufficiency of the native aortic valve. | Delivery performance of the AccuTrak delivery system was confirmed. | Delivery performance of the AccuTrak delivery system was confirmed. | | Valve Implant Location | Orthotopic position | Descending aorta | Orthotopic position | Orthotopic position | | Duration | 45 and 90 days | 150 ±10 days | Acute | Acute | | Major Endpoints | • To evaluate the hemodynamic performance of the Medtronic CoreValve System • To assess the in vivo response to the Medtronic CoreValve System | • Evaluate the safety and performance of the device in a sheep's descending aorta after creating sufficient aortic insufficiency (AI) of the native aortic valve (Hufnagel Model) • Identifying unanticipated or potential complications and adverse events associated with the use of the device • Assess morbidity or mortality of the study animals • Gross and microscopic examinations | • Accessibility of the intended vascular location • Trackability of the system over the recommended guidewire along the path of the vessel(s) to the intended location • Deployment of the TAV • Withdrawal of the catheter • Visualization of the system under fluoroscopy during access, placement, deployment, withdrawal, and after withdrawal • Hemostasis, or how effectively blood loss is minimized when using the system | • Accessibility of the intended vascular location • Trackability of the system over the recommended guidewire along the path of the vessel(s) to the intended location • Deployment of the TAV • Withdrawal of the catheter • Visualization of the system under fluoroscopy during access, placement, deployment, withdrawal, and after withdrawal • Hemostasis, or how effectively blood loss is minimized when using the system | | Results | Animals survived • Group 1: 45 days, 4 animals • Group 2: 90 days, 8 animals | Animals survived to 150 days: 7 • Test article (MCS): 6 animals • Control article: 1 animal | The AccuTrak DCS met all simulated use evaluation acceptance criteria. | The AccuTrak DCS and 23 mm CoreValve bioprosthesis met all simulated use evaluation acceptance criteria. | | Conclusion | The device performed as intended; thereby, demonstrating safety of the device. | The safety of the device was shown by adequate hemodynamic performance and in vivo healing response. | | | PMA P130021: FDA Summary of Safety and Effectiveness Data {13} C. Sterilization The Medtronic CoreValve System TAV undergoes liquid chemical sterilization in a glutaraldehyde solution. The terminal sterilization process involves incubation of the bioprosthesis in sterilant solution at elevated temperature for a defined period of time. The validated terminal liquid chemical sterilization process has demonstrated Sterility Assurance Levels (SAL) of 10⁻⁶. The AccuTrak DCS and the CLS are sterilized via Ethylene Oxide (EtO) in accordance with internal quality control procedures and ANSI/AAMI/ISO 11135:2007 Medical Device – Validation and Routine Control of Ethylene Oxide Sterilization. Residual testing was conducted per ISO 10993-7:2008 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals. The validated EtO sterilization process has demonstrated Sterility Assurance Levels (SAL) of 10⁻⁶. D. Packaging and Shelf Life The Medtronic CoreValve System components are all packaged separately. The TAV component is stored in glutaraldehyde in a glass jar and placed in a protective carton. Evaluations have demonstrated that packaging sterility and performance are maintained after sterilization and one year real time aging. The AccuTrak DCS is placed on a tray and then pouched. The pouched DCSs are then placed in their respective cartons. Evaluations have demonstrated packaging sterility and integrity are maintained after sterilization and one year real time aging. The CLS is also pouched and placed in a carton. Evaluations have demonstrated packaging sterility and performance are maintained after sterilization and one year real time aging. The shelf life of all components of the Medtronic CoreValve System is 1 year. Dimensional, functional, and biochemical testing, where applicable, was performed on aged components and compared to baseline performance to ensure the components meet specifications throughout the stated shelf life. X. SUMMARY OF PRIMARY CLINICAL STUDY Medtronic performed a clinical study to establish a reasonable assurance of safety and effectiveness of transcatheter aortic replacement with the Medtronic CoreValve System for iliofemoral or non-iliofemoral (i.e., subclavian and direct aortic) delivery in patients with severe symptomatic native aortic valve stenosis who have been determined by two cardiac surgeons to be at extreme risk for open aortic valve replacement and in whom existing co-morbidities would not preclude the expected benefit from correction of the aortic stenosis. The study was conducted in the U.S. under IDE G100012. A summary of the clinical study is presented below. PMA P130021: FDA Summary of Safety and Effectiveness Data Page 14 {14} # A. Study Design The CoreValve U.S. pivotal trial used to support this PMA was a prospective, non-randomized, unblinded, multi-center investigational study evaluating the safety and effectiveness of the Medtronic CoreValve System in a stratified population of patients unsuitable for cardiac surgery (referred to as the Extreme Risk study). Once the patient was determined as being at extreme risk for surgery, a determination of vascular access was made. All enrolled patients were assigned to transcatheter aortic valve replacement (TAVR) with the Medtronic CoreValve System (MCS). Patients received the CoreValve device through either an iliofemoral or a non-iliofemoral (subclavian or direct aortic) access route. The trial enrollment diagram is shown in Figure 4. ![img-3.jpeg](img-3.jpeg) Figure 4: CoreValve Extreme Risk Cohort Trial Enrollment Diagram PMA P130021: FDA Summary of Safety and Effectiveness Data {15} The trial was conducted at 41 investigational sites in the U. S. and a total of 656 iliofemoral and non-iliofemoral patients were enrolled between February 17, 2011 and August 23, 2012 in the Extreme Risk cohort. Five hundred (500) iliofemoral patients were enrolled to receive a 23, 26, 29, or 31 mm TAV and are included in the primary analysis. One hundred fifty-six (156) non-iliofemoral patients were enrolled to receive a 23, 26, 29, or 31 mm TAV and are not included in the primary analysis in accordance with the protocol. The database for this PMA reflected data from events through September 30, 2013. Contractors were utilized for monitoring and analysis of data for several aspects of the study, including: an independent Data Safety Monitoring Board (DSMB) that could contract an independent statistician; a Clinical Events Committee (CEC) that was responsible for adjudicating adverse events, an echocardiography core laboratory, and an economics quality of life core laboratory. ## 1. Clinical Inclusion and Exclusion Criteria Because tools such as the Society of Thoracic Surgeons (STS) risk calculator can only accommodate a limited number of risk factors and do not account for frailty, disabilities and anatomical characteristics which confer a prohibitive risk for surgical aortic valve replacement (e.g. porcelain aorta) these tools were not used as standalone mechanisms for identifying patients at extreme risk for cardiac surgery. Therefore, a team of two cardiac surgeons and one interventional cardiologist at each investigational site were required to assess patient suitability for inclusion in the study, taking into account risk factors not covered by the STS calculator. A central screening committee made a subsequent assessment of patient risk and agreed on patient eligibility or ineligibility. The inclusion and exclusion criteria for the Extreme Risk study are summarized below: ## Inclusion Criteria - Subject must have had co-morbidities such that one cardiologist and two cardiac surgeons agreed that medical factors preclude operation, based on a conclusion that the probability of death or serious morbidity exceeds the probability of meaningful improvement. Specifically, the predicted operative risk of death or serious, irreversible morbidity is ≥ 50% at 30 days (Extreme Risk) - Subject had senile degenerative aortic valve stenosis with: - Mean gradient &gt; 40 mmHg or jet velocity greater than 4.0 m/sec by either resting or dobutamine stress echocardiogram, or simultaneous pressure recordings at cardiac catheterization (either resting or dobutamine stress), AND - An initial aortic valve area of ≤ 0.8 cm² (or aortic valve area index ≤ 0.5 cm²/m²) by resting echocardiogram or simultaneous pressure recordings at cardiac catheterization PMA P130021: FDA Summary of Safety and Effectiveness Data Page 16 {16} - Subject was symptomatic from his/her aortic stenosis (AS), as demonstrated by New York Heart Association (NYHA) Functional Class II or greater - The subject or the subject's legal representative had been informed of the nature of the study, agreed to its provisions and had provided written informed consent as approved by the Institutional Review Board (IRB) of the respective clinical site - The subject and the treating physician agreed that the subject would return for all required post-procedure follow-up visits ## Exclusion Criteria - Evidence of an acute MI ≤ 30 days before the procedure - Any percutaneous coronary or peripheral interventional procedure performed within 30 days prior to the procedure - Blood dyscrasias as defined by: leukopenia (WBC &lt; 1000 mm³), thrombocytopenia (platelet count &lt; 50,000 cells/mm³), history of bleeding, diathesis or coagulopathy - Untreated clinically significant coronary artery disease (CAD) requiring revascularization - Cardiogenic shock manifested by low cardiac output, vasopressor dependence, or mechanical hemodynamic support - Need for emergency surgery for any reason - Severe ventricular dysfunction with left ventricular ejection fraction (LVEF) &lt; 20% as measured by resting echocardiogram - Recent (within 6 months) cerebrovascular accident (CVA) or transient ischemic attack (TIA) - End stage renal disease requiring chronic dialysis or creatinine clearance &lt; 20 cc/min. - Active Gastrointestinal (GI) bleeding within the past 3 months - A known hypersensitivity or contraindication to any of the following which cannot be adequately pre-medicated: - Aspirin - Heparin (HIT/HITTS) and bivalirudin - Nitinol (titanium or nickel) - Ticlopidine and clopidogrel - Contrast media - Ongoing sepsis, including active endocarditis - Subject refuses a blood transfusion - Life expectancy &lt; 12 months due to associated non-cardiac co-morbid conditions. - Other medical, social, or psychological conditions that in the opinion of an Investigator precludes the subject from appropriate consent - Severe dementia (resulting in either inability to provide informed consent for the trial/procedure, prevents independent lifestyle outside of a chronic care facility, or PMA P130021: FDA Summary of Safety and Effectiveness Data Page 17 {17} will fundamentally complicate rehabilitation from the procedure or compliance with follow-up visits) - Concurrently participating in an investigational drug or another device study - Symptomatic carotid or vertebral artery disease - Native aortic annulus size &lt; 18 mm or &gt; 29 mm per the baseline diagnostic imaging. - Pre-existing prosthetic heart valve in any position - Mixed aortic valve disease [AS and aortic regurgitation (AR) with severity (3-4+)] - Moderate to severe (3-4+) or severe (4+) mitral or severe (4+) tricuspid regurgitation - Moderate to severe mitral stenosis - Hypertrophic obstructive cardiomyopathy - New or untreated echocardiographic evidence of intracardiac mass, thrombus or vegetation - Severe basal septal hypertrophy with an outflow gradient - Aortic root angulation (angle between plane of aortic valve annulus and horizontal plane/vertebrae) &gt; 70° (for femoral and left subclavian/axillary access) and &gt; 30° (for right subclavian/axillary access) - Ascending aorta that exceeded the maximum diameter for any given native aortic annulus size - Congenital bicuspid or unicuspid valve verified by echocardiography - Sinus of valsalva anatomy that would prevent adequate coronary perfusion - Transarterial access not able to accommodate an 18Fr sheath ## 2. Follow-Up Schedule Follow-up periods were discharge or 7 days, whichever comes first, 30 days, 6 months, 12 months, and annually thereafter to a minimum of 5 years post procedure, and patients were followed for a minimum of 12 months prior to submission of the PMA. ## 3. Clinical Endpoints ### Primary Safety and Effectiveness Endpoints The primary endpoint of the study was to demonstrate the safety and effectiveness in transarterial delivery of the Medtronic CoreValve System (MCS), as measured by all-cause death or major stroke at 12 months, in the treatment of symptomatic severe aortic stenosis in patients necessitating aortic valve replacement, with predicted operative mortality or serious, irreversible morbidity risk ≥ 50% at 30 days (Extreme Risk). A performance goal of 43% was pre-specified for the 12-month rate of all-cause mortality or major stroke in TAVR patients with the Medtronic CoreValve System, which was based on review of literature for alternative treatments for extreme risk patients. The hypothesis for the primary endpoint was as follows: PMA P130021: FDA Summary of Safety and Effectiveness Data Page 18 {18} $$ H_0: \pi_{\mathrm{MCS\ TAVR}} \geq 43.0\% $$ $$ H_{\mathrm{A}}: \pi_{\mathrm{MCS\ TAVR}} &lt; 43.0\% $$ It was also developed a priori that the primary endpoint would be examined for the null hypothesis for the iliofemoral study cohort only and the results of the non-iliofemoral study cohort would be reported separately using descriptive statistics. This distinction must be borne in mind when viewing the results of the non-iliofemoral study cohort presented, for convenience only, alongside those of the iliofemoral study cohort later in this summary. ## Secondary Safety and Effectiveness Endpoints This study included the following secondary safety and effectiveness endpoints: 1. Major adverse cardiovascular or cerebrovascular events (MACCE)-free survival at 30 days, 6 months, 12 months and annually thereafter up to 5 years 2. The occurrence of individual MACCE components at 30 days, 6 months, 12 months and annually thereafter up to 5 years 3. Major adverse events (MAE) at 30 days, 6 months, 12 months and annually thereafter up to 5 years 4. Conduction disturbance requiring permanent pacemaker implantation (PPI) at 30 days, 6 months, 12 months and annually thereafter up to 5 years 5. Change in NYHA class from baseline at 30 days, 6 months, 12 months and annually thereafter up to 5 years. 6. Change in distance walked during 6-minute walk test (6MWT) from baseline to 30 days and baseline to 12 months 7. Ratio of days alive out of hospital versus total days alive assessed at 12 months follow-up 8. Quality of life (QoL) change from baseline at 30 days, 6 months, 12 months and annually thereafter up to 5 years 9. Echocardiographic assessment of valve performance at discharge, 30 days, 6 months, 12 months and annually thereafter up to 5 years using the following measures: - Transvalvular mean gradient - Effective orifice area (EOA) - Degree of aortic regurgitation (AR, transvalvular and paravalvular) 10. Aortic valve disease hospitalization 11. Cardiovascular deaths and valve-related deaths 12. Strokes 13. Index procedure related MAEs 14. Length of index procedure hospital stay 15. Device success defined as follows: - Successful vascular access, delivery and deployment of the device, and successful retrieval of the delivery system - Correct position of the device in the proper anatomical location (placement in the annulus with no impedance on device function) PMA P130021: FDA Summary of Safety and Effectiveness Data Page 19 {19} - Intended performance of the prosthetic valve (aortic valve area $&gt;1.2\mathrm{cm}^2$ for 26, 29 and $31\mathrm{mm}$ valves, $\geq 0.9\mathrm{cm}^2$ for $23\mathrm{mm}$ valve (by echocardiography using the continuity equation) and mean aortic valve gradient $&lt; 20\mathrm{mmHg}$ or peak velocity $&lt; 3\mathrm{m/sec}$ , without moderate or severe prosthetic valve AR) - assessed acutely in a resting state, either within 24-48 hours after the index procedure or before hospital discharge - Only one valve implanted in the proper anatomical location 16. Procedural success, defined as device success and absence of in-hospital MACCE 17. Evidence of prosthetic valve dysfunction at 30 days, 6 months, 12 months and annually thereafter up to 5 years Four (4) of the above secondary endpoints involve hierarchical hypothesis testing, which are changes from baseline to 12 months in transvalvular mean gradient, effective orifice area, NYHA classification, and KCCQ score. # B. Accountability of PMA Cohort At the time of database lock, 458 of the 656 patients enrolled were available for the analysis at the 1 year time point. Table 8 depicts the accountability at each follow-up period for the "All Enrolled" population (see Analysis Population section for definition). Table 8: Total Patient Accountability | Follow up Period | Variable | All Enrolled (N=656) | | --- | --- | --- | | 1 month | Expected | 583 | | | Number withdrew | 10 | | | Number died before visit | 60 | | | Lost to follow up | 0 | | | Other | 3 | | | Visit compliance | 572 (98.1%) | | 6 months | Expected | 503 | | | Number withdrew | 0 | | | Number died before visit | 80 | | | Lost to follow up | 0 | | | Other | 0 | | | Visit compliance | 485 (96.4%) | | 12 months | Expected | 462 | | | Number withdrew | 1 | | | Number died before visit | 40 | | | Lost to follow up | 0 | | | Other | 0 | | | Visit compliance | 458 (99.1%) | PMA P130021: FDA Summary of Safety and Effectiveness Data {20} # C. Study Population Demographics and Baseline Parameters The demographics of the study population are typical for an aortic stenosis valve replacement study performed in the U.S., as shown in Table 9. A high proportion of the patients had significant co-morbidities, frailties, or disabilities. The mean age for patients participating in the trial was approximately 83 years old, and slightly less than $50\%$ of patients were male. The mean STS score was approximately 10. Greater than $90\%$ of all patients were in NYHA classes III or IV. Additionally, coronary artery disease was present in approximately $80\%$ of patients, and greater than $30\%$ of patients had previous MI. Peripheral vascular disease, chronic obstructive pulmonary disease (COPD), and home oxygen use were more prevalent in non-iliofemoral patients. Table 9: Demographics of the Study Population (All Enrolled) | Demographic | Iliofemoral N=500 | Non-Iliofemoral N=156 | | --- | --- | --- | | Age (yrs) | 83.1 ± 8.6 | 81.6 ± 7.7 | | Gender (Male) | 48.0% (240/500) | 44.9% (70/156) | | NYHA Classification | | | | II | 8.6% (43/500) | 8.3% (13/156) | | III | 63.6% (318/500) | 66.0% (103/156) | | IV | 27.8% (139/500) | 25.6% (40/156) | | STS Score (Risk of Mortality, %) | 10.3 ± 5.5 | 10.5 ± 5.7 | | Coronary Artery Disease | 81.8% (409/500) | 78.8% (123/156) | | Previous MI | 31.0% (155/500) | 31.4% (49/156) | | Previous Interventions | | | | Coronary Artery Bypass Surgery | 39.0% (195/500) | 41.0% (64/156) | | Percutaneous Coronary Intervention | 37.4% (187/500) | 30.1% (47/156) | | Balloon Valvuloplasty | 20.4% (102/500) | 22.4% (35/156) | | Cerebral Vascular Disease | 24.0% (119/496) | 28.4% (44/155) | | Prior Stroke | 13.6% (68/499) | 14.2% (22/155) | | Peripheral Vascular Disease | 36.0% (179/497) | 59.0% (92/156) | | Chronic Lung Disease/COPD | 59.6% (298/500) | 69.9% (109/156) | | Home Oxygen | 30.8% (154/500) | 41.7% (65/156) | | Creatinine Level >2 mg/dl | 4.6% (23/500) | 2.6% (4/156) | | Atrial Fibrillation / Atrial Flutter | 47.4% (236/498) | 48.4% (75/155) | | Pre-Existing Permanent Pacemaker Placement / ICD | 25.8% (129/500) | 24.4% (38/156) | | Aorta Calcification1: Severe/Porcelain | | | | Severe | 16.6% (83/499) | 17.5% (27/154) | | Porcelain | 5.2% (26/499) | 7.8% (12/154) | | Chest Wall Deformity | 5.6% (28/500) | 1.9% (3/156) | | Hostile Mediastinum | 12.0% (60/499) | 9.0% (14/156) | | Cirrhosis of the Liver | 3.0% (15/500) | 1.3% (2/156) | | Wheelchair Bound | 16.6% (83/500) | 12.2% (19/156) | | Echocardiographic Findings | | | | Ejection Fraction (visual estimate, %) | 53.2 ± 13.6 (498) | 54.3 ± 15.3 (156) | | Aortic Valve Area (cm2) | 0.67 ± 0.25 (485) | 0.62 ± 0.23 (153) | PMA P130021: FDA Summary of Safety and Effectiveness Data {21} | Demographic | Iliofemoral N=500 | Non-Iliofemoral N=156 | | --- | --- | --- | | Mean Gradient across Aortic Valve (MGV_{2}, mmHg) | 47.72 ± 13.53 (498) | 49.67 ± 16.85 (156) | | Mitral Regurgitation: Moderate/Severe | 24.2% (120/496) | 23.2% (36/155) | | 1. Aorta Calcification is measured on screening CT Angiogram Plus-minus values present the mean ± standard deviation. | | | ## D. Safety and Effectiveness Results ### 1. Analysis Population The primary analysis was the “Attempted Implant” analysis. An attempted implant procedure was defined as when the patient was brought into the procedure room and any of the following had occurred: anesthesia administered, vascular line placed, TEE placed or any monitoring line placed. The “Attempted Implant” iliofemoral population (n=489) included all patients who were implanted via iliofemoral, had an attempted implant via iliofemoral, or were enrolled iliofemoral and no access site was reported during the attempted procedure (i.e., the patient had an attempted implant, but the procedure was aborted prior to obtaining access site). The “Attempted Implant” non-iliofemoral population (n=150) included all patients who were implanted via non-iliofemoral, had an attempted implant via non-iliofemoral, or were enrolled non-iliofemoral and no access site was reported during the attempted procedure. The “Implanted” population consisted of all “Attempted Implant” patients who were actually implanted with the CoreValve device. To be considered implanted, the patient’s device disposition form must have shown at least one device with a final disposition of “Implanted.” There were a total of 486 and 148 “Implanted” patients in the iliofemoral and non-iliofemoral cohorts, respectively. The “All Enrolled” population consisted of all patients who were enrolled, regardless of whether a CoreValve device was implanted. The number of “All Enrolled” iliofemoral and non-iliofemoral patients was 500 and 156, respectively. ### 2. Primary Safety and Effectiveness Endpoint The primary endpoint of all-cause mortality or major stroke at 12 months includes all deaths (cardiovascular and non-cardiovascular) from any cause after a valve intervention. Major stroke is a stroke causing clinically important disability (defined as a Modified Rankin score ≥ 2 at 90 days). Figure 5 and Table 10 show K-M rates of all-cause mortality or major stroke in the attempted implant population for the iliofemoral patients up to 12 months follow-up, which were 9.8% at 1 month, 19.8% PMA P130021: FDA Summary of Safety and Effectiveness Data {22} at 6 months and 26.0% at 12 months (Primary Endpoint). The primary endpoint was therefore met and the null hypothesis for the Primary Endpoint (K-M Rate ≥ 43%) rejected. ![img-4.jpeg](img-4.jpeg) Figure 5: Primary Endpoint: All-Cause Mortality or Major Stroke Kaplan-Meier Event Rate — Iliofemoral Attempted Implant Table 10: Primary Endpoint: All-Cause Mortality or Major Stroke – Iliofemoral Attempted Implant | Interval Post Procedure (months)* | Attempted Implant N=489 | | | | | --- | --- | --- | --- | --- | | | 0 | 1 | 6 | 12 | | # at start of interval | 489 | 441 | 392 | 360 | | # events in interval | 48 | 49 | 30 | 47 | | # event cumulative | 48 | 97 | 127 | 174 | | K-M Event Rate | 1.6 | 9.8 | 19.8 | 26.0 | | Lower 95% CI | 0.5 | 7.2 | 16.3 | 22.1 | | Upper 95% CI | 2.8 | 12.5 | 23.4 | 29.9 | | *0 = 0-29 days, 1 = 30-182 days, 6 = 183-364 days, 12 = ≥365 days. Cumulative probability of event estimate is based on the Kaplan-Meier method. | | | | | PMA P130021: FDA Summary of Safety and Effectiveness Data {23} 3. Key Secondary Safety and Effectiveness Endpoints Adverse Events that Occurred in the PMA Clinical Study Table 11 and Table 12 provide a summary of the adverse events (AEs) that occurred in this study for the iliofemoral and non-iliofemoral cohorts, respectively. Table 11: CEC Adjudicated Adverse Event Summary – Iliofemoral Attempted Implant | | Iliofemoral N=489 | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Event | 0-30 Days | | | 0-6 Months | | | 0-12 Months | | | | | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | | All-Cause Mortality or Major Stroke | 52 | 48 | 9.8% | 106 | 97 | 19.8% | 139 | 127 | 26.0% | | All-Cause Mortality | 41 | 41 | 8.4% | 91 | 91 | 18.6% | 119 | 119 | 24.3% | | Cardiovascular | 41 | 41 | 8.4% | 73 | 73 | 15.0% | 88 | 88 | 18.3% | | Valve-Related¹ | 12 | 12 | 2.5% | 19 | 19 | 4.1% | 23 | 23 | 5.1% | | Neurological Events | 80 | 74 | 15.5% | 120 | 101 | 21.5% | 141 | 117 | 25.3% | | All Stroke | 20 | 19 | 4.0% | 26 | 24 | 5.2% | 34 | 31 | 7.0% | | Major Stroke | 11 | 11 | 2.3% | 15 | 15 | 3.2% | 20 | 19 | 4.3% | | Bleed | 191 | 179 | 36.7% | 225 | 200 | 41.4% | 236 | 206 | 42.8% | | Life Threatening or Disabling | 63 | 62 | 12.7% | 81 | 77 | 16.1% | 88 | 83 | 17.6% | | Major Bleed | 128 | 121 | 24.9% | 144 | 133 | 27.7% | 148 | 136 | 28.5% | | Major Vascular Complication | 44 | 40 | 8.2% | 45 | 41 | 8.4% | 45 | 41 | 8.4% | | Acute Kidney Injury | 57 | 57 | 11.8% | 57 | 57 | 11.8% | 57 | 57 | 11.8% | | MI | 6 | 6 | 1.2% | 7 | 7 | 1.5% | 9 | 9 | 2.0% | | MACCE² | 72 | 60 | 12.3% | 131 | 110 | 22.5% | 171 | 143 | 29.2% | | Cardiogenic Shock | 13 | 13 | 2.7% | 13 | 13 | 2.7% | 13 | 13 | 2.7% | | Cardiogenic Tamponade | 9 | 9 | 1.9% | 10 | 10 | 2.1% | 10 | 10 | 2.1% | | Reintervention | 5 | 5 | 1.1% | 7 | 7 | 1.5% | 9 | 8 | 1.8% | | Surgical | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 0 | 0 | 0.0% | | Percutaneous | 5 | 5 | 1.1% | 7 | 7 | 1.5% | 9 | 8 | 1.8% | | Valve Endocarditis | 0 | 0 | 0.0% | 1 | 1 | 0.2% | 5 | 5 | 1.3% | | Valve Thrombosis | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 0 | 0 | 0.0% | | Valve Embolism/ Device Migration | 0 | 0 | 0.0% | 1 | 1 | 0.2% | 1 | 1 | 0.2% | | ¹ Valve-related death is any death caused by prosthetic valve dysfunction, valve thrombosis, embolism, bleeding event, or implanted valve endocarditis or related to reintervention on the operated valve.² MACCE includes all-cause death, myocardial infarction (MI), all stroke, and reintervention. | | | | | | | | | | PMA P130021: FDA Summary of Safety and Effectiveness Data Page 24 {24} Table 12: CEC Adjudicated Adverse Event Summary – Non-Iliofemoral Attempted Implant | | Non-Iliofemoral N=150 | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Event | 0-30 Days | | | 0-6 Months | | | 0-12 Months | | | | | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | | All-Cause Mortality or Major Stroke | 28 | 23 | 15.3% | 56 | 48 | 32.0% | 67 | 59 | 39.4% | | All-Cause Mortality | 17 | 17 | 11.3% | 43 | 43 | 28.7% | 54 | 54 | 36.0% | | Cardiovascular | 17 | 17 | 11.3% | 35 | 35 | 23.6% | 42 | 42 | 28.8% | | Valve-Related¹ | 4 | 4 | 2.8% | 6 | 6 | 4.5% | 7 | 7 | 5.4% | | Neurological Events | 36 | 32 | 21.8% | 43 | 38 | 26.6% | 46 | 40 | 28.5% | | All Stroke | 14 | 13 | 8.8% | 18 | 17 | 12.0% | 19 | 18 | 13.0% | | Major Stroke | 11 | 11 | 7.5% | 13 | 13 | 9.1% | 13 | 13 | 9.1% | | Bleed | 92 | 87 | 58.3% | 104 | 94 | 63.5% | 106 | 96 | 65.1% | | Life Threatening or Disabling | 36 | 36 | 24.2% | 42 | 42 | 28.5% | 43 | 43 | 29.4% | | Major Bleed | 56 | 55 | 37.1% | 62 | 59 | 40.8% | 63 | 60 | 41.9% | | Major Vascular Complication | 13 | 13 | 8.7% | 14 | 14 | 9.5% | 14 | 14 | 9.5% | | Acute Kidney Injury | 21 | 21 | 14.2% | 21 | 21 | 14.2% | 21 | 21 | 14.2% | | MI | 3 | 3 | 2.1% | 3 | 3 | 2.1% | 3 | 3 | 2.1% | | MACCE² | 34 | 26 | 17.3% | 64 | 52 | 34.7% | 77 | 62 | 41.4% | | Cardiogenic Shock | 9 | 9 | 6.0% | 9 | 9 | 6.0% | 9 | 9 | 6.0% | | Cardiogenic Tamponade | 2 | 2 | 1.3% | 2 | 2 | 1.3% | 2 | 2 | 1.3% | | Reintervention | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 1 | 1 | 1.0% | | Surgical | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 0 | 0 | 0.0% | | Percutaneous | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 1 | 1 | 1.0% | | Valve Endocarditis | 1 | 1 | 0.7% | 1 | 1 | 0.7% | 2 | 2 | 1.7% | | Valve Thrombosis | 0 | 0 | 0.0% | 1 | 1 | 0.8% | 2 | 1 | 0.8% | | Valve Embolism/ Device Migration | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 0 | 0 | 0.0% | | ¹ Valve-related death is any death caused by prosthetic valve dysfunction, valve thrombosis, embolism, bleeding event, or implanted valve endocarditis or related to reintervention on the operated valve.² MACCE includes all-cause death, myocardial infarction (MI), all stroke, and reintervention. | | | | | | | | | | PMA P130021: FDA Summary of Safety and Effectiveness Data Page 25 {25} # Neurological Events Table 13 and Table 14 provide a summary of the neurological events that occurred in this study for the iliofemoral and non-iliofemoral cohorts. Stroke and TIA were defined according to the Valve Academic Research Consortium I (VARC-I) definitions [1]. Table 13: CEC Adjudicated Neurological Events – Iliofemoral Attempted Implant | | Attempted Implant N=489 | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Event | 0-30 Days | | | 0-6 Months | | | 0-12 Months | | | | | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | | All Stroke | 20 | 19 | 4.0% | 26 | 24 | 5.2% | 34 | 31 | 7.0% | | Major Stroke | 11 | 11 | 2.3% | 15 | 15 | 3.2% | 20 | 19 | 4.3% | | Ischemic | 9 | 9 | 1.9% | 13 | 13 | 2.8% | 17 | 16 | 3.6% | | Hemorrhagic | 2 | 2 | 0.4% | 2 | 2 | 0.4% | 3 | 3 | 0.7% | | Minor Stroke | 9 | 9 | 1.9% | 11 | 11 | 2.4% | 14 | 14 | 3.2% | | Ischemic | 9 | 9 | 1.9% | 11 | 11 | 2.4% | 14 | 14 | 3.2% | | Hemorrhagic | 0 | 0 | 0.0% | 0 | 0 | 0.0% | 0 | 0 | 0.0% | | TIA | 3 | 3 | 0.6% | 4 | 4 | 0.9% | 5 | 5 | 1.1% | | Intracranial Hemorrhage | 1 | 1 | 0.2% | 2 | 2 | 0.4% | 2 | 2 | 0.4% | Table 14: CEC Adjudicated Neurological Events – Non-Iliofemoral Attempted Implant | | Attempted Implant N=150 | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Event | 0-30 Days | | | 0-6 Months | | | 0-12 Months | | | | | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | # Events | # Patients | K-M Rate (%) | | All Stroke | 14 | 13 | 8.8% | 18 | 17 | 12.0% | 19 | 18 | 13.0% | | Major Stroke | 11 | 11 | 7.5% | 13 | 13 | 9.1% | 13 | 13 | 9.1% | | Ischemic | 11 | 11 | 7.5% | 12 | 12 | 8.3% | 12 | 12 | 8.3% | | Hemorrhagic | 0 | 0 | 0.0% | 1 | 1 | 0.9% | 1 | 1 | 0.9% | | Minor Stroke | 3 | 3 | 2.1% | 5 | 5 | 3.7% | 6 | 6 | 4.7% | | Ischemic | 3 | 3 | 2.1% | 4 | 4 | 2.9% | 5 | 5 | 3.8% | | Hemorrhagic | 0 | 0 | 0.0% | 1 | 1 | 0.8% | 1 | 1 | 0.8% | | TIA | 2 | 2 | 1.4% | 3 | 3 | 2.3% | 3 | 3 | 2.3% | | Intracranial Hemorrhage | 0 | 0 | 0.0% | 1 | 1 | 0.9% | 1 | 1 | 0.9% | PMA P130021: FDA Summary of Safety and Effectiveness Data {26} Echocardiographic Assessment of Valve Performance (Total Aortic Regurgitation) Table 15 summarizes the total aortic regurgitation (AR) severity over time in the iliofemoral and non-iliofemoral cohorts. Table 15: Total Aortic Regurgitation by Visit – Implanted Population | | Screening/ Baseline | 1 month | 6 months | 12 months | | --- | --- | --- | --- | --- | | Iliofemoral (N=486) | | | | | | None | 11.7% (56/477) | 9.1% (38/419) | 19.9% (73/367) | 21.3% (70/329) | | Trivial | 36.5% (174/477) | 32.7% (137/419) | 33.5% (123/367) | 40.7% (134/329) | | Mild | 43.0% (205/477) | 43.0% (180/419) | 36.5% (134/367) | 31.6% (104/329) | | Moderate | 8.6% (41/477) | 14.1% (59/419) | 9.8% (36/367) | 6.4% (21/329) | | Severe | 0.2% (1/477) | 1.2% (5/419) | 0.3% (1/367) | 0.0% (0/329) | | Non-Iliofemoral (N=148) | | | | | | None | 12.2% (18/147) | 19.0% (23/121) | 33.3% (32/96) | 39.0% (32/82) | | Trivial | 28.6% (42/147) | 33.9% (41/121) | 27.1% (26/96) | 36.6% (30/82) | | Mild | 48.3% (71/147) | 34.7% (42/121) | 35.4% (34/96) | 20.7% (17/82) | | Moderate | 10.9% (16/147) | 10.7% (13/121) | 4.2% (4/96) | 2.4% (2/82) | | Severe | 0.0% (0/147) | 1.7% (2/121) | 0.0% (0/96) | 1.2% (1/82) | Echocardiographic Assessment of Valve Performance (Effective Orifice Area (EOA) and Mean Gradient) The effective orifice area (EOA) and mean gradient values obtained over time for the iliofemoral and non-iliofemoral patients in the Implanted population are shown in Table 16 and Table 17, respectively. Table 16: Effective Orifice Area (cm²) By Visit (Core Lab) – Implanted Population | | Baseline | 1 month | 12 months | | --- | --- | --- | --- | | Iliofemoral | 0.73 ± 0.23 (389) | 1.86 ± 0.56 (386) | 1.88 ± 0.54 (307) | | Non-Iliofemoral | 0.72 ± 0.27 (129) | 1.82 ± 0.64 (114) | 1.85 ± 0.51 (74) | | Plus-minus values present the mean ± standard deviation. | | | | Table 17: Mean Gradient (mmHg) By Visit (Core Lab) – Implanted Population | | Baseline | 1 month | 12 months | | --- | --- | --- | --- | | Iliofemoral | 47.3 ± 14.6 (481) | 8.7 ± 4.2 (418) | 8.9 ± 4.1 (330) | | Non-Iliofemoral | 49.5 ± 17.1 (143) | 9.7 ± 5.8 (126) | 9.5 ± 5.7 (83) | | Plus-minus values present the mean ± standard deviation. | | | | Conduction Disturbance Requiring Permanent Pacemaker Implantation Table 18 presents the pacemaker implantation rate for the iliofemoral and non-iliofemoral Attempted Implant cohorts. PMA P130021: FDA Summary of Safety and Effectiveness Data {27} Table 18: Conduction Disturbance Requiring Pacemaker – Attempted Implant | | Iliofemoral N=489 | | Non-Iliofemoral N=150 | | | --- | --- | --- | --- | --- | | | # of Patients | K-M Event Rate (%) | # of Patients | K-M Event Rate (%) | | New Permanent Pacemaker Implant¹ | | | | | | 0-30 Days | 104 | 21.6% | 24 | 16.4% | | 0-12 Months | 123 | 26.2% | 30 | 21.5% | | Permanent Pacemaker Implant² | | | | | | 0-30 Days | 104 | 29.4% | 24 | 22.0% | | 0-12 Months | 121 | 34.9% | 30 | 28.8% | | ¹ Patients with pacemaker or ICD at baseline are included in the denominator. ² Patients with pacemaker or ICD at baseline are excluded from the numerator and denominator. Note 2 patients with baseline pacemaker/ICD, received new pacemaker/ICD between 31-365 days. | | | | | Ratio of Days Alive out of Hospital versus Total Days Alive The total hospital days through 12 months (mean ± SD), including the days in hospital for the index procedure when the CoreValve was implanted or attempted, were 14.4 ± 15.1 days and 16.7 ± 13.0 days for the iliofemoral and non-iliofemoral cohorts, respectively. The ratio of days alive out of hospital versus total days alive assessed at 12 months was 0.86 ± 0.27 and 0.80 ± 0.31 for the iliofemoral and non-iliofemoral cohorts, respectively. The ratio of days alive is interpreted as on average subjects spent 86% of days alive after procedure out of the hospital. New York Heart Association (NYHA) Functional Class An evaluation of cardiac symptom severity based on NYHA classification was conducted at several evaluation time points through the first year of follow-up. Data at baseline and 1 year are presented in Table 19 for the iliofemoral and non-iliofemoral cohorts. Table 19: NYHA Classification By Visit – Attempted Implant | NYHA Classification | Iliofemoral N=489 | Non-Iliofemoral N=150 | | --- | --- | --- | | Baseline | | | | NYHA I | 0.0% (0/485) | 0.0% (0/148) | | NYHA II | 8.7% (42/485) | 8.1% (12/148) | | NYHA III | 64.7% (314/485) | 70.3% (104/148) | | NYHA IV | 26.6% (129/485) | 21.6% (32/148) | | Died prior to visit | 0.0% (0/485) | 0.0% (0/148) | | Exit prior to visit | 0 | 0 | | Visit occurred but NYHA not obtained | 4 | 2 | PMA P130021: FDA Summary of Safety and Effectiveness Data {28} | NYHA Classification | Iliofemoral N=489 | Non-Iliofemoral N=150 | | --- | --- | --- | | Visit missed | 0 | 0 | | 12 Month | | | | NYHA I | 43.3% (200/462) | 28.4% (40/141) | | NYHA II | 24.0% (111/462) | 24.1% (34/141) | | NYHA III | 5.4% (25/462) | 8.5% (12/141) | | NYHA IV | 1.1% (5/462) | 0.0% (0/141) | | Died prior to visit | 26.2% (121/462) | 39.0% (55/141) | | Exit prior to visit | 1 | 0 | | Visit occurred but NYHA not obtained | 21 | 8 | | Visit missed | 5 | 1 | ## Quality of Life (QoL) Change The QoL changes from baseline at 30 days and 12 months were evaluated using the Kansas City Cardiomyopathy Questionnaire (KCCQ), the QualityMetric’s SF-12v2® Health Survey (SF12), and the EuroQoL (EQ-5D), as shown in Table 20 and Table 21 for the iliofemoral and non-iliofemoral cohorts, respectively. The KCCQ is a validated self-administered 23-item questionnaire that quantifies physical limitations, symptoms, self-effectiveness, social interference and quality of life. These individual scales are incorporated into an Overall Summary Score which combines the domains of physical limitation, symptoms, QoL, and social limitation with values ranging from 0-100; higher scores indicate lesser symptoms and better quality of life. Previous studies have suggested that KCCQ Overall Summary scores correlate roughly with New York Heart Association Functional Class as follows: Class I ≈ KCCQ Summary Score 75-100; Class II ≈ 60-74; Class III ≈ 45-59; and Class IV ≈ 0-44. In addition, there is a Clinical Summary Score that combines the domains of physical limitation and symptoms. SF12 is a shorter version of the SF-36v2® Health Survey that uses 12 questions to measure functional health and well-being from the patient’s point of view and is generally reported in two summary scores which evaluate physical (the SF-12 Physical Summary Score) and mental (the SF-12 Mental Summary Score) health. Values range from 0-100; higher scores indicate better functional health and well-being. The EQ-5D is a measure of self-reported health outcomes that is applicable to a wide range of health conditions and treatments. It consists of 2 parts: a descriptive system (Part I) and a visual analogue scale (Part II). Part I of the scale consists of 5 single-item dimensions including: mobility, self care, usual activities, pain/discomfort, and anxiety/depression. Each dimension has a 3 point response scale designed to indicate the level of the problem. The overall EQ-5D score from Part is evaluated on a scale where 0.0 = death and 1.0 = perfect health. Part II uses a vertical graduated visual analogue scale (thermometer) to measure health status, ranging from worst imaginable health state to best imaginable health state. PMA P130021: FDA Summary of Safety and Effectiveness Data {29} Table 20: Quality of Life – Iliofemoral Attempted Implant | | Baseline | 1 month | 12 months | | --- | --- | --- | --- | | KCCQ | | | | | Overall Summary Score | 37.9 ± 22.1 (454) | 62.3 ± 25.5 (266) | 68.8 ± 23.6 (287) | | Change from Baseline | -- | 24.2 ± 28.9 (260) | 27.9 ± 27.1 (265) | | Clinical Summary Score | 42.0 ± 22.4 (454) | 62.3 ± 24.9 (266) | 66.3 ± 23.4 (287) | | Change from Baseline | -- | 20.2 ± 28.0 (260) | 20.8 ± 26.8 (265) | | SF12 | | | | | Physical Component | 28.5 ± 8.3 (422) | 34.9 ± 10.1 (245) | 34.3 ± 10.5 (259) | | Change from Baseline | -- | 5.9 ± 10.4 (223) | 5.5 ± 10.8 (229) | | Mental Component | 45.8 ± 12.3 (422) | 49.8 ± 12.0 (245) | 51.9 ± 11.8 (259) | | Change from Baseline | -- | 3.7 ± 14.2 (223) | 5.2 ± 13.7 (229) | | EQ-5D | 0.65 ± 0.23 (445) | 0.73 ± 0.24 (261) | 0.73 ± 0.21 (275) | | Change from Baseline | -- | 0.09 ± 0.29 (252) | 0.06 ± 0.25 (250) | | Plus-minus values present the mean ± standard deviation. | | | | Table 21: Quality of Life – Non-Iliofemoral Attempted Implant | | Baseline | 1 month | 12 months | | --- | --- | --- | --- | | KCCQ | | | | | Overall Summary Score | 42.5 ± 22.3 (141) | 51.0 ± 25.5 (74) | 65.1 ± 22.4 (81) | | Change from Baseline | -- | 7.9 ± 33.5 (71) | 21.9 ± 26.8 (76) | | Clinical Summary Score | 46.7 ± 23.0 (141) | 53.7 ± 24.6 (74) | 65.2 ± 21.3 (81) | | Change from Baseline | -- | 6.8 ± 32.0 (71) | 18.1 ± 24.9 (76) | | SF12 | | | | | Physical Component | 27.9 ± 8.0 (130) | 32.0 ± 9.2 (66) | 34.0 ± 9.4 (80) | | Change from Baseline | -- | 1.9 ± 10.4 (57) | 4.6 ± 10.0 (72) | | Mental Component | 47.6 ± 12.0 (130) | 45.1 ± 14.7 (66) | 49.0 ± 13.3 (80) | | Change from Baseline | -- | -1.7 ± 16.2 (57) | 2.4 ± 14.3 (72) | | EQ5D | 0.67 ± 0.23 (138) | 0.66 ± 0.25 (72) | 0.73 ± 0.20 (80) | | Change from Baseline | -- | -0.00 ± 0.30 (69) | 0.05 ± 0.25 (74) | | Plus-minus values present the mean ± standard deviation. | | | | ## Hierarchical Testing of Secondary Endpoints Four pre-specified secondary endpoints were explored for iliofemoral patients using a hierarchical test procedure, as shown in Table 24. Change from baseline to 12 months was evaluated for measures of forward flow hemodynamic performance (EOA and mean gradient) and the improvement in these parameters was found to be statistically significant (p&lt;0.0001). Similarly, improvement in NYHA functional classification was evaluated and found to be statistically significant (p&lt;0.0001). The Kansas City Cardiomyopathy Questionnaire (KCCQ) was used to evaluate changes from baseline in physical limitations, symptoms, self-effectiveness, social interference and quality of life and a statistically significant improvement was identified in the overall summary score (p&lt;0.0001). PMA P130021: FDA Summary of Safety and Effectiveness Data Page 30 {30} Table 22: Secondary Endpoints: Hierarchical Testing – Iliofemoral Attempted Implant | Secondary Endpoint | Paired Evaluations | Average Paired Difference (12 Month – Baseline) | Hypothesis Test Ho: μchange = 0 HA: μchange ≠ 0 | | | --- | --- | --- | --- | --- | | | | | P-value | Success | | #9 / Mean Gradient | 326 | -39.82 ± 14.83 | <0.0001 | PASS | | #9 / EOA | 245 | 1.16 ± 0.57 | <0.0001 | PASS | | #5 / NYHA | 338 | -1.6 ± 0.9 | <0.0001 | PASS | | #8 / KCCQ – Overall Summary Score | 265 | 27.9 ± 27.1 | <0.0001 | PASS | | Plus-minus values present the mean ± standard deviation. | | | | | # 4. Additional Study Observations # Procedure Data Table 25 provides a summary of the transcatheter valve implantation procedure for the iliofemoral and non-iliofemoral cohorts, respectively. Mean total time in the Catheterization Laboratory or Operating Room for patients in the iliofemoral cohort was approximately 3.5 hours while mean total procedure time (skin-to-skin) was on average slightly greater than 1 hour. Mean total time in the Catheterization Laboratory or Operating Room for the non-iliofemoral cohort was approximately 4 hours while mean total procedure time was slightly greater than 1 hour. Table 23: TAVR Procedure Data (Attempted Implant) | | Iliofemoral N=489 | Non-Iliofemoral N=150 | | --- | --- | --- | | Time to Procedure (days) | 8.9 ± 12.3 (489) | 10.2 ± 15.5 (150) | | Total Time in Cath Lab or OR (min) | 214.8 ± 64.9 (486) | 258.7 ± 72.5 (148) | | Total procedure time (min) (skin to skin) | 66.1 ± 39.0 (484) | 60.5 ± 46.5 (145) | | General Anesthesia | 94.4% (459/486) | 99.3% (147/148) | | Valve-in-Valve Procedure | 2.5% (12/486) | 0.7% (1/148) | | Emergent Operation Due to Device or Procedure | 0.0% (0/486) | 0.0% (0/148) | | Number of Devices Used | | | | 0 | 0.6% (3/489) | 1.3% (2/150) | | 1 | 93.3% (456/489) | 94.7% (142/150) | | 2 | 6.1% (30/489) | 4.0% (6/150) | | Valve Size Implanted | | | | 23mm | 2.5% (12/486) | 6.1% (9/148) | | 26mm | 35.0% (170/486) | 41.2% (61/148) | | 29mm | 58.4% (284/486) | 49.3% (73/148) | | 31mm | 4.1% (20/486) | 3.4% (5/148) | | Device Success1 | 84.6% (397/469) | 88.7% (125/141) | PMA P130021: FDA Summary of Safety and Effectiveness Data {31} | | Iliofemoral N=489 | Non-Iliofemoral N=150 | | --- | --- | --- | | Procedure Success² | 77.6% (370/477) | 77.5% (110/142) | | ¹ Device success is defined as deployment, only 1 valve implanted, only 1 valve in correct anatomic location, EOA >1.2cm² for 26, 29 and 31mm and ≥ 0.9 cm² for 23mm, mean gradient < 20mmHg, and aortic regurgitation < moderate. ² Procedure success is defined as device success and absence of in-hospital MACCE. Plus-minus values present the mean ± standard deviation. | | | ## Valve-in-Valve Experience In the "All Enrolled" population, a total of 17 patients had more than one CoreValve device implanted. Fourteen (14) patients had a CoreValve-in-CoreValve procedure (CViCV). All of the CViCV procedures were due to device malpositioning and/or aortic insufficiency; one of these patients received valve-in-valve due to native calcification causing under-expansion. Additionally, 3 patients had a non valve-in-valve implant of a second valve. ## Comparison between the Iliofemoral (IF) and Non-Iliofemoral (NIF) Cohorts Due to heterogeneity in the MCS procedure, patient characteristics (such as anatomy access, distinguishing differences not allowing for an iliofemoral approach) and potential clinical variability and outcome, the non-iliofemoral cohort is not included in the primary analysis. To provide contextual reference for the non-iliofemoral cohort, results of the subgroup analyses by iliofemoral and non-iliofemoral access sites for the primary endpoint and the key secondary endpoints #1-3 are presented in Figure 6, Table 26, and Table 27. The 12-month rate of all-cause mortality or major stroke for the "Attempted Implant" population of the non-iliofemoral cohort was 39.4% with an upper 95% CI of 47.2%, which was higher than that for the iliofemoral cohort. The non-iliofemoral cohort also had higher rates of MACCE, all-cause death, all-stroke, and MAE. PMA P130021: FDA Summary of Safety and Effectiveness Data {32} ![img-5.jpeg](img-5.jpeg) Figure 6: All-Cause Mortality or Major Stroke Kaplan-Meier Event Rate – Attempted Implant Table 24: All-Cause Mortality or Major Stroke – Attempted Implant | Interval Post Procedure (months)* | Iliofemoral (IF) N=489 | | | | Non-Iliofemoral (NIF) N=150 | | | | All N=639 | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | | # at start of interval | 489 | 441 | 392 | 360 | 150 | 127 | 102 | 90 | 639 | 568 | 494 | 450 | | # events in interval | 48 | 49 | 30 | 47 | 23 | 25 | 11 | 8 | 71 | 74 | 41 | 55 | | # event cumulative | 48 | 97 | 127 | 174 | 23 | 48 | 59 | 67 | 71 | 145 | 186 | 241 | | K-M Event Rate | 1.6 | 9.8 | 19.8 | 26.0 | 4.0 | 15.3 | 32.0 | 39.4 | 2.2 | 11.1 | 22.7 | 29.1 | | Lower 95% CI | 0.5 | 7.2 | 16.3 | 22.1 | 0.9 | 9.6 | 24.5 | 31.5 | 1.1 | 8.7 | 19.4 | 25.6 | | Upper 95% CI | 2.8 | 12.5 | 23.4 | 29.9 | 7.1 | 21.1 | 39.5 | 47.2 | 3.3 | 13.5 | 25.9 | 32.6 | | *0 = 0-29 days, 1 = 30-182 days, 6 = 183-364 days, 12 ≥ 365. Cumulative probability of event estimate is based on the Kaplan-Meier method. | | | | | | | | | | | | | Table 25: Kaplan-Meier Estimate of Event-Free Rates: Results by IF (N=489) and NIF (N=150) Cohorts | Secondary Objective | Event | Access Site | Days post Attempted Implant | | | p-value* | | --- | --- | --- | --- | --- | --- | --- | | | | | 30 days | 6 months (183 days) | 12 months (365 days) | | | #1 | MACCE | IF | 87.7% | 77.5% | 70.8% | 0.004 | | | | NIF | 82.7% | 65.3% | 58.6% | | | #2 | All-Cause Death | IF | 91.6% | 81.4% | 75.7% | 0.004 | | | | NIF | 88.7% | 71.3% | 64.0% | | PMA P130021: FDA Summary of Safety and Effectiveness Data {33} | Secondary Objective | Event | Access Site | Days post Attempted Implant | | | p-value* | | --- | --- | --- | --- | --- | --- | --- | | | | | 30 days | 6 months (183 days) | 12 months (365 days) | | | | Myocardial Infarction | IF | 98.8% | 98.5% | 98.0% | 0.861 | | | | NIF | 97.9% | 97.9% | 97.9% | | | | All Stroke | IF | 96.0% | 94.8% | 93.0% | 0.015 | | | | NIF | 91.2% | 88.0% | 87.0% | | | | Reintervention | IF | 98.9% | 98.5% | 98.2% | 0.408 | | | | NIF | 100.0% | 100.0% | 99.0% | | | #3 | MAE | IF | 46.2% | 40.1% | 37.2% | <0.001 | | | | NIF | 30.7% | 24.0% | 20.0% | | | *p-value from Log-Rank test comparing freedom from curves through 365 days | | | | | | | ## Gender Analysis The primary endpoint and secondary endpoints #1-3 (MACCE, individual MACCE components, and MAE) were examined for differences in outcome between genders. The 1-year all-cause mortality or major stroke K-M rate was 23.1% in the female group and 29.1% in the male group, as shown in Figure 7 and Table 28. No effect of gender on the primary endpoint was found. Additionally, no effect of gender on secondary endpoints #1-3 was found, as shown in Table 29. ![img-6.jpeg](img-6.jpeg) Figure 7: Primary Endpoint: All-Cause Mortality or Major Stroke by Gender – Iliofemoral Attempted Implant PMA P130021: FDA Summary of Safety and Effectiveness Data {34} Table 26: Primary Endpoint: All-Cause Mortality or Major Stroke by Gender – Iliofemoral Attempted Implant | Interval Post Procedure (months)* | Female N=255 | | | | Male N=234 | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | | # at start of interval | 255 | 234 | 209 | 194 | 234 | 207 | 183 | 166 | | # events in interval | 21 | 25 | 13 | 23 | 27 | 24 | 17 | 24 | | # event cumulative | 21 | 46 | 59 | 82 | 27 | 51 | 68 | 92 | | K-M Free From Event | 1.6 | 8.2 | 18.0 | 23.1 | 1.7 | 11.5 | 21.8 | 29.1 | | Lower 95% CI | 0.0 | 4.9 | 13.3 | 18.0 | 0.0 | 7.4 | 16.5 | 23.2 | | Upper 95% CI | 3.1 | 11.6 | 22.8 | 28.3 | 3.4 | 15.6 | 27.1 | 34.9 | | *0 = 0-29 days, 1 = 30-182 days, 6 = 183-364 days, 12 = ≥365 days. Cumulative probability of event estimate is based on the Kaplan-Meier method. | | | | | | | | | Table 27: Kaplan-Meier Estimate of Event-Free Rates: Results by Female (N=255) and Male (N=234) Cohorts | Secondary Endpoint | Event | Access Site | Days post Attempted Implant | | | p-value* | | --- | --- | --- | --- | --- | --- | --- | | | | | 30 days | 6 months (183 days) | 12 months (365 days) | | | #1 | MACCE | Female | 90.2% | 80.4% | 74.5% | 0.0521 | | | | Male | 85.0% | 74.4% | 66.7% | | | #2 | All-Cause Death | Female | 93.7% | 83.5% | 78.8% | 0.0855 | | | | Male | 89.3% | 79.1% | 72.2% | | | | Myocardial Infarction | Female | 99.6% | 99.6% | 98.7% | 0.2460 | | | | Male | 97.9% | 97.4% | 97.4% | | | | All Stroke | Female | 95.2% | 94.3% | 92.4% | 0.5562 | | | | Male | 97.0% | 95.4% | 93.8% | | | | Reintervention | Female | 100% | 100% | 99.5% | 0.0219 | | | | Male | 97.8% | 96.8% | 96.8% | | | #3 | MAE | Female | 43.1% | 38.4% | 35.3% | 0.1830 | | | | Male | 49.6% | 41.9% | 39.3% | | | *p-value from Log-Rank test comparing freedom from curves through 365 days | | | | | | | ## Mortality or Major Stroke Stratified by STS Score A post hoc analysis was conducted to compare the Kaplan-Meier (K-M) event rates for all-cause mortality or major stroke between Attempted Implant iliofemoral patients in different STS score categories (&lt;5%, 5-15%, &gt;15%), as shown in Figure 8 and Table 30. The majority of patients (n=341) had an STS score between 5-15 and the K-M rate of all-cause mortality or major stroke for these patients was similar to that for patients with an STS score of &lt;5 (23.5% and 25.0%, respectively, at 12 months). Patients with an STS score of &gt;15 had numerically higher event rates for all-cause mortality or major stroke at both 1 month (15.5%) and 12 months (36.9%) follow-up, indicating that very high STS scores did show predictive value in this patient population. The Log-rank p-value for the K-M analysis was 0.042, indicating a statistically significant difference in the event rate between the STS cohorts. PMA P130021: FDA Summary of Safety and Effectiveness Data {35} ![img-7.jpeg](img-7.jpeg) Figure 8: Primary Endpoint: All-Cause Mortality or Major Stroke Stratified by STS Score – Attempted Implant Iliofemoral Table 28: Primary Endpoint: All-Cause Mortality or Major Stroke Stratified by STS Score – Iliofemoral Attempted Implant | Interval Post Procedure | STS < 5% N=64 | | | | STS 5 - 15% N=341 | | | | STS > 15% N=84 | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | (months)* | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | | # at start of interval | 64 | 59 | 55 | 48 | 341 | 311 | 275 | 259 | 84 | 71 | 62 | 53 | | # events in interval | 5 | 4 | 7 | 5 | 30 | 36 | 14 | 34 | 13 | 9 | 9 | 8 | | # event cumulative | 5 | 9 | 16 | 21 | 30 | 66 | 80 | 114 | 13 | 22 | 31 | 39 | | Event Rate Estimate | 1.6 | 7.8 | 14.1 | 25.0 | 1.2 | 8.8 | 19.4 | 23.5 | 3.6 | 15.5 | 26.2 | 36.9 | | Lower 95% CI | 0.0 | 1.2 | 5.5 | 14.4 | 0.0 | 5.8 | 15.2 | 19.0 | 0.0 | 7.7 | 16.8 | 26.6 | | Upper 95% CI | 4.6 | 14.4 | 22.6 | 35.6 | 2.3 | 11.8 | 23.5 | 28.0 | 7.5 | 23.2 | 35.6 | 47.2 | | *0 = 0-29 days, 1 = 30-182 days, 6 = 183-364 days, 12 = ≥365 days. Cumulative probability of event estimate at the end of the interval (Pc) based on the Kaplan-Meier method. | | | | | | | | | | | | | PMA P130021: FDA Summary of Safety and Effectiveness Data {36} Post-Implant Aortic Regurgitation and All-Cause Mortality A post hoc sub-group analysis was performed for iliofemoral patients of the Implanted population to investigate the relationship between all-cause mortality and severity of aortic regurgitation at discharge (7 days post procedure or discharge, whichever is first). Four sub-groups of iliofemoral patients with none/trace, mild, moderate and severe total aortic regurgitation as assessed at discharge were analyzed. The results from the analysis are shown in Figure 9 and Table 31. All-cause mortality at 12 months was highest in the patients with severe aortic regurgitation (87.5%, note that only 8 patients were included in this subgroup) and was lowest in the patients with none/trace aortic regurgitation (17.8%). All-cause mortality in patients with mild aortic regurgitation (23.9%) was similar to freedom from mortality in patients with moderate aortic regurgitation (22.2%). These data indicate that aortic regurgitation up to mild in severity was not a strong driver of mortality in this study. ![img-8.jpeg](img-8.jpeg) Figure 9: All Cause Mortality Rate by Total Aortic Regurgitation at Discharge – Iliofemoral Implanted PMA P130021: FDA Summary of Safety and Effectiveness Data {37} Table 29: All Cause Mortality by Total Aortic Regurgitation at Discharge – Iliofemoral Implanted | Interval Post Procedure (months)* | None/Trace N=208 | | | | Mild N=180 | | | | Moderate N=54 | | | | Severe N=8 | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | 0 | 1 | 6 | 12 | | # at start of interval | 208 | 195 | 181 | 170 | 180 | 174 | 152 | 137 | 54 | 52 | 43 | 41 | 8 | 3 | 1 | 1 | | # events in interval | 13 | 14 | 10 | 19 | 6 | 22 | 15 | 21 | 2 | 9 | 1 | 6 | 5 | 2 | 0 | 0 | | # event cumulative | 13 | 27 | 37 | 56 | 6 | 28 | 43 | 64 | 2 | 11 | 12 | 18 | 5 | 7 | 7 | 7 | | Event Rate Estimate | 0.0 | 6.2 | 13.0 | 17.8 | 0.0 | 3.3 | 15.6 | 23.9 | 0.0 | 3.7 | 20.4 | 22.2 | 0.0 | 62.5 | 87.5 | 87.5 | | Lower 95% CI | NA | 3.7 | 9.1 | 13.2 | NA | 1.5 | 11.0 | 18.3 | NA | 0.9 | 11.8 | 13.2 | NA | 32.6 | 57.7…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...