← Product Code NQO · P110008

# COFLEX® INTERLAMINAR TECHNOLOGY (P110008)

_Companion Spine France · NQO · Oct 17, 2012 · Orthopedic · APPR_

**Canonical URL:** https://fda.innolitics.com/device/P110008

## Device Facts

- **Applicant:** Companion Spine France
- **Product Code:** NQO
- **Decision Date:** Oct 17, 2012
- **Decision:** APPR
- **Device Class:** Class 3
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The coflex® Interlaminar Technology is an interlaminar stabilization device indicated for use in one or two level lumbar stenosis from L1-L5 in skeletally mature patients with at least moderate impairment in function, who experience relief in flexion from their symptoms of leg/buttocks/groin pain, with or without back pain, and who have undergone at least 6 months of non-operative treatment. The coflex® is intended to be implanted midline between adjacent lamina of 1 or 2 contiguous lumbar motion segments. Interlaminar stabilization is performed after decompression of stenosis at the affected level(s).

## Device Story

The coflex® Interlaminar Technology is a U-shaped, titanium alloy (Ti6Al4V) implant designed for dynamic stabilization of the lumbar spine following surgical decompression. It is implanted midline between adjacent spinous processes at one or two contiguous levels (L1-L5). The device features ridged, bone-facing surfaces and vertical wings that are crimped intraoperatively to secure the implant to the spinous processes. By maintaining foraminal height and providing stabilization, the device aims to reduce pain and improve function. It is used in a clinical setting by surgeons. The device is MR Conditional. Clinical outcomes are assessed via patient-reported metrics (ODI, ZCQ, VAS) and radiographic evaluation of range of motion and foraminal height. The device provides an alternative to posterolateral fusion, potentially offering shorter operative times and reduced blood loss.

## Clinical Evidence

Prospective, randomized, multi-center, concurrently controlled study (IDE #G060059) comparing coflex® (n=215) to posterolateral fusion (n=107). Primary endpoint: Composite Clinical Success (CCS) at 24 months (ODI improvement ≥15, no reoperations/revisions/supplemental fixation, no major device-related complications, no epidural injections). Results: coflex® CCS 66.2% vs. fusion 57.7% (posterior probability of non-inferiority 0.999). coflex® showed shorter operative time (98.0 vs 153.2 min), less blood loss (109.7 vs 348.6 cc), and shorter hospital stay (1.90 vs 3.19 days). Spinous process fractures observed in 14.0% of coflex® patients; most were asymptomatic.

## Technological Characteristics

U-shaped interlaminar implant; material: Ti6Al4V titanium alloy (ASTM F136, ISO 5832-3). Five sizes (8-16mm). Features ridged bone-facing surfaces and vertical wings for spinous process fixation. MR Conditional (1.5T/3.0T). Sterilized via terminal sterilization (ISO 11137).

## Regulatory Identification

This device is an interspinous spacer or spinous process plate that is implanted between the spinous processes or attached to the spinous process.

## Reference Devices

- Expedium™ (Johnson and Johnson, Inc.)
- CD Horizon Legacy™ (Medtronic, Inc.)
- X-Stop®

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# Summary of Safety &amp; Effectiveness Data (SSED)

## I. GENERAL INFORMATION

Device Generic Name: Interlaminar Stabilization Device
Device Trade Name: coflex® Interlaminar Technology
Device Procode: NQO
Applicant’s Name and Address: Paradigm Spine, LLC
505 Park Ave, 14th Floor
New York, NY 10022
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P110008
Date of FDA Notice of Approval: October 17, 2012
Expedited: Not Applicable

## II. INDICATIONS FOR USE

The coflex® Interlaminar Technology is an interlaminar stabilization device indicated for use in one or two level lumbar stenosis from L1-L5 in skeletally mature patients with at least moderate impairment in function, who experience relief in flexion from their symptoms of leg/buttocks/groin pain, with or without back pain, and who have undergone at least 6 months of non-operative treatment. The coflex® is intended to be implanted midline between adjacent lamina of 1 or 2 contiguous lumbar motion segments. Interlaminar stabilization is performed after decompression of stenosis at the affected level(s).

## III. CONTRAINDICATIONS

- Prior fusion or decompressive laminectomy at any index lumbar level.
- Radiographically compromised vertebral bodies at any lumbar level(s) caused by current or past trauma or tumor (e.g., compression fracture).
- Severe facet hypertrophy that requires extensive bone removal which would cause instability.
- Grade II or greater spondylolisthesis.
- Isthmic spondylolisthesis or spondylolysis (pars fracture).
- Degenerative lumbar scoliosis (Cobb angle of greater than 25°).

PMA P110008: FDA Summary of Safety and Effectiveness Data

{1}

- Osteoporosis.
- Back or leg pain of unknown etiology.
- Axial back pain only, with no leg, buttock, or groin pain.
- Morbid obesity defined as a body mass index &gt; 40.
- Active or chronic infection – systemic or local.
- Known allergy to titanium alloys or magnetic resonance imaging (MRI) contrast agents.
- Cauda equina syndrome defined as neural compression causing neurogenic bowel or bladder dysfunction.

## IV. WARNING AND PRECAUTIONS

The warnings and precautions can be found in the coflex® Interlaminar Technology labeling.

## V. DEVICE DESCRIPTION

The coflex® Interlaminar Technology is an interlaminar functionally dynamic implant designed to impart a stabilization effect at the operative level(s). It consists of a single, U-shaped component, fabricated from medical grade titanium alloy (Ti6A14V, per ASTM F136 and ISO 5832-3). In clinical use, the “U” is positioned horizontally, with its apex oriented anteriorly and the two long arms of the “U” paralleling the long axis of the spinal processes. The bone-facing surfaces are ridged to provide resistance to migration.

![img-0.jpeg](img-0.jpeg)
Figure 1: coflex® Implanted in the Spine

A set of two wings extends vertically from the superior long arm of the “U”, with a second set of wings extending below the inferior long arm. Both sets of wings have

PMA P110008: FDA Summary of Safety and Effectiveness Data

{2}

serrated bone-facing surfaces, which are designed to further stabilize the coflex® device to the superior and inferior spinous processes, respectively, at the treated level. In addition, the opposing wing surfaces are spaced such that they surround the midportion of the spinous process between the base and the tip, but are more narrowly set (after intraoperative crimping, if necessary) than the flared posterior tip of the spinous process. Spacing of the superior and inferior wing sets is staggered, preventing overlapping of the wings if the coflex® device is implanted at adjacent levels.

To properly fit into the space between the spinous processes in a range of patient anatomies, the coflex® implant is manufactured in five sizes: 8, 10, 12, 14 and 16mm. The size corresponds to the size of the "U" as measured from opposing long arms. The number of teeth and the dimensions of the teeth are the same for all device sizes. The "gap" between the upper and lower arms of the "U" is 5mm for the size 8 device, 7mm for the size 10, 9mm for the size 12, 11mm for the size 14, and 13mm for the size 16.

## VI. ALTERNATIVE PRACTICES AND PROCEDURES

Non-surgical alternatives include non-steroidal anti-inflammatory medications, analgesics, oral and epidural steroids, an initial period of rest, physical therapy and bracing. Surgical alternatives to coflex® depend on the severity of the spinal stenosis, back pain, and instability and include various decompressive procedures (e.g., laminectomy, hemilaminectomy, foraminotomy etc.), interspinous process distraction devices (e.g., X-Stop®), and posterolateral fusion with pedicle screws. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his or her physician to select the method that best meets expectations and lifestyle.

## VII. MARKETING HISTORY

The coflex® has been commercially available in markets outside of United States since 2005. A listing of the countries in which the device has been commercially available is included below in Table 1. The coflex® has not been withdrawn from marketing in any of these markets.

Table 1: coflex® Marketing History

|  Argentina | Australia | Austria | Belgium  |
| --- | --- | --- | --- |
|  Bulgaria | Chile | China | Colombia  |
|  Czech Republic | Denmark | Egypt | Germany  |
|  Greece | Hong Kong | India | Indonesia  |
|  Israel | Italy | Jordan | Korea  |
|  Luxembourg | Malaysia | Mexico | Netherlands  |
|  New Zealand | Norway | Panama | Peru  |
|  Philippines | Poland | Portugal | Russia  |
|  Saudi Arabia | Singapore | Slovakia | Slovenia  |
|  South Africa | Spain | Sweden | Switzerland  |
|  Taiwan | Thailand | Turkey | UAE  |
|  UK | Ukraine | Venezuela | Vietnam  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{3}

# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

Below is a list of the potential adverse effects (e.g., complications) associated with the use of the coflex® Interlaminar Technology identified from the coflex® clinical study results, approved device labeling for other interlaminar devices, and published scientific literature including: (1) those associated with any surgical procedure; (2) those associated with decompressive procedures and posterolateral fusion for the treatment of spinal stenosis and instability; and (3) those associated with an interlaminar stabilization device, including the coflex® Interlaminar Technology. In addition to the risks listed below, there is also the risk that surgery may not be effective in relieving symptoms, or may cause worsening of symptoms. Additional surgery may be required to correct some of the adverse effects.

1. Risks associated with any surgical procedure include: infection; pneumonia; atelectasis; septicemia; injury to blood vessels; soft tissue damage; phlebitis, thromboembolus, or pulmonary embolus; hemorrhage; respiratory distress; pulmonary edema; reactions to the drugs or anesthetic agent used during and after surgery; reactions to transfused blood; failure of the tissue to heal properly (e.g., hematoma, seroma, dehiscence, etc.) which may require drainage, aspiration, or debridement or other intervention; incisional pain; heart attack; stroke; and death.

2. Risks associated with decompressive procedures and posterolateral fusion for treatment of spinal stenosis and instability include: damage to nerves leading to sensory or motor deficits; paralysis; parasthesia; cauda equina syndrome; damage to nerves, blood vessels, and nearby tissues; epidural bleeding, hematoma, or fibrosis; instability; blindness—secondary to pressure on the eye during surgery; surgery at incorrect level; osteolysis; injury to the spinal cord or the nerves leaving or entering the cord; loss of bowel or bladder function; retrograde ejaculation, sexual dysfunction, or sterility; disc herniation; injury to blood vessels; dural violation, with or without CSF leakage; impaired muscle or nerve function; hemorrhage; epidural injection reaction; epidural injection failure; fracture of the vertebrae, spinous process, or other damage to bony structures during or after surgery; postoperative muscle and tissue pain; surgery may not reduce the preoperative pain experienced; pain and discomfort associated with the presence of implants used to aid in the fusion surgery or reaction to the metal used in the implant, as well as the cutting and healing of tissues; failure of the fusion to heal or spontaneous fusion; the spine may undergo adverse changes or deterioration including loss of proper spinal curvature, correction, height, and/or reduction, or malalignment, and another surgery may be required; and adverse bone/implant interface reaction.

3. Risks associated with an interlaminar stabilization device, including the coflex® Interlaminar Technology, include: implant malposition or incorrect orientation; allergies to implant materials; possible wear debris, implantation at the wrong spinal level; fracture of the vertebrae, spinous process, or other damage to bony structures during or after surgery; the implant may loosen, deform, break, fatigue,

PMA P110008: FDA Summary of Safety and Effectiveness Data

{4}

or move, which may necessitate another surgery to correct the problem; and instruments also may break or malfunction in use, which may cause damage to the operative site or adjacent structures.

For the specific adverse events that occurred in the clinical studies, please see Section X below.

## IX. SUMMARY OF PRECLINICAL STUDIES

A variety of testing was conducted to characterize the performance of the coflex® Interlaminar Technology, as follows:

### Laboratory Studies

- Static Axial Compression
- Dynamic Axial Compression
- Dynamic Axial Compression following Static Axial Compression
- Dynamic Torsion
- Dynamic Torsion following Bending and Crimping Wings
- Biomechanical Wing Testing
- Expulsion
- Cadaveric Biomechanical Testing
- Functional Testing of Pliers

### Additional Studies

- Biocompatibility
- Sterilization Validation
- Shelf Life and Packaging Validation

### A. Laboratory Studies

Table 2: Laboratory Studies on coflex® Device

|  Test | Purpose | Method | Acceptance Criteria | Results  |
| --- | --- | --- | --- | --- |
|  Static Axial Compression | To evaluate the performance of the coflex® Interlaminar Technology under static axial compressive loading, under worst-case conditions. | Five (5) samples of the coflex® implant were tested under static compression. | The maximum compressive strength of the wire-EDM manufactured device (used in prior human clinical investigations), 218 N. | The mean yield load was 239 N.  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{5}

PMA P110008: FDA Summary of Safety and Effectiveness Data

|  Test | Purpose | Method | Acceptance Criteria | Results  |
| --- | --- | --- | --- | --- |
|  Dynamic Axial Compression | To evaluate the performance of the coflex® Interlaminar Technology under dynamic axial compressive loading, under worst-case conditions. | Fifteen (15) samples of the coflex® implant were tested under dynamic compression to 10 million cycles at 10 Hz (R=10). | The maximum dynamic compressive run-out load to 10 million cycles of the wire-EDM manufactured device (used in prior human clinical investigations), 150 N. | The maximum run-out load to 10 million cycles was 150 N.  |
|  Dynamic Axial Compression following Static Axial Compression | To evaluate the maximum load that the coflex® Interlaminar Technology can withstand under dynamic axial compressive loading, after the device is subjected to a static load exceeding the yield force. | Samples of the coflex® implant were initially statically loaded to 330 N, which caused plastic deformation. Subsequently, dynamic compression to 10 million cycles at 10 Hz was performed at multiple loads. | The maximum dynamic compressive run-out load of the milled coflex® device with no prior static loading, 150 N. | The maximum run-out load to 10 million cycles was 175 N.  |
|  Dynamic Torsion | To evaluate the maximum torsional load that the coflex® Interlaminar Technology can withstand. | Fifteen (15) samples of the coflex® implant were tested under dynamic torsion to 10 million cycles at 10 Hz. | The maximum dynamic torsion run-out load of the wire-EDM manufactured device (used in prior human clinical investigations), 75 N. | The maximum run-out load to 10 million cycles was 75 N.  |
|  Dynamic Torsion following Bending and Crimping Wings | To evaluate the maximum torsional load that the coflex® Interlaminar Technology can withstand after opening and closing the caudal wings to maximum displacement with the coflex® bending and crimping pliers. | Following bending and crimping of the wings, six (6) samples of the coflex® implant were tested under dynamic torsion to 10 million cycles at a frequency of 10 Hz. | The maximum dynamic torsion run-out load following bending and crimping of the wings is higher than the maximum forces on the wing in lateral bending (42N), axial rotation (27N), and flexion/extension (30N) [values from biomechanical wing test]. | The maximum run-out load to 10 million cycles was 50 N.  |

{6}

PMA P110008: FDA Summary of Safety and Effectiveness Data

|  Test | Purpose | Method | Acceptance Criteria | Results  |
| --- | --- | --- | --- | --- |
|  Biomechanical Wing Test | To evaluate the strain at various locations on the coflex® Interlaminar Technology when implanted in a cadaver spine and subjected to flexion/extension, lateral bending, and rotation. | Strain gauges were placed at 5 locations on the device to determine the maximum external moment in each loading mode. | In lateral bending, axial rotation, and flexion/extension the maximum bending force on the wing is less than or equal to the maximum dynamic torsion run-out load of 75N. | • In lateral bending the maximum wing bending moment was 0.42 Nm, and the maximum force on the wing was 42 N.
• In axial rotation, the maximum wing bending moment was 0.26 Nm, and the maximum force on the wing was 26 N.
• In flexion/extension, the maximum wing bending moment was 0.16 Nm, and the maximum force was 16 N.
• This test showed that the loads seen on the wings are low under worst case loading in flexion/extension, lateral bending, and axial rotation. Even with the worst case 10 Nm load, the test results showed that the device had adequate strength to survive 10 million cycles.  |
|  Expulsion | To evaluate the force required to cause expulsion of the coflex® Interlaminar Technology under worst-case conditions. Note: loading conditions in vivo present little or no potential for high axial push-out forces between the spinous processes that could cause device expulsion. | Five (5) samples of the coflex® implant were tested using standard polyurethane “sawbones” foam (grade 40). | • Indicate high expulsion resistance to withstand worst case in vivo forces as observed in similar devices (lumbar interbody cages) in the literature with 500N preload, 642-1033N. | • Initial testing under 100 N preload required 467 N force to cause expulsion.
• Second test to compare the milled devices to wire-EDM required 381 N force to cause expulsion.
• Third test performed under 500 N preload required 1318 N to cause expulsion.  |

{7}

|  Test | Purpose | Method | Acceptance Criteria | Results  |
| --- | --- | --- | --- | --- |
|  Cadaveric Biomechanical Testing | To evaluate the effect of the coflex® Interlaminar Technology on range of motion. | Eight (8) cadaveric specimens (L4/L5) were tested. The center of rotation was established to minimize off-axis bending. Specimens were preconditioned for 1,000 cycles at 500N. Compression (900N at 25cm/min); flexion/extension (±12 Nm); lateral bending (±12Nm); and axial rotation (±9 Nm) were applied under a 600 N preload. | Restoration of spine to normal flexion/extension, axial rotation and lateral bending ranges of motion. | • Partial destabilization increased flexion/extension relative to intact spine by 1.2:1. Implantation of the coflex® reduced this ratio to nearly 1:1, representing nearly complete restoration.
• In axial rotation a significant increase in ROM was observed after partial destabilization (~1.5:1) but it was restored to nearly normal (~1.2:1) after the implantation of coflex® device.
• In lateral bending the results of the cadaveric testing confirm that the coflex® device does not impose any significant limitation on lateral bending.  |
|  Functional Testing of Pliers | To ensure that the coflex® bending and crimping pliers allow the opening and closing of the device wings without damage to the implant. | Six (6) coflex® implants (8mm) were tested. All separation distances were measured. | Wings are required to be opened 50% compared to the baseline separation distance without damage to the implant following maximal spring back. | • Crimping produced an average decrease in wing separation distance of 42.4% (39-44%) compared to “open” dimension after bending.
• Compared to baseline (before bending) an increase of 5.3% (-0.3-9%) of the separation distance was produced.
• No distances showed evidence of damage.  |

Note that during the course of the clinical trial, the wings were modified slightly for ease of stacking two devices at adjacent levels. The holes in the wings were also removed. The modification was not the result of any clinical problems, safety issues or adverse events, product complaints, or surgeon requests from within or outside the United States. As this

PMA P110008: FDA Summary of Safety and Effectiveness Data

{8}

modification was minor, it did not affect the mechanical behavior of the device or the anticipated clinical outcome.

![img-1.jpeg](img-1.jpeg)
Figure 2: Original coflex® IDE Design (left) and Modified coflex® Design (right)

![img-2.jpeg](img-2.jpeg)

Table 3 provides a summary of the laboratory studies for the modified coflex® device.

Table 3: Laboratory Studies on Modified coflex® Device

|  Test | Purpose | Method | Acceptance Criteria | Results  |
| --- | --- | --- | --- | --- |
|  Static Axial Compression | To evaluate the performance of the modified coflex® Interlaminar Technology under static axial compressive loading, under worst-case conditions. | Five (5) samples of the coflex® implant were tested under static compression. | The maximum compressive strength of the original IDE device, 239 N. | The mean yield load was 309 N.  |
|  Dynamic Axial Compression | To evaluate the performance of the modified coflex® Interlaminar Technology under dynamic axial compressive loading, under worst-case conditions. | Fifteen (15) samples of the coflex® implant were tested under dynamic compression to 10 million cycles at 10 Hz. | Run out load equal to or greater than original IDE device, 150 N. | The maximum run-out load to 10 million cycles was 150 N.  |
|  Dynamic Torsion | To evaluate the maximum torsional load that the modified coflex® Interlaminar Technology can withstand. | Fifteen (15) samples of the coflex® implant were tested under dynamic torsion to 10 million cycles at 22 Hz and 18 Hz. | Run out load equal to or greater than original IDE device, 75 N. | The maximum run-out load to 10 million cycles was 75 N.  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{9}

PMA P110008: FDA Summary of Safety and Effectiveness Data

## B. Additional Studies

### 1. Biocompatibility

The coflex® implant is manufactured from standard medical grade Ti alloy, Ti6A14V, per ASTM F136 and ISO 5832-3. This is a standard material used in permanently implanted orthopaedic devices. However, cytotoxicity testing was performed to verify that the manufacturing process did not introduce any contaminants that could impact biocompatibility, in accordance with ISO 10993-5, Biological evaluation of medical devices; Part 5: Tests for in vitro cytotoxicity. The test results demonstrated that the finished device passed the cytotoxicity test with the score of ‘0’.

The coflex® surgical instruments are manufactured from titanium alloy per ASTM F136, stainless steel, and an acetal copolymer. Biocompatibility testing for the coflex® trials was performed as recommended in ISO 10993-1 for devices having limited duration contact with tissue or bone. Cytotoxicity, Sensitization, and Irritation testing was performed in accordance with ISO 10993-5 and ISO 10993-10. The test results demonstrate that all the trials (each with a different colorant) meet the acceptance criteria and are biocompatible for their intended use.

### 2. Sterilization

Sterilization validation according to EN556-1, Sterilization of medical devices - Requirements for terminally sterilized devices to be labeled ‘Sterile’ and ISO 11137, Sterilization of Health Care Products, Parts 1, 2 and 3 was conducted to confirm that the sterility of the device is maintained through a sterile barrier.

### 3. Shelf Life and Packaging Validation

Shelf life and packaging validation studies, including packaging seal and integrity, accelerated aging, and real-time aging testing, were conducted to demonstrate that the device packaging can maintain a sterile barrier, with a shelf life of 5 years.

### 4. MRI Compatibility

Non-clinical testing has demonstrated that the coflex® Interlaminar Technology is MR Conditional. It can be scanned safely under the following conditions:

- Static magnetic field of 1.5-Tesla (1.5T) or 3.0-Tesla (3.0T).
- Spatial gradient field of up to:
- 11,230 G/cm (112.3 T/m) for 1.5T systems
- 5,610 G/cm (56.1 T/m) for 3.0T systems.
- Maximum whole body averaged specific absorption rate (SAR) of:
- 2.0 W/kg for 15 minutes of scanning in Normal Operating Mode at 1.5T.
- 2.0 W/kg for 15 minutes of scanning in Normal Operating Mode at 3.0T.

{10}

PMA P110008: FDA Summary of Safety and Effectiveness Data

# 3.0T RF heating

In non-clinical testing with body coil excitation, the coflex® Interlaminar Technology produced a temperature rise of less than 3.5°C at a maximum whole body averaged specific absorption rate (SAR) of 2.0 W/kg, as assessed by calorimetry for 15 minutes of scanning in a 3.0T Siemens Trio (MRC20587) MR scanner with SYNGO MR A30 4VA30A software.

# 1.5T RF heating

In non-clinical testing with body coil excitation, the coflex® Interlaminar Technology produced a temperature rise of less than 3.5°C at a maximum whole body averaged specific absorption rate (SAR) of 2.0 W/kg, as assessed by calorimetry for 15 minutes of scanning in a 1.5T Siemens Espree (MRC30732) MR scanner with SYNGO MR B17 software.

Caution: The RF heating behavior does not scale with static field strength. Devices which do not exhibit detectable heating at one field strength may exhibit high values of localized heating at another field strength.

# MR Artifact

In testing using a 3.0T system with spin-echo sequencing, the shape of the image artifact follows the approximate contour of the device and extends radially up to 19 mm from the implant.

# X. SUMMARY OF CLINICAL STUDIES

The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of coflex® Interlaminar Technology for the treatment of moderate to severe spinal stenosis with back pain in the US under IDE #G060059. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.

## A. Study Design

Patients were treated between October 2006 and March 2010. The database for this PMA reflected data collected through March 2012. A total of 384 patients were enrolled consisting of up to 40 non-randomized "roll-in" patients and 344 randomized patients. Excluding 22 protocol violators, 215 randomized coflex® patients and 107 randomized control patients were enrolled. There were 21 investigational sites.

The study was a prospective, randomized, multi-center, concurrently controlled clinical study. Surgeons were blinded prior to patient randomization, and patients were blinded until after surgery. The control group was posterolateral fusion with autograft bone and pedicle screw fixation, following surgical decompression. Based on the well-established performance of posterolateral fusion in the medical literature, a 2:1 randomization ratio was applied with block randomization and a randomly

{11}

changing block size. A Bayesian statistical plan utilizing Jeffries non-informative priors and a single late-information time interim analysis was used to analyze the success of the device. After 70% of patients were evaluable for month 24 composite clinical success, the Bayesian posterior probability was to be computed and compared to 0.975. If larger than 0.975, the interim analysis sample was to be used to support approval. If not, the data on the remaining patients would be included in the analysis cohort after they complete 24 months of follow-up and again the posterior probability would be compared to 0.975 in a final analysis. Subsequently, FDA requested submission of the patient data for the entire cohort.

An independent Data Safety Monitoring Board (DSMB) evaluated all safety events on a quarterly basis during the course of the study to ensure patient safety was not compromised. All adverse events were independently reviewed and adjudicated by a Clinical Events Committee (CEC), with their decision binding on the study sponsor. All radiographs were analyzed by an independent core lab (Medical Metrics, Inc.).

The control group was the accepted standard of care for this indication, posterolateral fusion with pedicle screw fixation. The systems utilized were the Expedium™ (Johnson and Johnson, Inc.) and the CD Horizon Legacy™ (Medtronic, Inc.).

1. Clinical Inclusion and Exclusion Criteria

Enrollment in the coflex® study was limited to patients who met the following inclusion criteria.

- Radiographic confirmation of at least moderate lumbar stenosis, which narrows the central spinal canal at one or two contiguous levels from L1-L5 that require surgical decompression. Moderate stenosis is defined as &gt; 25% reduction of the antero-posterior dimension compared to the next adjacent normal level, with nerve root crowding compared to the normal level, as determined by the investigator on CT Scan or MRI. The patient may have, but is not required to have for inclusion in the study:

- Facet hypertrophy and subarticular recess stenosis at the affected level(s);
- Foraminal stenosis at the affected level(s);
- Up to Grade I stable degenerative spondylolisthesis (Meyerding classification) or equivalent retrolisthesis as determined by flexion/extension X-ray:

- For single level disease, there may be up to a Grade I stable spondylolisthesis or equivalent retrolisthesis at the affected level as determined on flexion/extension films by the investigator.
- For two level disease, there may be up to a Grade I stable spondylolisthesis or equivalent retrolisthesis at only one of the two contiguous affected levels as determined on flexion/extension films by the investigator. Patients with up to Grade I stable spondylolisthesis at two contiguous levels are excluded, but patients with up to Grade I stable spondylolisthesis at one level and equivalent retrolisthesis at the adjacent level may be included.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{12}

- Mild lumbar scoliosis (Cobb angle up to 25°)

- Radiographic confirmation of the absence of angular or translatory instability of the spine at index or adjacent levels (instability as defined by White &amp; Panjabi: Sagittal plane translation &gt;4.5mm or 15% or sagittal plane rotation &gt;15° at L1-L2, L2-L3, and L3-L4; &gt;20° at L4-L5 based on standing flexion/extension X-rays)

- VAS back pain score of at least 50 mm on a 100 mm scale.

- Neurogenic claudication as defined by leg/buttocks or groin pain that can be relieved by flexion such as sitting in a chair.

- Patient has undergone at least one epidural injection at any prior time point, AND at least 6 months of prior conservative care without adequate and sustained symptom relief.

- Age between 40 to 80 years.

- Oswestry Low Back Pain Disability Questionnaire score of at least 20/50 (40%).

- Appropriate candidate for treatment using posterior surgical approach.

- Psychosocially, mentally, and physically able to fully comply with this protocol, including adhering to scheduled visits, treatment plan, completing forms, and other study procedures.

- Personally signed and dated informed consent document prior to any study-related procedures indicating that the patient has been informed of all pertinent aspects of the trial.

Patients were not permitted to enroll in the coflex® study if they met any of the following exclusion criteria:

- More than two vertebral levels requiring surgical decompression.

- Prior surgical procedure that resulted in translatory instability of the lumbar spine [as defined by White &amp; Panjabi].

- More than one surgical procedure at any combination of lumbar levels.

- Prior fusion, implantation of a total disc replacement, complete laminectomy, or implantation of an interspinous process device at any lumbar level.

- Radiographically compromised vertebral bodies at any lumbar level(s) caused by current or past trauma or tumor (e.g., compression fracture).

- Severe facet hypertrophy that requires extensive bone removal which would cause instability.

- Isthmic spondylolisthesis or spondylolysis (pars fracture).

- Degenerative lumbar scoliosis (Cobb angle of greater than 25°).

- Disc herniation at any lumbar level requiring surgical intervention.

- Osteopenia: A screening questionnaire for osteopenia, SCORE (Simple Calculated Osteoporosis Risk Estimation), will be used to screen patients who require a DEXA bone mineral density measurement. If DEXA is required, exclusion will be defined as a DEXA bone density measured T score of ≤ -1.0 (The World Health Organization definition of osteopenia).

- Back or leg pain of unknown etiology.

- Axial back pain only, with no leg, buttock, or groin pain.

- Morbid obesity defined as a body mass index &gt; 40.

- Pregnant or interested in becoming pregnant in the next three years.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{13}

- Known allergy to titanium, titanium alloys, or MR contrast agents.
- Active or chronic infection – systemic or local.
- Chronically taking medications or any drug known to potentially interfere with bone/soft tissue healing (e.g., steroids), not including a medrol dose pack.
- History of significant peripheral neuropathy.
- Significant peripheral vascular disease (e.g., with diminished dorsalis pedis or posterior tibial pulses).
- Unremitting back pain in any position.
- Uncontrolled diabetes.
- Known history of Paget’s disease, osteomalacia, or any other metabolic bone disease (excluding osteopenia, which is addressed above).
- Cauda equina syndrome, defined as neural compression causing neurogenic bowel (rectal incontinence) or bladder (bladder retention or incontinence) dysfunction.
- Fixed and complete motor, sensory, or reflex deficit.
- Rheumatoid arthritis or other autoimmune diseases.
- Known or documented history of communicable disease, including AIDS, HIV, active Hepatitis
- Active malignancy: a patient with a history of any invasive malignancy (except nonmelanoma skin cancer), unless he/she has been treated with curative intent and there has been no clinical signs or symptoms of the malignancy for at least five years. Patients with a primary bony tumor are excluded as well.
- Prisoner or ward of the state.
- Subject has a history of substance abuse (e.g., recreational drugs, narcotics, or alcohol).
- Subject is currently involved in a study of another investigational product for similar purpose.
- Currently seeking or receiving workman’s compensation.
- In active spinal litigation.

## 2. Follow-up Schedule

All patients were scheduled to return for follow-up examinations at 6 weeks, 3 months, 6 months, 12 months, 18 months, and 24 months postoperatively.

Patients were evaluated for Oswestry Disability Index (ODI), Zurich Claudication Questionnaire (ZCQ), SF-12, back and leg pain (via visual analog scale (VAS)), and neurological assessment at preoperative visit and at all postoperative visits. Radiographic evaluation was performed at all timepoints. Adverse events and complications were recorded at all visits.

The key time points are shown below in the tables summarizing safety and effectiveness.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{14}

PMA P110008: FDA Summary of Safety and Effectiveness Data

# 3. Clinical Endpoints

The safety of the coflex® Interlaminar Technology was assessed by comparing adverse event incidence, epidural steroid injections, reoperations, revisions, and neurological function in comparison to the posterolateral fusion control group.

The effectiveness of the coflex® Interlaminar Technology was assessed by evaluating clinical pain and function (evaluated by ODI) compared to the posterolateral fusion control group.

Per the protocol, an individual patient was considered a Composite Clinical Success (CCS) if all of the following criteria were met at 24 months:

- Improvement of at least 15 points in the Oswestry Low Back Pain Disability Index (ODI) at 24 months compared to baseline;
- No reoperations, revisions, removals, or supplemental fixation; and
- No major device-related complications, including but not limited to permanent new or increasing sensory or motor deficit at 24 months; and
- No epidural steroid injections in the lumbar spine.

Overall study success criteria were based on a comparison of individual patient success rates, such that the patient success rate for the coflex® investigational group must be non-inferior to that of the posterolateral fusion control group. Bayesian statistical methods were used to obtain the posterior probabilities of non-inferiority and superiority. According to the statistical analysis plan, if non-inferiority was demonstrated, then superiority would be evaluated as defined more specifically in the analysis plan. The posterior probability threshold of 0.975 was used to determine non-inferiority.

Secondary effectiveness evaluations specified in the protocol included comparisons of the following: ZCQ Symptom Severity, ZCQ Physical Function, ZCQ Patient Satisfaction, Leg and Back Pain (via VAS), SF-12, time to recovery, and patient satisfaction.

In addition, several radiographic endpoints were considered in evaluating both safety and effectiveness, including index level and adjacent level range of motion, translation, instability, and device-related effects (e.g., device fracture or migration, fusion/non-fusion, spinous process fracture).

# B. Accountability of PMA Cohort

At the time of database lock (March 11, 2012), of 322 per protocol patients (215 coflex® and 107 fusion) enrolled in PMA study 95.7% (204 coflex® and 104 fusion) had data available for analysis at the completion of the study. Patient accountability is shown in Table 4, a patient accounting tree is shown in Figure 3, and a summary of data available at 24 months for each specific evaluation is provided in Table 5.

{15}

Table 4: Patient Accounting and Follow-Up Compliance Table – Efficacy Evaluable (PP) coflex® (I) and Fusion Control Patients (C)

|  Date of data transfer 03/11/2012 | Pre-Op |   | Week 6 |   | Month 3 |   | Month 6 |   | Month 12 |   | Month 18 |   | Month 24  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | I | C | I | C | I | C | I | C | I | C | I | C | I | C  |
|  (1) Theoretical follow-up | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107  |
|  (2) Cumulative deaths | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |
|  (3) Cumulative 'Study Failures' | 0 | 0 | 8 | 3 | 11 | 6 | 20 | 10 | 26 | 12 | 35 | 17 | 42 | 18  |
|  (4) Not Yet Overdue | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0  |
|  (5) Deaths+failures among theoretical due | 0 | 0 | 8 | 3 | 11 | 6 | 20 | 10 | 26 | 12 | 35 | 17 | 42 | 18  |
|  (6) Expected due for clinic visit^{a} | 215 | 107 | 207 | 104 | 204 | 101 | 195 | 97 | 189 | 95 | 180 | 90 | 172 | 89  |
|  (7) Failures among theoretical due | 0 | 0 | 8 | 3 | 11 | 6 | 20 | 10 | 26 | 12 | 35 | 17 | 42 | 18  |
|  (8) Expected due+failures among theoretical due | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107 | 215 | 107 | 214 | 107  |
|  All Evaluated Accounting (Actual^{b}) Among Expected Due Procedures  |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
|   | I | C | I | C | I | C | I | C | I | C | I | C | I | C  |
|  (9) # of procedures with any clinical data in interval | 215 | 107 | 205 | 104 | 200 | 99 | 189 | 95 | 176 | 94 | 163 | 83 | 162 | 86  |
|  (10) All Evaluated Visit Compliance (%) | 100.0% | 100.0% | 99.0% | 100.0% | 98.0% | 98.0% | 98.9% | 97.9% | 93.1% | 98.9% | 90.6% | 92.2% | 94.2% | 96.8%  |
|  (11) Change in Oswestry Disability Score | 215 | 107 | 202 | 102 | 196 | 96 | 187 | 95 | 176 | 92 | 163 | 83 | 162 | 86  |
|  (12) Radiographic evaluation | 215 | 107 | 202 | 102 | 196 | 98 | 186 | 95 | 171 | 93 | 149 | 79 | 139 | 68  |
|  (13) CCS at Month 24 |  |  |  |  |  |  |  |  |  |  |  |  | 204 | 104  |
|  (14) Actual^{a} % Follow-up for CCS at Month 24 or for change in ODI at other times. | 100.0% | 100.0% | 97.6% | 98.1% | 96.1% | 95.0% | 95.9% | 97.9% | 93.1% | 96.8% | 90.6% | 92.2% | 95.3% | 97.2%  |
|  Within Window Accounting (Actual^{b}) Among Expected Due  |   |   |   |   |   |   |   |   |   |   |   |   |   |   |
|   | I | C | I | C | I | C | I | C | I | C | I | C | I | C  |
|  (15) Change in Oswestry Disability Score | 215 | 107 | 184 | 93 | 187 | 92 | 165 | 82 | 168 | 88 | 151 | 72 | 149 | 78  |
|  (16) Radiographic evaluation | 215 | 107 | 183 | 94 | 188 | 94 | 162 | 82 | 164 | 88 | 137 | 69 | 131 | 63  |
|  (17) CCS at Mos. 24 |  |  |  |  |  |  |  |  |  |  |  |  | 191 | 95  |
|  (18) ActualA% Follow-up for CCS at Month 24 or and change in ODI at other times. | 100.0% | 100.0% | 88.9% | 89.4% | 91.7% | 91.1% | 84.6% | 84.5% | 88.9% | 92.6% | 83.9% | 80.0% | 89.3% | 88.8%  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{16}

Figure 3: Patient Accounting Tree for coflex® IDE Study
![img-3.jpeg](img-3.jpeg)
$^{1}$ Reasons for withdrawal prior to treatment: 17 patients failed to meet inclusion and exclusion criteria, 22 patients withdrew consent, and 13 patients elected not to have surgery.

Table 5: 24 Month Data Accounting for coflex IDE

|  Parameter | coflex® | Fusion Control  |
| --- | --- | --- |
|  Randomized | 262 | 136  |
|  Withdrawn Prior to Treatment | 32 | 22  |
|  Subjects Treated (mITT) | 230 | 114  |
|  Protocol Violators | 15 | 7  |
|  Per Protocol Cohort | 215 | 107  |
|  Radiologic Assessments: • Foraminal Height* • ROM • Translation • Fusion† | 180 (83.7%) 187 (87.0%) 185 (86.0%) n/a | n/a 102 (95.3%) 95 (88.8%) 102 (95.3%)  |
|  Clinical Failures Among Implanted¹ | 42 | 18  |
|  Expected (Per Protocol) | 172 | 89  |
|  ODI | 162 (94.2%) | 86 (96.6%)  |
|  ZCQ | 161 (93.6%) | 86 (96.6%)  |
|  VAS Leg and Back Pain | 162 (94.2%) | 85 (95.5%)  |
|  SF-12: • Physical Component Score • Mental Component Score | 132 (76.7%) 139 (80.8%) | 70 (78.7%) 75 (84.3%)  |

*This measurement taken only on coflex® patients
†This measurement taken only on fusion patients and defined as bridging bone
$^{1}$ Patients with Reoperations, Revisions, and Epidural Steroid Injection

PMA P110008: FDA Summary of Safety and Effectiveness Data

{17}

In the tables that follow throughout this summary, the randomized per protocol cohort is used for safety and efficacy analyses, unless otherwise indicated.

## C. Study Population Demographics and Baseline Parameters

The clinical study sites represent a mix between academic and community hospital settings, urban and regional settings of care, and were selected from varied geographic regions of the country.

Table 6: Summary of Baseline and Demographic Variables - coflex® and Fusion Control Efficacy Evaluable (PP) Cohorts

|   | coflex® |   |   | Fusion Control  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Demographics - All | N | Mean | SD | N | Mean | SD  |
|  Age at surgery (yrs) | 215 | 62.1 | 9.2 | 107 | 64.1 | 9.0  |
|  Height (inches) | 215 | 67.0 | 4.1 | 107 | 66.6 | 4.1  |
|  Weight (lbs) | 215 | 190.3 | 35.4 | 107 | 187.7 | 38.1  |
|  BMI (k/m²) | 215 | 29.7 | 4.5 | 107 | 29.6 | 4.9  |
|  Demographics - Male | N | Mean | SD | N | Mean | SD  |
|  Age at surgery (yrs) | 109 | 61.7 | 9.3 | 49 | 64.2 | 10.4  |
|  Height (inches) | 109 | 69.9 | 2.7 | 49 | 69.9 | 2.9  |
|  Weight (lbs) | 109 | 207.1 | 27.3 | 49 | 207.6 | 32.3  |
|  BMI (k/m²) | 109 | 29.8 | 3.7 | 49 | 29.7 | 4.4  |
|  Demographic - Female | N | Mean | SD | N | Mean | SD  |
|  Age at surgery (yrs) | 106 | 62.6 | 9.1 | 58 | 64.1 | 7.7  |
|  Height (inches) | 106 | 64.0 | 2.9 | 58 | 63.8 | 2.5  |
|  Weight (lbs) | 106 | 173.1 | 34.6 | 58 | 170.8 | 34.5  |
|  BMI (k/m²) | 106 | 29.6 | 5.2 | 58 | 29.5 | 5.4  |
|  Baseline Functional Status | N | Mean | SD | N | Mean | SD  |
|  Oswestry (ODI) | 215 | 60.8 | 11.8 | 107 | 60.7 | 11.5  |
|  Zurich Claudication Qx Severity | 214 | 3.6 | 0.6 | 107 | 3.6 | 0.6  |
|  Zurich Claudication Qx Physical | 214 | 2.7 | 0.4 | 107 | 2.8 | 0.4  |
|  SF-12 PCS (Physical) | 195 | 28.1 | 6.6 | 95 | 28.2 | 6.0  |
|  SF-12 MCS (Mental Health) | 195 | 45.5 | 13.0 | 95 | 44.9 | 12.2  |
|  VAS Back pain | 215 | 79.5 | 15.0 | 106 | 79.2 | 13.5  |
|  VAS Leg pain (worse leg) | 215 | 76.0 | 20.4 | 106 | 78.3 | 18.4  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{18}

Table 7: Summary of Baseline and Demographic Categorical Variables - coflex® and Fusion Control Efficacy Evaluable (PP) Cohorts

|   | coflex® |   | Control  |   |
| --- | --- | --- | --- | --- |
|   | n | % | n | %  |
|  Number of subjects | 215 |  | 107 |   |
|  Males | 109 | 50.7 | 49 | 45.8  |
|  Females | 106 | 49.3 | 58 | 54.2  |
|  Number of levels | n | % | n | %  |
|  1-level decompression | 138 | 64.2 | 68 | 63.6  |
|  2-level decompression | 77 | 35.8 | 39 | 36.4  |
|  Current smoker | n | % | n | %  |
|  Yes | 22 | 10.2 | 15 | 14.0  |
|  No | 193 | 89.8 | 92 | 86.0  |
|  Comorbidities | n | % | n | %  |
|  Cardiovascular | 137 | 63.7 | 74 | 69.2  |
|  Musculoskeletal | 112 | 52.1 | 61 | 57.0  |
|  Endocrine | 55 | 25.6 | 35 | 32.7  |
|  Duration of Back Pain | n | % | n | %  |
|  None | 0 | 0.0 | 0 | 0.0  |
|  Fewer than 6 months | 3 | 1.4 | 1 | 0.9  |
|  6 months to a year | 24 | 11.2 | 14 | 13.1  |
|  More than one year | 188 | 87.4 | 92 | 86.0  |
|  Duration of Leg Pain (maximum) | n | % | n | %  |
|  None | 1 | 0.5 | 1 | 0.9  |
|  Fewer than 6 months | 6 | 2.8 | 8 | 7.5  |
|  6 months to a year | 38 | 17.7 | 22 | 20.6  |
|  More than one year | 170 | 79.1 | 76 | 71.0  |
|  Duration of Buttock Pain | n | % | n | %  |
|  None | 32 | 14.9 | 21 | 19.6  |
|  Fewer than 6 months | 11 | 5.1 | 7 | 6.5  |
|  6 months to a year | 41 | 19.1 | 22 | 20.6  |
|  More than one year | 131 | 60.9 | 57 | 53.3  |
|  Duration of Groin Pain | n | % | n | %  |
|  None | 157 | 73.0 | 74 | 69.2  |
|  Fewer than 6 months | 6 | 2.8 | 5 | 4.7  |
|  6 months to a year | 13 | 6.0 | 12 | 11.2  |
|  More than one year | 39 | 18.1 | 16 | 15.0  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{19}

Table 8: Summary of Baseline and Demographic Categorical Variables - coflex® and Fusion Control Efficacy Evaluable (PP) Cohorts (Continued)

|   | coflex® |   | Control  |   |
| --- | --- | --- | --- | --- |
|  Previous Conservative Treatment of the Spine | n | % | n | %  |
|  None | 28 | 13.0 | 9 | 8.4  |
|  Physical therapy | 132 | 61.4 | 70 | 65.4  |
|  NSAIDs/ASA/Acetinomphen only | 121 | 56.3 | 65 | 60.7  |
|  Chiropractic | 82 | 38.1 | 41 | 38.3  |
|  Corset/Brace | 37 | 17.2 | 22 | 20.6  |
|  Any narcotic use | 107 | 49.8 | 55 | 51.4  |
|  Other | 34 | 15.8 | 15 | 14.0  |
|  Previous Surgical Treatment of the Spine | n | % | n | %  |
|  None | 0 | 0.0 | 0 | 0.0  |
|  Discectomy | 4 | 1.9 | 0 | 0.0  |
|  Fusion | 3 | 1.4 | 0 | 0.0  |
|  IDET | 1 | 0.5 | 1 | 0.9  |
|  Epidural injections | 210 | 97.7 | 105 | 98.1  |
|  Other injections | 35 | 16.3 | 18 | 16.8  |
|  Laminotomy | 10 | 4.7 | 2 | 1.9  |
|  Race | n | % | n | %  |
|  American Indian / Alaskan Native | 1 | 0.5 | 3 | 2.8  |
|  Asian | 4 | 1.9 | 3 | 2.8  |
|  Black or African American | 11 | 5.1 | 6 | 5.6  |
|  White | 191 | 88.8 | 93 | 86.9  |
|  Other | 8 | 3.7 | 2 | 1.9  |

D. Safety and Effectiveness Results

1. Safety Results

The analysis of safety was based on the per-protocol cohort of 322 patients (215 coflex® patients and 107 fusion patients). Adverse events reported by the investigating surgeons and adjudicated by the CEC are reported in Table 9 to Table 11. The key safety outcomes for this study are presented below in Table 12 through Table 16.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{20}

Table 9: Incidence of Adverse Events coflex® and Fusion Control Efficacy Evaluable (PP) Cohort

|   | coflex®
(N=215) |   | Control
(N=107)  |   |
| --- | --- | --- | --- | --- |
|   | n | % | n | %  |
|  Operative Site |  |  |  |   |
|  Pain; new, + frequency, worsening | 71 | 33.0% | 37 | 34.6%  |
|  Wound problems^{1} | 30 | 14.0% | 9 | 8.4%  |
|  Fracture^{2} | 11 | 5.1% | 2 | 1.9%  |
|  Other^{3} | 9 | 4.2% | 3 | 2.8%  |
|  Component loosening | 3 | 1.4% | 4 | 3.7%  |
|  Component migration | 3 | 1.4% | 1 | 0.9%  |
|  Component breakage | 2 | 0.9% | 2 | 1.9%  |
|  Infection (deep) | 2 | 0.9% | 0 | 0.0%  |
|  Component deformation | 0 | 0.0% | 0 | 0.0%  |
|  Incidental durotomy (<= 5 mm) | 0 | 0.0% | 0 | 0.0%  |
|  Tear >5mm | 0 | 0.0% | 0 | 0.0%  |
|  Heterotopic ossification | 0 | 0.0% | 0 | 0.0%  |
|  Hematoma requiring drainage | 0 | 0.0% | 1 | 0.9%  |
|  Non-Operative Site |  |  |  |   |
|  Musculoskeletal^{4} | 121 | 56.3% | 65 | 60.7%  |
|  Neurological^{5} | 51 | 23.7% | 23 | 21.5%  |
|  Other^{6} | 29 | 13.5% | 16 | 15.0%  |
|  Cardiovascular | 21 | 9.8% | 11 | 10.3%  |
|  Gastrointestinal | 15 | 7.0% | 12 | 11.2%  |
|  Skin and Subcutaneous Tissue | 14 | 6.5% | 9 | 8.4%  |
|  Genitourinary | 13 | 6.0% | 9 | 8.4%  |
|  Respiratory | 9 | 4.2% | 6 | 5.6%  |
|  Endocrine/Metabolic | 8 | 3.7% | 4 | 3.7%  |
|  Cancer/Neoplasm | 6 | 2.8% | 9 | 8.4%  |
|  EENT | 6 | 2.8% | 4 | 3.7%  |
|  Hematological | 5 | 2.3% | 4 | 3.7%  |
|  Immune | 1 | 0.5% | 0 | 0.0%  |
|  Psychiatric/Substance abuse | 1 | 0.5% | 7 | 6.5%  |

1. Wound problems: Include wound drainage, superficial infections, dehiscence, seroma, and delayed healing of incision
2. Fracture: Includes spinous process fracture, pars fracture, and other fractures of the vertebral bodies reported by investigators.
3. Other Operative Site: Includes events not placed into a specific category by investigators, including clicking sound, spondylolisthesis, drain complications, incisional pain, spinal swelling, and cellulitis.
4. Musculoskeletal: Includes weakness, cramping, joint pain, joint surgery or replacement, and other non-lumbar spinal musculoskeletal tissues.
5. Neurological: Includes balance problems, headaches, numbness and/or tingling, and changes in sensation.
6. Other Non-Operative Site: Includes psychological disorders, infectious diseases, insomnia, and fever.

Table 9 shows the comparison of percentages of complications between the coflex® and fusion Per Protocol cohorts at specific operative and non-operative

PMA P110008: FDA Summary of Safety and Effectiveness Data

{21}

sites. With the exception of wound problems, adverse events rates were comparable between coflex® and fusion control. The numerical difference of wound complications between coflex® 14.0% (30/215) and control 8.4% (9/107) was 5.6%. This difference was not statistically significant. Table 10 demonstrates the time course of all adverse events.

Table 10: Time Course of Adverse Events coflex® (I) and Fusion Control (C) Efficacy Evaluable (PP) Cohort

|   | Day of Surgery Relative Day 0 |   | Immed. Post-Op to Month 3 (RelDay 1-90) |   | >Mo. 3 to Mo 6 (RelDay 91-180) |   | >Mo. 6 to Mo.12 (RelDay 181-365) |   | >Mo. 12 to Mo. 24 (RelDay 365-730)  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | I | C | I | C | I | C | I | C | I | C  |
|  Expected Due | 215 | 107 | 204 | 101 | 195 | 97 | 189 | 95 | 172 | 89  |
|  Operative Site |  |  |  |  |  |  |  |  |  |   |
|  Pain; new, + frequency, w orsening | 0 | 0 | 21 | 10 | 13 | 11 | 25 | 7 | 24 | 17  |
|  Wound problems | 2 | 0 | 29 | 10 | 0 | 0 | 0 | 0 | 0 | 0  |
|  Fracture | 1 | 0 | 4 | 0 | 3 | 2 | 1 | 1 | 1 | 0  |
|  Other | 0 | 0 | 2 | 2 | 1 | 0 | 2 | 1 | 4 | 0  |
|  Device component loosening | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 2 | 2  |
|  Device component migration | 0 | 0 | 2 | 0 | 0 | 1 | 0 | 0 | 1 | 0  |
|  Device component breakage | 1 | 0 | 0 | 0 | 0 | 1 | 1 | 4 | 0 | 0  |
|  Infection (deep) | 0 | 0 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |
|  Hematoma requiring drainage | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |
|  Non-Operative Site |  |  |  |  |  |  |  |  |  |   |
|  Musculoskeletal | 1 | 1 | 61 | 27 | 26 | 27 | 59 | 24 | 72 | 34  |
|  Neurological | 0 | 0 | 25 | 7 | 11 | 9 | 16 | 3 | 25 | 11  |
|  Other | 0 | 0 | 12 | 3 | 3 | 2 | 1 | 2 | 14 | 6  |
|  Cardiovascular | 1 | 1 | 2 | 4 | 5 | 0 | 8 | 4 | 9 | 3  |
|  Gastrointestinal | 0 | 0 | 3 | 2 | 3 | 2 | 10 | 1 | 4 | 5  |
|  Skin and Subcutaneous Tissue | 0 | 1 | 4 | 5 | 1 | 1 | 6 | 2 | 4 | 2  |
|  Genitourinary | 0 | 2 | 4 | 4 | 1 | 1 | 0 | 0 | 5 | 2  |
|  Respiratory | 0 | 0 | 3 | 3 | 2 | 0 | 2 | 1 | 3 | 3  |
|  Endocrine/Metabolic | 0 | 0 | 1 | 0 | 0 | 1 | 0 | 0 | 5 | 1  |
|  Cancer/Neoplasm | 0 | 0 | 1 | 0 | 1 | 0 | 0 | 1 | 2 | 5  |
|  EENT | 0 | 0 | 0 | 0 | 2 | 0 | 0 | 0 | 2 | 1  |
|  Hematological | 0 | 1 | 2 | 1 | 1 | 0 | 0 | 1 | 2 | 2  |
|  Immune | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0  |
|  Psychiatric/Substance abuse | 0 | 0 | 0 | 3 | 1 | 1 | 0 | 0 | 0 | 2  |
|  Total | 6 | 7 | 178 | 81 | 74 | 59 | 132 | 53 | 180 | 96  |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{22}

Table 11: Numbers of Specific Device and Surgery Related Complications by Time of Occurrence
coflex® (I) and Fusion Control (C) Efficacy Evaluable (PP) Cohort

|  Type of Adverse Event/Complication | Day of Surgery Relative Day 0 |   | Immed. Post-Op to Mth 3 (Day 1-90) |   | >Mth 3 to Mth 6 (Day 91-180) |   | >Mth 6 to Mth 12 (Day 181-365) |   | >Mth 12 to Mth 24 (Day 365-730) |   | Overall  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Treatment Group (I = coflex®, C = control) | I | C | I | C | I | C | I | C | I | C | I (%) | C (%)  |
|  # Patients at each Follow-Up Interval | 215 | 107 | 204 | 101 | 195 | 97 | 189 | 95 | 172 | 89 | (N=215) | (N=107)  |
|  DEVICE-RELATED ADVERSE EVENTS¹  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Device migration | 0 | 0 | 2 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 2 (0.9%) | 1 (0.9%)  |
|  Device breakage | 1 | 0 | 0 | 0 | 0 | 1 | 1 | 2 | 0 | 0 | 2 (0.9%) | 3 (2.8%)  |
|  Device loosening | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 2 | 1 | 3 (1.4%) | 2 (1.9%)  |
|  Fracture | 0 | 0 | 3 | 0 | 1 | 0 | 1 | 0 | 0 | 0 | 5 (2.3%) | -  |
|  SUBTOTAL | 1 | 0 | 5 | 0 | 1 | 2 | 3 | 3 | 2 | 1 | 12 (5.6%) | 6 (5.6%)  |
|  SURGERY-RELATED ADVERSE EVENTS¹  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Wound problems | 2 | 0 | 28 | 7 | 0 | 0 | 0 | 0 | 0 | 0 | 30 (14.0%) | 7 (6.5%)  |
|  Decompression-Related Fracture | 0 | 0 | 1 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 2 (0.9%) | -  |
|  Hematoma requiring drainage | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | - | 1 (0.9%)  |
|  Infection (deep) | 0 | 0 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 2 (0.9%) | -  |
|  Pain, Back | 0 | 0 | 9 | 7 | 6 | 6 | 14 | 7 | 18 | 11 | 47 (21.9%) | 31 (29.0%)  |
|  Pain, Leg/Buttock and Back | 0 | 0 | 2 | 0 | 0 | 0 | 1 | 0 | 0 | 0 | 3 (1.4%) | -  |
|  Pain, Leg /Buttock | 0 | 0 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 2 (0.9%) | -  |
|  Pain, Back & Leg | 0 | 0 | 5 | 0 | 2 | 4 | 1 | 0 | 4 | 4 | 12 (5.6%) | 8 (7.5%)  |
|  Pain, Back & Buttock | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 0 | 0 | 1 (0.5%) | -  |
|  Pain, Buttock | 0 | 0 | 1 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 2 (0.9%) | -  |
|  Pain, Leg | 0 | 0 | 3 | 3 | 4 | 1 | 4 | 0 | 2 | 0 | 13 (6.0%) | 4 (3.7%)  |
|  Pain, Hip | 0 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | - | 1 (0.9%)  |
|  SUBTOTAL | 2 | 1 | 53 | 18 | 14 | 11 | 21 | 7 | 24 | 15 | 114 (53.0%) | 52 (48.6%)  |
|  TOTAL # of Events | 3 | 1 | 58 | 18 | 15 | 13 | 24 | 10 | 26 | 16 | 126 (58.6%) | 58 (54.2%)  |

¹Selected adverse events are described in more detail in Table 8.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{23}

# Spinous Process Fractures:

Spinous process fractures were observed by the core radiographic laboratory in 30 coflex® patients (14.0%) and 8 fusion patients (11.9% of patients with spinous processes retained by partial laminectomy). Spinous process fractures were also observed by the investigator surgeons. The incidence of fractures observed by the surgeons differed from that observed by the core radiographic laboratory, as 8 coflex® patients (3.7%) and no fusion patients (0.0%) had spinous process fractures noted by the investigational sites. 83% of patients in the coflex® group and 75% of patients in fusion group who had spinous process fractures observed by the radiographic laboratory did not have any associated symptoms at the time the fracture was observed. Table 12 and Table 13 detail the incidence of spinous process fractures in coflex® and fusion patients.

Table 12: Spinous Process Fracture Incidence in coflex® IDE Study

|   | coflex® |   | Fusion Control  |   |
| --- | --- | --- | --- | --- |
|   | n/N | % | n/N | %  |
|  Spinous Process Fracture | 30/215 | 14.0% | 8/67¹ | 11.9%  |

¹Fusion patients with spinous processes retained by partial laminectomy.

Table 13: Time Course of Spinous Process Fracture Incidence in coflex® IDE Study

|  Group | Time of Initial Fracture Observation |   |   |   |   |   |   | Total  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Post-op | 6 W | 3 M | 6 M | 12 M | 18 M | 24 M  |   |
|  coflex® | 5 | 13 | 6 | 1 | - | - | 5¹ | 30  |
|  Fusion Control | 4 | 2 | 2 | - | - | - | - | 8  |

¹ 3 out of the 5 observations at 24 months had unreadable or missing 6 week, 3 month, 6 month, 12 month, and 18 month X-rays.

By month 24, 48% of the coflex® spinous process fractures were resolved. Of the unresolved spinous process fractures, 75% were asymptomatic and resulted in no clinical sequelae or loss of foraminal height during the study. None (0%) of the fusion spinous process fractures were resolved by month 24, and 75% of these patients were asymptomatic.

The adverse event rate associated with spinous process fractures was not significantly higher than the patients without spinous process fractures. The long term effects of these spinous process fractures past 24 months are unknown.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{24}

# Surgery and Hospitalization Data:

Table 14: Summary of Operative Details Continuous Variables coflex® and Fusion Control Efficacy Evaluable (PP) Cohorts

|   | coflex® |   |   |   | Fusion Control  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  1- and 2-level procedures | N | Mean | SD | 95% CI (LB, UB) | N | Mean | SD | 95% CI (LB, UB)  |
|  Hospital LOS (days) | 215 | 1.90 | 1.08 | (1.75, 2.04) | 107 | 3.19 | 1.61 | (2.88, 3.50)  |
|  Estimated blood loss (cc) | 215 | 109.7 | 120.0 | (93.5, 125.8) | 105 | 348.6 | 281.8 | (294.0, 403.1)  |
|  Operative time (minutes) | 214 | 98.0 | 41.1 | (92.5, 103.6) | 107 | 153.2 | 55.5 | (142.5, 163.8)  |
|  1-level procedures | N | Mean | SD | 95% CI (LB, UB) | N | Mean | SD | 95% CI (LB, UB)  |
|  Hospital LOS (days) | 138 | 1.86 | 1.14 | (1.66, 2.05) | 68 | 2.87 | 1.45 | (2.52, 3.22)  |
|  Estimated blood loss (cc) | 138 | 98.0 | 96.3 | (81.8, 114.3) | 66 | 290.9 | 207.0 | (240.0, 341.8)  |
|  Operative time (minutes) | 137 | 90.8 | 44.0 | (83.4, 98.2) | 68 | 142.0 | 56.0 | (128.4, 155.5)  |
|  2-level procedures | N | Mean | SD | 95% CI (LB, UB) | N | Mean | SD | 95% CI (LB, UB)  |
|  Hospital LOS (days) | 77 | 1.97 | 0.95 | (1.76, 2.19) | 39 | 3.74 | 1.74 | (3.18, 4.31)  |
|  Estimated blood loss (cc) | 77 | 130.5 | 152.1 | (95.9, 165.0) | 39 | 446.2 | 358.4 | (330.0, 562.3)  |
|  Operative time (minutes) | 77 | 110.9 | 31.8 | (103.7, 118.1) | 39 | 172.7 | 49.3 | (156.7, 188.7)  |
|  The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

Table 14 demonstrates that the average operating time in the fusion patients was 55.2 minutes greater than the coflex® patients. Average blood loss in fusion patients was 238.9 cc greater in the fusion patients than in coflex® patients. The average hospital length of stay was 1.29 days longer in the fusion patients.

# Reoperations and Revisions:

Through 24 months of follow up, the overall reoperation rate was 10.7% in the coflex® group and 7.5% in the fusion control. Reoperations where the device was maintained are summarized in Table 15 and revision surgeries are summarized in Table 16.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{25}

Table 15: Reoperation Events in the coflex® Clinical Trial

|  Reoperation Type | Treatment Group | Event Time Course (months) |   |   |   |   |   |   | Total (events) | Reasons  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  <1.5 | 1.5-3 | 3-6 | 6-12 | 12-24 | 24-36 | 36-48  |   |   |
|  Irrigation and Debridement | coflex® | 4 | - | - | - | - | - | - | 4 | 2 wound dehiscence, 2 deep infections  |
|  Supplemental Decompression | coflex® | - | - | - | 1 | 1 | 1 | 1 | 4 | 3 leg and/or low back pain, 1 herniation  |
|  CSF Repair | coflex® | 1 | - | - | - | - | - | - | 1 | 1 CSF leak  |
|  Non-Index Lumbar Fusion | coflex® | - | - | - | - | - | 1 | 1 | 2 | 2 leg and/or low back pain  |
|  Hematoma Drainage | Fusion | 1 | - | - | - | - | - | - | 1 | 1 wound hematoma  |
|  Irrigation and Debridement | Fusion | - | - | - | - | - | 2 | - | 2 | 2 deep infections^{1}  |
|  Supplemental Decompression | Fusion | - | - | - | - | - | 1 | 1 | 2 | 1 synovial cyst, 1 herniation  |

¹A single fusion patient had 2 operations for deep infection

Table 16: Revision Events in the coflex® Clinical Trial

|  Revision Type | Treatment Group | Event Time Course (months) |   |   |   |   |   |   | Total (events) | Reasons  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  <1.5 | 1.5-3 | 3-6 | 6-12 | 12-24 | 24-36 | 36-48  |   |   |
|  Device replacement (with coflex®) | coflex® | - | 2 | - | - | - | - | - | 2 | 1 bone-related fracture, 1 seroma  |
|  Decompression and Device Removal | coflex® | - | - | - | 1 | 1 | - | - | 2 | 2 leg and/or low back pain  |
|  Transition to fusion | coflex® | - | - | 2 | 4 | 7 | 6 | 3 | 22 | 14 leg and/or low back pain^{2}, 4 bone-related fracture, 2 component loosening, 1 herniation, 1 synovial cyst  |
|  Debridement and Device Removal | coflex® | 1 | - | - | - | - | - | - | 1 | 1 deep infection  |
|  Device Removal | Fusion | - | - | - | - | - | - | 2 | 2 | 1 component loosening, 1 back and/or leg pain  |
|  Device replacement | Fusion | - | - | - | 1 | 3 | - | 1 | 5 | 2 broken pedicle screws^{1}, 3 component loosening  |
|  Adjacent level extension | Fusion |  | 1 | 1 | 1 | 2 | 3 | 2 | 10 | 7 back and/or leg pain, 2 pseudoarthrosis, 1 bone-related fracture  |

¹A single fusion patient had 2 revisions for broken pedicle screws
²Three coflex® patients had a transition to fusion after a previous reoperation or replacement of coflex®.

Through 24 months, the reoperations and revisions in the coflex® group included 5 irrigation and debridement procedures (including 1 cerebrospinal fluid leak), 2 supplemental decompression surgeries retaining the device, 2 revisions for coflex® removal &amp; replacement, 2 decompressions and device removal, 1 debridement and device removal, and 13 (6.0%, 13/215) conversions to primary fusion. Two patients had a reoperation prior to a revision. There were no revisions related to device breakage.

Through 24 months, the reoperations and revisions in the fusion control group included 1 reoperation due to post-operative hematoma, 4 revisions of the fusion system due to

PMA P110008: FDA Summary of Safety and Effectiveness Data

{26}

device breakage or component loosening, and 5 extensions of the fusion to an adjacent level.

Between 24 months and 48 months of follow up, there were 13 additional reoperations or revisions in 12 coflex® patients (6.3% (12/192)) and 12 additional reoperations or revisions in 10 fusion patients (10.1% (10/99)). One of each of the coflex® and fusion revisions was in a patient who had a reoperation prior to 2 years. Based on available patient data through 48 months, the coflex® revision rate is 15.8% and the fusion control revision rate is 15.9%.

## 2. Effectiveness Results

### Primary Effectiveness Analysis:

The analysis of effectiveness was based on the per protocol cohort of 322 patients (215 coflex® patients and 107 fusion patients) evaluable at the 24-month time point. Key effectiveness outcomes are presented in Table 17 through Table 32.

Table 17: Posterior Probabilities of Success at 24 Months in coflex® Clinical Trial

|   | Number and Percentage Achieving Month 24 CCS* |   |   |   |   |   | Posterior Probability of Non-Inferiority  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   | Fusion Control  |   |   |   |
|   | N | n | % | N | n | %  |   |
|  Month 24 | 204 | 135 | 66.2% | 104 | 60 | 57.7% | 0.999  |

*Composite Clinical Success

Non-inferiority of the coflex® group compared to the control group was demonstrated for the Composite Clinical Success (CCS) at 24 months.

Table 18: Posterior Means and 95% Credible Intervals for Month 24 CCS

|   | Mean¹ | SD | 95% Bayesian Credible Interval  |
| --- | --- | --- | --- |
|  coflex® | 66.2% | 3.3% | 59.5% to 72.4%  |
|  fusion | 57.7% | 4.8% | 48.1% to 66.9%  |
|  difference | 8.5% | 5.8% | -2.9% to 20.0%  |

¹ Mean, SD, and 95% Bayesian Credible Interval computed as the mean, standard deviation, 2.5th percentile, and 97.5th percentile of 10,000 draws from the posterior distributions

The Bayesian posterior means, standard deviations, and 95% credible intervals were determined from 10,000 draws from the posterior distributions based on the final per protocol population. The credible intervals are defined so that there is a 0.95 probability that the true success likelihoods are contained within the interval. The estimated difference is 8.5%. The lower bound of Bayesian posterior credible interval for the device group difference in success rates is equal to -2.9%, which is larger than the pre-specified non-inferiority margin of -10%.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{27}

The Statistical Analysis Plan specified that primary non-inferiority evaluation would be performed in a per protocol population. All protocol violations (PV) were confirmed by an Independent Clinical Events Committee. Among the 230 randomized patients receiving coflex®, 15 (6.5%) had a protocol violation leading to exclusion. Similarly, among the 114 randomized patients undergoing fusion, 7 (6.1%) had a protocol violation leading to exclusion. The primary efficacy variable was evaluable for all 22 PVs in this study. Among 15 coflex® PVs, 6 (40.0%) met the study success criterion. Similarly, among 7 fusion PVs, 3 (42.9%) met the study success criterion. The clinical results for the PVs were pooled with the per protocol population to construct a modified Intent-to-Treat (mITT) population defined as all randomized patients receiving a study procedure. The Bayesian posterior probability that coflex® is clinically non-inferior to fusion is 0.999, essentially the same as in the primary per protocol population

Table 19: Posterior Probabilities of Success at 24 Months in coflex® Clinical Trial (mITT Cohort)

|   | Number and Percentage Achieving Month 24 CCS |   |   |   |   |   | Posterior Probability of Non-Inferiority  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   | Fusion Control  |   |   |   |
|   | N | n | % | N | n | %  |   |
|  Month 24 | 219 | 141 | 64.4% | 111 | 63 | 56.8% | 0.999  |

Non-inferiority of the coflex® group compared to the control group was demonstrated for the CCS at 24 months in the mITT cohort.

Table 20: Posterior Means and 95% Credible Intervals for Month 24 CCS (mITT Cohort)

|   | Mean^{1} | SD | 95% Bayesian Credible Interval  |
| --- | --- | --- | --- |
|  coflex® | 64.4% | 3.2% | 57.9% to 70.5%  |
|  fusion | 56.8% | 4.7% | 47.4% to 65.7%  |
|  difference | 7.6% | 5.6% | -3.4% to 18.9%  |

1 Mean, SD, and 95% Bayesian Credible Interval computed as the mean, standard deviation, 2.5th percentile, and 97.5th percentile of 10,000 draws from the posterior distributions

For the per protocol population, Table 21 demonstrates the time course of success in the coflex® clinical trial.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{28}

Table 21: Time Course of Composite Clinical Success¹ in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI²  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   |   | Fusion Control  |   |   |   |
|   | N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  Week 6 | 210 | 172 | 81.9% | (76.7%, 87.1%) | 105 | 69 | 65.7% | (56.6%, 74.8%)  |
|  Month 3 | 207 | 171 | 82.6% | (77.4%, 87.8%) | 102 | 72 | 70.6% | (61.7%, 79.4%)  |
|  Month 6 | 207 | 162 | 78.3% | (72.6%, 83.9%) | 105 | 81 | 77.1% | (69.1%, 85.2%)  |
|  Month 12 | 202 | 151 | 74.8% | (68.8%, 80.7%) | 104 | 74 | 71.2% | (62.4%, 79.9%)  |
|  Month 18 | 198 | 135 | 68.2% | (61.7%, 74.7%) | 100 | 68 | 68.0% | (58.9%, 77.1%)  |
|  Month 24 | 204 | 135 | 66.2% | (59.7%, 72.7%) | 104 | 60 | 57.7% | (48.2%, 67.2%)  |
|  Notes:
¹ The composite clinical success criteria at times points prior to Month 24 did not include the 'no persistent new or worsening sensory or motor deficit' since 'persistence' was established by identifying new or worsening deficits at Month 18 that did not resolve by Month 24; otherwise the CCS criteria at earlier time points were consistent with the primary Month 24 CCS.
²The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

Table 21 demonstrates the CCS at each timepoint. The CCS at 24 months is determined by the ODI improvement compared to baseline, absence of secondary surgeries or epidural pain management and neurologic success. It should be noted that neurologic success endpoint is based on comparing changes from baseline to both Month 18 and Month 24, and thus is not definable prior to the 24 month timepoint. ODI measurements and success may fluctuate over time, while discrete events endpoints such as secondary surgeries and epidural injections were assessed as time to event variables.

Patients in the coflex® group demonstrated a 81.9% CCS at 6 weeks which increased to 82.6% at 3 months and gradually fell to 66.2% at 24 months. Patients in the control group demonstrated 65.7% CCS at 6 weeks which rose gradually from 6 Weeks to 6 Months to 77.1%. CCS fell to 57.7% at 24 months. At every assessment time period, the percentage of coflex® patients achieving CCS was greater than fusion, with the largest differences occurring at week 6 and month 3, demonstrating statistical significance at those time points. The final CCS at 24 months demonstrates numerical success that is 8.5% higher in the coflex® group when compared to the fusion control.

PMA P110008: FDA Summary of Safety and Effectiveness Data

{29}

Table 22: Treatment Success at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   | Fusion Control  |   |   |
|   | N | n | % | N | n | %  |
|  Improvement of at least 15 points in ODI at Month 24 compared to baseline | 162 | 139 | 85.8 | 86 | 66 | 76.7  |
|  No reop or epidural (Up to Day 730) | 215 | 173 | 80.5 | 107 | 89 | 83.2  |
|  No reoperations, revisions, removals, or supplemental fixation | 215 | 192 | 89.3 | 107 | 99 | 92.5  |
|  No epidural injection at any lumbar level | 215 | 190 | 88.4 | 107 | 94 | 87.9  |
|  No persistent new or increasing sensory or motor deficit at 24 months | 179 | 169 | 94.4 | 97 | 89 | 91.8  |
|  No persistent new or increasing sensory deficit at 24 mo. | 199 | 191 | 96.0 | 99 | 96 | 97.0  |
|  No persistent new or increasing motor deficit at 24 mo. | 180 | 177 | 98.3 | 97 | 91 | 93.8  |
|  No major device-related complications | 215 | 212 | 98.6 | 107 | 103 | 96.3  |
|  Composite Clinical Success | 204 | 135 | 66.2 | 104 | 60 | 57.7  |

With regard to the functional parameter of the CCS, the coflex® device group demonstrated a greater proportion of patients with a clinically significant improvement in ODI score compared to the fusion control. In the neurological and device related complications components of the primary endpoint, the coflex® group demonstrated similar or higher patient success percentages compared to the fusion control. Success in the reoperations and revisions component of the primary endpoint is higher in the fusion control group than in the coflex® group.

## Sensitivity Analysis:

Table 23: Posterior Probabilities of Success at 24 Months in coflex® Clinical Trial

|   | Number and Percentage Achieving Month 24 CCS |   |   |   |   |   | Posterior Probability of Non-Inferiority  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   | Fusion Control  |   |   |   |
|   | N | n | % | N | n | %  |   |
|  Per Protocol Analysis | 204 | 135 | 66.2% | 104 | 60 | 57.7% | 0.999  |
|  Unresolved Spinous Process Fractures as Failures¹ | 204 | 119 | 58.3% | 104 | 56 | 53.8% | 0.993  |

¹Unresolved Spinous Process fractures counted as failures regardless of clinical significance. 83% of patients in the coflex® group and 75% of patients in fusion group who had spinous process fractures observed by the radiographic laboratory did not have any associated symptoms at the time the fracture was observed.

In sensitivity analyses, the 24 Month Composite Clinical Success endpoint was modified to include as failures patients with an unresolved spinous process fracture at 24 months. Review of the spinous process fractures and the resolution of these fractures were performed by an independent radiographic core laboratory for the purpose of this analysis. With this

PMA P110008: FDA Summary of Safety and Effectiveness Data

{30}

modification in the success definition, the Composite Clinical Success rate decreased from 66% (135 of 204) to 58% (119 of 204) in the coflex® group and from 58% (60 of 104) to 54% (56 of 104) in the fusion group, and the Bayesian posterior probability changed from 0.999 to 0.993, still meeting the a priori defined criterion for success. Therefore, including unresolved spinous process fractures in the failure definition had no appreciable impact on the comparison between the devices.

A tipping point analysis was also performed to determine the effect on the primary endpoint of missing Month 24 data. Results of the tipping point analysis demonstrated that the finding of non-inferiority was insensitive to missing data at Month 24.

## Poolability Analysis:

Analyses were conducted to assess poolability of data across sites and between patients with 1 versus 2 level implants. There was no statistical evidence of site-to-site differences in the comparisons between coflex® and fusion. Similarly, patients receiving 2 level implants had clinical outcomes that were generally comparable to those receiving a 1 level implant.

## Secondary Effectiveness Analysis:

In addition to the components of the primary endpoint presented above, secondary effectiveness variables were also assessed and the results are provided below. The following secondary endpoints were specified:

- ZCQ Symptom Severity
- ZCQ Physical Function
- ZCQ Composite Success
- VAS Leg Pain
- VAS Back Pain
- SF-12

ZCQ Symptom Severity
Table 24: ZCQ Symptom Severity at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI¹  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   |   | Fusion Control  |   |   |   |
|   | N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  ZCQ Symptom Severity Improvement >0.5 points | 161 | 142 | 88.2% | (83.2%, 93.2%) | 86 | 67 | 77.9% | (69.1%, 86.7%)  |
|  ¹The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

PMA P110008: FDA Summary of Safety and Effectiveness Data

{31}

Table 24 shows the subjects achieving success, defined as a decrease in ZCQ Symptom Severity of at least 0.5 points, in the Per Protocol cohort. Month 24 data demonstrates a higher percentage of coflex® patients meeting the success threshold compared to the fusion control (88.2% vs. 77.9%).

## ZCQ Physical Function

Table 25: ZCQ Physical Function at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI¹  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   |   | Fusion Control  |   |   |   |
|   | N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  ZCQ Physical Function Improvement >0.5 points | 161 | 138 | 85.7% | (80.3%, 91.1%) | 86 | 63 | 73.3% | (63.9%, 82.6%)  |
|  ¹The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

Table 25 shows the subjects achieving success, defined as a decrease in ZCQ Physical Function of at least 0.5 points, in the Per Protocol cohort. Month 24 data demonstrates a higher percentage of coflex® patients meeting the success threshold compared to fusion (85.7 vs. 73.3%).

## ZCQ Composite Success

Table 26: ZCQ Composite Success at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI¹  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | coflex® |   |   |   | Fusion Control  |   |   |   |
|   |  N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  ZCQ Composite Success at Month 24 | 161 | 126 | 78.3% | (71.9%, 84.7%) | 86 | 58 | 67.4% | (57.5%, 77.3%)  |
|  ¹The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

Table 26 shows the subjects achieving a Composite ZCQ Success in the Per Protocol cohort, defined as a decrease in ZCQ Physical Function of at least 0.5 points, a decrease in ZCQ Symptom Severity of at least 0.5 points, and ZCQ Satisfaction score &gt;2.5. Month 24 data demonstrates a higher percentage of coflex® patients meeting the success threshold compared to the fusion control (78.3% vs. 67.4%).

PMA P110008: FDA Summary of Safety and Effectiveness Data

{32}

# VAS Leg Pain

Table 27: VAS Leg Pain Success at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI¹  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   |   | Fusion Control  |   |   |   |
|   |  N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  Decrease of at least 20 mm VAS leg Pain (Max) | 162 | 134 | 82.7% | (76.9%, 88.5%) | 85 | 67 | 78.8% | (70.1%, 87.5%)  |
|  ¹The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

Table 27 shows the subjects achieving success, defined as a decrease in VAS Leg Pain of at least 20mm in the Per Protocol cohort. Month 24 data demonstrates a higher percentage of coflex® patients meeting the success threshold compared to the fusion control (82.7% vs. 78.8%).

# VAS Back Pain

Table 28: VAS Back Pain at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI¹  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   |   | Fusion Control  |   |   |   |
|   |  N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  Decrease of at least 20 mm VAS Back Pain | 162 | 143 | 88.3% | (83.3%, 93.2%) | 85 | 68 | 80.0% | (71.5%, 88.5%)  |
|  ¹The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage. Non-overlapping confidence intervals imply statistically reliable device group differences.  |   |   |   |   |   |   |   |   |

Table 28 shows the subjects achieving success, defined as a decrease in VAS Back Pain of at least 20mm, in the Per Protocol cohort. Month 24 data demonstrates a higher percentage of coflex® patients meeting the success threshold compared to the fusion control (88.3% vs. 80.0%).

PMA P110008: FDA Summary of Safety and Effectiveness Data

{33}

SF-12

Table 29: SF-12 Success at 24 Month Follow-Up in coflex® Clinical Trial

|   | Number and Percentage Meeting Criteria with 95% CI¹  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  coflex® |   |   |   | Fusion Control  |   |   |   |
|   |  N | n | % | 95% CI (LB, UB) | N | n | % | 95% CI (LB, UB)  |
|  Maintenance or improvement in SF-12 MCS | 132 | 92 | 69.7% | (61.9%, 77.5%) | 70 | 48 | 68.6% | (57.7%, 79.4%)  |
|  Maintenance or improvement in SF-12 PCS | 132 | 121 | 91.7% | (87.0%, 96.4%) | 70 | 58 | 82.9% | (74.0%, 91.7%)  |
|  ¹The 95% confidence interval is provided as a measure of the statistical precision of the estimated treatment group mean or percentage  |   |   |   |   |   |   |   |   |

Table 29 shows the percentages of subjects meeting success, defined as maintaining or improving in the SF-12 Physical Function and Mental Health components of the per protocol cohort. The percentage of patients meeting SF-12 Physical Function success criterion is higher for coflex® at month 24 compared to the fusion control (91.7% vs. 82.9%).

## Radiographic Assessments

Maintenance or improvement of foraminal height was a radiographic endpoint in the study. This is a measure of the mechanism of action of the coflex® device which is to maintain foraminal height. coflex® was able to improve or maintain foraminal height in 100% of patients measured at 24 months. This measurement was taken only on the coflex® patients.

Range of motion at the index level was measured at 24 m…

---

**Source:** [https://fda.innolitics.com/device/P110008](https://fda.innolitics.com/device/P110008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
