The Rayner C-flex™ intraocular lens is indicated for primary implantation for the visual correction of aphakia in adults in whom a cataractous lens has been removed by phacoemulsification. The lens is intended to be placed in the capsular bag.
Device Story
C-flex™ (model 570C) is a single-piece, UV-absorbing posterior chamber intraocular lens (IOL) designed for surgical implantation into the capsular bag following phacoemulsification. The lens is constructed from Rayacryl™, a hydrophilic/hydrophobic methacrylate copolymer, which reduces silicone oil adhesion. It features an overall diameter of 12.0 mm and an optic diameter of 5.0 to 5.75 mm. The device is implanted by an ophthalmologist during cataract surgery. By replacing the natural crystalline lens, the IOL restores visual acuity in aphakic patients. Clinical outcomes demonstrate high patient satisfaction and visual acuity results exceeding historical controls. The device is provided sterile for single use.
Clinical Evidence
Multi-center clinical trial with 301 C-flex™ patients and 182 Centerflex patients (pooled N=483). Primary endpoints: Best Corrected Visual Acuity (BCVA) at 12 months and cumulative/persistent adverse event rates compared to FDA historical controls. Results: 99.5% of pooled patients achieved BCVA ≥20/40 (vs 96.7% historical control). Adverse event rates for cumulative macular edema (2.3%), persistent macular edema (1.1%), and surgical reintervention (0.6%) were not statistically significantly different from historical controls.
Technological Characteristics
Material: Rayacryl™ (copolymer of hydrophilic and hydrophobic methacrylates). Refractive index: 1.46. Dimensions: 12.0 mm overall diameter, 5.0-5.75 mm optic diameter. Power range: +8.0 to +30.0 D (0.5 D steps). Sterilization: Moist heat. Standards: ISO 11979 (Ophthalmic implants – IOLs), ISO 10993 (Biocompatibility).
Indications for Use
Indicated for primary implantation for visual correction of aphakia in adults (≥21 years) following phacoemulsification cataract removal. Contraindicated in patients with microphthalmia, corneal decomposition/endothelial insufficiency, active ocular disease (e.g., chronic severe uveitis, proliferative diabetic retinopathy, chronic glaucoma), or those who are pregnant or nursing.
Regulatory Classification
Identification
An intraocular lens is a device made of materials such as glass or plastic intended to be implanted to replace the natural lens of an eye.
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
Device Generic Name: Ultraviolet-absorbing posterior chamber intraocular lens (IOL)
Device Trade Name: C-flex™ intraocular lens (model 570C)
Applicant's Name and Address: Rayner Surgical Inc.
6654 Church Street
Los Angeles, CA 90042
USA
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P060011
Date of Notice of Approval to Applicant: May 3, 2007
## II. INDICATIONS FOR USE
The Rayner C-flex™ intraocular lens is indicated for primary implantation for the visual correction of aphakia in adults in whom a cataractous lens has been removed by phacoemulsification. The lens is intended to be placed in the capsular bag.
## III. CONTRAINDICATIONS
The C-flex™ IOL is contraindicated in patients:
- Who are less than 21 years
- Who are microphthalmic
- Who have corneal decomposition or corneal endothelial cell insufficiency
- Who have active ocular disease (e.g., chronic severe uveitis, proliferative diabetic retinopathy, chronic glaucoma not responsive to medication)
- Who are pregnant or nursing
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the C-flex™ labeling (Attachment 1).
## V. DEVICE DESCRIPTION
The single-piece C-flex™ ultraviolet-absorbing posterior chamber intraocular lens (IOL) is designed to be surgically implanted into the capsular bag of the human eye as a replacement for the crystalline lens following phacoemulsification with an anterior continuous curvilinear capsulorhexis just covering 360° the anterior edge of the IOL optic by 0.5 to 1.0 mm. The overall diameter is 12.0 mm, the optic diameter 5.0 to 5.75 mm, with a refractive index of 1.46. The C-flex™ is made from Rayacryl™ which is a
{1}
copolymer of hydrophilic and hydrophobic methacrylates. The hydrophilic nature of the material reduces silicone oil adhesion. C-flex™ IOLs are available from +8.0 to +30.0 Diopters (D) with 0.5 D steps.
## VI. ALTERNATIVE PRACTICES OR PROCEDURES
Other approved IOLs may be used in the treatment for visual correction.
## VII. MARKETING HISTORY
The C-flex™ IOL has been marketed in over 40 countries worldwide to date. The C-flex™ IOL has not been withdrawn from any market for reasons relating to safety and effectiveness of the device.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
A multi-center clinical trial with a historical control was conducted to determine the safety and effectiveness of the C-flex™ IOL. Three hundred one (301) patients were enrolled and implanted with the C-flex™ IOL. Data from an additional 182 patients implanted with the Centerflex IOL (identical to the C-flex™ except for minor differences in edge design that were determined not to affect the evaluation of safety and effectiveness) were used as supporting clinical data. Adverse events reported in the study are presented in Table 1. The adverse event rates were not statistically significantly different from the historical control population rate for all of the listed cumulative and persistent adverse events. The major adverse events experienced during the clinical trial were cumulative macular edema (2.3%), persistent macular edema (1.1%), and cumulative surgical reintervention (0.6%).
TABLE 1: Specific Cumulative and Persistent Adverse Events in the Investigational Model and FDA Historical Control at 11-14 Months
| | Posterior Chamber Historical Control^{A} | | Centerflex IOL | | C-flex™ IOL | | Pooled Centerflex and C-flex™ Data | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | % | n/N | % | n/N | % | n/N | % |
| Cumulative^{B} Hyphema | 91 | 2.2 | 0/182 | 0.0 | 0/301 | 0.0 | 0/483 | 0.0 |
| Cumulative Macular Edema | 124 | 3.0 | 5/182 | 2.7 | 6/301 | 2.0 | 11/483 | 2.3 |
| Cumulative Retinal Detachment | 11 | 0.3 | 0/182 | 0.0 | 0/301 | 0.0 | 0/483 | 0.0 |
| Cumulative Pupillary Block | 5 | 0.1 | 0/182 | 0.0 | 1/301 | 0.3 | 1/483 | 0.2 |
| Cumulative Lens Dislocation | 5 | 0.1 | 0/182 | 0.0 | 0/301 | 0.0 | 0/483 | 0.0 |
| Cumulative Endophthalmitis | 4 | 0.1 | 0/182 | 0.0 | 0/301 | 0.0 | 0/483 | 0.0 |
| Cumulative Hypopyon | 16 | 0.3 | 0/182 | 0.0 | 0/301 | 0.0 | 0/483 | 0.0 |
| Cumulative Surgical Reintervention | 46 | 0.8 | 0/182 | 0.0 | 3/301 | 1.0 | 3/483 | 0.6 |
| Persistent^{C} Macular Edema | 19 | 0.5 | 3/166 | 1.8 | 2/284 | 0.7 | 5/450 | 1.1 |
| Persistent Corneal Edema | 11 | 0.3 | 1/166 | 0.6 | 1/284 | 0.4 | 2/450 | 0.4 |
| Persistent Iritis | 11 | 0.3 | 0/166 | 0.0 | 0/284 | 0.0 | 0/450 | 0.0 |
| Persistent Raised Intraocular pressure (IOP) Requiring Treatment | 17 | 0.4 | 0/166 | 0.0 | 0/284 | 0.0 | 0/450 | 0.0 |
P060011 SSED
Page 2 of 9
{2}
Notes:
a. Annex B – FDA grid of historical controls: Best Corrected Visual Acuity (BCVA) at one year, FDA Intraocular Lens Guidelines, 1999.
b. Cumulative Adverse Event (AE) for the investigational model intraocular lens (IOL) is computed as any occurrence up to and including the current interval. Historical control cumulative values are up to and including 1 year.
c. Persistent Adverse Event (AE) for the investigational model is defined as an AE remaining unresolved at the start of the current evaluation interval.
Other potential complications of cataract or implant surgery include, but are not limited to the following: Endophthalmitis, retinal detachment, cyclitic membrane, iris prolapse, hypopyon, corneal edema, corneal endothelial damage, uveitis, hyphema, lens epithelial cell on-growth, secondary glaucoma and precipitates on the lens surface. Secondary surgical interventions may be required for, but is not limited to the following: Vitreous aspiration or iridectomy for papillary block, wound leak repair, retinal detachment repair, lens repositioning, and lens replacement due to refractive error or severe inflammation.
IX. SUMMARY OF PRECLINICAL STUDIES
Rayner performed preclinical studies on the device in accordance with the International Organization for Standardization (ISO) document 10993 Ophthalmic implants – intraocular lens.
Biocompatibility Studies:
The objective of the biocompatibility studies was to establish a complete biocompatibility profile for the IOL material. Summaries of the biocompatibility tests conducted are listed below (Table 2):
TABLE 2: Biocompatibility Tests
| Biocompatibility Test | Results and Conclusions |
| --- | --- |
| Cytotoxicity: Cell Inhibition | The results showed the lenses were mildly inhibitory |
| Cytotoxicity: Direct Contact Agarose Overlay | No toxic effects occurred in the treated cells. |
| Cytotoxicity: Indirect Contact Agarose Overlay | No toxic effects occurred in the treated cells. |
| Cytotoxicity: Indirect Contact | No toxic effects occurred in the extract treated cells |
| Cytotoxicity: Modified Eagle’s Media Elution | No cytotoxic reaction was observed on the treated cells |
| Genotoxicity: Bacterial Reverse Mutation | The lens extract did not induce mutagenic activity |
| Genotoxicity: Mammalian Chromosomal Aberration | There was no evidence that the extracts induced structural or numerical chromosomal alternation in the Chinese Hamster Ovary cells |
| Genotoxicity: Mammalian Chromosomal Aberration | There was no evidence that the extracts induced structural or numerical chromosomal alternation the mouse lymphoma cells |
| Acute Systemic Toxicity | No mortality, no weight loss, and evidence of significant systemic toxicity were observed |
| Maximization Sensitization | The extractants showed no evidence of causing delayed dermal contact sensitization in the guinea pigs |
| Ocular Irritation and Sensitization | There was no significant irritation in the test eye of any rabbit |
Physiochemical and Engineering
Rayner conducted non-clinical studies on the C-flex™ IOL in accordance with the international standard series ISO 11979 (Ophthalmic implants – Intraocular lenses). Additionally, an on-site facility inspection was used to establish the adequacy of the manufacturing process. Non-clinical testing demonstrated the C-flex™ lens’ safety and effectiveness from an engineering and manufacturing perspectives. Summaries of the physiochemical and engineering tests conducted are listed below:
P060011 SSED
Page 3 of 9
{3}
TABLE 3: Physiochemical and Engineering Tests
| Physicochemical & Engineering | Test and Conclusions |
| --- | --- |
| Extractables and hydrolytic stability (Aged sample) | The aged IOL exhibited no significant residuals, weight change, or UV spectra change. There was no significant change in the formation of bubbles, breaks, dendrites or fissures in the aged IOL. |
| Extractables | The total level of residual monomer extracted was less than 0.04%. |
| Hydrolytic Stability | Lenses were extracted in water for real-time equivalent of 9 months. No significant changes to the lenses were detected. These are acceptable hydrolytic stability results. |
| Photostability | The spectral transmittance through the lenses in the visible regions was slightly decreased after the 20-year equivalent UV exposure (as per ISO 11979-5) but the change was not significant. No significant degradation products were found in solution. |
| Nd:YAG Laser Damage | The lenses were not cytotoxic after being treated by the Nd:YAG laser. |
| Lens Discoloration | No evidence of discoloration was observed after 24 hour exposure to in vivo conditions. |
| Surface Contamination | Residuals from manufacturing processes were not detected above the detection limits of the methods used. |
| Optical | The MTF (modulation transfer function) values at 100 lp/mm were greater or equal to 70% of that calculated as the maximum attainable for the system of model eye with the IOL. |
| Visual Inspection | Following simulated injection using injector packaged with lens, lenses were inspected at 10x magnification. Lenses injected with the viscoelastic passed; lenses injected with saline failed because of broken haptics and torn optics. |
| Power and MTF Measurement | Power and resolution were within tolerance after injection when the viscoelastic was used and failed when only saline was used as the lubricant. |
| Optic Tilt Measurement | All lenses were less than 0.4 degrees after injection. |
| Optic Decentration Measurement | All were less than 0.3 mm decentered after injection. |
| Dynamic Fatigue Durability Measurement | All lenses passed after injection. |
| Limb Pull Strength Measurement | All were greater than 0.9 N after injection |
| Compression force measurement | Compression force averaged about 1.6 mN and the lenses were not significantly changed by the injection process. |
| Axial Displacement Measurement | Axial displacement averaged less than 0.1 mm and the lenses were not significantly changed by the injection process. |
| Shipping Study | All lenses were undamaged after simulated shipping. |
Sterilization, Packaging, Shelf-life and Transport Stability:
The objective of the sterilization, shelf-life and transport stability studies was to establish a complete microbiological profile for the finished IOL. The IOL is packaged in a blister tray with a foil lid. The blister tray is placed in a paper pouch and the package is then terminally sterilized by moist heat. Rayner conducted sterilization revalidation, package integrity, shelf life and transport stability studies on the C-flex™ IOL in accordance with the following standards and pharmacopoeial chapters: ISO 11134 Sterilization of health care products – Requirements for validation and routine control-Industrial moist heat sterilization; ISO 11737-1:1995 Sterilization of medical devices-Microbiological methods – Part 1: Estimation of population of microorganisms on products; BS EN 556 Sterilization of medical-devices-Requirements for medical devices to be designated ‘STERILE’; United States Pharmacopoeia (USP) 29 <71> and 26 <85>; British Pharmacopoeia (BP) 2005 A and A308; the European Pharmacopoeia (EP) Chapter 5.0, Sections 2.6.1 and 2.6.14; ISO 8362-2 Annex C Injection containers for injectables and accessories – Part 2: Closures for injection vials; ASTM F 1929-8
P060011 SSED
Page 4 of 9
{4}
Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration; and ASTM F 1140-96 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages for Medical Applications and ISO 11979-6 Ophthalmic implants-Intraocular lenses – Part 6: Shelf-life and transport stability. The following tests were conducted to establish the microbiological profile (Table 4):
TABLE 4: Sterilization, Packaging, Shelf-Life & Transport Tests
| Sterilization, Packaging, Shelf-Life & Transport | Test Conclusions |
| --- | --- |
| Presterilization Bioburden Test | The bioburden percent recovery was within acceptable limits. |
| Poststerilization Sterility Test | No microbial growth was detected. |
| Bacterial Endotoxin Testing | Endotoxin levels were below the agency’s recommended limit for medical devices. |
| Bacteriostasis and Fungistasis Test | No antimicrobial inhibition demonstrated. |
| Seal/Closure Integrity Test | No visible penetration of dye after application of test; thus no leaks were detected in the package or seal. |
| Dye Penetration Test | No visible penetration of dye into the packaging. |
| Burst Test | Burst test results were within acceptable limits. |
| Peel Strength Test | Peel strength test results were within acceptable limits. |
| Aging Studies | Results were satisfactory to support a 12 month shelf life |
The results of these tests were deemed acceptable from a microbiology standpoint. Adequate data was provided to justify a 12 month shelf-life.
## Preclinical Results Conclusions:
The overall results of these tests were acceptable from a microbiology, biocompatibility, physiochemical and engineering standpoint.
## X. SUMMARY OF CLINICAL STUDIES
### Objectives:
The objective of the clinical studies was to assess the safety and effectiveness of the C-flex intraocular lens in the visual correction of aphakia in adults in whom a cataractous lens has been removed.
### Study Design:
Primary efficacy analyses are based on Best Case Visual Acuity (BCVA) at one year (11-14 months) post-implantation as determined in the sample of procedures with no pre-existing macular degeneration or with macular degeneration developing at any time during the study, or with a clinically significant violation of the exclusion or inclusion criteria.
Safety was evaluated with regards to specific cumulative adverse event rates and persistent adverse events rates as specified in the FDA Intraocular Lens Guidelines, 1996 and ISO 11979-7 (Ophthalmic implants – Intraocular lenses – Part 7: Clinical investigations). Primary safety analyses are based on data from all enrolled patients with follow-up to at least one-year post implantation. The FDA historical control is derived from weighted averages of the data from 13 large clinical investigations of IOLs (anterior and posterior chamber) between March 1988 and June 1991. The pooled sample size for these clinical investigations was 5162 adverse events.
P060011 SSED
Page 5 of 9
{5}
Study Population:
Three hundred and one (301) C-flex™ patients were enrolled (unilateral implants) in the investigation. Additional data from 182 patients implanted with the Centerflex IOL was included in the analysis. The Centerflex and C-flex™ IOLs are identical except for minor differences in edge design that were determined not to affect the evaluation of safety and effectiveness.
Analysis of the patient demographic reveal an approximately equal number of left (47.4%; 229/483) and right eyes (52.6%; 254/483), a preponderance towards the female gender (63.8%; 308/483), although there was no difference in the safety and effectiveness of the lens based on gender. Corneal status was generally normal and the pre-operative pathology was a low percentage of the total population enrolled. Cataract etiology was 100% senile. The mean age of the patients was 72.8 years. Ethinicity was 99.4% (480/483) Caucasians, 0.4% (2/483) Hispanic and 0.2% (1/483) Asian.
Patient Assessments:
The following table summarizes the study patient’s visit schedule and tests performed at each visit window (see Table 5).
TABLE 5: Schedule of Visit Examinations for the C-flex™ Study
| Visit Examination | Form 0 | Form 1 | Form 2 | Form 3 | Form 4 | Form 5 | Form 6 | Form 7 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Pre-Op/Op | 1-2 days | 1-2 weeks | 1-2 months | 4-6 months | 11-14 months | 2 years | 3 years |
| Inclusion Criteria | X | | | | | | | |
| Exclusion Criteria | X | | | | | | | |
| Informed Consent | X | | | | | | | |
| Biometry | X | | | | | | | |
| Corneal Status | X | | | | | | | |
| Targeted Post Operative Refraction | X | | | | | | | |
| Cataract etiology | X | | | | | | | |
| Patient Demographics | X | X | X | X | X | X | X | X |
| IOP op eye/fellow eye | X | X | X | X | X | X | X | X |
| Medication used | X | X | X | X | X | X | X | X |
| UCVA op. eye/fellow eye | X | N/R | X | X | X | X | X | X |
| BCVA op eye/fellow eye | X | N/R | X | X | X | X | X | X |
| Keratometry | X | N/R | X | X | X | X | X | X |
| Refraction op eye/fellow eye | | N/R | X | X | X | X | X | X |
| Complete Ophthalmic Examination (under mydriasis) | X | N/R | X | X | X | X | X | X |
| Dilated Fundus Exam | X | N/R | X | X | X | X | X | X |
| Pathology/complication/surgery | X | X | X | X | X | X | X | X |
| Macular Degeneration | X | N/R | X | X | X | X | X | X |
| Dilated IOL evaluation | | N/R | X | X | X | X | X | X |
| Fibrosis/Elschnig’s Pearls | | N/R | X | X | X | X | X | X |
| Anterior capsular opacification | | N/R | X | X | X | X | X | X |
| Opacities on/within IOL | | N/R | X | X | X | X | X | X |
| IOL discoloration/tilt/decentration | | N/R | X | X | X | X | X | X |
| Glare Questionnaire | | N/R | N/R | X | X | X | X | X |
N/R – not recorded
P060011 SSED
Page 6 of 9
{6}
# Data Analysis and Results:
Visual Acuity – The C-flex™/Centerflex met or exceeded historical controls for posterior chamber IOLs in all areas for best corrected visual acuity (BCVA) at the 12-month post-operative examination. Best case visual acuity and overall visual acuity greater than 20/40 was 98.2% and 99.5% (see Table 6 and 7). The historical control from the “FDA grid” was 92.5% and 96.7%.
TABLE 6: The Best Case Visual Acuity (% with at least 20/40) of the C-flex™-Centerflex IOL and the FDA Historical Control at 12-month post-operative
| Years of Age | Posterior Chamber Historical Control^{B} | | Centerflex IOL | | C-flex™ IOL | | Pooled Centerflex and C-flex™ Data | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n/N | % | n/N | % | n/N | % | n/N | % |
| <60 | 203/206 | 98.5 | 6/6 | 100 | 24/24 | 100 | 30/30 | 100 |
| 60-69 | 793/822 | 96.5 | 41/41 | 100 | 61/61 | 100 | 102/102 | 100 |
| 70-79 | 1338/1372 | 97.5 | 71/72 | 98.6 | 116/117 | 99.1 | 187/189 | 98.9 |
| ≥80 | 601/634 | 94.8 | 17/17 | 100 | 36/36 | 100 | 53/53 | 100 |
| Overall | 2935/3034 | 96.7 | 135/136 | 99.4 | 237/238 | 99.6 | 372/374 | 99.5 |
Notes:
A: Best Case Visual Acuity is summarized for the Primary Efficacy Sample that excludes patients with preoperative ocular pathologies and those with macular degeneration developing at any time during the study.
B: Annex B – FDA grid of historical controls: BCVA at one year, FDA Intraocular Lens Guidelines, 1999.
TABLE 7: The Overall Visual Acuity (% with at least 20/40) of the C-flex™-Centerflex IOL and the FDA Historical Control at 11-14 months post-operative
| Years of Age | Posterior Chamber Historical Control^{B} | | Centerflex IOL | | C-flex™ IOL | | Pooled Centerflex and C-flex™ Data | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n/N | % | n/N | % | n/N | % | n/N | % |
| <60 | 230/235 | 95.7 | 6/6 | 100 | 24/24 | 100 | 30/30 | 100 |
| 60-69 | 968/1012 | 93.4 | 42/42 | 100 | 66/66 | 100 | 108/108 | 100 |
| 70-79 | 1793/1920 | 86.5 | 83/84 | 98.8 | 139/140 | 99.3 | 222/224 | 99.1 |
| ≥80 | 901/1042 | 92.5 | 30/34 | 88.2 | 51/53 | 96.2 | 81/87 | 93.1 |
| Overall | 3893/4210 | 92.5 | 161/166 | 97.0 | 280/283 | 98.9 | 441/449 | 98.2 |
Notes:
A: Overall visual acuity is summarized for All Enrolled Procedures sample that only excludes second implants for any patient implanted bilaterally.
B: Annex B – FDA grid of historical controls: BCVA at one year, FDA Intraocular Lens Guidelines, 1999.
Patient Satisfaction Survey – A modification of the patient satisfaction questionnaire used by Tester, Pace, Samore and Olson (2000) to assess patient reports of dysphotopsia and patient satisfaction with the C-flex™ IOL was added to the clinical follow-up assessments for patients implanted with the C-flex™. The Centerflex study protocol did not include this assessment. For this questionnaire, patients were asked to rate the severity of symptoms present in their operative and fellow eye. Analyses revealed substantial differences in results between patients in whom their fellow eye had a prior implant. These prior implants were not the C-flex™ lens and were present prior to the patient’s enrollment into this study.
P060011 SSED
Page 7 of 9
{7}
Of the 301 C-flex™ implanted patients, 104 patients (34.6%) were previously implanted in their fellow eye and 197 patients (64.5%) did not have a previous implant in their fellow eye. To determine if a comparison between the C-flex™ operative eye and fellow eye depended on the fellow eye’s implant status, three analyses were conducted:
1. All patients regardless of fellow eye implant status
2. Patients with an implant in the fellow eye
3. Patients with no implant in the fellow eye
Satisfaction with corrected vision was larger for the C-flex™ treated eye compared to the fellow eye, a finding driven mostly by the patient subset with no fellow eye implant. Overall patient satisfaction was 89% (268/301).
## XI. CONCLUSIONS DRAWN FROM THE STUDIES
### RISK BENEFIT ANALYSIS
The Rayner C-flex™ intraocular lens is indicated for primary implantation for the visual correction of aphakia in adults in whom a cataractous lens has been removed by phacoemulsification. The lens is intended to be placed in the capsular bag. The risks associated with eye surgery and this lens included: retinal detachment, endophthalmitis, raised intraocular pressure, uvetis and corneal decompensation. It is reasonable to conclude that the benefits of use of the lens for the target population outweigh the risk of illness or injury when used as indicated in accordance with the directions for use.
### SAFETY
The pooled C-flex™ and Centerflex accumulative and persistent adverse event rates at 12-months are lower than the FDA historical control grid in all areas except persistent macular edema, persistent corneal edema and cumulative papillary block. These rates were higher than the grid rate but the differences were not statistically significant.
### EFFECTIVENESS
The C-flex™ met or exceeded the effectiveness criteria in all areas for visual acuity at the 12-month post-operative examination.
## XII. PANEL RECOMMENDATION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIII. CDRH DECISION
CDRH issued an approval order to Rayner Surgical on May 3, 2007. The applicant's manufacturing facility was inspected and was found to be in compliance with the Quality System Regulation (21 CFR 820).
P060011 SSED
Page 8 of 9
{8}
XIV. **APPROVAL SPECIFICATIONS**
Directions for use: See the labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.
XV. **REFERENCES**
Tester R, Pace NL, Samore M, Olson RJ (2000) Dysphotopsia in phakic and pseudophakic patients: Incidence and relation to intraocular lens type. Journal of Cataract Refract Surgery 26:810-816.
P060011 SSED
Page 9 of 9
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.