Reagent Pack For the in vitro qualitative detection of IgM antibody to hepatitis B core antigen (anti-HBc IgM) in human adult and pediatric serum and plasma (heparin, EDTA and citrate) and neonate serum using the VITROS ECi Immunodiagnostic System. Assay results, in conjunction with other serological and clinical information, may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B. Calibrator For use in the calibration of the Vitros ECi Immunodiagnostic System when used for the in vitro qualitative detection of IgM antibody to hepatitis B virus core antigen (anti-HBc IgM) in human adult and pediatric serum and plasma (EDTA, heparin and citrate) and neonate serum using Vitros Anti-HBc IgM Reagent Packs.
Device Story
Device is an in vitro diagnostic assay for anti-HBc IgM detection using the Vitros ECi Immunodiagnostic System. Input: human serum/plasma samples. Principle: antibody class capture technique; sample diluted and reacted with biotinylated mouse monoclonal anti-human IgM antibody; immune complex captured by streptavidin-coated wells. Conjugate (HRP-labeled mouse monoclonal anti-HBc antibody complexed with recombinant HBc antigen) binds to captured IgM. Bound HRP measured via enhanced chemiluminescence; light signal proportional to anti-HBc IgM concentration. System is fully automated, random access, refrigerated storage, touch-screen interface. Used in clinical laboratories by technicians. Output: qualitative reactive/non-reactive result. Results assist physicians in differential diagnosis of hepatitis B, distinguishing acute/chronic infection, and guiding therapy decisions.
Clinical Evidence
Prospective multi-center study (N=1691 US, N=315 India). Compared Vitros Anti-HBc IgM to reference serological markers. Population I (US): 63.16% positive agreement, 99.76% negative agreement. Population II (India): 78.65% positive agreement, 100% negative agreement. Evaluated 6 seroconversion panels, cross-reactivity (HAV, HCV, etc.), and interfering substances (hemoglobin, bilirubin, triolein).
Indicated for in vitro qualitative detection of anti-HBc IgM in human adult, pediatric, and neonate serum and plasma (heparin, EDTA, citrate) to aid laboratory diagnosis of acute or chronic hepatitis B.
Regulatory Classification
Identification
A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected.
(B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include:
(A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following:
(i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration.
(ii) Documentation and characterization (
*e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate.
(ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses.
(x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas.
(3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply:
(i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate.
(ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.
Reference Devices
Abbott Anti-HBc IgM assay
Related Devices
P110031 — ELECSYS ANTI-HBC IGM IMMUNOASSAY AND ELECSYS PRECICONTROL ANTI-HBC IGM · Roche Diagnostics Corp. · Jan 3, 2012
P110022 — ELECSYS ANTI-HBC IGM IMMUNOASSAY AND ELECSYS PRECICONTROL ANTI-HBC IGM · Roche Diagnostics Corp. · Oct 26, 2011
P110025 — ELECSYS ANTI-HBC IGM IMMUNOASSAY & ELECSYS PREICONTROL ANTI-HBC IGM FOR USE ON THE MODULAR ANAYTICS E170 IMMUNOASSAY ANA · Roche Diagnostics Corp. · Dec 14, 2011
P990044 — DIASORIN ETI-CORE-IGMK PLUS ASSAY · DiaSorin, Inc. · Mar 30, 2001
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
I. General Information
Device Generic Name: IgM antibody to Hepatitis B virus core antigen assay (Anti-HBc IgM assay) and IgM antibody to Hepatitis B virus core antigen assay calibrator (Anti-HBc IgM assay calibrator)
Device Trade Name: Vitros Immunodiagnostic Products Anti-HBc IgM Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc IgM Calibrator
Name and Address of Applicant: Ortho-Clinical Diagnostics, Inc, 100 Indigo Creek Drive, Rochester, NY 14626-5101
Date of Panel Recommendation: None
Premarket Approval Application Number: P030026
Date of Notice of Approval to Applicant: March 4, 2004
II. Indications for Use
Reagent Pack
For the in vitro qualitative detection of IgM antibody to hepatitis B core antigen (anti-HBc IgM) in human adult and pediatric serum and plasma (heparin, EDTA and citrate) and neonate serum using the VITROS ECi Immunodiagnostic System. Assay results, in conjunction with other serological and clinical information, may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B.
Calibrator
For use in the calibration of the Vitros ECi Immunodiagnostic System when used for the in vitro qualitative detection of IgM antibody to hepatitis B virus core antigen (anti-HBc IgM) in human adult and pediatric serum and plasma (EDTA, heparin and citrate) and neonate serum using Vitros Anti-HBc IgM Reagent Packs.
III. Contraindications: None known
IV. Warnings and Precautions
The warnings and precautions can be found in the Vitros Anti-IgM Assay Reagent Pack and Vitros Anti-IgM Assay Calibrator labeling.
The Vitros Anti-HBc IgM Calibrator has been validated for use only on the Vitros System with Vitros Immunodiagnostic Products anti-HBc IgM Reagent Packs. Refer to the Vitros Anti-HBc IgM Reagent Pack instructions for use for further details.
{1}
Page 2 of 21 – P030026 Summary of Safety and Effectiveness Data
## V. Device Description
Kit Configuration and Components
For detection of anti-HBc IgM, the Vitros System is comprised of the following:
- Vitros Immunodiagnostic Products Anti-HBc IgM Reagent Pack (Vitros Anti-HBc IgM Reagent Pack) and Vitros Immunodiagnostic Products Anti-HBc IgM Calibrator (Vitros Anti-HBc IgM Calibrator) together comprise the Vitros Anti-HBc IgM assay.
The Vitros Anti-HBc IgM Reagent Pack is composed of three components:
- **Biotinylated Antibody Reagent**
Biotin-mouse monoclonal anti-human IgM antibody in buffer with bovine serum albumin and anti-microbial agent (Kathon).
- **Conjugate Reagent**
Horse Radish Peroxidase - mouse monoclonal anti-HBc antibody with recombinant HBc antigen derived from *E. coli* in buffer with bovine serum albumin and anti-microbial agent (Kathon).
- Streptavidin coated microwells (52 wells per pack).
The Vitros Anti-HBc IgM Calibrator contains:
- Anti-HBc IgM positive human defibrinated plasma (inactivated) with anti-microbial agent (Kathon). The calibrator is supplied ready for use.
In addition, the following components are required:
- Vitros ECi Immunodiagnostic System (Vitros Analyzer) – is dedicated random access instrumentation, which provides automated analysis of the Vitros assays.
- Vitros Immunodiagnostic Products Signal Reagent and Vitros Immunodiagnostic Products Universal Wash Reagent are Common Reagents used in all Vitros System assays.
- Vitros Immunodiagnostic Products High Sample Diluent B – a buffer with bovine serum albumin and antimicrobial agent (Bronidox-L). High Sample Diluent B is a Common Reagent used with select Vitros System assays.
## Principle of Device Methodology
The Vitros ECi Immunodiagnostic System (Vitros Analyzer) allows for the determination of analytes in human samples (for example, serum and plasma). All assays on the Vitros Analyzer employ an enhanced chemiluminescence detection reaction. The analyzer is fully
2
{2}
Page 3 of 21 – P030026 Summary of Safety and Effectiveness Data
automated with a refrigerated on board assay storage system. All standard bar code symbologies are supported by the analyzer, which has a throughput of up to 90 assays per hour. The analyzer also provides menu driven software, which can be accessed from a high-resolution touch screen monitor.
The Vitros Anti-HBc IgM assay is performed using the Vitros Anti-HBc IgM Reagent Pack and the Vitros Anti-HBc IgM Calibrator on the Vitros Immunodiagnostic System. The assay uses an antibody class capture technique which involves dilution of the sample and the simultaneous reaction of IgM in the diluted sample with biotinylated mouse monoclonal anti-human IgM antibody. The immune complex is captured by streptavidin on the wells; unbound materials are removed by washing. Horseradish peroxidase (HRP)-labeled mouse monoclonal anti-HBc antibody which has been complexed with recombinant HBc antigen (conjugate) is then captured by anti-HBc specific IgM bound to the wells. Unbound material is removed by washing.
The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron-transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The Vitros Analyzer reads the light signals. The amount of HRP conjugate bound is directly proportional to the concentration of anti-HBc IgM present in the sample.
The Vitros Anti-HBc IgM Calibrator allows calibration of the Vitros Anti-HBc IgM assay setting a boundary for interpretation of positive and negative results. Calibration is lot specific. A Master Calibration is established for each new reagent lot by performing multiple assays. The lot-specific parameter, the expected calibrator signal and the data that enables the Vitros System to calculate the cutoff value are encoded on the calibration card. Scanning the lot calibration card loads the encoded data into the Vitros System. When the calibrator is processed, the validity of the calibration is assessed against a quality parameter that compares the actual signal of the calibrator with the expected signal. If the calibration is acceptable the cutoff value is calculated and stored for use with any reagent pack of that lot. Recalibration is required after a predetermined calibration interval, or when a different reagent pack lot is loaded.
## VI. Alternate Practices and Procedures
Determining the presence of anti-HBc IgM in patients may be achieved by using a variety of commercially available, FDA licensed or approved serological tests. Additionally, when test results are used in conjunction with a physician’s assessment and other laboratory test results, infection with HBV can be identified.
{3}
Page 4 of 21 – P030026 Summary of Safety and Effectiveness Data
## VII. Marketing History
The following table lists the countries where the Vitros Immunodiagnostic Products Anti-HBc IgM Assay may be marketed as of March 2003.
| Argentina | Australia |
| --- | --- |
| Austria | Belgium |
| Brazil | Canada |
| Chile | Colombia |
| Czech Republic | Denmark |
| France | Germany |
| Greece | Holland |
| India | Indonesia |
| Italy | Japan |
| Korea | Luxembourg |
| Malaysia | Norway |
| Panama | Philippines |
| Poland | Romania |
| Portugal | Russia |
| Saudi Arabia | Singapore |
| Slovak Republic | Slovenia |
| Spain | Switzerland |
| Taiwan | Thailand |
| Turkey | United Kingdom |
| Venezuela | |
The Vitros Immunodiagnostic Products Anti-HBc Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc Calibrator have not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device.
## VIII. Potential Adverse Effects of the Device on Health
Since the Vitros Immunodiagnostic Products Anti-HBc IgM Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc IgM Calibrator are for in vitro diagnostic use, there is no direct adverse effect on the health of the patient.
However, failure of the product to perform as indicated, or human error in use of the product may lead to a false result.
A false reactive or negative result may be considered a patient or public health concern because false results in the diagnostic setting may lead to diagnostic confusion and inappropriate therapy. IgM class antibodies against HBc are detected in high concentrations soon after infection with HBV and are an indicative marker for acute hepatitis. The detection of anti-HBc IgM is also used in the differential diagnosis of hepatitis B from other forms of viral hepatitis. A false negative result may lead to a diagnosis of autoimmune hepatitis (AH), and administration of immunosuppressive drugs potentially harmful to the patient. A false
4
{4}
Page 5 of 21 – P030026 Summary of Safety and Effectiveness Data
positive result might delay the diagnosis of AH, which requires early aggressive therapy. The risk of unnecessary liver biopsy might also be increased with erroneous anti-HBc IgM results.
## IX. Summary of Non-Clinical Studies
### Instrumentation
Software and hardware verification testing was performed for the Vitros ECi Immunodiagnostic System (Vitros Analyzer). Appropriate information and study results were furnished demonstrating that the Vitros Analyzer hardware and software, used with the Vitros Immunodiagnostic Products Anti-HBc IgM Reagent Pack, functioned as described and have appropriate safeguards.
### Comparison of Fresh Serum/Plasma Samples
To determine the acceptability of using the Vitros Anti-HBc IgM assay for testing serum or plasma specimens, fifty fresh blood samples (25 unspiked and 25 spiked with anti-HBc IgM to give a target result of $2.0 \pm 1.0$ signal/cut-off $(\mathrm{s} / \mathrm{c})$)¹, were collected by syringe and dispensed serum collection tubes, and collection tubes containing sodium heparin, $\mathbf{K}_2$ EDTA, and sodium citrate anticoagulants. Testing with the Vitros Anti-HBc IgM assay was conducted on the same day blood was drawn.
For matched anti-HBc IgM spiked samples, EDTA plasma compared with serum showed $0.5\%$ mean difference $(n = 25)$ in anti-HBc IgM s/c ratio, heparin plasma compared with serum showed $1.6\%$ mean difference $(n = 25)$ in anti-HBc IgM s/c ratio and citrate plasma compared with serum showed $-7.7\%$ mean difference $(n = 25)$ in anti-HBc IgM s/c ratio. The mean s/c ratios for EDTA and heparin plasma were not statistically different from the mean s/c ratio for serum. The mean s/c ratio $(n = 25)$ for citrate plasma was observed to be statistically different from the mean s/c ratio for serum. Part of this difference may be attributed to the dilution of the blood into the citrate buffer. Though there was a statistically significant difference, there was no clinical difference. A statement is included in the Limitation of Procedure section of the labeling cautioning the user concerning low positive or high negative results from citrated plasma. All of the anti-HBc IgM spiked samples tested maintained reactivity in the Vitros Anti-HBc IgM assay regardless of the sample type preparation. The within sample precision estimates were equivalent for anti-HBc IgM spiked serum and EDTA, heparin and citrate plasma.
All unspiked samples, with serum and plasma (EDTA, heparin, and citrate) preparations, were classified correctly as negative in the Vitros Anti-HBc IgM assay. Although the EDTA plasma samples had higher estimates of within sample precision than the other matrices, differences in precision were mostly $(99.79\%)$ accounted for by differences in mean s/c ratio results.
The results indicate that serum and plasma (EDTA, heparin, and citrate) are suitable for use in
¹ Where $\geq 1.1$ s/c is reactive for anti-HBc IgM and $< 0.90$ s/c is nonreactive for anti-HBc IgM. The range $\geq 0.9$ to $\leq 1.1$ s/co denotes an indeterminate result and the specimen requires duplicate retesting.
{5}
Page 6 of 21 – P030026 Summary of Safety and Effectiveness Data
the Vitros Anti-HBc IgM assay, if the user is aware that there will be a slight dilutional factor when sodium citrate is used.
## Comparison of Stability of Serum/Plasma Samples
The stability of serum and plasma samples for use in the Vitros Anti-HBc IgM assay was investigated. Twenty fresh blood samples (10 unspiked and 10 spiked with Anti-HBc IgM to give a target result of $2.0 \pm 1.0 \, \text{s/c}$), were collected by syringe and dispensed into serum and plasma (heparin, sodium citrate, and $\mathbf{K}_2$ EDTA) collection tubes. Testing with the Vitros Anti-HBc IgM assay was conducted on the same day blood was drawn, and again after 5 and 7 days storage at $2-8^{\circ}\text{C}$ ($36-46^{\circ}\text{F}$) and after 28 days at $-20^{\circ}\text{C}$ ($-4^{\circ}\text{F}$).
Statistical analysis of the data resulted in significantly different positive slopes for negative samples stored at $2-8^{\circ}\mathrm{C}$ ($36-46^{\circ}\mathrm{F}$) and negative slopes for positive samples stored $2-8^{\circ}\mathrm{C}$ ($36-46^{\circ}\mathrm{F}$). Though statistically significant, all samples were classified correctly.
For samples stored at $-20^{\circ}\mathrm{C}$ ($-4^{\circ}\mathrm{F}$), statistically significant differences were seen for negative heparin plasma and serum samples over the 28 day time period. Similar analysis for the anti-HBc IgM spiked samples resulted in statistically significant mean decreases among all sample types. However, none of the observed differences over time changed the classification of the specimens.
These data show that storage of serum or plasma (heparin, EDTA or citrate) samples for up to 5 days at $2 - 8^{\circ}\mathrm{C}$ ($36 - 46^{\circ}\mathrm{F}$), or 28 days at $-20^{\circ}\mathrm{C}$ ($-4^{\circ}\mathrm{F}$) would not have a significant clinical effect on the final test results with the Vitros Anti-HBc IgM assay.
## Potentially Cross-reacting Subgroups
The specificity of the Vitros Anti-HBc IgM assay was evaluated by testing 244 samples from 16 potentially cross-reacting sub-groups. Patient samples from the following sub-groups were tested: HAV, HEV, non-viral liver disease, autoimmune disease (rheumatoid arthritis and systemic lupus erythematosis), CMV, EBV, HSV, parvovirus B19 infection, rubella, syphilis, toxoplasmosis, HIV 1/2 antibody positive, HTLV 1/2 antibody positive, heterophilic antibodies and anti-HBc Positive/anti-HBc IgM Negative.
None of the 244 samples from 16 sub-groups were found to be anti-HBc IgM reactive with the comparator method (Abbott) or the Vitros Anti-HBc assay.
The specificity of the Vitros Anti-HBc IgM assay was evaluated further by testing anti-HBc IgM spiked and unspiked samples with an additional spike of *Staphylococcus aureus*, *Escherichia coli* or *Pseudomonas aeruginosa*.
None of the anti-HBc IgM unspiked (negative) samples were found to be false reactive and none of the anti-HBc IgM spiked samples were observed to be false negative in the presence or absence of the addition of potential cross reacting microorganisms in the Vitros Anti-HBc IgM assay.
{6}
Page 7 of 21 – P030026 Summary of Safety and Effectiveness Data
## Interfering Substances
The potentially interfering effects of hemoglobin, bilirubin and triolein were evaluated using samples from 10 blood donors. The results (test results at each level of interferent) demonstrate that samples spiked with hemoglobin (up to 500 mg/dL), bilirubin (up to 20 mg/dL), and triolein (up to 3000 mg/dL), should cause no misclassification of qualitative results.
Anti-HBc IgM spiked samples were tested near the cut-off (reactive (s/c) ≥1.00). No change in qualitative interpretation was observed for spiked samples at all levels tested with each potential interferent. Similarly no interference was observed in samples not spiked with anti-HBc IgM (negative), with anti-HBc IgM values remaining < 1.00 s/c.
## Stability
*Vitros* Anti-HBc IgM Reagent Packs, Calibrators and Controls that were subjected to a period of simulated transport to mimic effects of shipment were tested at various time points up to 26 weeks after storage at 2 – 8 °C (36 – 46 °F). All results obtained were within acceptability limits, and overall no trends were evident.
In addition, a commercially obtained performance panel was tested using transported, stored materials at week 0 and week 26. Materials stored for 26 weeks yielded results that indicated no change in the qualitative result interpretation of the samples from the qualitative results obtained at the initial time point.
This data supports the storage of the *Vitros* Anti-HBc IgM Reagent Pack, Calibrators and Controls for 26 weeks at 2 – 8 °C (36 – 46 °F).
## Open On-Board Storage for the *Vitros* Anti-HBc IgM Reagent Pack
*Vitros* Anti-HBc IgM Reagent Packs that were subjected to a period of simulated transport to mimic effects of shipment were opened and placed in an environmental chamber (4 – 8 °C, ≤ 40% humidity) for a period of 8 weeks to simulate the storage on board the *Vitros* Analyzer. These Reagent Packs were tested at various timepoints within the 8 week time period. In addition, a single transported, opened Reagent Pack from each Kit Lot was removed from the chamber on six different occasions, and brought to room temperature over the 8 week period to simulate typical customer usage. Results of testing were within acceptability limits and overall no trend was observed between Reagent Packs stored at 2 – 8 °C (36 – 46 °F) and freshly opened, and Reagent Packs stored opened on board for 8 weeks.
These data supports the on-board storage of the *Vitros* Anti-HBc IgM Reagent Packs for up to 8 weeks.
{7}
Page 8 of 21 – P030026 Summary of Safety and Effectiveness Data
## Open Off-Board Storage for Vitros Anti-HBc IgM Calibrators
*Vitros* Anti-HBc IgM Calibrators that were subjected to a period of simulated transport to mimic effects of shipment were opened, pooled, sub-aliquoted and stored at $2 - 8^{\circ}\mathrm{C}$ (36 – 46 °F) and $-20^{\circ}\mathrm{C}$ (-4 °F) for 13 weeks. Results of testing these calibrators at various timepoints up to 13 weeks indicated no observable trends and met all acceptance criteria.
The data supports the storage of the calibrators at $2 - 8^{\circ}\mathrm{C}$ (36 – 46 °F) and $-20^{\circ}\mathrm{C}$ (-4 °F) after opening for up to 13 weeks (with no more than one freeze-thaw cycle).
## Vitros Universal Wash Reagent Study
*Vitros* Anti-HBc IgM Reagent Packs, calibrators, and controls that were subjected to a period of simulated transport to mimic the effects of shipment were tested with 3 lots of *Vitros* Universal Wash Reagent at months 0 and 6 to determine the effect of aged *Vitros* Universal Wash Reagent.
The data presented in this report show that the performance of the *Vitros* Anti-HBc IgM assay is acceptable when used with *Vitros* Universal Wash Reagent which is either fresh or up to 6 months old.
## Vitros Signal Reagent Study
*Vitros* Anti-HBc IgM Reagent Packs, calibrators, and controls that were subjected to a period of simulated transport to mimic the effects of shipment were tested with 4 lots of *Vitros* Signal Reagent to determine the effect of aged *Vitros* Signal Reagent.
The data presented in this report show that the performance of the *Vitros* Anti-HBc IgM assay is acceptable when used with *Vitros* Signal Reagent which is either fresh or up to 6 months old.
## Temperature Stressing Study (30 °C and 37 °C (86 °F and 99 °F))
*Vitros* Anti-HBc IgM Reagent Packs and calibrators were subjected to 5 days at $30^{\circ}\mathrm{C}$ (86 °F) or 1 day at $37^{\circ}\mathrm{C}$ (99 °F) and the performance compared with Reagent Packs and calibrators stored at $2 - 8^{\circ}\mathrm{C}$ (36 – 46 °F).
Exposing *Vitros* Anti-HBc IgM Reagent Packs and calibrators to a temperature of $30^{\circ}\mathrm{C}$ (86 °F) for 5 days or $37^{\circ}\mathrm{C}$ (99 °F) for 1 day had no adverse effect on calibration quality parameters or control results. Exposure of the Reagent Packs or calibrators up to these temperatures for the times stated does not appear to significantly compromise the performance of the *Vitros* Anti-HBc IgM assay.
## Microbiology
*Vitros* Anti-HBc IgM reagents are formulated with an anti-microbial agent (Kathon) that provides protection against adventitious contamination by microorganisms. A study conducted
8
{8}
Page 9 of 21 – P030026 Summary of Safety and Effectiveness Data
according to US Pharmacopoeia (USP) 23/NF 18, general chapter 51, assessed the ability of the reagents to withstand or control microbial contamination. Results indicate that the preservative systems for Assay Reagent and calibrator, conjugate reagent and high sample diluent B met the requirements of the USP 23 at 0 and 26 weeks.
## Precision
Precision was evaluated on a different *Vitros* ECi Immunodiagnostic System at each of three external sites, using one reagent pack and calibrator kit lot. At least two replicates each of a three-member panel were assayed on a single occasion per day on 20 different days. The data shown in the table were rounded following all calculations.
| | | Within day * | | Between day | | Total ‡ | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical Site | Mean Vitros Anti-HBc IgM S/C (Ratio) | SD | CV (%) | SD† | CV (%) | SD | CV (%) | No. Obs. | No. Days |
| Site 1 | 0.01 | 0.001 | 11.0 | 0.001 | 10.5 | 0.001 | 15.2 | 40 | 20 |
| | 1.95 | 0.021 | 1.1 | 0.033 | 1.7 | 0.039 | 2.0 | 40 | 20 |
| | 0.98 | 0.021 | 2.2 | 0.012 | 1.3 | 0.025 | 2.5 | 40 | 20 |
| | | | | | | | | | |
| Site 2 | 0.01 | 0.001 | 16.1 | 0.000 | 0.0 | 0.001 | 16.1 | 40 | 20 |
| | 1.89 | 0.023 | 1.2 | 0.047 | 2.5 | 0.052 | 2.8 | 40 | 20 |
| | 0.96 | 0.014 | 1.4 | 0.025 | 2.6 | 0.028 | 2.9 | 40 | 20 |
| | | | | | | | | | |
| Site 3 | 0.01 | 0.001 | 8.6 | 0.000 | 6.0 | 0.001 | 10.5 | 40 | 20 |
| | 1.95 | 0.021 | 1.1 | 0.032 | 1.6 | 0.038 | 2.0 | 40 | 20 |
| | 0.97 | 0.015 | 1.6 | 0.017 | 1.8 | 0.023 | 2.4 | 40 | 20 |
* Within Day: Variability of the assay performance from replicate to replicate.
† Between Day: Variability of the assay performance from day to day.
‡ Total: Variability of the assay performance combining the effects of within day and between days.
## Reproducibility
Precision was further evaluated incorporating between site and between lot variations. The study was performed at three external sites using three reagent lots. At least three replicates each of a four-member panel were assayed on a single occasion per day on six different days. The between site, between lot, and total precision estimates (CV (%)) were derived from a variance component analysis. The data shown in the table were rounded following all calculations.
9
{9}
Page 10 of 21 – P030026 Summary of Safety and Effectiveness Data
| Mean Vitros Anti-HBc IgM S/C (Ratio) | Between Site * | | Between Lot † | | Total ‡ | | No. Obs. |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | CV (%) | SD | CV (%) | SD | CV (%) | |
| 2.49 | 0.076 | 3.0 | 0.097 | 3.9 | 0.135 | 5.4 | 162 |
| 1.06 | 0.031 | 3.0 | 0.025 | 2.3 | 0.047 | 4.5 | 162 |
| 0.97 | 0.027 | 2.8 | 0.016 | 1.7 | 0.040 | 4.1 | 162 |
| 0.23 | 0.008 | 3.4 | 0.008 | 3.8 | 0.013 | 5.7 | 162 |
* Between site: Variability of the assay performance from site to site.
† Between lot: Variability of the assay performance from lot to lot, calculated using data across all sites.
‡ Total: Variability of the assay incorporating factors of site, lot and day.
## Instrument Calibration Interval
The performance of the Vitros Anti-HBc IgM assay within and beyond one instrument calibration interval (28 days) was evaluated at three sites by testing a three-member sera panel with one kit lot. One panel member was close to the Vitros Anti-HBc IgM assay cutoff. Additional testing was performed on Day 29 and 30 of the calibration cycle to show that the Analyzer would still yield valid results beyond the end of a 28-day cycle. Two replicates of each panel member were assayed per day at each clinical site. Appropriate calibration was performed and verified on Day 1 of the study, and the testing was performed for a total of 20 study days over a 28-day period.
The least squares regression analyses were performed within site and across sites. For analyses within site, although the slopes were statistically significant for each of the panel members at one site, the changes in s/c ratios over the entire testing period were either not clinically relevant, i.e., did not change the qualitative result interpretation, or were so small that they would not have any clinical implications.
The Vitros Anti-HBc IgM assay demonstrated adequate performance throughout the entire calibration interval, and continued to perform successfully two days beyond the expiration of calibration.
## Seroconversion Panels
Six commercially available HBV seroconversion panels were tested. The Vitros and reference anti-HBc IgM assay results are summarized below. The table lists the first bleed of each panel that tested reactive with the Vitros and the reference assays as well as the difference between the two assays in identifying the first reactive panel member by number of days.
{10}
Page 11 of 21 – P030026 Summary of Safety and Effectiveness Data
## Anti-HBc IgM Seroconversion Panel Study - Summary Results
| Days to Reactive Anti-HBc IgM Result | | | | | |
| --- | --- | --- | --- | --- | --- |
| Panel ID | Reference Anti-HBc IgM Assay | | Vitros Anti-HBc IgM Assay | | Difference in Days to Anti-HBc IgM Reactive Result |
| | -* | + ** | -* | + ** | Reference - Vitros |
| 6278 | 33 | 37 | 37 | 41 | -4 |
| 6281 | 41 | 43 | 41 | 43 | 0 |
| PHM935A | 50 | 66 | 66 | 68 | -2 |
| RP009 | 36 | 43 | 36 | 43 | 0 |
| RP016 | 56 | 59 | 56 | 59 | 0 |
| RP017 | 76 | 78 | 76 | 78 | 0 |
* Post bleed day of last nonreactive result, usually denotes previous bleed from first reactive.
** Post bleed day of first reactive result.
Based on the results of the testing, the Vitros Anti-HBc IgM assay demonstrated agreement with the time of anti-HBc IgM comparable to the reference anti-HBc IgM assay.
## Cord Blood Testing
A total of 20 cord blood patient samples were tested in the Vitros Anti-HBc IgM assay.
In testing the cord blood samples, 0 out of 20 samples were found to give a repeat reactive result in the Vitros Anti-HBc IgM assay. Results of the reference method were also all negative.
Recovery of anti-HBc IgM from serum and cord blood was assessed by calculating the difference in result for each specimen when spiked with anti-HBc IgM reactive plasma or anti-HBc IgM negative plasma. The mean difference for serum specimens was 1.98 s/c. The mean difference for cord blood specimens was 2.04 s/c. It appears that anti-HBc IgM is recovered from the cord blood matrix to a similar extent as serum.
The above information shows that cord blood should be an acceptable sample type for use with the Vitros Anti-HBc IgM assay. Due to the low sample numbers and a single matrix being tested OCD has stated in the indications for use statement that serum may only be used and placed a warning in the Interpretations of Results stating that low level results should not be repeated. The warning implies that another specimen should be obtained and tested.
## X. Summary of Clinical Studies
A multi-center prospective study was conducted to evaluate the clinical performance of the Vitros Anti-HBc IgM assay among individuals with signs or symptoms or biochemical manifestations (elevated liver function tests) of hepatitis and those at high risk of hepatitis infection due to lifestyle, behavior, occupation, or known exposure events. Specimens were obtained from 1691 subjects prospectively enrolled at three geographically separated collection sites within the United States (Population I) located in Miami, FL (37.0%), Dallas, TX (28.1%) and Chicago, IL (34.9%). Specimens were also obtained from 315 subjects prospectively
{11}
Page 12 of 21 – P030026 Summary of Safety and Effectiveness Data
enrolled in an area in India with a high prevalence of viral hepatitis (Population II). Statistical testing performed to evaluate the homogeneity of the distribution of *Vitros* Anti-HBc IgM s/c values across the four collection sites indicated that the data from Population I and Population II could not be pooled for statistical analysis.
The HBV disease classification for each subject was determined by a single point serological assessment using a hepatitis marker profile consisting of reference assays (previously licensed or approved by the FDA) for the detection of HBsAg, HBeAg, anti-HBc, anti-HBc IgM, anti-HBe, and anti-HBs (quantitative). The reference assays' procedures were adhered to during the clinical laboratory study.
The subjects in Population I were Caucasian (24.9%), African American (44.1%), Hispanic (22.4%) and Asian (3.7%), with the remaining 4.9% represented by other ethnic groups. The group was 52.4% male and 47.6% female, and ranged in age from 5 to 89 years. Testing of these specimens with the *Vitros* Anti-HBc IgM assay occurred at diagnostic laboratories located in Miami, FL (37.0%), Port Jefferson, NY (34.9%) and Minneapolis MN, (28.1%). Agreement of the *Vitros* Anti-HBc IgM assay was assessed relative to the reference anti-HBc IgM assay and HBV disease classification using serum samples from the 1691 subjects in Population I.
The subjects in Population II were Asian Indian (100.0%). The group was 73.0% male and 27.0% female, and ranged in age from 18 to 90 years. Testing of specimens with the *Vitros* Anti-HBc IgM assay occurred at diagnostic laboratories located in Miami, FL (33.0%), Minneapolis MN, (32.4%) and Los Angeles, CA (34.6%). Agreement of the *Vitros* Anti-HBc IgM assay was assessed relative to the reference anti-HBc IgM assay and HBV disease classification using serum samples from the 315 subjects in Population II.
## XI. Results by Specimen Classification
The data were analyzed following the assignment of HBV disease classifications based upon the positive (+) / negative (-) patterns for the six HBV serological reference markers. The table below summarizes how these classifications were derived. There were 28 unique reference marker profiles observed among the subjects in Populations I and II (24 unique patterns in Population I and 18 unique patterns in Population II) during the *Vitros* Anti-HBc IgM clinical study.
{12}
Page 13 of 21 – P030026 Summary of Safety and Effectiveness Data
## HBV Reference Marker Profiles and HBV Disease Classification
| Reference HBsAg^{1,2} | Reference HBeAg | Reference IgM aHBc | Reference Total aHBc | Reference aHBe | Reference aHBs a10 mIU/mL | HBV Disease Classification |
| --- | --- | --- | --- | --- | --- | --- |
| + | + | + | + | + | - | Acute |
| + | + | + | + | - | - | Acute |
| + | - | + | + | + | + | Acute |
| + | - | + | + | + | - | Acute |
| + | - | + | + | - | - | Acute |
| + | - | - | - | - | - | Acute |
| + | + | - | + | + | - | Chronic |
| + | + | - | + | - | + | Chronic |
| + | + | - | + | - | - | Chronic |
| + | - | - | + | + | + | Chronic |
| + | - | - | + | + | - | Chronic |
| - | - | + | + | + | + | Early Recovery |
| - | - | + | + | + | - | Early Recovery |
| - | - | + | + | - | + | Early Recovery |
| - | - | + | + | - | - | Early Recovery |
| - | - | - | + | + | - | Early Recovery |
| - | - | - | + | + | + | Recovery |
| - | - | - | + | - | + | Recovered |
| - | - | - | + | - | - | Recovered |
| - | - | - | - | - | + | HBV Vaccine Response |
| - | - | - | - | - | - | Not Previously Infected with HBV |
| + | + | - | - | + | + | Uninterpretable |
| + | - | - | - | - | + | Uninterpretable |
| - | + | - | + | - | - | Uninterpretable |
| - | + | - | - | - | + | Uninterpretable |
| - | + | - | - | - | - | Uninterpretable |
1. Positive = Reference HBsAg assay reactive and confirmed by neutralization.
2. Negative = Reference HBsAg assay negative or not confirmed by neutralization.
## XII. Comparison of Results
The table below compares the Vitros Anti-HBc IgM results with the reference anti-HBc IgM results by specimen classification for the subjects in Population I.
{13}
Page 14 of 21 – P030026 Summary of Safety and Effectiveness Data
## Comparison of Vitros Anti-HBc IgM Results with Reference Anti-HBc IgM Results by HBV Disease Classification - Population I (N=1691)
| | Reference Anti-HBc IgM Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | VITROS Anti-HBc IgM Result | | VITROS Anti-HBc IgM Result | | |
| | Reactive | Negative * | Reactive ** | Negative | Total |
| Acute | 8 | 0 | 0 | 9 | 17 |
| Chronic | 0 | 0 | 2 | 41 | 43 |
| Early Recovery | 4 | 6 | 1 | 36 | 47 |
| Recovery | 0 | 0 | 0 | 138 | 138 |
| Recovered | 0 | 0 | 1 | 195 | 196 |
| HBV Vaccine Response | 0 | 0 | 0 | 169 | 169 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 1074 | 1074 |
| Uninterpretable | 0 | 1 | 0 | 6 | 7 |
| Overall | 12 | 7 | 4 | 1668 | 1691 |
* These samples were tested with a second FDA approved anti-HBc IgM assay with the following results:
- Early recovery: 2/6 negative
- Uninterpretable: 1/1 negative
- Overall: 3/7 (42.9%) negative agreement with VITROS
** These samples were tested with a second FDA approved anti-HBc IgM assay with the following results:
- Chronic: 2/2 positive
- Early Recovery: 1/1 positive
- Recovered: 1/1 positive
- Overall: 4/4 (100%) positive agreement with VITROS
The table below compares the Vitros Anti-HBc IgM results with the reference anti-HBc IgM results by specimen classification for the subjects in Population II.
## Comparison of Vitros Anti-HBc IgM Results with Reference Anti-HBc IgM Results by HBV Disease Classification - Population II (N=315)
| | Reference Anti-HBc IgM Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | VITROS Anti-HBc IgM Result | | VITROS Anti-HBc IgM Result | | |
| | Reactive | Negative * | Reactive | Negative | Total |
| Acute | 69 | 19 | 0 | 16 | 104 |
| Chronic | 0 | 0 | 0 | 185 | 185 |
| Early Recovery | 1 | 0 | 0 | 0 | 1 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 0 | 0 | 0 | 3 | 3 |
| HBV Vaccine Response | 0 | 0 | 0 | 3 | 3 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 17 | 17 |
| Uninterpretable | 0 | 0 | 0 | 2 | 2 |
| Overall | 70 | 19 | 0 | 226 | 315 |
* These samples were tested with a second FDA approved anti-HBc IgM assay with the following results:
- Acute: 8/19 negative; 3/19 indeterminate
- Overall: 11/19 (57.9%) negative or indeterminate
14
{14}
Page 15 of 21 – P030026 Summary of Safety and Effectiveness Data
## XIII. Percent Agreement
Positive and negative percent agreement between the Vitros Anti-HBc IgM assay and the reference anti-HBc IgM assay were calculated for subjects in Population I (N=1691) with various HBV disease classifications, and for the overall study population. The table below summarizes these calculations and provides the upper and lower 95% exact confidence intervals.
Positive and Negative Percent Agreement between the Vitros Anti-HBc IgM and Reference Anti-HBc IgM Assays in Population I
| HBV Disease Classification | Positive Percent Agreement (%) | 95% Exact Confidence Interval | Negative Percent Agreement (%) | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Overall | 12/19 (63.16%) | 38.36, 83.71 | 1668/1672 (99.76%) | 99.39, 99.93 |
| Acute | 8/8 (100%) | 63.06, 100 | 9/9 (100%) | 66.37, 100 |
| Chronic | 0/0 (N/A) | N/A | 41/43 (95.35%) | 84.19, 99.43 |
| Early Recovery | 4/10 (40%) | 12.16, 73.76 | 36/37 (97.3%) | 85.84, 99.93 |
| Recovery | 0/0 (N/A) | N/A | 138/138 (100%) | 97.36, 100 |
| Recovered | 0/0 (N/A) | N/A | 195/196 (99.49%) | 97.19, 99.99 |
| HBV Vaccine Response | 0/0 (N/A) | N/A | 169/169 (100%) | 97.84, 100 |
| Not Previously Infected with HBV | 0/0 (N/A) | NA | 1074/1074 (100%) | 99.66, 100 |
| Uninterpretable | 0/1 (0%) | N/A | 6/6 (100%) | 54.07, 100 |
In Population I, overall positive percent agreement of the Vitros Anti-HBc IgM assay with the reference anti-HBc IgM assay was estimated to be 63.16% (12/19, with a 95% exact confidence interval of 38.36% to 83.71%). Overall negative percent agreement of the Vitros Anti-HBc IgM assay with the reference anti-HBc IgM assay in Population I was estimated to be 99.76% (1668/1672, with a 95% exact confidence interval of 99.39, 99.93).
Positive and negative percent agreement between the Vitros Anti-HBc IgM assay and the reference anti-HBc IgM assay were also calculated for subjects in Population II (N=315) with various HBV disease classifications, and for the overall study population. The table below summarizes these calculations and provides the upper and lower 95% exact confidence intervals.
{15}
Page 16 of 21 – P030026 Summary of Safety and Effectiveness Data
## Positive and Negative Percent Agreement between the Vitros Anti-HBc IgM and Reference Anti-HBc IgM Assays in Population II
| HBV Disease Classification | Positive Percent Agreement (%) | 95% Exact Confidence Interval | Negative Percent Agreement (%) | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Overall | 70/89 (78.65%) | 68.69 - 86.63 | 226/226 (100%) | 98.38, 100 |
| Acute | 69/88 (78.41%) | 68.35 - 86.47 | 16/16 (100%) | 79.41, 100 |
| Chronic | 0/0 (N/A) | N/A | 185/185 (100%) | 98.03, 100 |
| Early Recovery | 1/1 (100%) | 2.5 - 100 | 0/0 (N/A) | N/A |
| Recovered | 0/0 (N/A) | N/A | 3/3 (100%) | 29.24, 100 |
| HBV Vaccine Response | 0/0 (N/A) | N/A | 3/3 (100%) | 29.24, 100 |
| Not Previously Infected with HBV | 0/0 (N/A) | N/A | 17/17 (100%) | 80.49, 100 |
| Uninterpretable | 0/0 (N/A) | N/A | 2/2 (100%) | 15.81, 100 |
The overall positive percent agreement of the Vitros Anti-HBc IgM assay with the reference anti-HBc IgM assay in Population II was estimated to be 78.65% (70/89, with a 95% exact confidence interval of 68.69% to 86.63%). Overall negative percent agreement of the Vitros Anti-HBc IgM assay with the reference anti-HBc IgM assay in Population II was estimated to be 100.0% (226/226, with a 95% exact confidence interval of 98.38, 100.0).
## XIV. Percent Agreement of the Vitros Anti-HBc IgM Assay with Clinical Status for Subjects with Clinically Diagnosed Acute or Chronic HBV Infection
The performance of the Vitros Anti-HBc IgM assay was further evaluated among archived serum samples from subjects with clinical and laboratory documentation of acute or chronic (HBsAg present for ≥ 6 months) HBV infection. The table below summarizes the performance of the Vitros Anti-HBc IgM assay in samples from subjects with documented acute or chronic HBV infection.
## Overall Clinical Performance of the Vitros Anti-HBc IgM Assay in Samples from Subjects with Clinically Documented Acute or Chronic HBV Infection
| HBV Infection | Number of Samples | Number (%) of Vitros Anti-HBc IgM Reactive Samples | 95% Exact Confidence Interval | Number (%) of Reference Anti-HBc IgM Reactive Samples | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- | --- |
| Acute | 8 | 8 (100) | 63.1, 100 | 8 (100) | 63.1, 100 |
| Chronic | 76 | 11 (14.5) | 7.5, 24.4 | 30 (39.5) | 28.4, 51.3 |
{16}
Page 17 of 21 – P030026 Summary of Safety and Effectiveness Data
For the information contained in Sections XII through XIV, although percent agreement for the *Vitros Anti-HBc IgM Assay* is less than 90% (lower 95% confidence interval) the above information shows that the *Vitros Anti-HBc IgM Assay* has similar performance compared to a reference anti-HBc IgM assay in the HBV disease categories listed. The lower percent agreement may be due to the low numbers tested in the disease classifications or differences between the two assays, e.g., antigen or detector antibody differences. It is believed that with the above performance the *Vitros Anti-HBc IgM Assay* will furnish useful and meaningful results when used as indicated.
## XV. Potentially Cross-Reacting Subgroups
Samples with evidence of hepatitis A virus infection (HAV) or hepatitis C virus infection (HCV) were identified in a population of 1691 samples prospectively collected from subjects in the U.S with signs or symptoms of, or at risk for, viral hepatitis (Population I). The tables below compare *Vitros Anti-HBc IgM* results with reference anti-HBc IgM results according to the HBV disease classifications assigned to the study subjects.
{17}
Page 18 of 21 – P030026 Summary of Safety and Effectiveness Data
## Comparison of Vitros and Reference Anti-HBc IgM Results and HBV Disease Classification among Anti-HAV IgM Reactive Study Subjects - Population I (N=7)
| | Reference Anti-HBc IgM Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | Vitros Anti-HBc IgM Result | | Vitros Anti-HBc IgM Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 0 | 0 | 0 | 0 | 0 |
| Chronic | 0 | 0 | 0 | 0 | 0 |
| Early Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 0 | 0 | 0 | 2 | 2 |
| HBV Vaccine Response | 0 | 0 | 0 | 0 | 0 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 5 | 5 |
| Uninterpretable | 0 | 0 | 0 | 0 | 0 |
| Overall | 0 | 0 | 0 | 7 | 7 |
## Comparison of Vitros and Reference Anti-HBc IgM Results and HBV Disease Classification among Anti-HCV Reactive Study Subjects - Population I (N=353)
| | Reference Anti-HBc IgM Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | VITROS Anti-HBc IgM Result | | VITROS Anti-HBc IgM Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 1 | 0 | 0 | 3 | 4 |
| Chronic | 0 | 0 | 0 | 9 | 9 |
| Early Recovery | 1 | 3 | 0 | 21 | 25 |
| Recovery | 0 | 0 | 0 | 43 | 43 |
| Recovered | 0 | 0 | 1 | 99 | 100 |
| HBV Vaccine Response | 0 | 0 | 0 | 22 | 22 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 148 | 148 |
| Uninterpretable | 0 | 0 | 0 | 2 | 2 |
| Overall | 2 | 3 | 1 | 347 | 353 |
Samples with evidence of hepatitis A virus infection (HAV) or hepatitis C virus infection (HCV) were identified in a population of 315 samples prospectively collected from subjects in an area in India with a high prevalence of viral hepatitis (Population II). The tables below compare Vitros Anti-HBc IgM results with reference anti-HBc IgM results according to the HBV disease classifications assigned to the study subjects.
18
{18}
Page 19 of 21 – P030026 Summary of Safety and Effectiveness Data
## Comparison of Vitros and Reference Anti-HBc IgM Results and HBV Disease Classification among Anti-HAV IgM Reactive Study Subjects - Population II (N=29)
| | Reference Anti-HBc IgM Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | Vitros Anti-HBc IgM Result | | Vitros Anti-HBc IgM Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 3 | 8 | 0 | 7 | 18 |
| Chronic | 0 | 0 | 0 | 1 | 1 |
| Early Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 0 | 0 | 0 | 0 | 0 |
| HBV Vaccine Response | 0 | 0 | 0 | 3 | 3 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 6 | 6 |
| Uninterpretable | 0 | 0 | 0 | 1 | 1 |
| Overall | 3 | 8 | 0 | 18 | 29 |
## Comparison of Vitros and Reference Anti-HBc IgM Results and HBV Disease Classification among Anti-HCV Reactive Study Subjects - Population II (N=90)
| | Reference Anti-HBc IgM Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | VITROS Anti-HBc IgM Result | | VITROS Anti-HBc IgM Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 45 | 13 | 0 | 0 | 58 |
| Chronic | 0 | 0 | 0 | 32 | 32 |
| Early Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 0 | 0 | 0 | 0 | 0 |
| HBV Vaccine Response | 0 | 0 | 0 | 0 | 0 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 0 | 0 |
| Uninterpretable | 0 | 0 | 0 | 0 | 0 |
| Overall | 45 | 13 | 0 | 32 | 90 |
## XVI. Conclusions Drawn from Studies
The Vitros Anti-HBc IgM Reagent Pack and Calibrator can be stored for up to 26 weeks at 2 – 8 °C (36 – 46 °F). After opening, the Reagent Pack can be stored on-board the Vitros Analyzer (4 – 8 °C, ≤40% relative humidity) for up to 8 weeks, and the Calibrator stored for up to 13 weeks at 2 – 8 °C (36 – 46 °F) or -20 °C (-4 °F) (with no more than one freeze-thaw cycle).
The preservative systems with which the Vitros Anti-HBc IgM assay reagents are formulated have been shown to meet USP 23 requirements at 26 weeks for the assay Reagent, Conjugate Reagent, Calibrator, and High Sample Diluent B.
{19}
Page 20 of 21 -- P030026 Summary of Safety and Effectiveness Data
The *Vitros* Anti-HBc IgM assay demonstrated adequate precision estimates within day and between day for each site as well as across all sites, and between replicate, between day, between site and between lots when these variables were introduced.
The *Vitros* Anti-HBc IgM assay has been shown to perform adequately over a 28-day calibration interval.
The *Vitros* Anti-HBc IgM assay agreed with a reference assay for four out of six seroconversion panels in determining the point at which the panel went from negative to repeatedly reactive. Two seroconversion panels were determined by the *Vitros* Anti-HBc IgM assay to become repeatedly reactive one bleed after a reference assay determined the panel to be repeatedly reactive. For one panel member, both the reference and the *Vitros* Anti-HBc IgM assay identified the point where the panel went from being negative to repeatedly reactive, but the *Vitros* Anti-HBc IgM assay identified that panel to then become repeatedly negative again before the reference assay showed consistent non-reactive results.
The *Vitros* Anti-HBc IgM Assay does not detect interference with cord blood samples (from newborn subjects). Both the reference and the *Vitros* Anti-HBc IgM assay detected 0 out of twenty reactive samples. Cord blood spiked with anti-HBc IgM positive serum reacts similarly to spiked adult serum.
The results of the clinical laboratory studies provide reasonable assurance the *Vitros* Anti-HBc IgM assay, when used according to the provided instructions for use, is safe and effective for the laboratory diagnosis of individuals with acute or chronic hepatitis B.
## XVII. PANEL RECOMMENDATION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XVIII. CDRH DECISION
FDA issued an approval order on March 4, 2004.
The applicant’s manufacturing facility was inspected on June 14, 2002 and found to be in compliance with the Quality Systems Regulation (21 CFR 820).
## XIX. APPROVAL SPECIFICATIONS
Directions for use: See the labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, precautions and Adverse Events in the labeling.
20
{20}
Page 21 of 21 – P030026 Summary of Safety and Effectiveness Data
Postapproval Requirements and Restrictions: See approval order.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.