ABBOTT ARCHITECT GLYCATED HEMOGLOBIN

K991483 · Abbott Laboratories · LCP · Nov 1, 1999 · Hematology

Device Facts

Record IDK991483
Device NameABBOTT ARCHITECT GLYCATED HEMOGLOBIN
ApplicantAbbott Laboratories
Product CodeLCP · Hematology
Decision DateNov 1, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Intended Use

The Abbott ARCHITECT™ Glycated Hemoglobin assay is a boronate affinity binding assay intended to measure glycated hemoglobin in human anticoagulated whole blood. Measurement of glycated hemoglobin is used for the management of diabetes mellitus.

Device Story

The ARCHITECT™ Glycated Hemoglobin assay is a Chemiluminescent Microparticle Immunoassay (CMIA) used on the Abbott ARCHITECT™ i System. It processes human anticoagulated whole blood samples to provide quantitative glycated hemoglobin measurements. The device utilizes boronate affinity binding to isolate glycated hemoglobin. It is intended for use in clinical laboratory settings by trained personnel. The output, provided as a percentage of A1c, assists clinicians in the management of diabetes mellitus. The system includes dedicated calibrators and controls to ensure analytical accuracy and precision.

Clinical Evidence

Clinical performance was evaluated by comparing the ARCHITECT™ Glycated Hemoglobin assay to the Bio-Rad DIAMAT™ HPLC system using 249 human anticoagulated whole blood specimens. Results showed a correlation coefficient (r) of 0.961. Regression analysis yielded a slope of 0.96 (95% CI: 0.93, 1.00) and intercept of 0.42 (95% CI: 0.15, 0.68) via Least Squares, and a slope of 1.00 (95% CI: 1.00, 1.06) and intercept of 0.20 (95% CI: -0.27, 0.20) via Passing-Bablok. Tested ranges were 4.3-14.3% A1c for the subject device and 4.4-14.5% A1c for the predicate.

Technological Characteristics

Chemiluminescent Microparticle Immunoassay (CMIA) utilizing boronate affinity binding. Designed for use on the Abbott ARCHITECT™ i System. Requires specific ARCHITECT™ Glycated Hemoglobin Calibrators and Controls for verification of accuracy and precision. Compatible with EDTA, lithium heparin, or sodium heparin anticoagulated whole blood.

Indications for Use

Indicated for the quantitative determination of glycated hemoglobin in human anticoagulated whole blood (EDTA, lithium heparin, or sodium heparin) for the management of diabetes mellitus.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summarv Abbott ARCHITECTTM Glycated Hemoglobin ## Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination The following information as presented in the Premarket Notification [510(k)] for Abbott ARCHITECT™ Glycated Hemoglobin constitutes data supporting a substantially equivalent determination. The ARCHITECT™ Glycated Hemoglobin assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of glycated hemoglobin in human anticoagulated whole blood (EDTA, lithium heparin, or sodium heparin). The ARCHITECTTM Glycated Hemoglobin assay is calibrated with Abbott ARCHITECT™ Glycated Hemoglobin Calibrators. Abbott ARCHITECT™ Glycated Hemoglobin Controls are assayed for the verification of the accuracy and precision of the Abbott ARCHITECT™ i System. Substantial equivalence has been demonstrated between the ARCHITECT™ Glycated Hemoglobin assay and the Bio-Rad DIAMAT™ Glycosylated Hemoglobin Analyzer System (K851636). The intended use of the ARCHITECTTM Glycated Hemoglobin assay is for the quantitative determination of glycated hemoglobin in human anticoagulated whole blood using a boronate affinity binding immunoassay. The intended use of the Bio-Rad DIAMAT™ Glycosylated Hemoglobin Analyzer System is for the quantitative determination of glycated hemoglobin in human anticoagulated whole blood using ion exchange high performance liquid chromatography (HPLC). | Sample | Regression Method | n | r | Slope (95%CI) | Intercept (95% CI) | |-------------|-------------------|-----|-------|-------------------|--------------------| | Whole Blood | Least Squares | 249 | 0.961 | 0.96 (0.93, 1.00) | 0.42 (0.15, 0.68) | | | Passing-Bablok | 249 | 0.961 | 1.00 (1.00, 1.06) | 0.20 (-0.27, 0.20) | {1}------------------------------------------------ In this evaluation, human anticoagulated whole blood specimens tested ranged from 4.3 to 14.3 %A1c eq with the ARCHITECT™ Glycated Hemoglobin assay and from 4.4 to 14.5 %A1c with the Bio-Rad DIAMAT™ Glycosylated Hemoglobin Analyzer System. In conclusion, these data demonstrate that the ARCHITECT™ Glycated Hemoglobin assay is as safe and effective as, and is substantially equivalent to, the Bio-Rad DIAMATTM Glycosylated Hemoglobin Analyzer System. Prepared and Submitted April 27, 1999 by: Katherine M. Wortley, Ph.D. Regulatory Specialist ADD Regulatory Affairs (847) 938-7707 Bldg. AP31, Dept. 9V6 Abbott Laboratories 200 Abbott Park Road Abbott Park, IL 60064-6200 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The symbol is black and the background is white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 ## NOV - 1 1999 Katherine M. Wortley, Ph.D. Regulatory specialist ADD Regulatory Affairs Abbott Laboratories Bldg. AP31. Dept. 9V6 200 Abbott Park Road Abbott Park, Illinois 60064-3537 K991483 Re: > Trade Name: Abbott ARCHITECT™ Glycated Hemoglobin Regulatory Class: II Product Code: LCP Dated: August 2, 1999 Received: August 3, 1999 Dear Dr. Wortley: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 510(k) Number (if known): \$991483 Device Name: Abbott ARCHITECT™ Glycated Hemoglobin Indications For Use: The Abbott ARCHITECT™ Glycated Hemoglobin assay is a boronate affinity binding assay intended to measure glycated hemoglobin in human anticoagulated whole blood. Measurement of glycated hemoglobin is used for the management of diabetes mellitus. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K991483 Prescription Use (Per 21 CFR 801.109) (Optional Format 1-2-96) OR Over-The-Counter Use*_*_