K984581 · Ab Biodisk · JWY · Mar 9, 1999 · Microbiology
Device Facts
Record ID
K984581
Device Name
SYSTEM U3 FOR CEPHALOTHIN
Applicant
Ab Biodisk
Product Code
JWY · Microbiology
Decision Date
Mar 9, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
System U3™ is a qualitative technique for antimicrobial susceptibility testing of nonfastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species. The system comprises an antibiotic gradient with MIC break-point concentrations and is used to determine the susceptibility categories of microorganisms to different antimicrobial agents, as tested on agar media using overnight incubation. This 510(k) application is for System U3/Cephalothin susceptibility testing based on the MIC breakpoints of Susceptible ≤ 8 µg/ml, Intermediate 16 µg/ml and Resistant ≥ 32 µg/ml for use with non-fastidious Gram negative and Gram positive aerobic bacteria.
Device Story
System U3 is an in vitro diagnostic device for antimicrobial susceptibility testing. It utilizes an antibiotic gradient on agar media to determine the Minimum Inhibitory Concentration (MIC) of Cephalothin against nonfastidious Gram-negative and Gram-positive aerobic bacteria. The system is used in clinical laboratories by trained personnel. Following overnight incubation, the device provides qualitative susceptibility categories (Susceptible, Intermediate, Resistant) based on established MIC breakpoints. These results assist clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
No clinical data provided; device relies on established MIC breakpoint methodology for antimicrobial susceptibility testing.
Technological Characteristics
The device consists of an antibiotic gradient system for use on agar media. It functions via diffusion-based MIC determination. No electronic, software, or mechanical components are described.
Indications for Use
Indicated for in vitro diagnostic antimicrobial susceptibility testing of nonfastidious Gram-negative and Gram-positive aerobic bacteria (e.g., Enterobacteriaceae, Pseudomonas, Staphylococcus, Enterococcus species) to determine susceptibility to Cephalothin.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
K984491 — SYSTEM U3 FOR CEFUROXIME · Ab Biodisk · Mar 9, 1999
K984087 — SYSTEM U3 FOR AMPICILLIN · Ab Biodisk · Feb 8, 1999
K983486 — SYSTEM U3 FOR NORFLOXACIN · Ab Biodisk · Dec 23, 1998
K983252 — SYSTEM U3 FOR GENTAMICIN · Ab Biodisk North America, Inc. · Nov 20, 1998
K981639 — SYSTEM U3 FOR CIPROFLOXACIN · Ab Biodisk · Sep 3, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The eagle is composed of three thick, curved lines that form the body and wings. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR - 9 1999 Anne Bolmström President AB BIODISK Dalvägen 10 S-169 56 Solna Sweden
Re: K984581
> Trade Name: System U3TM for Cephalothin Regulatory Class: II Product Code: JWY Dated: February 25, 1999 Received: March 1, 1999
Dear Ms. Bolmström:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (2) CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely vours.
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known):
Device Name: System U3™ for Cephalothin
Indications For Use:
For in vitro diagnostic use:
System U3™ is a qualitative technique for antimicrobial susceptibility testing of nonfastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species. The system comprises an antibiotic gradient with MIC break-point concentrations and is used to determine the susceptibility categories of microorganisms to different antimicrobial agents, as tested on agar media using overnight incubation.
This 510(k) application is for System U3/Cephalothin susceptibility testing based on the MIC breakpoints of Susceptible ≤ 8 µg/ml, Intermediate 16 µg/ml and Resistant ≥ 32 µg/ml for use with non-fastidious Gram negative and Gram positive aerobic bacteria.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | Woody Dabors |
|-----------------------------------------|--------------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K984581 |
| Prescription Use | X | OR | Over-The Counter Use |
|------------------|---|----|----------------------|
|------------------|---|----|----------------------|
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use *_*_
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.