JMS DIALYZER PRIMING SET
K983707 · Jms Co., Ltd. · KOC · Jan 8, 1999 · Gastroenterology, Urology
Device Facts
| Record ID | K983707 |
| Device Name | JMS DIALYZER PRIMING SET |
| Applicant | Jms Co., Ltd. |
| Product Code | KOC · Gastroenterology, Urology |
| Decision Date | Jan 8, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
Intended Use
This device is connected with Blood Tubing Set and used to prime Dialyzer and Blood Tubing Set. JMS Dialyzer Priming Set must be discarded after one time use.
Device Story
JMS Dialyzer Priming Set is a single-use accessory device used in hemodialysis procedures. It connects to blood tubing sets to facilitate the priming of the dialyzer and tubing circuit prior to patient connection. The device is intended for one-time use and must be discarded after the procedure. It is operated by clinical staff in a dialysis or clinical environment to prepare the extracorporeal circuit, ensuring the system is flushed and ready for patient treatment.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Single-use, disposable tubing set designed for fluid priming of hemodialysis circuits. Materials and construction are consistent with standard medical-grade tubing sets for extracorporeal blood circuits.
Indications for Use
Indicated for priming of dialyzers and blood tubing sets in clinical settings. For prescription use only.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Related Devices
- K100364 — GAMBRO CARTRIDGE BLOOD SET LOW WEIGHT- LOW VOLUME, MODEL 003422-520 · Gambro Renal Products, Inc. · Nov 3, 2010
- K971313 — NMC HEMODIALYSIS ARTERIAL BLOOD TUBING SET · National Medical Care, Inc. · Oct 27, 1997
- K112628 — NIPRO BLOOD TUBING SET WITH TRANSDUCER PROTECTOR AND PRIMING SET · Nipro Medical Corporation · Sep 28, 2012
- K113023 — STREAMLINE AIRLESS SYSTEM SET WITH PRE-ATTACHED DIALYZER · Nxstage Medical, Inc. · Dec 21, 2012
- K063290 — GAMBRO QUICKSET BLOODLINES · Gambro Renal Products · Jan 30, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized symbol. The symbol consists of three abstract human profiles facing to the right, stacked on top of each other. The profiles are depicted with simple, curved lines, creating a sense of unity and connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 8 1999
Tetsuharu Hirose JMS CO., LTD. 12-17 Kako -- Machi, Naka-Ku Hiroshima 730 JAPAN
Re: K983707 JMS Dialyzer Priming Set Dated: December 17, 1998 Received: December 21, 1998 Regulatory Class: II 21 CFR 876.5820/Procode: 78 KOC
Dear Mr. Hirose:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for vour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely vours.
~~
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{1}------------------------------------------------
## INDICATION FOR USE
510(k) Number (if known) :
Device Name : JMS Dialyzer Priming Set
:This device is connected with Blood Tubing Set and used to prime Indication for Use Dialyzer and Blood Tubing Set. JMS Dialyzer Priming Set must be discarded after one time use.
(PLEASE NOT WRITE THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use \/
or
Over - The - Counter Use______________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
Thind G. Sesson
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological D 510(k) Number
(Optional Format 1-2-96