K Number
K983501

Validate with FDA (Live)

Date Cleared
1998-12-18

(73 days)

Product Code
Regulation Number
862.3610
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the qualitative detection of methamphetamine, opiates. cocaine metabolites, THC metabolites, phencyclidine, benzodiazepines, barbiturates, tricyclic antidepressants, methadone, and medionein, phonoyoneans, on assist in screening of drug of abuse samples. The detecting cut-off concentrations are as follows:

METD-Methamphetamine1000 ng/mL
OPIMorphine300 ng/mL
COCBenzoylecgonine300 ng/mL
THC11-nor-Δ 9-9-carboxylic acid50 ng/mL
PCPPhencyclidine25 ng/mL
BenzodiazepineOxazepam300 ng/mL
BarbiturateSecobarbital300 ng/mL
MethadoneMethadone300 ng/mL
TCANortriptyline1000 ng/mL
Device Description

Not Found

AI/ML Overview

The provided document is a 510(k) premarket notification letter from the FDA for a device called "AccuSign® DOA 10" and related products. It outlines the FDA's determination of substantial equivalence to legally marketed predicate devices. However, this document primarily grants market authorization and does not contain the detailed study information, acceptance criteria, or performance data typically found in a clinical study report or a summary of safety and effectiveness data (SSED).

Therefore, I cannot fulfill all parts of your request with the information provided. Specifically, the document does not include:

  • A table of acceptance criteria and reported device performance (beyond the cut-off concentrations listed for each substance).
  • Sample sizes used for test sets.
  • Data provenance.
  • Number of experts and their qualifications for ground truth.
  • Adjudication method for the test set.
  • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
  • Information on standalone algorithm performance.
  • Type of ground truth used (beyond implying chemical analysis for cut-off concentrations).
  • Sample size for the training set.
  • How ground truth for the training set was established.

What can be extracted from the provided document is the stated indications for use and the cut-off concentrations, which serve as a foundational aspect of the device's intended performance.

Here's the closest I can get to your request using the available information:


Device Description and Intended Use:

The AccuSign® DOA 10 (and related AccuSign DOA Panel, AccuSign MET/OPI/COC/THC/PCP/BZO/BAR/MTD/TCA/AMP) is an immunoassay for the qualitative detection of various drugs of abuse and their metabolites in samples. It is intended to assist in the screening of drug of abuse samples.


Acceptance Criteria and Reported Device Performance

The core "acceptance criteria" presented in this document relate to the cut-off concentrations at which the device is designed to qualitatively detect the various substances. These cut-off concentrations define the threshold for a positive result.

SubstanceDetecting Cut-off Concentration (Acceptance Criteria)
MET (Methamphetamine)1000 ng/mL
OPI (Morphine)300 ng/mL
COC (Benzoylecgonine)300 ng/mL
THC (11-nor-Δ9-THC-carboxylic acid)50 ng/mL
PCP (Phencyclidine)25 ng/mL
Benzodiazepine (Oxazepam)300 ng/mL
Barbiturate (Secobarbital)300 ng/mL
Methadone300 ng/mL
TCA (Nortriptyline)1000 ng/mL

Note: The document states "The detecting cut-off concentrations are as follows:" implying these are the performance targets the device aims to meet. Actual reported performance metrics like sensitivity, specificity, accuracy, or concordance with a reference method are not detailed in this document. Such information would typically be found in the device's intended use instructions or 510(k) summary (which is not this letter).


Missing Information and Limitations:

Based solely on the provided FDA 510(k) letter, the following information required by your request is not available:

  1. Sample size used for the test set: Not mentioned.
  2. Data provenance (e.g., country of origin, retrospective/prospective): Not mentioned.
  3. Number of experts used to establish ground truth & qualifications: Immunoassay performance is typically validated against laboratory reference methods (e.g., GC/MS), not through expert consensus in the same way as image interpretation. The concept of "experts" for ground truth as traditionally applied to image analysis or clinical diagnosis is not directly applicable here. The "ground truth" for these devices is usually the precise chemical concentration determined by a more definitive laboratory method.
  4. Adjudication method for the test set: Not applicable for this type of test (validation against a reference method).
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable, as this is a diagnostic assay, not an AI interpretation of complex data by human readers.
  6. Standalone (algorithm only) performance: This device is a standalone diagnostic assay. Its performance would be assessed directly without human interpretation as part of the "device performance" (e.g., accuracy against a gold standard). However, the specific metrics (sensitivity, specificity) are not provided in this letter.
  7. Type of ground truth used: Implied to be quantitative chemical analysis (e.g., GC/MS) to establish the true concentration of drug metabolites, which then determines if it's above or below the cut-off. This document does not explicitly state the method.
  8. Sample size for the training set: Not applicable in the context of an immunoassay. The "training" for such devices involves assay development and calibration, not machine learning on a distinct training set.
  9. How the ground truth for the training set was established: See point 8.

Conclusion:

This FDA 510(k) letter confirms the market clearance of the AccuSign® DOA 10 device based on substantial equivalence to existing devices. It defines the substances detected and their respective cut-off concentrations, which are fundamental to the device's intended performance. However, it does not include the detailed performance study results or methodologies (like expert adjudication, sample sizes, provenance, or specific validation metrics) that would typically be found in a comprehensive technical report or FDA's decision summary for substantial equivalence. For those details, one would need to consult the full 510(k) submission summary, if publicly available.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, composed of three curved lines that suggest wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.

DEC 1 8 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Jemo Kang, Ph.D., M.B.A. Director Princeton BioMeditech Corporation P.O. Box 7139 Princeton, NJ 08543-7139

Re: K983201

Trade Name: AccuSign® DOA 10, AccuSign DOA Panel, AccuSign MET/OPI/ COC/THC/PCP/BZO/BAR/MTD/TCA/AMP Regulatory Class: II Product Code: LAG,DJG,DIO,DKE,LCM,DIS,DKN,DJR,LFI and DKZ Dated: October 5, 1998 Received: October 6, 1998

Dear Dr. Kang:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act). or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{1}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K9835 510(k) Number (if known): __

ice Name:__AccuSign®DOA 10 (MET/QPI/COC/THC/PCP/BZO/BARMTD/TCA/AMP)

Indications For Use:

Immunoassay for the qualitative detection of methamphetamine, opiates. cocaine metabolites, THC metabolites, phencyclidine, benzodiazepines, barbiturates, tricyclic antidepressants, methadone, and medionein, phonoyoneans, on assist in screening of drug of abuse samples. The detecting cut-off concentrations are as follows:

METD-Methamphetamine1000 ng/mL
OPIMorphine300 ng/mL
COCBenzoylecgonine300 ng/mL
THC11-nor-Δ 9-9-carboxylic acid50 ng/mL
PCPPhencyclidine25 ng/mL
BenzodiazepineOxazepam300 ng/mL
BarbiturateSecobarbital300 ng/mL
MethadoneMethadone300 ng/mL
TCANortriptyline1000 ng/mL

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANN)THER PAGE IF NEEDED)



Concurrence of CDRH Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Clinical Laboratory Devices
510(k) Number____________________
Professional Use:X
Prescription Use:X
(Per 21 CFR 801.109)

OR

Over The Counter Use:____________________
---------------------------------------------

(Optional Format 1-2-96)

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).