IMPLEX SCREW AND WASHER SYSTEM
K982751 · Implex Corp. · MBI · Feb 1, 1999 · Orthopedic
Device Facts
| Record ID | K982751 |
| Device Name | IMPLEX SCREW AND WASHER SYSTEM |
| Applicant | Implex Corp. |
| Product Code | MBI · Orthopedic |
| Decision Date | Feb 1, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Implex Screw and Washer System is indicated for use in the attachment of ligaments, tendons and other soft tissue to bone.
Device Story
Implex Screw and Washer System consists of Implex Continuum Bone Screw (5 mm diameter; 15-65 mm lengths) and Hedrocel®/CP titanium washer. Washer available in two versions: flat tissue-contacting surface or flat surface with recessed grooves; three outer diameters (14, 17, 20 mm). Device used by surgeons for soft tissue-to-bone fixation. Mechanical testing confirms strength of CP Titanium/Hedrocel® interface. System provides secure attachment for ligaments and tendons.
Clinical Evidence
Bench testing only. Mechanical testing demonstrated strength of the CP Titanium/Hedrocel® interface. Material and structural properties verified via MAF#920.
Technological Characteristics
Components: CP Titanium and Hedrocel®. Screw diameter: 5 mm. Washer diameters: 14 mm, 17 mm, 20 mm. Mechanical fixation principle. No software or electronic components.
Indications for Use
Indicated for attachment of ligaments, tendons, and other soft tissue to bone in patients requiring surgical fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Implex Continuum® Bone Screw
- MDLI Soft Tissue Screw and Washer System
- Mitek Screw and Washer System
Related Devices
- K994202 — ACUFEX SPIKED WASHER SYSTEM, ACUFEX SUTURE WASHER SYSTEM, ACUFEX CANCELLOUS SCREW AND SPIKED WASHER SYSTEM · Smith & Nephew, Inc. · Feb 25, 2000
- K981967 — WASHERLOC SCREW SYSTEM, LIGAMENT SCREW SYSTEM, NO-PROFILE BI-CORTICAL SCREW SYSTEM, CHANNEL LIGAMENT CLAMP, HECKMAN SCRE · Arthrotek/Biomet, Inc. · Jul 17, 1998
- K080447 — SURGICRAFT SCREW FIXATION SYSTEM · Surgicraft , Ltd. · Jun 13, 2008
- K012572 — SOFT TISSUE SCREW AND WASHER · Biomet, Inc. · Dec 3, 2001
- K961649 — ACUFEX TIBIAL ANCHOR SCREW & SPIKED WASHER · Smith & Nephew Endoscopy, Inc. · Jul 16, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
FEB
## 510(k) SUMMARY Implex Screw and Washer System
| Submitter Name: | Implex Corp. |
|-----------------------|-------------------------------------------------------|
| Submitter Address: | 80 Commerce Drive<br>Allendale, New Jersey 07401-1600 |
| Contact Person: | John Schalago, RAC or Robert Poggie, Ph.D. |
| Phone Number: | (201) 818-1800 |
| Fax Number: | (201) 818-0567 |
| Date Prepared: | January 27, 1999 |
| Device Trade Name: | Implex Screw and Washer System |
| Device Common Name: | Washer, bolt, nut |
| Classification Number | 21 CFR § 878.3030 |
| Substantial<br>Equivalence: | The term "substantial equivalence" as used in this<br>510(k) notification is limited to the definition of<br>substantial equivalence found in the Federal Food,<br>Drug and Cosmetic Act, as amended and as applied<br>under 21 CFR 807, Subpart E under which a device<br>can be marketed without premarket approval or<br>reclassification. A determination of substantial<br>equivalency under this notification is not intended<br>to have any bearing whatsoever on the resolution of<br>patent infringement suits or any other patent<br>matters. No statements related to, or in support of<br>substantial equivalence herein shall be construed<br>as an admission against interest under the US<br>Patent Laws or their application by the courts. |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
- Device Description: The Implex Screw and Washer System consists of the Implex Continuum Bone Screw configured with a Hedrocel®/CP titanium Washer. The Implex Continuum Bone Screw has a 5 mm diameter and is offered in lengths from 15 to 65 mm, in 5 mm increments. The Washer is offered in two versions. Washer-A and Washer-A has a flat tissue-contacting Washer-B. surface and Washer-B has a flat tissue-contacting surface with recessed grooves. Both versions are round, and offered in three outer diameters, 14 mm, 17 mm, and 20 mm.
{1}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness (continued)
,
| Indications for Use: | The Implex Screw and Washer System is indicated<br>for use in the attachment of ligaments, tendons and<br>other soft tissue to bone. |
|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Technological<br>Characteristics and<br>Comparison to<br>Predicate Device: | The Hedrocel® Screw and Washer System has the<br>same intended use and indications for use, and is<br>offered in equivalent geometries and sizes as the<br>predicate devices. |
| Performance Data: | Performance data provided in MAF#920<br>demonstrates that material and structural properties<br>of the washer component of the Implex Screw and<br>Washer System are suitable for the intended use of<br>the device. The screw is a currently legally marketed<br>bone screw. Mechanical testing demonstrating the<br>strength of the CP Titanium/Hedrocel® interface is<br>included in this 510(k) Premarket Notification. |
| Predicate Device<br>Information: | The predicate device cited in this premarket<br>notification to support a determination of substantial<br>equivalence are the Implex Continuum® Bone Screw,<br>the MDLI Soft Tissue Screw and Washer System, and<br>the Mitek Screw and Washer System. |
| Conclusion: | The Implex Screw and Washer System is<br>substantially equivalent to the identified predicate<br>devices. |
Pagc 2 of 2
ਾਰ ਦੇ ਸਾਹਿਤ ਕਿ ਇਹ ਸਿੰਘ
1
2016/08/14/10
1
ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺪﺍﻓﻊ
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized human profiles incorporated into the design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB | 1 1999
John A. Schalago, RAC Requlatory Affairs Manager Implex Corporation 80 Commerce Drive 07401-1600 Allendale, New Jersey
Re: K982751 Implex Corporation Screw and Washer System Trade Name: Regulatory Class: II Product Codes: MBI and HWC Dated: November 3, 1998 November 4, 1998 Received:
Dear Mr. Schalago:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2 - John A. Schalago, RAC
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Neil R.P. Ogden for
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): 2 1 5 1
Implex Screw and Washer System Device Name:
Indications For Use:
The Implex Screw and Washer System is indicated for use in the ligaments, tendons and other soft tissue to attachment of bone .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
X Prescription Use (Per 21 CFR 801.109)
OR...
Over-The-Counter Use
(Optional Format 1-2-96)
fo
(Division Sign-Off) Division of Ceneral Restorative Devices 982 510(k) Number_ ﮨﮯ۔ 5