K Number
K982332

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-01-20

(202 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STERIS Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture- stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.

The STERIS Ten Sixteen Tabletop Steam Sterilizer is equipped with the following factoryprogrammed set sterilization cycles and cycle values:

CYCLESRECOMMENDED LOADSTERILIZETEMP.STERILIZETIMEDRY TIME
UNWRAPPEDUnwrapped Instrument tray witha single instrument (No lumensand no dental handpieces).270°F (132°C)3 minutes0 minute
WRAPPEDWrapped instrument pack: onepack, maximum weight 1300 g(2.87 lbs). Paper-peel Pouches:maximum 5 pouches, oneinstrument per pack, totalmaximum weight 500 g (1.1 lb).(No lumens and no dentalhandpieces).270°F (132°C)10 minutes45 minutes
LIQUIDVented Borosilicate glasscontainers (500ml or smaller).250°F (121°C)40 minutes10 minutes(Cool Time)
PACKEDFabric pack(s): maximum weight1050 g (2.32 lbs).250°F (121°C)30 minutes30 minutes
Device Description

The STERIS Ten Sixteen Steam Sterilizer is a Class II medical device as defined by 21 CFR §880.6880. The Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture-stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.

AI/ML Overview

The STERIS Ten Sixteen Steam Sterilizer operates with several factory-programmed cycles, each with specific acceptance criteria related to sterilization temperature, time, and dry time. The device's performance is demonstrated through studies showing the complete kill of biological indicators.

1. Table of Acceptance Criteria and Reported Device Performance:

CyclesRecommended LoadSterilize Temp. (Acceptance Criteria)Sterilize Time (Acceptance Criteria)Dry Time (Acceptance Criteria)Reported Device Performance
UNWRAPPEDUnwrapped Instrument tray with a single instrument (No lumens and no dental handpieces).270°F (132°C)3 minutes0 minuteMeets Criteria: "complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10-6"
WRAPPEDWrapped instrument pack; Paper-peel Pouches. (No lumens and no dental handpieces).270°F (132°C)10 minutes45 minutesMeets Criteria: "complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10-6"
LIQUIDVented borosilicate glass containers (500ml or smaller).250°F (121°C)40 minutes10 minutes (Cool Time)Meets Criteria: "complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10-6"
PACKEDFabric packs.250°F (121°C)30 minutes30 minutesMeets Criteria: "complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10-6"

Study Proving Device Meets Acceptance Criteria:

The efficacy of the STERIS Ten Sixteen Steam Sterilizer is demonstrated through a rigorous qualification and validation process that utilizes Bacillus stearothermophilus spores as the Most Resistant Organism (MRO) to steam. This organism is used to ensure a worst-case test of the sterilizer's efficacy across all forms of microbial and viral contamination.

The core of the study's findings is "showing complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10^-6" (probability of less than one chance out of one million of a non-sterile indicator).

Additional Information Not Available in the Provided Text:

  1. Sample size used for the test set and the data provenance: The document does not specify the exact sample size for the test set (number of biological indicators or test runs). It broadly states that "qualification studies were in compliance with the Association for the Advancement of Medical Instrumentation (AAMI) standard for biological indicators (AAMI-ST19)." The data provenance is not explicitly stated in terms of country of origin, but it is implied to be from studies conducted by STERIS during product development and validation for the US market. The studies are prospective in nature, as they are part of the initial development and qualification of the device before market release.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided. The ground truth (sterilization efficacy) is established by the scientific principle of killing Bacillus stearothermophilus spores and achieving a specific sterility assurance level, rather than expert consensus on individual results.

  3. Adjudication method for the test set: Not applicable in the context of biological indicator testing. The outcome (kill or no-kill of biological indicators) is a direct, objective measurement.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not applicable. The device is a steam sterilizer, not an AI-powered diagnostic tool requiring human reader interpretation.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is not applicable to a steam sterilizer. The device functions standalone in achieving sterilization, with human interaction primarily for loading, cycle selection, and monitoring, not for interpreting results in a way that an algorithm would.

  6. The type of ground truth used: The ground truth is objective biological efficacy – the complete kill of Bacillus stearothermophilus spores within biological indicators, demonstrating a Sterility Assurance Level (SAL) of less than 10^-6. This is a scientific and microbiological ground truth, not pathology slides or outcomes data.

  7. The sample size for the training set: The document does not provide a specific sample size for a "training set" in the context of sterilizer validation. Instead, it refers to "development, qualification and validation" studies, which imply repeated testing to establish the parameters and efficacy of the cycles.

  8. How the ground truth for the training set was established: The ground truth for establishing the sterilization cycles (analogous to a "training set" for physical parameters) was established by using B. stearothermophilus spores as the MRO under worst-case loading conditions. The process involves systematically testing different temperature and time combinations until "complete kill of biological indicators" and the required SAL (<10^-6) are consistently achieved for each cycle type. This iterative testing under controlled conditions establishes the validated cycle parameters.

{0}------------------------------------------------

JAN 20 1999

K 982332

A 510(k) SUMMARY PERTAINING TO THE SAFETY AND EFFECTIVENESS OF THE STERIS TEN SIXTEEN STEAM STERILIZER

Submitter Information.

Raymond Ursick Senior Director, Regulatory Affairs and Quality Systems 5960 Heisley Road Mentor, Ohio 44060 (440) 354-2600 Phone Date Summary Prepared: June 26, 1998

Introduction. Moist heat in the form of saturated steam under pressure is the oldest, safest and most dependable form of sterilization. Steam sterilization is universally recognized as the method of choice, except when high temperatures, pressure and/or moist heat would damage or degrade items to be sterilized. Steam sterilizers in a wide variety of sizes are used in hospitals, clinics, doctors' and dentists' offices, clinical and research laboratories and industry. Uses include sterilization of reusable medical devices such as surgical instruments, gowns and drapes; sterilization of single-use medical devices such as syringes and catheters; decontamination of potentially infectious waste generated during surgery or research; sterilization of other materials, instruments and goods used in biotechnology and pharmaceutical houses, and in other applications whenever biological contamination and infection control are essential.

The STERIS Ten Sixteen Steam Sterilizer is a Class II medical device as defined by 21 CFR §880.6880. The Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture-stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.

STERIS's manufacturing facility for the STERIS Ten Sixteen Steam Sterilizer is located in Erie, Pennsylvania and is registered with and regulated by the FDA. STERIS's Erie facility is inspected by the FDA to assure that the requirements of Quality System Regulation are met. The Erie manufacturing facility is certified to the ISO 9001 Quality System Standard and the EN Medical Device Directive.

{1}------------------------------------------------

The STERIS Ten Sixteen Steam Sterilizer cycle descriptions are as follows:

CYCLESRECOMMENDEDLOADSTERILIZETEMP.STERILIZETIMEDRYTIME
UNWRAPPEDUnwrapped Instrumenttray with a singleinstrument (No lumensand no dentalhandpieces).270°F (132°C)3 minutes0 minute
WRAPPEDWrapped instrumentpack. Paper-peelPouches. (No lumensand no dentalhandpieces).270°F (132°C)10 minutes45minutes
LIQUIDVented borosilicate glasscontainers (500ml orsmaller).250°F (121°C)40 minutes10minutes(CoolTime)
PACKEDFabric packs.250°F (121°C)30 minutes30minutes

Effectiveness. Under controlled conditions of temperature and pressure, steam is capable of killing all forms of sporulating and vegetative bacteria, viruses, fungi, yeast and bacterial and fungal endospores. It has been well established that Bacillus stearothermophilus in the form of spores is the Most Resistant Organism (MRO) to steam. Thus, during the development, qualification and validation of STERIS steam sterilizers, B. stearothermophilus spores are used in order to ensure a worst-case test of the steam sterilizer's efficacy in killing all forms of microbial and viral contamination.

Efficacy of sterilizer function and exposure time recommendations are ultimately demonstrated by showing complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10* (probability of less than one chance out of one million of a non-sterile indicator). STERIS qualifies its recommended sterilization cycles through the application of rigorous performance standards and recommended practices developed by various independent organizations that have specified worst-case loads. The STERIS Ten Sixteen Steam Sterilizer meet the applicable requirements of the AAMI-ST55 "Tabletop Steam Sterilizers" performance standard. The STERIS Ten Sixteen Unwrapped Cycle meets the applicable requirements of AAMI-ST37 standard for flash sterilization. Biological indicators used in qualification studies were in compliance with the Association for the Advancement of Medical Instrumentation (AAMI) standard for biological indicators (AAMI-ST19).

{2}------------------------------------------------

Safety. STERIS sterilizers have been designed, constructed and tested to meet the safety and performance requirements of various national codes and standards. Before release to market, STERIS sterilizers are tested and certified by independent testing agents to the following requirements:

  • Underwriters Laboratory (UL) 544 as certified by ETL Testing Laboratories, Inc.; 1.
  • American Society of Mechanical Engineers (ASME), Section VII, Division 1; and 2.
  • Canadian Standards Association (CSA) C22.2 No. 125 as certified by ETL Testing 3. Laboratories, Inc.

A Failure Modes Effect Criticality Analysis was also conducted on the STERIS Ten Sixteen Steam Sterilizer's electrical, mechanical and piping system.

Hazards-Failure of Performance. Failure of the sterilization process can lead to incidents of cross contamination, the transmission of potentially infectious organisms from an infected person to another who was not otherwise infected prior to the incident.

To avoid failure, the operator must ensure the following:

  • That the materials, instruments and devices to be sterilized are thoroughly 1. cleaned and decontaminated;
  • That the manufacturer's instructions for use are followed; 2.
    1. That the operator must select the validated cycle applicable for the load;
  • That the sterilizer has been maintained in accordance with the manufacturer's 4. recommended maintenance schedule and is operating properly; and
  • That each sterilizer load is monitored with available biological and chemical 5. sterilization process indicators.

Today there are thousands of steam sterilizers in daily use in hospitals and other healthcare environments throughout the United States running several million cycles per year. Despite its widespread use, the incidence of sterilizer malfunction or sterilization process failure is relatively rare. Further, there are no known reports in the literature of patient infections that have resulted from steam sterilizer failure. Although the ability to prove that nosocomial infections result from failure of the sterilization process or a malfunction of the sterilizer is at best difficult, the technology designed in today's STERIS sterilizers provides numerous builtin safeguards that abort the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.

User Information. STERIS conducts in-house customer training and has developed a series of user training videos that provide helpful information about the appropriate use of steam sterilizers. Through its careful attention to detail in instructions for use and other labeling, STERIS further provides information to the user that is intended to ensure safe and effective use of steam sterilization. Numerous recommended practice documents have been developed by AAMI and AORN to give guidance in techniques that are intended to reduce the risks of cross-contamination. The use and periodic review of, as well as adherence to these materials, give further assistance of the safe and effective use of steam sterilization equipment in healthcare environments.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three profiles, representing health, human services, and the USA.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 20 1999

Mr. Raymond Ursick Senior Director Regulatory Affairs and Quality Systems Steris® Corporation 5960 Heisley Road Mentor, Ohio 44060

K982332 Re : STERIS Ten Sixteen Steam Sterilizer Trade Name: Requlatory Class: II FLE Product Code: October 21, 1998 Dated: Received: October 22, 1998

Dear Mr. Ursick:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਕੇ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Mr. Ursick

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Tim Director

Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

STERIS TEN SIXTEEN TABLETOP STEAM STERILIZER

INDICATIONS FOR USE:

The STERIS Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture- stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.

The STERIS Ten Sixteen Tabletop Steam Sterilizer is equipped with the following factoryprogrammed set sterilization cycles and cycle values:

CYCLESRECOMMENDED LOADSTERILIZETEMP.STERILIZETIMEDRY TIME
UNWRAPPEDUnwrapped Instrument tray witha single instrument (No lumensand no dental handpieces).270°F (132°C)3 minutes0 minute
WRAPPEDWrapped instrument pack: onepack, maximum weight 1300 g(2.87 lbs). Paper-peel Pouches:maximum 5 pouches, oneinstrument per pack, totalmaximum weight 500 g (1.1 lb).(No lumens and no dentalhandpieces).270°F (132°C)10 minutes45 minutes
LIQUIDVented Borosilicate glasscontainers (500ml or smaller).250°F (121°C)40 minutes10 minutes(Cool Time)
PACKEDFabric pack(s): maximum weight1050 g (2.32 lbs).250°F (121°C)30 minutes30 minutes

(Please Do Not Write Below This Line - Continue on Another Page If Needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

Qam Silm
(Optional Format 1-2-96)

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).