ETHYL ALCOHOL CLINICAL STANDARD SOLUTIONS
K981926 · Emd Chemicals, Inc. · DLJ · Jul 22, 1998 · Clinical Toxicology
Device Facts
| Record ID | K981926 |
| Device Name | ETHYL ALCOHOL CLINICAL STANDARD SOLUTIONS |
| Applicant | Emd Chemicals, Inc. |
| Product Code | DLJ · Clinical Toxicology |
| Decision Date | Jul 22, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3200 |
| Device Class | Class 2 |
Intended Use
The Ammonia assay is used for the quantitation of ammonia in human plasma.
Device Story
Ammonia assay is an in vitro clinical chemistry diagnostic test for human plasma. Principle of operation involves enzymatic amination of alpha-ketoglutarate by ammonia, catalyzed by glutamate dehydrogenase (GLDH), with concomitant oxidation of NADPH. The reaction produces glutamate and NADP+; the resulting decrease in absorbance at 340 nm is measured spectrophotometrically. The magnitude of absorbance decrease is directly proportional to ammonia concentration in the sample. The assay is performed on clinical chemistry analyzers (e.g., AEROSET System) in a laboratory setting by trained technicians. Healthcare providers use the quantitative results to assess patient metabolic status related to ammonia levels. The device provides objective data to assist in clinical decision-making regarding conditions associated with hyperammonemia.
Clinical Evidence
Bench testing only. Comparative performance study conducted on the AEROSET System against the predicate A-GENT Ammonia assay on the ABBOTT SPECTRUM Series II System. Results: correlation coefficient 0.9957, slope 0.971, Y-intercept -2.083 umol/L. Precision studies (within-run, between-run, between-day) yielded total %CVs of 11.7% (Level 1), 4.0% (Level 2), and 5.5% (Level 3). Linearity established up to 1574.9 umol/L. Limit of quantitation (sensitivity) is 20.6 umol/L.
Technological Characteristics
In vitro clinical chemistry assay; enzymatic method using glutamate dehydrogenase (GLDH) and NADPH. Measures absorbance decrease at 340 nm. Designed for use on automated clinical chemistry analyzers (e.g., AEROSET System).
Indications for Use
Indicated for the quantitative determination of ammonia in human plasma for clinical diagnostic purposes.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- A-GENT® Ammonia assay (K870787)
Reference Devices
- ABBOTT SPECTRUM® Series II™ System
Related Devices
- K981920 — AMM · Abbott Laboratories · Jul 22, 1998
- K981467 — AMM · Abbott Laboratories · Jun 22, 1998
- K030873 — RANDOX AMMONIA · Randox Laboratories, Ltd. · May 12, 2003
- K183517 — Ammonia II · Roche Diagnostics Operations (Rdo) · Feb 8, 2019
- K123320 — DIMENSION AMMONIA FLEX REAGENT CARTRIDGE · Siemens Healthcare Diagnostics · Feb 15, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
K98 1926
### Submitter's Name/Address
Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038 Contact Person Mark Littlefield Section Manager MS 1-8 Regulatory Affairs (972) 518-6062 Fax (972) 753-3367
| Date of Preparation of this Summary: | May 29, 1998 |
|--------------------------------------------------|---------------|
| Device Trade or Proprietary Name: | Amm |
| Device Common/Usual Name or Classification Name: | Ammonia |
| Classification Number/Class: | 75JIX/Class I |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
## Test Description:
Ammonia is an in vitro diagnostic assay for the quantitative determination of ammonia in human plasma. The Ammonia assay is a clinical chemistry assay which utilizes the amination of x-ketoglutarate by ammonia (NH3) along with the concomitant oxidation of NHzDPH. These reactions, catalyzed by glutamate dehydrogenase (GLDH), produce glutamate and NH2DP*. The oxidation of NHxDPH produces a decrease in absorbance at 340 nm which is directly proportional to the concentration of ammonia in the sample.
#### Substantial Equivalence:
The Ammonia assay is substantially equivalent to the A-GENT® Ammonia assay (K870787) on the ABBOTT SPECTRUM® Series II™ System.
Ammonia 510(k) May 29, 1998 Ammf.lwp
Section II Page 1
0000012
{1}------------------------------------------------
Both assays yield similar Performance Characteristics.
Similarities:
- Both assays are in vitro clinical chemistry methods. .
- Both assays can be used for the quantitative determination of ammonia. .
- Both assays yield similar clinical results. .
## Intended Use:
The Ammonia assay is used for the quantitation of ammonia in human plasma.
#### Performance Characteristics:
Comparative performance studies were conducted using the AEROSET™ System. The Ammonia assay method comparison vielded acceptable correlation with the A-GENT Ammonia assay on the ABBOTT SPECTRUM Series II System. The correlation coefficient = 0.9957, slope = 0.971, and Y-intercept = -2.083 umol/L. Precision studies were conducted using the Ammonia assay. Within-run, between-run, and between-day studies were performed using three levels of control material. The total %CV for Level 1/Panel 103 is 11.7%, 4.0% for Level 2/Panel 104, and 5.5% for Level 3/Panel 105. The Ammonia assay is linear up to 1.574.9 umol/L. The limit of quantitation (sensitivity) of the Ammonia assay is 20.6 umol/L. These data demonstrate that the performance of the Ammonia assay is substantially equivalent to the performance of the A-GENT Ammonia assay on the ABBOTT SPECTRUM Series II System.
#### Conclusion:
The Ammonia assay is substantially equivalent to the A-GENT Ammonia assay on the ABBOTT SPECTRUM Series II System as demonstrated by results obtained in the studies.
Ammonia 510(k) May 29, 1998 Ammf.lwp
Section II Page 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's role in protecting and promoting the health and well-being of the nation.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 22 1998
Anna G. Bentley Director, Regulatory and Quality Affairs EM Science 480 S. Democrat Road Gibbstown, New Jersey 08027
Re : K981926 Ethyl Alcohol Calibrators Requlatory Class: II Product Code: DLJ Dated: June 1, 1998 Received: June 2, 1998
Dear Ms. Bentley:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
0(k) Number (if Known): _ K 98 1926
Device Name: Ethyl Alcohol Clinical Standard Solutions
Indications for Use:
The Ethyl Alcohol Clinical Standard Solutions are intended for use in the calibration and standardization of instruments and methods for the determination of ethyl alcohol in whole blood or serum. The Standard Solutions may also be used for validation of the laboratory's routine standards and/or control materials.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109
OR
Over-The-Counter Use _
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Clinical Laboratory Devices,
510(k) Number k 981426