K Number
K981604

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-06-25

(58 days)

Product Code
Regulation Number
862.1775
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use with the CARESIDE™ Analyzer to quantitatively measure uric acid from whole blood, plasma or serum specimens to aid in the diagnosis and treatment of patients with numerous renal and metabolic disorders including renal failure, gout, leukemia, psoriasis, starvation or wasting conditions, and of patients receiving cytotoxic drugs. It is intended for professional laboratory use: not for point of care or physician office laboratory use.

Device Description

CARESIDE™ Uric Acid cartridges are used with the CARESIDE™ Analyzer to quantitatively measure uric acid concentration in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE™ Uric Acid cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of uric acid concentration. The film cartridge (patent pending) contains all reagents necessary to measure uric acid concentration.

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study information:

CARESIDE™ Uric Acid Device: Acceptance Criteria and Study Details

The provided 510(k) summary and FDA letter do not explicitly list formal "acceptance criteria" in a quantitative sense as might be found in a later clinical trial report (e.g., "sensitivity must be >X%, specificity >Y%"). Instead, the document focuses on demonstrating substantial equivalence to a predicate device, the Vitros URIC Slides for Johnson and Johnson's Vitros DT 60. The "acceptance criteria" are implicitly met by demonstrating comparable clinical performance.

The study described is an equivalence study aiming to show that the CARESIDE™ Uric Acid test is "substantially equivalent in principle, intended use, and clinical performance" to the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

As stated, explicit quantitative acceptance criteria are not provided. The performance is assessed relative to the predicate device.

Acceptance Criteria (Implicit)Reported CARESIDE™ Uric Acid Device Performance (Clinical Performance Relative to Predicate)
Preamble: The CARESIDE™ Uric Acid test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of uric acid on the Vitros DT 60 II.Implied based on FDA clearance: The FDA has determined the device is substantially equivalent to the predicate device (Vitros URIC Slides, K912844/A), indicating that its clinical performance is considered comparable and acceptable for its intended use.
Specific Performance Aspects (Implicit): Accuracy, precision, linearity, and other performance characteristics expected of a uric acid test system, comparable to a legally marketed predicate device. This would include accurate quantification of uric acid concentration in the specified sample types (whole blood, plasma, serum).The document describes the device's mechanism of action (uricase-peroxidase reaction leading to colorimetric measurement at 655 nm, calculated using a lot-specific standard curve). While specific numerical performance values are not given in this summary, the FDA clearance hinges on the submission containing sufficient data to support these aspects relative to the predicate.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for the Test Set: Not explicitly stated in the provided text. The document is a 510(k) summary, which often provides an overview rather than a detailed data breakdown.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This information would typically be found in the detailed study report submitted with the 510(k).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable/Not stated. For a quantitative in vitro diagnostic (IVD) device like a uric acid test, the "ground truth" is typically established by reference methods or comparison to a cleared predicate device, not by expert consensus interpreting images or clinical cases. The comparison would be against a known, accurate uric acid measurement.
  • Qualifications of Experts: Not applicable/Not stated.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable/Not stated. This concept (e.g., 2+1, 3+1) is typically relevant for studies involving human interpretation of data (e.g., radiology reads) where disagreements need resolution. For a quantitative IVD, the "adjudication" is inherent in the comparison of numerical results from the new device against the reference/predicate method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. MRMC studies are specific to evaluating how different human readers perform on multiple cases, often with and without AI assistance, and are common in image-based diagnostics. This device is a quantitative in vitro diagnostic, not an imaging AI diagnostic.
  • Effect Size of Human Readers with/without AI: Not applicable, as it's not an AI-assisted diagnostic for human readers.

6. If a Standalone (Algorithm only without Human-in-the-Loop Performance) was done

  • Standalone Performance: Yes, the study inherently evaluates the standalone performance of the device. As an in vitro diagnostic (IVD) test, the CARESIDE™ Uric Acid system (cartridge + analyzer) operates independently to produce a quantitative uric acid result. Its performance is measured as the accuracy and precision of these direct numerical outputs. The comparison is between the CARESIDE™ Analyzer's results and those of the predicate Vitros DT 60 II. Human intervention is limited to sample loading and initiating the test, not in interpreting the primary measurement.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for this type of IVD device is typically established by:
    • Comparison to a legally marketed predicate device: The text explicitly states "Substantial Equivalency Claim... to the currently marketed Vitros slides for the quantitative measurement of uric acid on the Vitros DT 60 II." This implies that the Vitros system serves as the primary comparator.
    • Reference methods: While not explicitly mentioned in this summary, such studies often involve comparison to established laboratory reference methods for uric acid measurement, which are considered highly accurate.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not explicitly stated in the provided text. For IVDs, "training set" might refer to samples used for initial calibration, method development, or establishing lot-specific standard curves. The document mentions the analyzer uses "the lot-specific standard curve to calculate uric acid concentration," implying some form of calibration or training data is generated for each production lot, but the size is not given.

9. How the Ground Truth for the Training Set was Established

  • How Ground Truth for Training Set was Established: Not explicitly stated. For the purpose of establishing a "lot-specific standard curve" (which functions similarly to a model training for IVDs), ground truth would be established through:
    • Calibrator solutions: Using precisely known concentrations of uric acid, often traceable to international reference materials. These known concentrations are run on the device to generate the standard curve that relates measured signal (reflectance at 655 nm) to uric acid concentration.
    • Internal reference materials: Using well-characterized control materials with established uric acid values.

{0}------------------------------------------------

JUN 2 5 1998

510(K) SUMMARY CARESIDE™ URIC ACID SAFETY AND EFFECTIVENESS

Applicant Information I.

  • A. Applicant Name
  • Applicant/Manufacturer Address B.
  • Telephone Number C.
  • D. Contact Person
  • FAX Number E.
  • e-Mail Address F.
  • Date 510(k) Summary prepared G.

Device Information II.

  • Device Name (Trade) A.
  • B. Device Name (Classification)
  • Device Classification C.

CARESIDE, Inc.

6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 asarchk @ worldnet.att.net April 27, 1998

CARESIDE™ Uric Acid

Uric acid test system Clinical chemistry panel Uric acid test system Regulation Number: 21 CFR 862.1775 Regulatory Class I Classification Number: 75JHB None applicable

  • Special controls and D. performance standards

III. Substantial Equivalence Claim

  • General equivalency claim A.
    The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.

Uric acid in vitro diagnostic products, in both dry film and other formats, are already on the U.S. market, including uric acid products which utilize uricase catalyzed generation of hydrogen peroxide which reacts with chromogens in a peroxidase catalyzed reaction to form a blue dye.

B. Specific equivalency claim

This CARESIDE™ Uric Acid test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of uric acid on the Vitros DT 60 II.

Name of Predicate Device:Johnson and Johnson's (formerly Eastman Kodak, Inc.) Vitros URIC Slides for Johnson and Johnson's Vitros DT 60 (formerly Eastman Kodak's DT 60 II).
Predicate Device 510K number:K912844/A
Product Code:75JHB

{1}------------------------------------------------

IV. Device Description

CARESIDE™ Uric Acid cartridges are used with the CARESIDE™ Analyzer to quantitatively measure uric acid concentration in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE™ Uric Acid cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of uric acid concentration. The film cartridge (patent pending) contains all reagents necessary to measure uric acid concentration.

A. Explanation of Device Function

Each CARESIDE™ Uric Acid cartridge consists of a uric acid-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the whole blood, serum, or plasma specimen into the cartridge sample deposition well, closes the lid and inserts the cartridge into the CARESIDE™ Analyzer. Once loaded, the CARESIDE™ analyzer scans the cartridge barcode, spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers, and brings the cartridge and the contained specimen to 37℃. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately ten microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well. The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer reagent film. The spreading layer distributes the uric acid containing specimen uniformly and removes large molecular weight components, such as protein, as the specimen moves into the underlying reaction layer. Uric acid is hydrolyzed by uricase to generate hydrogen peroxide which in turn oxidizes the leuco dye in a peroxidase catalyzed reaction. The color intensity of the resulting blue dye, as measured by the amount of reflected light at 655 nanometers, is directly related to the specimen uric acid concentration.

Test Reaction Sequence:

Uric Acid + 2 H20 + O2 - Unicase > Allantoin + H2O2+ CO2 Peroxidase H2O2 + Leuco Dye Blue Dye + 2 H20

As the cartridges spin, photodiodes measure reflectance of light emitted by wavelengthspecific light emitting diodes (LEDs) at a fixed time. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate uric acid concentration.

B. Test Summary

Uric acid is the major product of the catabolism of purine nucleosides, adenosine and guanosine. Purines from catabolism of dietary nucleic acid are converted to uric acid directly. Approximately 75 percent of uric acid excreted is lost in the urine. Most of the remainder is secreted into the gastrointestinal tract, where it is degraded to allantoin and other compounds by bacterial enzymes'. Hyperuricemia is caused when serum or plasma uric acid concentration levels rise above 7 mg/dL in men and 6 mg/dL in women's Disorders of hyperuricemia can be divided into those due to increased intake, decreased intake excretion, and increased production". Hyperuricemia can result from uric acid overproduction or under excretion, renal retention due to renal failure, drug or chemical toxicity, hypothyroidism, hyperparthyroidism, increased turnover of nucleic acid due to myeloproliferative syndromes or cancer chemotherapy, or specific enzyme deficiencies such as hypoxanthine-guanine phosphoribosyl transferase or phosphoribosyl pyrophosphate synthetase.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 25 1998

Kenneth B. Asarch, Ph.D. . VP Quality Systems and Regulatory Affairs CareSide Inc. 6100 Bristol Parkway Culver City, California 90230

Re: K981604 CareSide™ Uric Acid Regulatory Class: I Product Code: JHB Dated: April 27, 1998 Received: April 28, 1998

Dear Dr. Asarch:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE VI.

510(k) Number: K981604

CARESIDE™ Uric Acid Device Name:

For in vitro diagnostic use with the CARESIDE™ Analyzer to Indications for use: quantitatively measure uric acid from whole blood, plasma or serum specimens to aid in the diagnosis and treatment of patients with numerous renal and metabolic disorders including renal failure, gout, leukemia, psoriasis, starvation or wasting conditions, and of patients receiving cytotoxic drugs. It is intended for professional laboratory use: not for point of care or physician office laboratory use.

Division Sign-Off) Division of Clinical Laboratory Devices
510(k) NumberK981604

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

§ 862.1775 Uric acid test system.

(a)
Identification. A uric acid test system is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.(b)
Classification. Class I (general controls). The device, when it is solely intended for use as an acid reduction of ferric ion test, a phosphotungstate reduction test, a gasometric uricase test, an ultraviolet uricase test, or an oxygen rate uricase test, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.