ONE STEP URINE DRUG OF ABUSE: PHENCYCLIDINE (PCP) TEST

K981363 · Tcpi, Inc. · LCM · Dec 22, 1998 · TX

Device Facts

Record IDK981363
Device NameONE STEP URINE DRUG OF ABUSE: PHENCYCLIDINE (PCP) TEST
ApplicantTcpi, Inc.
Product CodeLCM · TX
Decision DateDec 22, 1998
DecisionSESE
Submission TypeTraditional
Device ClassClass U

Indications for Use

TCPI's One Step™ Urine Drug of Abuse: Phencyclidine (PCP) Test is a rapid, qualitative, competitive binding immunoassay for the determination of Phencyclidine (PCP) in urine at the cutoff level of 25 ng/ml. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. TCPI's One Step™ Urine Drug of Abuse: Phencyclidine (PCP) Test is not intended to monitor drug levels, but only to screen urines for the presence of Phencyclidine (PCP) and its metabolites.

Device Story

Device is a chromatographic absorbent test strip for qualitative screening of Phencyclidine (PCP) in urine. Principle of operation: competitive binding immunoassay; drug/metabolites in sample compete with immobilized drug conjugate on porous membrane for limited antibody binding sites. Input: urine sample. Process: sample flows up strip; labeled antibody-dye conjugate binds to free drug forming antibody-antigen complex. Output: visual color bands. Negative result: two magenta bands (Test and Control zones). Positive result: one magenta band (Control zone only). Used in professional medical/forensic laboratories; operated by laboratory personnel. Results are preliminary; require confirmation by GC/MS. Clinical utility: rapid screening for presence of PCP; not for monitoring drug levels. Benefits: provides quick preliminary data to guide clinical decision-making and identify need for confirmatory testing.

Clinical Evidence

Bench testing and clinical trial conducted. In-house testing on 227 samples; clinical trial on 286 samples. Total 513 samples compared against Emit® II screening method (cutoff 25 ng/ml). Results: 100% accuracy (513/513), 100% relative sensitivity (223/223), 100% relative specificity (290/290). All 223 positive samples confirmed by GC/MS. Two false positives reported by both the subject device and Emit® II, confirmed by GC/MS as containing 21 and 23 ng/ml of PCP.

Technological Characteristics

Chromatographic absorbent test strip; porous membrane support; competitive binding immunoassay format. Detects PCP at 25 ng/ml cutoff. Visual readout via magenta color bands. No electronic components, energy source, or software.

Indications for Use

Indicated for the qualitative screening of Phencyclidine (PCP) and its metabolites in human urine at a cutoff concentration of 25 ng/ml in professional laboratory settings.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K981363: One Step Urine Drug of Abuse: Phencyclidine™ (PCP) Test Technical Chemicals and Products, Inc. (TCPD) Page 82 of 82 ### Revision B - 12/14/98 # Summary of Safety and Effectiveness The Sponsor, TCPI, Inc. (3341 S.W. 15th Street, Pompano Beach, FL 33069), has developed, manufactured, and tested under GMP/GLP guidelines a device for the qualitative testing of urine for the presence of Phencyclidine (PCP) and its metabolites in a screening format. The trade name of the device is One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test, having a designated common name of Phencyclidine (PCP) Test System and a classification as a Class II device per FDA .. TCPI's One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test Strip is a rapid, qualitative, competitive binding immunoassay for the determination of Phencyclidine (PCP) in urine. The One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test Strip is intended to be used in professional laboratories (medical & forensic) where preliminary screening is essential. The test provides only preliminary data which should be confirmed by other methods in particular, preliminary positive results must be substantiated with an approved confirmatory method, such as gas chromatography/mass spectrometry (GC/MS) Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are positive. This test is not intended to be used for monitoring drug levels, but only to screen urines for the presence of Phencyclidine (PCP) TCPI's One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test Strip is a chromatographic absorbent device in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for a limited number of antibody binding sites. As a test sample flows up through the absorbent device, labeled antibody-dye conjugate binds to free drug in the specimen, forming an antibody-antigen complex. This complex with immobilized antigen conjugate in the Test Zone of the strip, and will not produce a magenta color band when the drug concentration in the specimen is above the detection level of 25 ng/ml. Unbound dye conjugate birds to the reagent in the Control Zone of the strip, producing a magenta color band, demonstrating that the reagents and device are functioning correctly. A negative specimen produces two distinct color bands, one in the Test Zone and one in the Control Zone A positive specimen produces only one color band in the Control Zone. TCPI's One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test Strip was tested in-house on 227 individual urine samples, and was subsequently tested in a clinical trial on 286 individual urine samples which were submitted to a NDA certified laboratory. In both studies, the laboratories used Emit® II Phencyclidine (PCP) as their screening procedure, with a cutoff of 25 ng/ml. The accuracy of the One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test Strip was 513/513 or 100%, when compared to Emit® II. Compared to Emit® II, the relative sensitivity or agreement between positive samples was 223/223 or 100%, and the relative specificity or agreement between negative samples was 290/290 or 100%. All 223 samples which tested positive by either screening method (Emit® II or One Step Urine Drug of Abuse Phencyclidine™ (PCP) Test Strip) were confirmed by GC/MS. When compared to GC/MS, the Emit® II screening method reported two (2) false positives and the One Step strip reported two (2) false positives GC/MS results for these samples indicated that they contained phencyclidine levels of 21 and 23 ng/ml. Additional information on this submission may be obtained by contacting Dr. Jeffrey Bolts, Technical Chemicals and Products, Inc., at 954-979-0400 (FL) or by fax at 954-979-0009. {1}------------------------------------------------ Public Health Service Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the logo. DEC 22 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Dr. Jeffrey Bolts Director of Quality Assurance Technical Chemicals & Products, Inc. P.O. Box 8726 Ft. Lauderdale, FL 33310 Re: K981363 Trade Name: One Step Urine Drug of Abuse: Phencyclidine (PCP) Test Regulatory Class: II Product Code: LCM Dated: November 20, 1998 Received: November 27, 1998 Dear Dr. Bolts: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections. the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ #### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # 510(k) Number (if Known): Not Yet Issued Device Name: One Step™ Urine Drug of Abuse: Phencyclidine (PCP) Test ## Indications For Use: # INTENDED USE TCPI's One Step™ Urine Drug of Abuse: Phencyclidine (PCP) Test is a rapid, qualitative, competitive binding immunoassay for the determination of Phencyclidine (PCP) in urine at the cutoff level of 25 ng/ml. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. TCPI's One Step™ Urine Drug of Abuse: Phencyclidine (PCP) Test is not intended to monitor drug levels, but only to screen urines for the presence of Phencyclidine (PCP) and its metabolites. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANDOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) or Ston L. Leach (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K981363 **Prescription Use:** (Per 21 CFR 801.109 Over The Counter Use: (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...