K Number
K974038

Validate with FDA (Live)

Date Cleared
1998-05-11

(199 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of this pump is general suction. This is best described under 21 CFR 878.4780, which states that the pump is intended to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter

Device Description

" The Aspirator " is a suction pump with a 350mm Hg capacity. It is of low maintenance by design. It provides low noise level. It has an oversized vacuum gauge to allow easy reading. It has a detachable power cord with externally replaceable fuses, and positive detented vacuum lines for fewer leaks. It is available in both 110 and 230 volts. Only UL approved parts are used in the construction.

AI/ML Overview

The provided text is a 510(k) summary for "The Aspirator" general aspiration pump, claiming substantial equivalence to a predicate device. This type of submission relies on comparative data rather than a detailed study with acceptance criteria and performance metrics typically seen for novel medical devices or AI-powered systems.

Therefore, many of the requested points cannot be found in the provided text, as this medical device (a general aspiration pump) does not involve AI, machine learning, or complex algorithmic performance that would necessitate acceptance criteria, ground truth determination, or MRMC studies in the way an AI diagnostic tool would.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

This information is not explicitly stated in the provided 510(k) summary in the format of specific, quantifiable acceptance criteria. Instead, the submission relies on demonstrating substantial equivalence to a predicate device, meaning its performance is considered acceptable if it is comparable to a legally marketed device.

The table below summarizes the comparison with the predicate device, which serves as the implicit "performance" comparison. The "acceptance criteria" are effectively that the new device's characteristics are "Equivalent" or "Similar" to the predicate.

Characteristic"The Aspirator" Performance (compared to Cook Aspiration Pump)Implicit Acceptance Criteria (based on substantial equivalence)
Indications for UseE (Equivalent)Must be Equivalent to predicate
Target PopulationE (Equivalent)Must be Equivalent to predicate
DesignE (Equivalent)Must be Equivalent to predicate
MaterialsS (Similar)Must be Similar or Equivalent to predicate
PerformanceS (Similar)Must be Similar or Equivalent to predicate
SterilityS (Similar)Must be Similar or Equivalent to predicate
BiocompatibilityN/A (Not Applicable)Not applicable (implied by predicate)
Mechanical SafetyS (Similar)Must be Similar or Equivalent to predicate
Chemical SafetyN/A (Not Applicable)Not applicable (implied by predicate)
Anatomical SitesE (Equivalent)Must be Equivalent to predicate
Human FactorsS (Similar)Must be Similar or Equivalent to predicate
Energy used and/or deliveredE (Equivalent)Must be Equivalent to predicate
Compatibility w/ environment & other devicesE (Equivalent)Must be Equivalent to predicate
Where usedE (Equivalent)Must be Equivalent to predicate
Standards metS (Similar)Must be Similar or Equivalent to predicate
Electrical SafetyS (Similar)Must be Similar or Equivalent to predicate
Thermal SafetyS (Similar)Must be Similar or Equivalent to predicate
Radiation SafetyS (Similar)Must be Similar or Equivalent to predicate

The device further states:

  • "The Aspirator " is safe and effective as other predicate devices cited above.
  • It is made out of only UL approved parts.
  • It has a 350mm Hg capacity.
  • Provides low noise level.
  • Has an oversized vacuum gauge.
  • Detachable power cord with externally replaceable fuses.
  • Positive detented vacuum lines for fewer leaks.
  • Available in both 110 and 230 volts.

2. Sample size used for the test set and the data provenance

Not applicable/Not provided. This submission is a 510(k) for a physical medical device (a pump), not an AI/ML algorithm. There isn't a "test set" of data in the sense of patient cases to evaluate algorithm performance. The "study" here is primarily a comparison of technological characteristics to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable/Not provided. As stated above, there is no "test set" and thus no ground truth established by experts for algorithmic evaluation.

4. Adjudication method for the test set

Not applicable/Not provided.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC study was not done. This device is a general aspiration pump, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable/Not provided. There is no standalone algorithm.

7. The type of ground truth used

Not applicable/Not provided. Performance is based on similarity to a legally marketed predicate device, rather than a "ground truth" derived from patient outcomes, pathology, or expert consensus on clinical cases. The ground truth for this type of device is implicitly "functions as a general aspiration pump with similar characteristics to existing approved devices."

8. The sample size for the training set

Not applicable/Not provided. There is no training set as it's not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable/Not provided.

{0}------------------------------------------------

510(k) Summary MAY | | | 1998 As Require 4 by 21 section 807.92 (c)

K974038

1-Submitter Name:A & A Medical, Inc.
2-Address:4100 Nine McFarland Drive, suite BAlpharetta, GA 30004
3-Phone:(770) 343- 8400
4-Fax:(770) 343- 8985

5-Contact Person: Jihad Mansour

6-Date summary prepared: October 24th, 1997

7-Device Trade or Proprietary Name: "The Aspirator "

8-Device Common or usual name: General Aspiration Pump

9-Device Classification Name: Powered Suction Pump

10-Substantial Equivalency is claimed against the following device:

Substantially equivalent product has been found as approved while searching the

FDA web site under product code JCX

Our pump is substantially equivalent to

*K930900 Cook Aspiration Pump from Cook Ob/Gyn It is classified under 79JCX: "Powered Suction Pump"

"The Aspirator" is a preammendment medical device introduced into the U.S interstate commerce prior to 28 May, 1976. It was and is still advertised/labeled rather as an "endometrial aspirator" that follows the product classification of (85HFF) CLASS III device

11-Description of the Device:

" The Aspirator " is a suction pump with a 350mm Hg capacity. It is of low maintenance by design. It provides low noise level. It has an oversized vacuum gauge to allow easy reading. It has a detachable power cord with externally replaceable fuses, and positive detented vacuum lines for fewer leaks. It is available in both 110 and 230 volts. Only UL approved parts are used in the construction.

Recommended accessories are listed below:

Foot Pedal, Silicone Tubing 6 ft, Spare Luer Lock, Metal Tubing Adapter, Spare Glass Reservoir. Neoprene Stopper, and Angled Tube Mounts.

This pump is intended to be used as a general purpose suction pump.

12-Intended use of the device: (Also typed on a separate FDA form as required)

The intended use of this pump is general suction. This is best described under 21 CFR 878,4780, which states that the pump is intended to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter

{1}------------------------------------------------

13-Safety and Effectiveness of the device:

« The Aspirator " is safe and effective as other predicate devices cited above. It is made out of only UL approved parts. It is made our of only OE approved be parison with the other predicate devices 1 ms is expreboom 14 below ) and also documented in the Appendix

14-Summary comparing technological characteristics with other predicate devices:

Comparison should be done with other predicate devices having ONLY same intended use.

use.
Accordingly, a tabulated comparison with Cook Aspiration Pump is found below. Also, Equivalency overview chart path is attached

Image /page/1/Figure/5 description: The image provides a list of abbreviations used in the document. The abbreviations are defined as follows: E=Equivalent, S=Similar, D=Different, N/A= Not Applicable, DES=Description available, N/I=No Information available, 510(k) Sum=510(k)Summary available, 510(k)=510(k) available, web=fda web printout enclosed. The abbreviations are used to provide a concise way to refer to the different terms.

FDA file reference numberAttachments inside NotificationIndications for UseTarget PopulationDesignMaterialsPerformanceSterilityBiocompatibilityMechanical SafetyChemical SafetyAnatomical SitesHuman FactorsEnergy used and/or deliveredCompatibility w/ environment & other devicesWhere usedStandards metElectrical SafetyThermal SafetyRadiation Safety
Cook Aspiration Pumpfrom Cook Ob/GynK930900web 510k approvalEEESSSN/ASN/AESEEESSSS

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the three branches of government. The eagle is surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

MAY | | 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jihad Mansour Quality and Requlatory Manager A&A Medical, Incorporated 4100 Nine McFarland Drive, Suite B Alpharetta, Georgia 30004-3386

Re: K974038 Trade Name: The Aspirator, 6FT Silicone Tubing, Metal Tubing Adapter, Glass Requlatory Class: II Product Code: JCX Dated: April 1, 1998 Received: April 16, 1998

Dear Mr. Mansour:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) ; it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਚੇ substantially equivalent determination assumes compliance with… the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{3}------------------------------------------------

Page 2 - Mr. Mansour

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Whitten, M.D.

M. Witten, Ph.D., M.D. Celia Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Dy------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Donaﺏ ﺍﻟ1
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

510(k) Number (if known):

The Aspirator Device Name:_

Indications For Use:

The intended use of this pump is general suction. This is best described under 21 CFR 878.4780, which states that the pump is intended to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

..............................................................................................................................................................................

ાવ

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Page 7 of 14

P. Collyer

(Division Sign-Off) Division of General Restorative Device 510(k) Number

.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.