K973915 · Kci New Technologies, Inc. · IZH · Nov 10, 1997 · Radiology
Device Facts
Record ID
K973915
Device Name
INFLATABLE POSITIONING CUSHIONS
Applicant
Kci New Technologies, Inc.
Product Code
IZH · Radiology
Decision Date
Nov 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1710
Device Class
Class 2
Indications for Use
The Inflatable Positioning Cushion device is intended to be used on nuclear imaging tables for Scintimammography procedures. The Inflatable Positioning Cushion allows for comfortable, anatomically correct, prone positioning of the patient during Scintimammography.
Device Story
Inflatable Positioning Cushion designed for use on standard nuclear imaging tables (e.g., SPECT tables). Device consists of head, abdominal, and lower body sections containing foam and inflatable cushions. Secured to table via Velcro and inflatable straps; external blower unit provides inflation. Used in clinical nuclear medicine settings to facilitate comfortable, anatomically correct prone patient positioning during Scintimammography. By stabilizing the patient and providing proper anatomical alignment, the device helps maintain position, potentially reducing motion artifacts and improving image quality. Operated by clinical staff.
Clinical Evidence
No clinical data provided for this specific device. The document references a clinical study conducted by the University of Montreal regarding the sensitivity and specificity of 99mTc-sestamibi scintimammography, but this is cited in relation to a separate Scintimammography Prone Breast Cushion product, not the subject Inflatable Positioning Cushions.
Technological Characteristics
Materials include foam cushions, inflatable cushions, and nylon/vinyl covers. Features Velcro and inflatable straps for table attachment. Includes an external blower unit for inflation. Designed for standard nuclear imaging tables (e.g., 80" x 14" cradle). Non-electronic, mechanical positioning support.
Indications for Use
Indicated for patients undergoing Scintimammography procedures requiring prone positioning on nuclear imaging tables.
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
General Purpose T-Foam Pads
Scintimammography Prone Breast Cushion
Scintimammography Table Overlay
Scinti Mammography Positioning Pads
Related Devices
K980235 — REMOTE HAND CONTROL TO THE INFLATABLE POSITIONING CUSHION (A.K.A. IMAGEAIR) · Kci New Technologies, Inc. · Apr 9, 1998
K101373 — EMBRACE · Izi Medical Products, Inc. · Sep 15, 2010
K211001 — C-Qual M Breastboard; Monarch Overhead Arm Positioner · Medtec, Inc. Dba Civco Medical Solutions and Civco · May 20, 2021
K080072 — MRI PATIENT POSITIONING DEVICES · CIVCO Medical Instruments Co., Inc. · Jun 4, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
### 510 (k) Summary
NOV 1 0 1997
| I. | Name of Device: | Inflatable Positioning Cushions |
|------|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| II. | Classification Name: | Patient supports, accessories to<br>Mammographic x-ray system. |
| III. | Substantial Equivalence: | General Purpose T-Foam Pads<br>Scintimammography Prone Breast Cushion<br>Scintimammography Table Overlay<br>Scinti Mammography Positioning Pads |
#### Device Description: IV.
The Inflatable Positioning Cushion is designed to be used on a standard nuclear imaging table with cradle dimensions of 80" long by 14" wide, such as the SPECT Table manufactured by Medical Imaging Electronics. The Inflatable Positioning Cushion is comprised of a Head Section and Abdominal Section, containing foam cushions and inflatable cushions, and a Lower Body Section of foam cushions. The cushions are secured to the nuclear imaging table with Velcro straps and inflatable straps that wrap underneath the table. The blower for the inflatable cushions is in a separate enclosure that can hang at the end of the nuclear imaging table or sit on the floor.
#### V. Indications
The Inflatable Positioning Cushion device is intended to be used on nuclear imaging tables for Scintimammography procedures. The Inflatable Positioning Cushion device allows for comfortable, anatomically correct, prone positioning of the patient during Scintimammography.
#### VI. Comparative Information
See attached Promotional Material.
{1}------------------------------------------------
# AliMed Diagnostic Imaging Products
Image /page/1/Picture/1 description: The image shows the text "20 new replacement table pads!" in white text on a black background. The text is arranged in two lines, with "20 new replacement" on the first line and "table pads!" on the second line. The font is bold and sans-serif, and the text is centered in the image.
- · Substantially reduces lower back pain during long procedures!
- Helps reduce patient movement—you work faster!
- Radiolucent and artifact free
20am represents a unique breakthrough in pressure distribution! Immobilized patients often experience lower back discomfort and painful pressure during long procedures, such as cardiac cath, angiographies and hypertensive IVPs. The lumbar region is not supported well by ordinary table pads. Patients react by moving to relieve the pain and pressure-blurring images and complicating procedures. T-Foam precisely molds to each patient's contours to provide a total contact resting surface-eliminating lower back discomfort and painful pressure. Patient comfort soars. Procedures are completed quickly and efficiently.
Image /page/1/Picture/6 description: The image shows an advertisement for T-Foam. The text in the advertisement says "T-Foam fits every need in Imaging". The background of the advertisement is dark and blurry, but it appears to be a medical image.
Image /page/1/Picture/7 description: The image shows a comparison of pressure distribution with and without a T-Foam Pad. The top half of the image shows dangerous high-pressure peaks, while the bottom half shows safe, distributed pressure with the T-Foam Pad. The text indicates that the T-Foam Pad helps to reduce peak pressure safely.
T-Foam pads are radiolucent. They are 11/2"thick and are covered wi. blue nylon cover with water resistant seams.
Image /page/1/Figure/9 description: This image shows a variety of general-purpose T-foam pads with different dimensions and prices. The pads are labeled with codes such as #E9-213, #E9-211, #E9-210, #E91-370, #E9-677, #E9-546, #E9-545, #E920177, #E920176, #E920161, #E920162, #E920159, and #E920158. The prices range from \$299.95 to \$644.0, and the dimensions vary, including lengths like 77", 87", 75", 95", 101", 118", 79", 82", 94", and 98".
{2}------------------------------------------------
Image /page/2/Figure/0 description: This image shows a collection of medical equipment parts with their corresponding names, ID numbers, dimensions, and prices. The parts include models from Phillips, Picker, Siemens, and Toshiba, such as the Nuclear Medicine Spect, EP 94, 1200X CT, and Somatom Plus. Prices range from \$183.00 for a Toshiba TCT 500S to \$642.60 for an XRE-ADC, with dimensions varying for each part.
# lf you don't see your pad above, we'll make one to your specifications!
### stom Table Pad Order Form
Custom Table Pad Order Form
Fill in the information below and provide either a dimensional sketch on the graph form at right or a full-scale tracing of the original pad. Include location and description of any Velcro
<sup>®</sup>
fasteners or zippers desired.
| MAIL TO: | OR FAX TO: (617) 329-8392 |
|------------------------------------------------------------------------------------|---------------------------|
| Custom Table Pad Department<br>AliMed, inc.<br>297 High Street<br>Dedham, MA 02026 | Each square=2" |
| Name | | Title | |
|------------------|--|------------|--|
| Facility | | Customer # | |
| Bill to Address: | | | |
| <input type="checkbox"/> Blue Nylon (Radiolucent) | | | | | |
|-------------------------------------------------------------------|----------------------------------------------------------|-------|--|--------|--|
| <input type="checkbox"/> T-Foam <input type="checkbox"/> Polyfoam | | | | | |
| Length | | Width | | Height | |
| Water Resistant Seams | <input type="checkbox"/> Yes <input type="checkbox"/> No | | | | |
{3}------------------------------------------------
NUCLEAR ASSOCIATES
New!
## SCINTIMAMMOGRAPHY PRONE BREAST CUSHION
Now you can effortlessly and affordably perform scintimammography by using Nuclear Associates' Scintimammography Prone Breast Cushion and the FREE "Guide to Scintimammography."
The Scintimammography Prone Breast Cushion is an Innovative Product That:
- Provides optimal separation of the breast from the chest wall, making it easy to get THE BEST RESULTS!
- · Allows lateral breast compression for larger imaging surfaces.
- · Is free of attenuating materials, unlike expensive, bulky imaging tables.
- · Allows you to image a full range of breast sizes-from an A cup to a double D.
- · Is made of top-quality, medical-grade materials.
- · Features an impermeable surface that is easy to clean.
- · Can be used with any camera system; it's universal.
- · Comes with an invaluable, FREE "GUIDE TO SCINTIMAMMOGRAPHY"!
Was clinically proven in a study conducted by Departments of Nuclear Medicine, Surgery and Radiology, Hotel-Dieu de Montreal, University of Montreal, Canada, to evaluate the sensitivity and the specificity of 99mTc-sestamibi scintimammography in the detection of breast cancer and axillary lymph node involvement. (See next page for FREE reprint information.)
Image /page/3/Picture/15 description: The image shows a rectangular box divided into three sections. The bottom section contains the word "VICTOREEN" in bold, sans-serif font. The top two sections are blank.
DO VOICE BOAD . PO BOY 349 CARLE PLACE, NY 11514-0349 US (516) 741-6360 · FAX (516) 741 ail: NclrAssoc@aol.cor
Image /page/3/Picture/17 description: The image shows a logo for Visa. The logo is a rectangle with the word "VISA" in white letters on a blue background. There are two horizontal lines, one above and one below the word "VISA", that are black. The logo is surrounded by a white border.
8:00 AM to 5:00 PM Monday thru Friday,
FAX (24 hours) (516) 741-5414,
E-Mail: NclrAssoc@aol.com or write.
{4}------------------------------------------------
Image /page/4/Picture/3 description: The image shows a black and white photograph of an object that appears to be made of two rectangular blocks connected by a narrower section. The top and bottom blocks are wider than the middle section, creating a distinct shape. The object is placed on a surface, and the background is dark, providing contrast to the lighter object.
### SCINTIMAMMOGRAPHY TABLE OVERLAY
- · Simplifies patient positioning for Scintimammography imaging
- · Converts standard SPECT imaging tables to Scintimammography tables
- · Also available for standard flat bed planar imaging tables
- · Locking guide rails facilitate fast setup and removal of table overlay
The Scintimammography Table Overlay is designed to improve the guality of providing by providing a stable, more comfortable platform for your patient during a scintimammography procedure. Prone imaging allows the breast to be positioned from the chess wall, which can greatly increase the target-to-background ration. Your scanning table can now be fitted with the Scintimammography Table Overlay to provide the necessary indentations for prone imaging.
The Scintimating graphy Table Overlay is fabricated light a ½" thick polymer plastic material. The normal overall dimensions are 81" x 24" but will vary according to the size of your inaging table. Each table overlay is built to the specifications of your existing intelle using locking guide rails. The svstem includes a 5/8" foam core heavy-duty vinyl pad Please specify the model of your camera and table with your order.
JRT-STO/S Scintimammography Table Overlay for Spect Tables 5 1,330 Scintimammography Table Overlay for Planar Tables JRT-STO/P S 1.430
### JRT Associates
124 Saw Mill River Road Elmsford, NY 10523 800-221-0111 FAX 914-592-3167
{5}------------------------------------------------
# Scinti Mammography Positioning Pads
Image /page/5/Picture/1 description: The image shows a collection of various 3D shapes and objects. There are several rectangular prisms of different sizes, along with a few curved shapes that resemble ribbons or decorative elements. Additionally, there is a long, rectangular object that appears to be broken or segmented into two parts, and a more complex, irregular shape that could be a decorative molding or architectural detail.
Scinti Mammographv Positioning Pads.
Part #PBI-3 849900
### · Universal Imaging System Application · Comfortable, Anatomically Correct Positioning · · Durable Seamless Foam Construction ·
ie specialized design of the positioning pads allow comfortable, anatomically correct, prone patient positioning on virtually any imaging system. The unique design of the pads makes right, left, or bilateral breast image acquisition easy.
Compatible with flat or concave imaging tables, the PBI-3 pad set comes with all the accessories needed to facilitate Scinti Mammography on your imaging system.
- KIT INCLUDES: 1 Upper Body Pad 2 Breast Compression Pads 1 Lower Body Pad 1 Ankle/Foot Rest Pad 1 Roll of Patient Safety Strap 4 Velcro Brand Mounting Strips
Call today for complete details or a catalog of our entire product line.
Image /page/5/Picture/9 description: The image is a black and white logo with a star and the letters "PTI". The star is a five-pointed star and is located at the top of the logo. The letters "PTI" are located below the star. The background of the logo is black, and the star and letters are white. The logo is surrounded by a black border.
Technology. Inc.
P.O. Box 824, Greenville, PA 16125
Your complete source for Nuclear Medicine Supplies and Accessories.
| Toll Free: | 800-682-2226 |
|------------|------------------|
| Phone: | 412-932-2121 |
| Fax: | 412-932-3176 |
| Email: | pti@nauticom.net |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters.
Re:
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 10 1997
William H. Quirk Director. Regulatory Affairs KCI New Technologies Inc. 3440 East Houston San Antonio, TX 78219
K973915 Inflatable Positioning Cushions Dated: October 9, 1997 Received: October 14, 1997 Regulatory class: II 21 CFR 892.1710/Procode: 90 IZH
Dear Mr. Quirk:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce. prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours
William Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
Page
510(k) Number (if known):
Device Name: Inflatable Positioning Cushions__________________________________________________________________________________________________________________________________
Indications for Use:
The Inflatable Positioning Cushion device is intended to be used on nuclear imaging tables for Scintimammography procedures. The Inflatable Positioning Cushion allows for comfortable, anatomically correct, prone positioning of the patient during Scintimammography.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Timith le. Sygm
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K973915
Prescription Use / (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.