VIRGO PANCO KIT

K973824 · Hemagen Diagnostics, Inc. · MOB · Nov 13, 1997 · Immunology

Device Facts

Record IDK973824
Device NameVIRGO PANCO KIT
ApplicantHemagen Diagnostics, Inc.
Product CodeMOB · Immunology
Decision DateNov 13, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Intended Use

This enzyme-linked immunosorbent assay (ELISA) is indicated for the detection of altoantilian to the antigen Myeloperoxidase (MPO) in human serum. The presence of MPO antibodies, in combination with dinical observations and other serological tests, can aid in the dlagnosis of polyarterits, necrotizing glomerulonephritis and other conditions associated with elevated anti-neutrophil cytoplasmic antibodies (ANCA).

Device Story

VIRGO pANCA Kit is an ELISA for detecting autoantibodies to myeloperoxidase (MPO) in human serum. Input: patient serum sample; process: serum antibodies bind to MPO-coated microwells; secondary HRP-conjugated antibody binds patient antibodies; TMB substrate added; HRP catalyzes color change; output: optical density (OD) measured by EIA plate reader. Used in clinical laboratories by technicians. Results aid physicians in diagnosing autoimmune-mediated vasculitides, polyarteritis, and necrotizing glomerulonephritis. Benefits: provides objective serological evidence to support clinical diagnosis.

Clinical Evidence

Bench testing only. Comparative study of 132 serum specimens (22 pANCA positive, 40 glomerulonephritis, 70 normal) against predicate device. Results: 100% relative sensitivity (61/61 positive samples) and 100% relative specificity (62/62 normal samples, with 8 discrepant samples confirmed negative by third-party IFA). Precision studies (inter-assay and intra-assay) showed acceptable CVs. Interference testing (lipemic, icteric, hemolytic) showed <15% variation.

Technological Characteristics

ELISA-based immunoassay. Components: MPO-coated microwell plates, HRP-conjugated secondary antibody, TMB substrate. Sensing principle: colorimetric optical density measurement. Connectivity: standalone (requires external EIA plate reader). Sterilization: N/A (in vitro diagnostic kit).

Indications for Use

Indicated for detection of MPO autoantibodies in human serum to aid in diagnosis of polyarteritis, necrotizing glomerulonephritis, and other conditions associated with elevated ANCA in patients presenting with clinical symptoms of autoimmune mediated vasculitides.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K973824 ## 510(k) Summary # NOV 1 3 1997 ### Submitter's Name/Contact Person 1. Joseph M. Califano Manager, Regulatory Affairs ## Address Hemagen Diagnostics, Inc. 34-40 Bear Hill Road Waltham, MA, 02154 (781) 890-3766 Phone: (781) 890-3748 Fax: jcalifano@hemagen.com email: ## Date Prepared 30 September 1997 ### 2. Device Name | Trade Name: | VIRGO ® pANCA Kit (EIA method) | |----------------------|---------------------------------------------------| | Common Name: | MPO Antibody Kit | | Classification Name: | Antineutrophil Cytoplasmic Antibodies test system | #### 3. Predicate Device Scimedx MPO Antibody EIA Kit {510 (k) Docket No. K 954062} {1}------------------------------------------------ ### Description of Device 3. An enzyme-linked immunosorbent assay (ELISA) designed for the detection and measurement of autoantibodies to the antigen myeloperoxidase in human serum. The ELISA methodology is commonly used for serum antibody evaluations. Purified MPO antigen has been attached to the inner surfaces of the microwell plate. During the initial incubation step, antibodies in patient serum bind specifically to the immobilized antigen and remain in place after a wash step. A second antibody which is conjugated to horseradish peroxidase (HRP) is used to recognize the "heavy + light" chain regions of the patient's antibodies remaining after the wash step. In the wells where the second antibody remains bound, the conjugated HRP catalyzes a color change in the substrate, tetramethyl benzidine (TMB). After the reaction is stopped, the color is read in an EIA Plate reader. ## 4. Intended Use of Device An enzyme-linked immunosorbent assay (ELISA) designed for the detection and measurement of autoantibodies to the antigen myeloperoxidase in human serum. The test is intended as an aid in the diagnosis of current or past autoimmune mediated vasculitides. ## 5.(A) Technological Characteristics ## Proposed Device The VIRGO ® pANCA Kit is an enzyme-linked immunosorbent assay. The device utilizes optical density as a measure of antibody presence, with an established cutoff between a positive and a negative reaction. ## Predicate Device The Scimedx MPO ANTIBODY EIA is also an enzyme-linked immunosorbent assay. The device utilizes optical density as a measure of antibody presence, with an established cutoff between a positive and a negative reaction. {2}------------------------------------------------ ## 5.(B) Performance Data ## Precision To evaluate precision, inter-assay and intra-assay studies were conducted. ## A. Inter-assay Seven samples were assayed in duplicate twice a day for five different days. | Sample | Mean OD | Std. Dev. | % CV | Mean Units | Std. Dev | % CV | |--------|---------|-----------|------|------------|----------|------| | 1 | 3.338 | 0.142 | 4.2 | 12.6 | 0.828 | 6.6 | | 2 | 2.333 | 0.142 | 6.1 | 8.6 | 0.542 | 6.3 | | 3 | 1.376 | 0.111 | 8.1 | 5.1 | 0.370 | 7.3 | | 4 | 0.712 | 0.071 | 10.0 | 2.6 | 0.221 | 8.4 | | 5 | 0.362 | 0.040 | 11.0 | 1.4 | 0.162 | 11.7 | | 6 | 0.047 | 0.005 | N/A | 0.2 | 0.015 | N/A | | 7 | 0.038 | 0.006 | N/A | 0.1 | 0.023 | N/A | The assay Cutoff Serum, and Positive Control were assayed concurrently twice a day for each of the five days. | | Mean OD | Std. Dev. | % CV | |------------------|---------|-----------|------| | Cutoff Serum | 0.271 | 0.027 | 10.1 | | Positive Control | 4.027 | 0.096 | 2.4 | ## B. Intra-assay The same seven serum samples were assayed 20 consecutive times in a single run. | Sample | Mean OD | Std. Dev. | % CV | |--------|---------|-----------|------| | 1 | 3.274 | 0.131 | 4.0 | | 2 | 2.239 | 0.121 | 5.4 | | 3 | 1.315 | 0.112 | 8.5 | | 4 | 0.570 | 0.034 | 5.9 | | 5 | 0.374 | 0.024 | 6.5 | | 6 | 0.066 | 0.003 | 4.3 | | 7 | 0.058 | 0.004 | 7.1 | The assay Cutoff Serum, and Positive Control were assayed concurrently 20 consecutive times in a single run | | Mean OD | Std. Dev. | % CV | |------------------|---------|-----------|------| | Cutoff Serum | 0.264 | 0.017 | 6.4 | | Positive Control | 3.895 | 0.166 | 4.3 | . {3}------------------------------------------------ ## Comparison Testing A total of 132 serum specimens (22 pANCA positive specimens taken from throughout the United States, 40 specimens from individuals demonstrating pauci-immune necrotizing and/or crescentic glomerulonephritis, and, 70 specimens from normal apparently healthy donors} were concurrently assayed by both the predicate device and the proposed device. The results are presented in the tables that follow: | Table 1.1 pANCA Panel 1. n = 22<br>Predicate Device | | | | |-----------------------------------------------------|----------|----------|-------| | Proposed Device | Positive | Negative | Total | | Positive | 21 | 0 | 21 | | Negative | 0 | 1 | 1 | | Total | 21 | 1 | 22 | Relative Sensitivity = 100.0 % {21/21}, 0.85 confidence interval = 84.5 % to 100 % | Table 1.2 pANCA Panel 2. n = 40<br>Predicate Device | | | | |-----------------------------------------------------|----------|----------|-------| | Proposed Device | Positive | Negative | Total | | Positive | 40 | 0 | 40 | | Negative | 0 | 0 | 0 | | Total | 40 | 0 | 40 | Relative Sensitivity = 100.0 % {40/40}. 0.88 confidence interval = 91.2 % to 100 % | Table 1.3 Normals, n = 70<br>Predicate Device | | | | |-----------------------------------------------|----------|----------|-------| | Proposed Device | Positive | Negative | Total | | Positive | 0 | 0 | 0 | | Negative | 81 | 62 | 70 | | Total | 8 | 62 | 70 | Relative Specificity = 100 % {62/62}, 4.8 confidence interval = 94.2 % to 100 % 1. The eight discrepant samples were assayed with a third party IFA kit and were found to be negative. The range of AU values for the positive specimens was 1.2 to 18. The average AU value for the normal specimens was 0.3 with a range of 0.1 to 0.8 {4}------------------------------------------------ ## Interfering Substances Libemic, icteric, and hemolytic samples were evaluated with the assay following NCCLS Document EP7-P Proposed Guideline, Interference Testing in Clinical Chemistry. The results indicate that there is no significant effect (<15 % variation) on the assay for samples with: Hemoglobin concentration: Bilirubin concentration: Lipid concentration: ≤ 500 mg/dL < 20 maldL ≤ 3000 ma/dL ## Prozone The VIRGO ® pANCA Kit was used to assay several high titered serum samples to determine if the kit would return unexpectedly low values. The results of this evaluation indicate that the kit gives appropriately high positive results with high titered sera. ### Conclusions The results of the comparative studies support the claim that the proposed device is substantially equivalent to the predicate device and performs as an effective screening assay. ## 510(k) Summary Page 5 {5}------------------------------------------------ Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract design resembling a bird or a wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 NOV 1 3 1997 Mr. Joseph M. Califano Manager, Requlatory Affairs Hemagen Diagnostics, Inc. 34-40 Bear Hill Road Waltham, Massachusetts 02154 Re : K973824 Trade Name: VIRGO® pANCA Kit (EIA method) Tier: II Regulatory Class: II Product Code: мов Dated: September 30, 1997 Received: October 7, 1997 Dear Mr. Califano: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {6}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4588. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ K973824 Device Name: VIRGO ® pANCA Kit Indication(s) For Use This enzyme-linked immunosorbent assay (ELISA) is indicated for the detection of altoantilian to the antigen Myeloperoxidase (MPO) in human serum. The presence of MPO antibodies, in combination with dinical observations and other serological tests, can aid in the dlagnosis of polyarterits, necrotizing glomerulonephritis and other conditions associated with elevated anti-neutrophil cytoplasmic antibodies (ANCA). ## (PLEASE DO NOT WRITE BELOW THIS LINE) Concurrence of CDRH, Office of Device Evaluation (ODE) Peter E. Mater (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number . Prescription Use (Fer 21 CFR 801.109) OR Over-The-Counter-Use
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%