EASYSAFE RETRACTING NEEDLE SYRINGE
K973792 · Saf-T-Med, Inc. · MEG · Jan 29, 1998 · General Hospital
Device Facts
| Record ID | K973792 |
| Device Name | EASYSAFE RETRACTING NEEDLE SYRINGE |
| Applicant | Saf-T-Med, Inc. |
| Product Code | MEG · General Hospital |
| Decision Date | Jan 29, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5860 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The primary intended use of the EasySafe™ Retracting Needle Syringe is to inject fluid intramuscularly (IM) into the body. The secondary intended uses are: 1. to help prevent sharps injuries when using the product for its primary intended use, and; 2. to help reduce wasted medication when using the product for its primary intended use.
Device Story
EasySafe™ Retracting Needle Syringe is a manual medical device designed for intramuscular (IM) fluid injection. It incorporates a retracting needle mechanism to minimize sharps injuries and reduce medication waste. Operated by clinicians, the device functions as a standard syringe with an integrated safety feature that retracts the needle after use. It is used in clinical settings for routine medication delivery. The device benefits patients and clinicians by lowering the risk of accidental needle sticks and ensuring efficient medication dosing. No complex electronics, software, or external power sources are involved.
Clinical Evidence
Simulated use study involving over 1,000 IM injections performed by a wide variety of clinicians; 100% success rate in injection completion; zero injuries reported. Focus group of 20 clinicians reported no impediments to use. Biocompatibility testing performed by contract labs confirmed safety. Sterilization validated to SAL 10^-6; pyrogenicity testing yielded acceptable results.
Technological Characteristics
Manual retracting needle syringe. Materials biocompatibility tested. Complies with ISO performance standards for syringes. Features integrated needle retraction mechanism for sharps injury prevention. Sterilized to SAL 10^-6.
Indications for Use
Indicated for intramuscular (IM) injection of fluids in patients requiring medication administration.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
- USMI SafeSnap™ Syringe (K925039)
- BD syringe (pre-Amendment device)
Related Devices
- K991039 — EASYSAFE RETRACTING NEEDLE SYRINGE · Saf-T-Med, Inc. · May 21, 1999
- K051762 — SAF-T-SYRINGE/SAF-T-NEEDLE · Safety Medical International, Incorporated · Sep 23, 2005
- K063340 — SAF-T-SYRINGE- 0.5 AND 1ML/CC · Safety Medical International, Incorporated · Mar 6, 2007
- K213077 — Automatically Retractable Safety Syringe · Zhejiang Lingyang Medical Apparatus Co., Ltd. · Apr 27, 2022
- K113673 — MYSAFETY SYRINGE · Medicalchain Internartional Corporation · Aug 3, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 9 1998
Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters and numbers. The sequence starts with the letter 'K', followed by the numbers '973792'. The numbers are written in a simple, slightly irregular style, typical of handwriting.
## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
| 510(k) Number: | K973792 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | EasySafe™ Retracting Needle Syringe |
| Common Name: | Retracting Needle Syringe |
| Manufacturer: | Saf-T-Med, Inc.<br>1250 South Grove, Suite 200<br>Barrington, IL 60010 |
| Predicate Device Information: | A claim of substantial equivalence is made to:<br>USMI SafeSnap™ Syringe, K925039<br>BD syringe (pre-Ammendment device) |
| Device Claims: | For the injection of fluid intramuscularly (IM) into<br>the body, while helping to reduce the risk of sharps<br>injuries and reducing the amount of wasted<br>medication. |
In compliance with the requirements of the Safe Medical Device Act (SMDA) of 1990, the following information is submitted as a summary of safety and effectiveness information for this 510(k) premarket notification.
## Effectiveness
In a recent simulated use study, over 1,000 IM injections were administered by a wide variety of clinicians to test the effectiveness of the product in injecting fluids intramuscularly into a patient. All injections were successfully completed.
Bench testing of the product confirms its compliance with the applicable performance standards established by the International Standards Organization (ISO).
In focus groups twenty clinicians indicated that they saw no impediments to being able to effectively administer injections to patients using the product.
## Safety
In a recent simulated use study, over 1,000 IM injections were administered by a wide variety of clinicians to test the effectiveness of the product in injecting fluids into a patient. No injuries of any kind were reported.
Bench testing of the product confirms its comparability with legally-marketed predicate devices for a variety of safety factors, including but not limited to graduation accuracy and needle guard puncture resistance.
{1}------------------------------------------------
Biocompatability testing was performed by outside contract laboratories, with the results indicating that, from a biocompatability standpoint, the product is safe for its intended purpose.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
A variety of design factors are intended to improve upon patient and clinician safety during use of the product. While sample size requirements have precluded a broad statistical study of the impact of such factors on injury rates, clinicians have reacted in a unanimous fashion to those factors.
A detailed risk analysis was performed on the product and, based on information currently available, the risk of harm from the product appears to be relatively low.
Sterilization validation was conducted by an outside contract laboratory. Sterilization to a SAL of 10° was confirmed during the tests. Pyrogenicity was also evaluated and acceptable results obtained.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center is a stylized eagle emblem, which is a common symbol associated with the U.S. government.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 9 1998
Mr. James Erbs President Saf-T-Med, Incorporated 1250 South Grove, Suite 200 Barrington, Illinois 60010
Re : K973792 Easysafe Retracting Needle Syringe Trade Name: Requlatory Class: II Product Code: MEG Dated: December 30, 1997 January 2, 1998 Received:
Dear Mr. Erbs:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labelinq, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does
{3}------------------------------------------------
Page 2 - Mr. Erbs
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
ny A Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## K973792 510(k) Number (if knowf)): EasySafe" Retracting Needle Syringe Device Name:
Indications For Use:
The primary intended use of the EasySafe™ Retracting Needle Syringe is to inject fluid intramuscularly (IM) into the body. The secondary intended uses are:
- 1. to help prevent sharps injuries when using the product for its primary intended use, and;
- 2. to help reduce wasted medication when using the product for its primary intended use.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Palatine Cossinote
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 792 510(k) Number /
Prescription Use (Per 21 CFR 801.109)
ଠନ
Over-The-Counter Use
(Optional Format 1-2-96)