MEDI-TRACE 1510M COMBINATION DEFIBRILLATION, MONITORING AND PACING ELECTRODE
K972970 · Graphic Controls Corp. · MKJ · Aug 27, 1997 · Cardiovascular
Device Facts
| Record ID | K972970 |
| Device Name | MEDI-TRACE 1510M COMBINATION DEFIBRILLATION, MONITORING AND PACING ELECTRODE |
| Applicant | Graphic Controls Corp. |
| Product Code | MKJ · Cardiovascular |
| Decision Date | Aug 27, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.5310 |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Intended Use
Intended for use in defibrillation procedures, cardioversion and pacing. This device is intended for use on adults. Not for use on children or infants.
Device Story
Pre-gelled conductive electrode; consists of cable attachment, vinyl insulating layer, metal foil, conductive adhesive hydrogel, pressure-sensitive adhesive ring, and release liner. Used for defibrillation, cardioversion, and pacing. Applied to adult patients by clinicians. Device functions as interface between patient and defibrillator/monitor/pacer; transmits electrical energy for therapy and signals for ECG monitoring. Benefits include reliable electrical contact for life-saving cardiac procedures.
Clinical Evidence
Bench testing only. Device met AAMI DF-39 specifications for electrical performance, energy throughput, and simulated use (multiple shocks, maximum current pacing, ECG monitoring). Biocompatibility testing per ISO-10993 confirmed device is non-irritating, non-cytotoxic, and non-sensitizing. Accelerated shelf-life studies performed.
Technological Characteristics
Pre-gelled conductive electrode. Materials: vinyl insulating layer, metal foil, conductive adhesive hydrogel, pressure-sensitive adhesive ring. Connectivity: cable interface. Standards: AAMI DF-39 (electrical/performance), ISO-10993 (biocompatibility). Packaging: heat-sealed pouches.
Indications for Use
Indicated for use in defibrillation, cardioversion, and pacing procedures in adult patients. Contraindicated for use on children or infants.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
- Marquette Electronics Defibrillation/Pacing Pads
Related Devices
- K955882 — MEDI-TRACE COMBINATION DEFIBRILLATION & ECG ELECTRODE FOR HEWLETT PACKARD DEFIBRILLATORS · Graphic Controls Corp. · Oct 24, 1996
- K983287 — MEDI-TRACE 1610C COMBINATION DEFIBRILLATION, MONITORING AND PACING ELECTRODE · Graphic Controls Corp. · Nov 5, 1998
- K964469 — 700-F SERIES STIMULATION ELECTRODES · Cardiotronics Systems, Inc. · Dec 11, 1996
- K103159 — CARDIO MED MODEL A102, H104, O101, P103, AND Z100 · Cardio Medical Products, Inc. · Jan 11, 2011
- K980857 — MEDI-TRACE 1310P COMBINATION DEFIBRILLATION, MONITORING & PACING ELECTRODE · Graphic Controls Corp. · May 12, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
189 Van Rensselaer Street PO Box 1271
Buffalo, New York 14240-1271 USA
716 853-7500
Fax: 716 849-6499
K972970
Image /page/0/Picture/3 description: The image shows a white lowercase letter 'g' inside of a black circle. The letter 'g' is stylized with a curved tail extending upwards. The letters 'TM' are located in the bottom right corner of the circle, indicating a trademark.
## 510(k) SUMMARY
AUG 27 1997
Graphic Controls
| Date: | August 6, 1997 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Manufacturing Facility: | Graphic Controls Corporation<br>189 Van Rensselaer Street<br>PO Box 1271<br>Buffalo, NY 14240<br>Registration Number 1317188 |
| Telephone: | (716) 853-7500 |
| Contract Person: | Kathleen Selover<br>Regulatory Affairs Specialist<br>(71 6) 853-7500, ext 7630<br>Fax: (716) 847-7531 |
| Device Trade Name: | Medi-Trace® 1510M Combination<br>Defibrillation, Monitoring and Pacing<br>Electrode |
| Device Common Name: | Self Adhesive Electrode for Monitoring<br>and Defibrillation, optional Pacing |
| Classification Name: | Multi-function Electrocardiograph<br>Electrode |
| Regulatory Reference: | 74 MLN |
| Predicate Device | Marquette Electronics<br>Defibrillation/Pacing Pads |
{1}------------------------------------------------
#### Description:
A pre-gelled conductive elecrode consisting of a means of attaching the electrode to the cable, an insulating layer of vinyl, a layer of metal foil, a conductive adhesive hydrogel. a pressure sensitive adhesive ring and a release liner. Device is packaged in pairs in a heat sealed pouch, 10 pouches are packaged into one shelf-box/shipper
#### Intended/Indications for Use:
Intended for use in defibrillation procedures, cardioversion and This device is intended for use on adults. Not for use on pacing. children or infants.
#### Physical/Technical Comparison
Medi-Trace® 1510M Electrode is replaceable with Marquette Electronics Defibrillation/Pacing Pads. Physical and technical characteristics, including materials used in construction, size, intended use and conductive gel type of these electrodes are comparable. Since the formulation of conductive gel is considered to be proprietary, exact chemical comparison could not be made.
## Performance Summary:
The device and the predicate were subjected to AAMI electrical tests as described in DF-39, 3.3.19 and energy throughput testing. Test results for both the device and the predicate met the specifications as established in DF-39 for self-adhesive electrodes for mointoring, defibrillation and pacing.
In addition, the device was subjected to simulated use testing consisting of multiple defibrillation shocks, maximum current pacing Test results for the device met the and ECG monitoring. specifications as established in DF-39 for self adhesive electrodes for moinitoring, defibrillation and pacing.
## Biocompatibility Testing:
The device was subjected to biocompatibility testing as recommended in the May 1, 1995 FDA memorandum entitled Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. The device was found to be nonirritating, non-cytotoxic and non-sensitizing.
#### Shelf Life:
Data obtained in accelerated shelf life studies was reviewed and found to substantiate our claimed shelf life.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
AUG 27 1997
Susan A. Krasny, Ph.D. Graphic Controls Corporation 189 Van Rensselaer Street P.O. Box 1271 14240-1271 Buffalo, New York
Re: K972970 Medi-Trace® 1510M Combination Defibrillation, Monitoring and Pacing Electrodes Regulatory Class: III (three) Product Code: 74 MLN Dated: August 6, 1997 Received: August 11, 1997
Dear Dr. Krasny:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GME regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Susan A. Krasny, Ph.D.
・
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97).
Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
A.A. Licht
As Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page _________________________________________________________________________________________________________________________________________________________________________
510(k) Number, if known: _____________________________________________________________________________________________________________________________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# Device Name: MediTrace® 1510M Combination Defibrillation, Monitoring and Pacing Electrode
Indications for Use:
Medi-Trace 1510M Combination Defibrillation, Monitoring and Pacing Electrode is intended for use on adults. Not for use on children or infants.
(PLEASE DP MPT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div> (Division Sign-Off) <div>Division of Cardiovascular, Respiratory,</div> <div>and Neurological Devices</div> </div> |
|---------------|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K972970 |
Prescription Use (Per 21CFR 801.109) OR Over-the-Counter
Use