PLAYTEX TAMPONS #14897/PLAYTEX GENTLE GLIDE,SILK GLIDE, SLIMFITS,SOFT COMFORT AND PORTABLES TAMPONS
K971979 · Playtex Products, Inc. · HIB · Sep 26, 1997 · Obstetrics/Gynecology
Device Facts
| Record ID | K971979 |
| Device Name | PLAYTEX TAMPONS #14897/PLAYTEX GENTLE GLIDE,SILK GLIDE, SLIMFITS,SOFT COMFORT AND PORTABLES TAMPONS |
| Applicant | Playtex Products, Inc. |
| Product Code | HIB · Obstetrics/Gynecology |
| Decision Date | Sep 26, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.4530 |
| Device Class | Class 2 |
Intended Use
Scented, Scented Deodorant and Unscented menstrual tampons for absorption of menstrual filud: Part 21 C. F. R., Section 884.5460 and Section 884.5470 respectively.
Device Story
Playtex Tampons are menstrual hygiene products designed for the absorption of menstrual fluid. The device consists of absorbent fibers and a withdrawal string. It is intended for over-the-counter use by individuals during menstruation. The device functions by being inserted into the vaginal canal to absorb fluid. It is available in various product lines including Gentle Glide, Silk Glide, Slimfits, Soft Comfort, and Portables, offered in Regular, Super, and Super Plus absorbency levels.
Clinical Evidence
No clinical data provided. Evidence consists of nonclinical bench testing including human sensitization, dermal irritation, acute oral toxicity, subacute vaginal irritation, agar diffusion, and TSST-1 toxin testing.
Technological Characteristics
Menstrual tampons composed of absorbent fibers and a withdrawal string. Materials in contact with the vaginal wall are identical to predicate devices. Operates via passive absorption of menstrual fluid.
Indications for Use
Indicated for individuals requiring absorption of menstrual fluid. Available in scented, scented deodorant, and unscented varieties.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
- Playtex Gentle Glide (K971979)
- Silk Glide (K971979)
- Slimfits (K971979)
- Soft Comfort (K971979)
- Portables (K971979)
Related Devices
- K982383 — PLATEX GENTLE GLIDE, SILK GLIDE, SLIMFITS, SOFT COMFORT AND PORTABLES TAMPONS · Playtex Products, Inc. · Aug 19, 1998
- K971547 — PLAYTEX TAMPONS #11197, PLAYTEX GENTLE GLIDE, SOFT COMFORT AND SLIMFITS TAMPONS · Playtex Products, Inc. · Jun 19, 1997
- K993794 — PLATEX ODER ABSORBING GENTLE GLIDE, SLIMFITS AND SILK GLIDE TAMPONS · Playtime Products, Inc. · Feb 7, 2000
- K072376 — PLAYTEX GENTLE GLIDE, PLAYTEX GENTLE GLIDE MULTIPACK TAMPONS · Playtex Products, Inc. · Dec 12, 2007
- K020202 — PLAYTEX DEODORANT TAMPONS, MODEL 34702 · Playtex Products, Inc. · Mar 12, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY 9.
1.
K 971979
fi,91
## SUMMARY OF SAFETY AND EFFECTIVENESS SEP 26 1997 A.
| 1. | COMPANY NAME | - | Playtex Products Inc. |
|----|-----------------|---|----------------------------------------------------------------|
| | ADDRESS | - | 215 College Road<br>P. O. Box 728<br>Paramus, New Jersey 07652 |
| | TELEPHONE | - | 201-265-8000 |
| | CONTACT PERSON | - | M. Rosengarten<br>Director of Regulatory Affairs |
| | DATE OF SUMMARY | - | May 28, 1997 |
| 2. | DEVICE NAME | - | Playtex Tampons |
- The new Playtex tampons are substantially equivalent to previously 3. cleared Playtex Gentle Glide®, Silk Glide®, Slimfits™, Soft Comfort™ and Portables® Tampons, Regular, Super and Super Plus.
CLASSIFICATION NAME - - Unscented Menstrual Tampons
- The device description is: Scented or scented deodorant and unscented 4. menstrual tampons for the absorption of menstrual fluid.
- Playtex tampons are intended to be used as scented, scented deodorant and న. unscented menstrual tampons for the absorption of menstrual fluid.
- The new tampon has the same technological characteristics as the predicate 6. device. The fiber, string and materials in contact with the vaginal wall are the same or have the same mode of action.
- 1. Nonclinical testing referenced for the determination of substantial equivalence B. includes:
Human sensitization, dermal irritation, acute oral toxicity, subacute vaginal irritations, agar diffusion and TSST-1 toxin testing.
- Based on the review of the data referenced in this "510(k) Summary," the 3. Playtex Tampons are substantially equivalent to the predicate device in terms of safety and effectiveness.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a bird. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 26 1997
Re: K971979
Mr. Mark E. Rosengarten Director Playtex Products, Inc. 215 College Road, P.O. Box 728 Paramus, New Jersey 07652
Playtex Gentle Glide®, Silk Glide®, Slimfits™ Soft Comfort™ and Portable® Tampons Dated: August 14, 1997 Received: August 18, 1997 Regulatory class: II 21 CFR §884.5460/Product code: 85 HIB 21 CFR §884.5470/Product code: 85 HEB
Dear Mr. Rosengarten:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## STATEMENT OF INDICATIONS OF USE PAGE
Page 1 _ of _ 1 _
510(k) Number (if known): _ K971979
Device Name: Playtex Gentle Glide®, Silk Glide®, Slimfits™ Soft Comfort™ and Portable® Tampons
Indications For Use:
Scented, Scented Deodorant and Unscented menstrual tampons for absorption of menstrual filud: Part 21 C. F. R., Section 884.5460 and Section 884.5470 respectively.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rher D. Mathing/
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 1971979 510(k) Number_
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)