BROWNE STEAM INDICATOR

K970918 · Board of Directors, Albert Browne , Ltd. · JOJ · Jun 24, 1997 · General Hospital

Device Facts

Record IDK970918
Device NameBROWNE STEAM INDICATOR
ApplicantBoard of Directors, Albert Browne , Ltd.
Product CodeJOJ · General Hospital
Decision DateJun 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The Browne Steam Indicator is a process indicator to be used in a steam autoclave with a working range of 121-134°C to distinguish between processed and unprocessed units.

Device Story

Browne Steam Indicator is a paper strip with chemical indicator ink pads at each end. Designed for use in gravity and vacuum-assisted steam autoclaves (121-134°C). Principle of operation involves a temperature-dependent chemical reaction causing ink pads to change color from white to black via a brown intermediate. Used by healthcare personnel to confirm exposure to a steam sterilization cycle. Provides visual confirmation to distinguish processed from unprocessed items, ensuring sterilization cycle completion.

Clinical Evidence

Bench testing only. Performance evaluated in a BIER vessel conforming to ANSI/AAMI ST45-1992. Strips tested in partial cycles at 121°C and 134°C; color change from white to black occurred in ≥8 minutes at 121°C and ≥3 minutes at 134°C. Stability testing confirmed a 2-year shelf life.

Technological Characteristics

Paper strip with chemical indicator ink pads. Operates via temperature-dependent chemical reaction. Compatible with gravity and vacuum-assisted steam autoclaves (121-134°C). Performance testing conducted per ANSI/AAMI ST45-1992.

Indications for Use

Indicated for use in steam autoclaves (121-134°C) to distinguish between processed and unprocessed units. Suitable for over-the-counter use.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} JUN 24 1999 510(k) Summary Albert Browne Ltd. Browne Steam Indicator K970918 1. SUBMITTED BY Albert Browne Ltd. Chancery House Rosebery Road Anstey Leicester LE7 7EL United Kingdom CONTACT PERSON Alan Charlton Chancery House Rosebery Road Anstey Leicester LE7 7EL United Kingdom DATE PREPARED March 11, 1997 2. DEVICE NAME Browne Steam Indicator CLASSIFICATION NAME Physical/chemical sterilization process indicator Browne Steam Indicator Additional Information - K970918 5/1/97 Page B-1 {1} # CLASSIFICATION STATUS Physical/chemical process indicator is classified as Class II under Sterilization process indicator in 21 CFR 880.2800 by the General Hospital and Personal Use Devices Panel. ## 3. PREDICATE DEVICE 3M Comply 1250 Steam Integrator Strips, 3M Health Care. ## 4. INTENDED USE The Browne Steam Indicator is a process indicator to be used in a steam autoclave with a working range of 121-134°C to distinguish between processed and unprocessed units. ## 5. DEVICE DESCRIPTION The Browne Steam Indicator is a paper strip with indicator ink pads on each end. The indicator ink pads change color from white to black through a brown intermediate in a steam autoclave working at 121-134°C (250-273°F). ## 6. TECHNOLOGICAL CHARACTERISTICS The Browne Steam Indicator consists of a paper strip with chemical indicator ink pads located on each end. The device is designed to be used in gravity and vacuum-assisted steam autoclaves with a working range of 121-134°C. The color change in both devices is produced by a temperature-dependent chemical reaction. ## 7. PERFORMANCE TESTING All performance testing was conducted in a BIER vessel/prototype which conforms to the performance requirements for BIER/Steam vessels described in ANSI/AAMI ST45-1992. Browne Steam Indicator Additional Information - K970918 5/1/97 Page B-2 {2} Testing was conducted to evaluate the performance of the strips in partial cycles at $121^{\circ}\mathrm{C}$ and $134^{\circ}\mathrm{C}$. The data showed that the strips changed color from white to black with a brown intermediate. The time required for complete development of an end point response was $\geq 8$ minutes at $121^{\circ}\mathrm{C}$ and $\geq 3$ minutes at $134^{\circ}\mathrm{C}$. The data demonstrates that the device can be used for the confirmation of exposure to a processing cycle in a steam autoclave with a working range of $121 - 134^{\circ}\mathrm{C}$. The performance of the Browne Steam Indicator was equivalent to that of the predicate device. All strips used for testing were $\geq 2$ years from the date of manufacture, demonstrating that the 2-year expiration date is adequate to ensure accurate, reproducible performance. Browne Steam Indicator Additional Information - K970918 5/1/97 Page B-3 {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 24 1997 Cynthia J.M. Nolte, Ph.D. Associate Consultant Medical Device Consultants, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760-4153 Re: K970918 Trade Name: Browne Steam Indicator Regulatory Class: II Product Code: JOJ Dated: May 1, 1997 Received: May 2, 1997 Dear Dr. Nolte: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 - Dr. Nolte This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597. Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} 510(k) Number (if known): K970918 Device Name: Browne Steam Indicator Indications For Use: The Browne Steam Indicator is a process indicator to be used in a steam autoclave with a working range of 121-134°C to distinguish between processed and unprocessed units. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-0.jpeg](img-0.jpeg) Prescription Use ☐ (Per 21 CFR 801.109) OR Over-The-Counter Use ☑ Browne Steam Indicator Additional Information - K970918 5/1/97 Page A-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...