QUICKPAC II ONE STEP METHAMPHETAMINE TEST
Device Facts
| Record ID | K970724 |
|---|---|
| Device Name | QUICKPAC II ONE STEP METHAMPHETAMINE TEST |
| Applicant | Drial Consultants, Inc. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | May 13, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Intended Use
The QuikPac II One Step Methamphetamine Test is intended for medical/forensic screening of urine.
Device Story
Chromatographic absorbent device for qualitative urine drug screening; utilizes competitive immunoassay principle. Drug/metabolites in urine sample compete with immobilized drug conjugate on porous membrane for limited antibody sites. Labeled antibody-dye conjugate binds to free drug; complex competes with immobilized antigen in positive reaction zone. Absence of magenta band indicates drug presence above 500 ng/ml; presence of magenta band in control zone confirms reagent/device function. Used in clinical/forensic settings; operated by laboratory personnel. Provides visual qualitative result; assists in identifying potential methamphetamine exposure; positive results require confirmation via GC/MS.
Clinical Evidence
Clinical trial of 305 urine samples. Compared against Syva EMIT® II with GC/MS confirmation for discrepant results. Reported relative sensitivity 100%, relative specificity 97.7%, and overall accuracy 98.68%. In-house testing showed 100% accuracy against Syva EMIT® II.
Technological Characteristics
Chromatographic absorbent device; porous membrane support; antibody-dye conjugate; competitive immunoassay format. Qualitative visual readout. No electronic components or software.
Indications for Use
Indicated for the qualitative screening of human urine for the presence of Methamphetamine and its metabolites at a cutoff concentration of 500 ng/ml. Intended for medical and forensic use.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Reference Devices
- Syva EMIT® II
Related Devices
- K970265 — ONE STEP URINE DRUG OF ABUSE METHAMPHETAMINE TEST · Drial Consultants, Inc. · Apr 8, 1997
- K970395 — QUIKSTRIP ONE STEP METHAMPHETAMINE · Drial Consultants, Inc. · Jun 26, 1997
- K971109 — QUIKPAC II ONE STEP AMPHETAMINE TEST · Drial Consultants, Inc. · May 14, 1997
- K973269 — QUIKSTRIP ONE STEP BENZODIAZEPINE TEST · Drial Consultants, Inc. · Feb 25, 1998
- K971218 — QUIKSTRIP ONE STEP AMPHETAMINE TEST · Drial Consultants, Inc. · May 30, 1997