K Number
K970525

Validate with FDA (Live)

Device Name
STATNET
Date Cleared
1997-04-16

(64 days)

Product Code
Regulation Number
892.1600
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

StatNet is intended for the storage, distribution to any of several work stations, viewing, interpretation and annotation of X-ray of coronary anatomy that have been obtained in coronary catheterization laboratories and written on DICOM-3 CD-ROM media.

Device Description

StatNet is a network of image display, processing, and archiving modules composed of general purpose, personal computer level, digital hardware; commercial, general purpose digital communication hardware; general purpose operating systems and communications software; medical image analysis and processing software developed by others; and proprietary software modules developed by GORCA that integrate the preceding components into a complete system. Images enter the network on DICOM-3 CD-ROM media.

AI/ML Overview

The provided text describes a PACS system called StatNet and compares its features to predicate devices to establish substantial equivalence. However, it does not contain information about acceptance criteria for device performance, nor a study proving it meets such criteria. The document is a 510(k) summary, which focuses on device description and comparison to predicates for regulatory clearance, not performance validation studies against specific acceptance criteria.

Therefore, I cannot provide the requested table or answer the specific questions about sample sizes, ground truth, expert qualifications, adjudication, MRMC studies, or standalone performance, as this information is not present in the provided text.

{0}------------------------------------------------

K 970525 GORCA MEDICAL TECHNOLOGIES

510(k) Summary for GORCA Medical Technologies StatNet Prepared February 6, 1997

APR 1 6 1997

GORCA Medical Technologies Submitted by: 300 West Route 38 Moorestown, NJ 08057

  • Donald G. Herzog Contact Person: Vice President, Medical Technologies
    Device Trade Name: StatNet.

  • Picture Archiving and Communication System (PACS) Common name: for cardiology departments.

  • Accessory to angiographic X-ray system (892.1600) or PACS, product code Classification: LLZ, consisting of a combination of medical image storage devices, product code LMB, and medical image communication devices, product code LMD, to which image processing and display functions have been added.

  • Predicate Device: Camtronics Archium manufactured by Camtronics Ltd., 900 Walnut Ridge Drive, Hartland, WI (K934496).

Description of Device:

StatNet is a network of image display, processing, and archiving modules composed of general purpose, personal computer level, digital hardware; commercial, general purpose digital communication hardware; general purpose operating systems and communications software; medical image analysis and processing software developed by others; and proprietary software modules developed by GORCA that integrate the preceding components into a complete system. Images enter the network on DICOM-3 CD-ROM media.

Intended Use of Device:

StatNet is intended for the storage, distribution to any of several work stations, viewing, interpretation and annotation of X-ray of coronary anatomy that have been obtained in coronary catheterization laboratories and written on DICOM-3 CD-ROM media.

{1}------------------------------------------------

Statement of Substantial Equivalence

(

Function or VariableSTATnetPredicate
IMACSEIGENCamtronics
1Electronic Equivalent ofCine ReviewYesYesYesYes
2Cine-like ControlTechniquesPan and Zoom, Frame RateControl, Window and LevelAdjustments, Image InversionPan and Zoom, Frame RateControl, Window and LevelAdjustments, Image InversionPan and Zoom, Frame RateControl, Window and LevelAdjustments, Image InversionPan and Zoom, Frame RateControl, Window and LevelAdjustments, Image Inversion
3Vessel MeasurementsYes, Using NIH Image SoftwareYes, Under Proprietary MethodYes, Under Proprietary MethodYes, Under Proprietary Method
4Display Resolution512x512, 512x1024, and1024x1024 Pixels512x512 and 512x1024 Pixels512x512, 512x1024, and1024x1024 Pixels512x512, 512x1024, and1024x1024 Pixels
5Image EnhancementReal-Time Digital EdgeEnhancement While DisplayingReal-Time Digital EdgeEnhancement While DisplayingReal-Time Digital EdgeEnhancement While DisplayingReal-Time Digital EdgeEnhancement While Displaying
6Case Storage TechniqueAs a Single Patient RecordAs a Single Patient RecordAs a Single Patient RecordAs a Single Patient Record
7Interchange of CardiacAngiographic StudiesCDR DICOM 3.0 FormattedDiskCDR DICOM 3.0 FormattedDiskCDR DICOM 3.0 FormattedDiskCDR DICOM 3.0 FormattedDisk
8Network LinkDigital FiberDigital FiberNone (Dual-Path Digital OpticalDisk as Input to all ReviewStations and Archives)Digital Fiber
9Image Acquisition intoSystemDICOM 3 Formatted CDR DiskCDR DICOM 3.0 FormattedDisk and Analog AcquisitionFrom Cath Lab CameraAnalog Acquisition From CathLab CameraCDR DICOM 3.0 FormattedDisk and Analog AcquisitionFrom Cath Lab Camera
10Demographics-BasedCase Selection MenuYesYesYesYes
11Archive Storage FormatDigitalDigitalDigitalDigital
12Selected FramesSeparately Stored for Usein Report GenerationYesYesYesYes

空室

(

(

§ 892.1600 Angiographic x-ray system.

(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.