K Number
K965139

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-04-23

(121 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Microscope Navigator (SMN) and the Surgical Tool Navigator (STN) System are a modular product group intended to provide the surgeon with navigational support during surgery They permit the surgical team to load radiological imaging studies onto a workstation equipped with image processing and visualization software. After correlation of the patient position with the images, (the registration procedure) the system displays the position, trajectory and other localization information of the current "tool" interactively throughout the surgical procedure.

Device Description

The Surgical Microscope Navigator (SMN) and the Surgical Tool Navigator (STN) System are a modular product group intended to provide the surgeon with navigational support during surgery. The SMN's primary "tool" is a surgical microscope. The STN uses a variety of "Pointers" and instruments as tools. Both systems localize the tools in space via an industry standard infra-red LED based camera system. The LED's coordinates are then transferred to the workstation which in turn converts them to the coordinate system of the imaging data and interactively displays them.

AI/ML Overview

The provided 510(k) summary for K965739 does not contain the information required to answer your questions regarding acceptance criteria and the study that proves the device meets the acceptance criteria.

This 510(k) summary focuses primarily on:

  • Device Description and Intended Use: Explaining what the SMN and STN systems are and how they function.
  • Substantial Equivalence Claims: Comparing the new devices to previously cleared devices (MKM system and StealthStation) based on their intended use, technology, and operating principles.
  • Legal Basis for Submission: Referencing an exemption for surgical microscopes or substantial equivalence to existing devices.

Missing Information:

To answer your questions, the following information would typically be found in a more detailed technical report or study summary submitted as part of the 510(k) application, which is not present in this publicly available summary:

  1. Acceptance Criteria and Reported Device Performance Table: This would require specific performance metrics (e.g., accuracy, precision) and defined thresholds for "acceptable" performance, along with the measured results from a validation study.
  2. Sample Sizes and Data Provenance for Test Set: Details about the number of cases/patients used to evaluate the device's performance, where the data came from (e.g., specific hospitals, geographic regions), and whether it was collected retrospectively or prospectively.
  3. Number and Qualifications of Experts for Ground Truth: Information about the professionals (e.g., surgeons, radiologists) who reviewed the test cases to establish the reference standard, including their number and years of experience.
  4. Adjudication Method: How disagreements among experts were resolved (e.g., 2+1, 3+1, consensus).
  5. MRMC Comparative Effectiveness Study: Whether a study comparing human performance with and without the device's assistance was conducted, and the quantitative effect size.
  6. Standalone Performance Study: Data on the algorithm's performance without human interaction.
  7. Type of Ground Truth Used: The specific method used to establish the truth for the test cases (e.g., histopathology, clinical outcome, expert consensus).
  8. Sample Size for Training Set: The amount of data used to develop and train the device's algorithms (if applicable, though this device appears hardware-focused with localization algorithms).
  9. Ground Truth Establishment for Training Set: How the reference standard was created for the training data.

Conclusion:

Based solely on the provided 510(k) summary, it is not possible to provide the requested details regarding acceptance criteria, study design, and performance metrics. The summary focuses on establishing substantial equivalence rather than detailing the specifics of a performance validation study against defined acceptance criteria.

{0}------------------------------------------------

510-k Summary

K965739

APR 23 1997

Pursuant to 21 CFR 807.92 the following summary is submitted.

  1. Submitter's name-

Carl Zeiss, Inc. 1 Zeiss Drive Thornwood, NY 10594 (914) 681-7761 Contact Person-Kenneth M. Nicoll December 19, 1996

  1. SMN System with STN system and permit individual sale of STN Surgical Microscope/ Image Guided System and Instrument Based Image Guided System

  2. We are claiming an exemption as a Surgical Microscope due to the exemption set forth in the Federal Register Notice dated December 7, 1994 which exempted surgical microscopes. Or in the alternative we are also claiming substantial equivalence to our MKM system which is used in a similar manner as our SMN system. Finally, we are claiming substantial equivalence to the Stealth Station sold by Sofamor Danek which is very similar to our STN system.

    1. The Surgical Microscope Navigator (SMN) and the Surgical Tool Navigator (STN) System are a modular product group intended to provide the surgeon with navigational support during surgery They permit the surgical team to load radiological imaging studies onto a workstation equipped with image processing and visualization software. After correlation of the patient position with the images, (the registration procedure) the system displays the position, trajectory and other localization information of the current "tool" interactively throughout the surgical procedure.
      The SMN's primary "tool" is a surgical microscope that, through features such as an autofocus and heads-up display, permit the high accuracy localization of the critical focus and hence an area of interest in the patients anatomy. The STN by contrast uses a variety of "Pointers" and instruments as tools. Here the surgeon touches the area of interest with the pointer or instrument instead of virtually probing it with the microscope autofocus. The functionality of the SMN can be added to the STN as an upgrade and conversely the microscope functionality (i.e. SMN) can be added to an STN. Furthermore a system having both "modules" can easily be used in either configuration in a given surgical procedure.

Both systems localize the tools in space, to a high degree of accuracy, via an industry standard infra-red LED based camera system. The LED's coordinates are then transferred to the workstation which in turn converts them to the coordinate system of the imaging data and interactively displays them.

{1}------------------------------------------------

510-k Summary Page 2

  1. These devices will be used in the same manner as all Surgical Microscope/ Image Guided Systems are used and as all Instrument Based Image Guided Systems are used. These devices are used to aid the surgeon in visualizing the surgical site and guiding the surgeon through the surgical course. The systems do not take the place of the surgeon they are only a tool used to assist the surgeon during the procedure. Decisions about what to do and how to perform a procedure still rest firmly with the surgeon. These systems incorporate the original intended uses of our MKM system and now the STN system can be used as a stand alone navigational tool.

  2. The SMN is substantially equivalent to our MKM system ( Multiple Coordinate Manipulator) which is already cleared. Both of these systems are used in surgical procedures where magnification and illumination of the site is required. They are both ideally suited for stereotactically guided surgical procedures in any location of the body. Both systems utilize either identical or very similar microscopes. They both operate on the same stereotactic surgical planning system, which has been specially adapted for use with these systems and can be used to plan stereotactic-guided procedures.

The primary differences between the variations of the SMN and the MKM are that the MKM is capable of moving and tracking its position with internal servo-motors. While all versions of the SMN are passive and require manual positioning of the microscope, relying on an external light based localization system.

The STN system is virtually identical to the medical device known as the StealthStation® sold by Sofamor Danek. The primary difference is in the image processing software that runs on the workstation. Otherwise the localization system of both instruments are manufactured and supplied by the same company.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).