K Number
K965115
Date Cleared
1997-03-10

(80 days)

Product Code
Regulation Number
874.3880
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Round Window u Cath™ (RWuC) is intended to be used by any board certified otolaryngologist or otologist to irrigate the entire middle ear or the round window area of the middle ear. During the course of the irrigation treatment, it is intended that the vent tube through which the RW (C's dual lumen passes, will aerate the middle ear. The treating physician will choose, based on what hels will be most beneficial for his/her patient, the following: { }) the exact fluids to be delivered through the RWAC; (2) the proper quantities; and (3) the duration and interval of treatment (up to 14 days). It is intended that the physician load the RWuC with the initial fluid to be delivered prior to placement. The RWAC is intended to remain in place for up to 14 days. Neuro-Biometrix, Inc. recommends the physician use only fluids indicated for this use. The physician should refer to the fluid labelling prior to use with the RWC.

Device Description

The RWAC consists of a double lumen (two parallel and attached Tecoffex 80A (polyurethane) extruded tubes) opening into a common fenestrated reservoir tip on the distal end. The soft, flexible reservoir tip of the RWuC will be available in three sizes: 2mm, 3mm, and 4mm diameters. It is intended that the surgeon will choose an appropriate sized tip to fit the anatomy of the patient. The double lumen is bent at approximately a 60-75 degree angle slightly proximal to the reservoir tip. One lumen is intended to serve as an in-flow and the other an out-flow. The outflow lumen is intended to (1) relieve or avoid a build up of air and/or fluid pressure in the device; and (2) facilitate the removal of fluid from the device. A silicone, paparella-type vent tube will be placed around the RWAC during manufacturing.

AI/ML Overview

The provided text describes a medical device, the Round Window u Cath™ (RWuC), and its intended use, but it does not contain specific acceptance criteria or a study proving that the device meets such criteria in terms of performance metrics like sensitivity, specificity, accuracy, or clinical outcomes.

The "Performance Testing" section outlines what aspects will be tested (pressure vs. flow, bond strength, visual inspection, dimensional compliance) but does not provide the actual results of these tests or explicit acceptance criteria (e.g., "flow rate shall be X ml/minute at Y pressure"). It only states that "Sufficient samples will be used for each test" and that these tests "will be conducted." This indicates a plan for testing, not completed tests with established criteria and results.

Therefore, most of the information requested in your prompt cannot be extracted from this document.

However, I can provide the following based on the text:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not specifiedNo performance metrics reported in the provided text.
Pressure vs. flowTesting planned, but no results or specific criteria provided.
Distal tip bond strengthTesting planned, but no results or specific criteria provided.
Proximal hub bond strengthTesting planned, but no results or specific criteria provided.
WorkmanshipVisual inspection planned, but no results or specific criteria provided.
Dimensional complianceVisual inspection planned, but no results or specific criteria provided.
BiocompatibilityDevice materials were subjected to biocompatibility testing, deemed Class VI biocompatible. (Although specific criteria and results are not detailed, this implies meeting some biocompatibility standards.)

Regarding the study proving the device meets the acceptance criteria, the document states the plan for "Performance Testing" but does not present the study results.

Therefore, the following information is not available in the provided text:

  • Sample size used for the test set and the data provenance.
  • Number of experts used to establish the ground truth for the test set and their qualifications.
  • Adjudication method for the test set.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers improving with AI vs without AI assistance.
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  • The sample size for the training set.
  • How the ground truth for the training set was established.

{0}------------------------------------------------

Neuro-Biometrix, Inc.

Image /page/0/Picture/1 description: The image shows a black and white abstract design. The design appears to be composed of interconnected loops or curves, creating a sense of movement and complexity. The thick black lines contrast sharply against the white background, emphasizing the shapes and forms within the design. The overall impression is modern and artistic.

,3151 South Albion Street; Denver, Colorado 80222 (303) 806-0543

Fax (303) 788-4234

18. 510(K) Summary.

Contact Person:I. Kaufman Arenberg, MD
Date Summary Prepared: February 13, 1997
Trade Name:Round Window u Cath™ (RWuC)
Common Name:Middle Ear Catheter
Classification Name:Class II

Equivalence: The RWyC is substantially equivalent to the combination of the middle ear catheter and the tympanostomy tube. Both the RWgC and the middle ear catheters with a fenestrated reservoir tip on the distal end; both are designed to be attached to a syringe at the proximal end; and both are intended to be used to irrigate the middle ear. Both the RWAC and the vent tube allow air to pass between the middle ear and ear canal. This air passage serves to aerate the middle ear, thereby reducing the risk of middle ear infection. The RW4C and the vent tube are also similar in that they are intended to be left in place for longer periods of time than the middle ear catheter (the vent tube for up to a year, and the for up to 14 days).

Description: The RWAC consists of a double lumen (two parallel and attached Tecoffex 80A (polyurethane) extruded tubes) opening into a common fenestrated reservoir tip on the distal end. The soft, flexible reservoir tip of the RWuC will be available in three sizes: 2mm, 3mm, and 4mm diameters. It is intended that the surgeon will choose an appropriate sized tip to fit the anatomy of the patient (for background information about the anatomy in the Round Window Niche See Appendix 7). The double lumen is bent at approximately a 60-75 degree angle slightly proximal to the reservoir tip. One lumen is intended to serve as an in-flow and the other an out-flow. The outflow lumen is intended to (1) relieve or avoid a build up of air and/or fluid pressure in the device; and (2) facilitate the removal of fluid from the device. A silicone, paparella-type vent tube will be placed around the RWAC during manufacturing.

Intended use: The Round Window u Cath™ (RWuC) is intended to be used by any board certified otolaryngologist or otologist to irrigate the entire middle ear or the round window area of the middle ear. During the course of the irrigation treatment, it is intended that the vent tube through which the RW (C's dual lumen passes, will aerate the middle ear. The treating physician will choose, based on what hels will be most beneficial for his/her patient, the following: { }) the exact fluids to be delivered through the RWAC; (2) the proper quantities; and (3) the duration and interval of treatment (up to 14 days). It is intended that the physician load the RWuC with the initial fluid to be delivered prior to placement. The RWAC is intended to remain in place for up to 14 days. Neuro-Biometrix, Inc. recommends the physician use only fluids indicated for this use. The physician should refer to the fluid labelling prior to use with the RWC.

Principles of Operation: The distal end of the RWLC is a soft, flexible, fenestrated-reservoir tip which is designed to allow the surgeon to irrigate various locations in the middle ear. For example, the surgeon can (1) irrigate the middle ear generally by placing the tip in the middle ear space and applying the fluid with a syringe, (2) irrigate the round window area specifically by placing the appropriately sized tip in the round window niche; or (3) place the tip anywhere else in the middle ear he/she feels is appropriate. The RWpC's double lumen runs from the reservoir tip back through a ventilation tube in the ear drum and into the external ear canal where the proximal end will be accessible to the treating physician.

D:\wp51\Mike\IEMDS\510(K)2 February 13, 1997 2:17pm

Inner Ear Innovations for the 21st Century

{1}------------------------------------------------

Technological Characteristics: The RWgC and the predicate devices, the Microtek Irrigation System and the tympanostomy tube are constructed of class VI biocompatible materials (even though the materials are different). Each has been subjected to biocompatibility testing.

Performance Testing: A bench test protocol was developed to assess the performance of the RWAC in terms of pressure vs. flow, distal tip bond strength of the catheter body, and proximal hub bond strength. All tests will be conducted on final sterile product. Sufficient samples will be used for each test. Each sample will also be visually inspected for workmanship and dimensional compliance.

D:\wp51\Mike\IEMDS\510(K)2 February 13, 1997 - 2:17pm

§ 874.3880 Tympanostomy tube.

(a)
Identification. A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.(b)
Classification. Class II.