DEPUY/ACE MODIFIED MEDICAL ASPECT (MMA) FRACTURE STEM
Device Facts
| Record ID | K964704 |
|---|---|
| Device Name | DEPUY/ACE MODIFIED MEDICAL ASPECT (MMA) FRACTURE STEM |
| Applicant | Depuy, Inc. |
| Product Code | LZO · Orthopedic |
| Decision Date | Jan 24, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3353 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The DePuy/Ace MMA Fracture Stem is indicated for cemented or press-fit cementless use as the femoral component in total hip arthroplasty (THA) for replacing the hip joints of patients whose hip joint has been damaged by fracture. The prosthesis may also be used for total hip arthroplasty for patients requiring total hip replacement for inflammatory or non-inflammatory degenerative joint disease or failure of a previous arthroplasty.
Device Story
Cast Co-Cr-Mo hip stem; triangular proximal body; rounded shoulder; straight distal stem; anterior/posterior grooves; 9-16.5mm diameter range. Used as femoral component in total hip arthroplasty (THA). Operates via cemented or press-fit cementless fixation. Features Articul-eze (12/14) taper for modular femoral head attachment. Used by orthopedic surgeons in clinical/OR settings. Replaces damaged hip joints; restores joint function; addresses degenerative disease or previous arthroplasty failure.
Clinical Evidence
Bench testing only; fatigue strength testing performed to demonstrate equivalence to predicate devices.
Technological Characteristics
Material: Cast Co-Cr-Mo alloy. Finish: Bead blasted. Geometry: Triangular proximal body, rounded shoulder, straight distal stem with anterior/posterior grooves. Fixation: Cemented or press-fit cementless. Interface: Articul-eze (12/14) taper for modular femoral heads. Sizes: 9mm to 16.5mm diameter.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to hip joint fracture, inflammatory or non-inflammatory degenerative joint disease, or failure of a previous arthroplasty.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
- DePuy Vision AML Hip Prosthesis
- DePuy Quantum Hip Stem
Reference Devices
- DePuy Co-Cr-Mo or Zirconia Articul-eze taper modular femoral heads (K883460, K952088)
Related Devices
- K030079 — VERSYS BEADED FULLCOAT BOWED REVISION HIP PROSTHESIS · Zimmer, Inc. · Feb 5, 2003
- K191056 — Conformity stem, cemented · United Orthopedic Corporation · Jan 3, 2020
- K141001 — ALPINE HIP STEM; ALPINE HA HIP STEM · Ortho Development · Jul 30, 2014
- K120362 — METAFIX FEMORAL STEM FOR HEMI-ARTHROPLASTY · Corin USA · May 4, 2012
- K992058 — BI-METRIC CCOCR HEAD/NECK REPLACEMENT SYSTEM · Biomet, Inc. · Jul 1, 1999