PROSELECT 3 PERIDONTAL THERAPY SYSTEM
Device Facts
| Record ID | K964596 |
|---|---|
| Device Name | PROSELECT 3 PERIDONTAL THERAPY SYSTEM |
| Applicant | Professional Dental Mfg., Inc. |
| Product Code | ELC · Dental |
| Decision Date | Apr 22, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.4850 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The device for which 510(k) recognition of substantial equivalence is sought is an ultrasonic scaler intended for use during dental and periodontal therapy treatments to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Device Story
Ultrasonic scaler; removes calculus deposits from teeth via vibrating scaler tip. Used in dental/periodontal therapy. Operated by dental professionals in clinical settings. Device applies ultrasonic energy to tip to mechanically dislodge calculus. Output is mechanical vibration; aids clinician in cleaning procedures; improves periodontal health.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Ultrasonic scaler; utilizes ultrasonic vibration for calculus removal. Form factor: dental handpiece with interchangeable tips. Energy source: electrical power converted to ultrasonic mechanical energy. No software or digital connectivity.
Indications for Use
Indicated for dental and periodontal therapy to remove calculus deposits from teeth in patients requiring professional dental cleaning.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Predicate Devices
- Dentsply Cavitron® Ultrasonic 3000
- Dentsply Cavitron® 1010
- Dentsply Cavitron® 660
- Johnson & Johnson Alfa-Sonic Hygienic Scaler Model 300
- Healthco Air Flow II Scaler
Related Devices
- K983727 — BIOSONIC ULTRASONIC SCALER SYSTEM · Coltene/Whaledent, Inc. · Jan 20, 1999
- K051910 — MAXISONIC FAMILY AND SWIFT ULTRASONIC SCALER · D.B.I. America Corp. · Sep 2, 2005
- K062791 — USI DENTAL SCALER, 25XXX SERIES · Ultrasonic Services, Inc. · Dec 11, 2006
- K061448 — BONART ART-P3II & P4 PIEZO ELECTRIC ULTRASONIC SCALER UNITS WITH ACCESSORIES · Bonart Co., Ltd. · Aug 17, 2006
- K051547 — POWERMAX 25, MODEL PM25; POWERMAX 30, MODEL PM30 · Tpc Advanced Technology, Inc. · Jul 14, 2005