UNITRON MODEL ICON AOHP +4 (ICON AOHP +4)
Device Facts
| Record ID | K964508 |
|---|---|
| Device Name | UNITRON MODEL ICON AOHP +4 (ICON AOHP +4) |
| Applicant | Unitron Industries, Inc. |
| Product Code | ESD · Ear, Nose, Throat |
| Decision Date | Nov 27, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 874.3300 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Intended Use
To amplify and transmit sound via air conduction to the ear.
Device Story
Behind-the-ear (BTE) hearing aid; amplifies and transmits sound via air conduction. Features four fitter controls (power, compression limiting, active low cut tone, high cut tone); Class D circuitry for battery efficiency; telecoil with preamplifier. User-operated volume control and microphone/telecoil/off switch. Powered by size 13 battery. Assembled from standard industry components. Used by patients to improve hearing.
Clinical Evidence
Bench testing only; compliance with ANSI S3.22-1987 standard.
Technological Characteristics
Behind-the-ear form factor; Class D circuitry; telecoil with preamplifier; four fitter controls (power, compression, low/high cut tone); size 13 battery power; ANSI S3.22-1987 compliant.
Indications for Use
Indicated for individuals requiring sound amplification via air conduction to the ear.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
Related Devices
- K964504 — UNITRON MODEL ICON AOHPL +4 · Unitron Industries, Inc. · Nov 27, 1996
- K964510 — UNITRON MODEL ICON AOHPD +4 (ICON AOHPD +4) · Unitron Industries, Inc. · Nov 27, 1996
- K960155 — UNITRON MODEL SOUND F/XP · Unitron Industries, Inc. · Mar 7, 1996
- K964511 — UNITRON MODEL SOUND F/XD +4 (SOUND F/XD +4) · Unitron Industries, Inc. · Nov 27, 1996
- K964502 — UNITRON MODEL SOUND F/XP +4 · Unitron Industries, Inc. · Nov 27, 1996