K Number
K964485

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-02-04

(88 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Not Found

Device Description

The PORTpro is a free-standing, digital camera based electronic portal imaging device (EPID) for use as an accessory to existing x-radiation therapy systems operating in the megavoltage range (1 - 15 MeV). The PORTpro displays high quality portal images on a computer monitor in virtually real time and also provides archiving and image processing capabilities.

AI/ML Overview

This document describes the PORTpro Electronic Portal Imaging System, a device used as an accessory to x-radiation therapy systems. However, the provided text does not contain any information regarding acceptance criteria, studies, sample sizes, ground truth establishment, expert qualifications, or adjudication methods.

Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as this information is missing from the provided input.

The document primarily focuses on the device's description, its predicate devices, and a comparison of its technological characteristics. It mentions that "The differences between the systems raise no new issues of safety or effectiveness," but does not provide any data or studies to support this claim or any other performance claim.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENE

K964485

i his summary of safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

FEB - 4 1997

Submitter:

Eliav Ltd., POB 10403, Haifa 26115, Israel Tel: +972.4.842.1541 Fax: +972.2.841.1129

Name of the Device: PORTpro Electronic Portal Imaging System

Predicate Devices: The PORTpro is substantially equivalent to a combination of the SRI 100 Radiotherapy Imaging System manufactured by Phillips Medical Systems and the Huestis Flexi-Holder manufactured by Huestis Medical Products.

Description of the Device: The PORTpro is a free-standing, digital camera based electronic portal imaging device (EPID) for use as an accessory to existing x-radiation therapy systems operating in the megavoltage range (1 - 15 MeV). The PORTpro displays high quality portal images on a computer monitor in virtually real time and also provides archiving and image processing capabilities. The real time display of images is advantageous insofar as treating physicians and therapy technicians are able to quickly produce position verification images without time consuming set up and development of radiographic films. This is helpful in radiation therapy since frequent monitoring of patient positioning can reduce the frequency of discrepancies in (x-ray) field placement. and improve geometric accuracy of the treatments.

Because the PORTpro has a free-standing detector support system, it offers advantages over other EPID's without this feature, i.e., easy positioning of the imaging system, adjustable and relatively large effective field of view, and the ability to do "non-couch" total body irradiation (TBI) and total lymphatic irradiation (TLI).

TechnologicalCharacteristicPORTproSRI 100Flexi-holder
Acquisition regimeFull fieldFull fieldNA
Acquisition matrix746 x 582512 x 256NA
Spatial resolution1.0 mm2.0 mmNA
Pixel resolution (bits)1616NA
High VoltageNoNoNo
CCD/VideoCCDCCDNA
Height75"NA75"
Weight136 lbs.-----110 lbs.
Base14" x 30"NA19" x 22"

Comparison of Technological Characteristics:

The differences between the systems raise no new issues of safety or effectiveness.

Date

Mr. Reuven Levinson, President

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.