K Number
K964459

Validate with FDA (Live)

Date Cleared
1996-12-18

(41 days)

Product Code
Regulation Number
864.7290
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated use - Factor deficient plasma, Factor - X is a human plasma immunodepleted of the specific factor and intended for use in the quantitative determination of the specific factor levels in patients suspected of congenital or acquired deficiency of this specific coagulation protein and is performed by clotting assay.

Device Description

Factor deficient plasma to be free of antigen of Factor X utilized in in vitro diagnostic use.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the "Factor deficient coagulation plasma - X" device:

It is important to note that the provided documents (K964459) are from 1996, and the regulatory and scientific standards for device studies have significantly evolved since then. The information available in these summaries is quite limited compared to what would be required for a modern submission.

Based on the provided text, the device in question is a "Factor deficient coagulation plasma - X" intended for in vitro diagnostic use in quantitative determination of Factor X levels. The summary primarily focuses on comparing the new device to a predicate device (Sigma - K843310 - Factor X) to establish substantial equivalence.


1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state numerical acceptance criteria in the way modern device submissions do (e.g., a specific sensitivity or specificity threshold, or an agreement percentage). Instead, acceptance is implied by demonstrating substantial equivalence to a predicate device across various attributes.

Here's an interpretation of the performance-related comparisons, framed as implicit criteria and reported performance:

Acceptance Criteria (Implied)Reported Device Performance (vs. Predicate)
Use: Indicated for determining coagulation of plasmaYes (Matches predicate)
Use: In vitro diagnostic useYes (Matches predicate)
Use: Used as a quantitative assayYes (Matches predicate)
Design: Packaging either frozen or dry/lyophilizedYes (Matches predicate)
Design: Usable with different instruments/reagentsYes (Matches predicate)
Materials: Donor human plasmaYes (Matches predicate)
Materials: Various buffersYes (Matches predicate)
Performance Testing: Compare assay to known sampleYes (Matches predicate)
Performance Testing: Negative for HIV 1/2 and HBsAG (FDA approved test)Yes (Matches predicate)
Performance Testing: Negative for HCV and HIV-1ag (FDA approved test)Yes (Differences cited, but claimed for new device)
Donor Criteria: Deficiency of relevant factor less than 1%Yes (Predicate "not known")
Donor Criteria: Negative for HIV and HBsAGYes (Matches predicate)
Donor Criteria: Negative for HCV and HIV-1agYes (Predicate "not known")
Donor Criteria: No inhibitor presentYes (Predicate "not known")

Key takeaway for Acceptance Criteria: The primary acceptance criterion appears to be "substantially equivalent to the predicate device, Sigma - K843310 - Factor X, for stated attributes, with additional safety features (HCV/HIV-1ag testing) as an improvement."


2. Sample Size Used for the Test Set and Data Provenance

The document does not specify sample sizes for any test set. It mentions "Performance Testing: Compare assay to known sample" under the attributes, but no details on how many samples, what type of samples, or their origin are provided.

  • Sample Size (Test Set): Not specified.
  • Data Provenance: Not specified. It can be inferred that samples would be of human origin, related to coagulation testing, but no specific country or whether it was retrospective/prospective is mentioned.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not provide any information about experts or their qualifications for establishing ground truth for any test set.


4. Adjudication Method (for the Test Set)

The document does not specify any adjudication method. Given the nature of a laboratory diagnostic reagent comparison, it's unlikely that a multi-reader adjudication process (like 2+1) would be directly applicable in the same way it is for image-based diagnostic AI. The "ground truth" for a Factor X level would typically be established by a reference method or known concentration.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study (comparing human reader performance with and without AI assistance) is specific to AI/CAD devices, especially in imaging, where human interpretation is a key part of the diagnostic process. The "Factor deficient coagulation plasma - X" is a laboratory reagent, not an AI or CAD system, so such a study would not be relevant.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The concept of "standalone performance" for an algorithm is not directly applicable here because the device is a reagent, not an algorithm. The performance of the reagent would be assessed in a laboratory setting when used with instruments and other reagents to perform a coagulation assay. The performance data mentioned (e.g., negative for certain viruses, deficiency of relevant factor) relates to the quality and characteristics of the plasma itself, not an algorithm's output.

However, if we interpret "standalone" as the performance of the reagent itself (before human interpretation of the final assay results), then the performance attributes listed under "Performance Testing" and "Donor Criteria" could be considered an assessment of the reagent's inherent characteristics. The document implies that the device performs in a manner consistent with a factor-deficient plasma, but details on how that performance was measured (e.g., specific quantitative results, precision, accuracy) are absent.


7. The Type of Ground Truth Used

The document does not explicitly define the "ground truth" in detail. However, based on the context of the device:

  • For "compare assay to known sample": The ground truth would likely be the known concentration/activity of Factor X in reference samples or spiked samples.
  • For "Negative by FDA approved test for HIV 1/2 and HBsAG" etc.: The ground truth would be the results from the FDA-approved reference tests for these viral markers.
  • For "Deficiency of relevant factor less than 1%": The ground truth would be established through a specific laboratory assay (presumably a reference method) that quantifies Factor X levels in donor plasma.

Essentially, it relies on established laboratory methodologies and reference materials as ground truth.


8. The Sample Size for the Training Set

The document does not mention a training set. This is because the device is a laboratory reagent and not an AI/machine learning algorithm that requires a training phase.


9. How the Ground Truth for the Training Set Was Established

Since there is no training set mentioned (as it's not an AI device), this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a world map projection. The map is divided into several lobes, each representing a different region of the world. The map is also covered with a grid of lines, which are used to show the latitude and longitude of different locations. The map appears to be a type of interrupted projection, which is designed to minimize distortion.

UNIVERSAL REAGENTS, INC. 2858 North Pennsylvania Street Indianapolis, Indiana 46205 PHONE: (317) 926-0015 FAX: (317) 926-0014

K964459

DEC 1 8 1996

Non-Confidential Summary of Safety and Effectiveness

November 1, 1996

page 1 of 2

Universal Reagents, Inc.
2858 N. Pennsylvania St.
Indianapolis, IN 46205
Tel - (317) 926-0006
Fax - (317) 926-0014

Official contact:Jorge Miller, Director, Coagulation Products
Proprietary or Trade Name:Factor deficient coagulation plasma - X
Common/Usual Name:Qualitative and Quantitative Factor Deficiency Test - X
Classification Name:Qualitative and Quantitative Factor Deficiency Test
Intended device:Factor deficient coagulation plasma - X
Predicate devices:Sigma - K843310 - Factor X
Device description:Factor deficient plasma to be free of antigen of Factor X utilized in in vitro diagnostic use.

Intended use:

Indicated use - Factor deficient plasma, Factor - X is a human plasma immunodepleted of the specific factor and intended for use in the quantitative determination of the specific factor levels in patients suspected of congenital or acquired deficiency of this specific coagulation protein and is performed by clotting assay.

Environment of use: Clinical laboratories

Comparison to predicate devices:

AttributeIntended productSigma
Use
Indicated for use in determinationof coagulation of plasmaYesYes
In vitro diagnostic useYesYes
Used as a quantitative assayYesYes

{1}------------------------------------------------

Non-Confidential Summary of Safety and Effectiveness (continued)

November 1, 1996

page 2 of 2

Comparison to predicate devices: (continued)

AttributeIntended productsSigma
Design
Packaging either -Frozen or Dry / lyophilizedYesYes
Can be used with differentinstruments and reagents permanufacturer instructionsYesYes
Materials
Donor human plasmaYesYes
Various buffersYesYes
Performance Testing
Compare assay to known sampleYesYes
Negative by FDA approved test forHIV 1/2 and HBsAGYesYes
Negative by FDA approved test forHCV and HIV-1agYesnot known
Donor criteria
Deficiency of relevant factorless than 1%Yesnot known
Negative for HIV and HBsAGYesYes
Negative for HCV, HIV-1agYesnot known
No inhibitor presentYesnot known

Differences

The only difference is that the intended product is claimed to be negative for HCV and HIV-1ag by an FDA approved test.

.

Any other differences that do exist would not have a significant effect on the safety or effectiveness of the device.

§ 864.7290 Factor deficiency test.

(a)
Identification. A factor deficiency test is a device used to diagnose specific coagulation defects, to monitor certain types of therapy, to detect coagulation inhibitors, and to detect a carrier state (a person carrying both a recessive gene for a coagulation factor deficiency such as hemophilia and the corresponding normal gene).(b)
Classification. Class II (performance standards).