K Number
K964256

Validate with FDA (Live)

Date Cleared
1997-05-06

(193 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The primary purpose of the device is to protect surgical instruments during sterilization, storage and transportation to the surgical suite. This product is intended to be used by hospital staff during the preparation of the sterile instrument kits.

Device Description

The "Cygnus Instrument Protectors" liners and pouches are constructed with reticulated polyurethane ester foam. The material's open cell structure allows for the free flow of air, moisture and heat and is virtually free of particulate matter. The variety of sizes and shapes allows for greater flexibility in padding the instruments. Tray liners are available in precut sheets and rolls. uthe pouches are available in three sizes of single pouches and two sizes of multi pouch rolls. The pouches are formed by heat sealing the edges and seams. Each product is available in individual boxes as well as bulk case packaging. The products are shipped and marked as "non sterile" and "disposable". Each box is marked with a control lot number and universal bar code.

AI/ML Overview

This device is for protecting surgical instruments during sterilization, storage, and transportation. The provided text is a 510(k) Summary from 1996, which is a premarket submission made to the FDA to demonstrate that the device to be marketed is at least as safe and effective, that is, substantially equivalent, to a legally marketed predicate device. This type of document does not typically contain detailed acceptance criteria and study data in the way a clinical trial report would. Instead, it focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and intended use.

Therefore, many of the requested sections (e.g., sample size for test/training sets, experts for ground truth, MRMC study, ground truth type) are not applicable or cannot be extracted from this specific document.

Here's an analysis based on the provided text, addressing the applicable points:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a tabular format with specific thresholds. Instead, it compares the technological characteristics of the "Cygnus Instrument Protectors" to those of predicate devices to establish substantial equivalence. The "performance" here relates to these physical characteristics.

CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance ("Cygnus Instrument Protectors")
Tensile StrengthAppropriate and consistent with intended use30 lbs. psi.
ElongationAppropriate and consistent with intended use200%
Tear StrengthAppropriate and consistent with intended use4.0 lbs. psi.
DensityAppropriate and consistent with intended use1.9 lb./cu. ft.
PorosityAppropriate and consistent with intended use50 Pore per sq. in.
MaterialPolyurethane ester foamReticulated polyurethane ester foam
Cell StructureOpen cell structureOpen cell structure
Particulate MatterVirtually free of particulate matterVirtually free of The variety of sizes and shapes allows particulate matter.
Sterilization MethodSteam autoclave (121°C to 134°C)Steam autoclave (121°C to 134°C)

Explanation of "Acceptance Criteria" (Implied): The document states: "The technological characteristics of the Cygrass Instrument Protectors are identical to the predicate devices, as well as a number of other manufacturers producing this type of product. The tensile streagth [sic] (30 lbs. psi.), elongution [sic] (200%), tour [sic] strength (4.0 lbs. psi), density ( 1.9 lb./cu. ft.) and porosity (50Pore per sa. in.) all define this product as being appropriate and consistent with the intended use of the product." This indicates that the reported performance values are considered "accepted" because they are "identical to the predicate devices" and "appropriate and consistent with the intended use."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified. The document describes physical characteristics, not a clinical test set with a specific number of samples or patients. It refers to material properties.
  • Data Provenance: Not specified. This information pertains to the material's inherent properties and manufacturing specifications rather than patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not applicable. The "ground truth" for the physical and material characteristics would be established through standard engineering and materials testing, not expert consensus in the clinical sense.

4. Adjudication Method for the Test Set

  • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for clinical image interpretation or diagnostic performance assessments, not for verifying material properties.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No, an MRMC comparative effectiveness study was not done. This type of study assesses the impact of a device (often AI-assisted) on human reader performance, which is not relevant for a passive instrument protector in a 510(k) summary focused on material properties and substantial equivalence.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

  • No, a standalone performance study was not done. This device is not an algorithm; it's a physical product.

7. The Type of Ground Truth Used

  • The "ground truth" for this device's performance is based on engineering and material specifications (tensile strength, elongation, tear strength, density, porosity) and the established performance characteristics of legally marketed predicate devices.

8. The Sample Size for the Training Set

  • Not applicable. This device is a physical product, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for an algorithm, there is no ground truth established for it.

{0}------------------------------------------------

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

M. SAFE MEDICAL DEVICES ACT OF 1990 STATEMENT OF SAFETY AND EFFECTIVENESS.

196425 "510(k) Summary" MAY - 6 1997 October 1, 1996 Contact person: Shaun Sweeney (203) 265-9550 phone (203) 265-9961 fax "Cygnus Instrument Protectors" Trade Name : Tray liners, instrument pouches Common Name: Sterilization wrap, pack Classification Name: sterilization wrapper, bag or accessories (per 21 CFR section 880.6850).

Predicate device: Manufacturer Devon Industries, 14569 Titus Street, Panorama City, California 91402 9530 Desoto Drive, Chatsworth, California 91311

Product - Devon Tiprotectors, tray pads and liners 510(k) Number - K771123 General and Plastic Surgery Device Panel Classification - Class II Regulatory Citation - 21 CFR ch 1 (4-1-95)

Product Description

The "Cygnus Instrument Protectors" liners and pouches are constructed with reticulated polyurethane ester foam. The material's open cell structure allows for the free flow of air, moisture and heat and is virtually free of The variety of sizes and shapes allows particulate matter. for greater flexibility in padding the instruments. Tray liners are available in precut sheets and rolls. uthe pouches are available in three sizes of single pouches and two sizes of multi pouch rolls. The pouches are formed by heat sealing the edges and seams. Each product is available in individual boxes as well as bulk case The products are shipped and marked as "non packaging. sterile" and "disposable". Each box is marked with a control lot number and universal bar code.

{1}------------------------------------------------

Summery: Intended Use

The primary purpose of the device is to protect surgical instruments during sterilization, storage and transportation to the surgical suite. This product is intended to be used by hospital staff during the preparation of the sterile instrument kits.

  • The liners are first placed into the empty sterilization tray. ●
  • The instruments are then placed onto the Cygnus liner (or in the pouch) and the trays . are wrapped in CSR wrap using conventional wrapping methods.
  • Cygnus tray liners are to be used with steam autoclave sterilization only at . temperature settings ranging between - 121°C to 134 ℃.
  • The sterile wrapped kit remains closed until arrival within the sterile prepped . operating room.
  • . The liners may continue to be used during the procedure as a sterile field on which to place instruments.
  • Upon the completion of the procedure the liners are disposed of in accordance with . applicable medical waste regulations.

Technological Characteristics

The technological characteristics of the Cygrass Instrument Protectors are identical to the predicate devices, as well as a number of other massafisturers producing this type of product. The tensile streagth (30 lbs. psi.), elongution (200%), tour strength (4.0 lbs. psi), density ( 1.9 lb./cu. ft.) and porosity (50Pore per sa. in.) all define this product as being appropriate and consistent with the intended use of the product. ..............................................................................................................................................................................

20

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).