METAL BACKED ACETABULAR COMPONENT
Device Facts
| Record ID | K963998 |
|---|---|
| Device Name | METAL BACKED ACETABULAR COMPONENT |
| Applicant | Encore Orthopedics, Inc. |
| Product Code | LPH · Orthopedic |
| Decision Date | Jan 27, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3358 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The metal backed acetabular component is intended for treatment of patients who are candidates for total hip arthroplasty because the natural femoral head and neck and/or acetabulum have been affected by osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis or femoral neck fracture, and revision arthroplasty where bone loss is minimal. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.
Device Story
Metal backed acetabular component for total hip arthroplasty; hemispherical Ti-6Al-4V shell with plasma-sprayed exterior for press-fit fixation; includes optional screw holes for secondary fixation; ultra-high molecular weight polyethylene (UHMWPE) liners available in neutral or 10° hooded designs; hooded design addresses subluxation risk from soft tissue laxity; used by orthopedic surgeons in clinical settings; device relieves hip pain and restores motion; provides stable prosthetic interface for femoral head.
Clinical Evidence
Bench testing only. Attachment strength of liners to shell tested; results sufficient for in-vivo loading.
Technological Characteristics
Materials: Ti-6Al-4V (wrought/forged or cast) shell, UHMWPE liners. Surface: Plasma-sprayed roughened exterior. Geometry: Hemispherical shell with rim flare. Fixation: Press-fit with optional cancellous bone screws. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, femoral neck fracture, or revision arthroplasty with minimal bone loss.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Related Devices
- K964227 — METAL ACETABULAR COMPONENT · Encore Orthopedics, Inc. · Jan 14, 1997
- K960633 — PRESS-FIT FLARED SHELL · Encore Orthopedics, Inc. · Oct 1, 1996
- K961526 — PRESS-FIT POROUS SHELL · Encore Orthopedics, Inc. · Sep 17, 1996
- K141043 — CS2 MULTI-HOLE ACETABULAR SHELL · Consensus Orthopedics, Inc. · May 19, 2014
- K071784 — DEPUY PINNACLE WITH GRIPTON ACETABULAR CUPS · DePuy Orthopaedics, Inc. · Jul 25, 2007