K Number
K963662

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-02-26

(166 days)

Product Code
Regulation Number
882.1870
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Galileo Electroencephalograph with Evoked Potential option is indicated for use in the detection and display of evoked responses of electrical activity of the brain and the peripheral nervous system.

Device Description

ACQ board: Analog input 32 channels, Analog output 32 channels, A/D conversion 12 bit, 10 micro sec/point, 4 converter, Accumulator 32 bit, Processor Motorola MC68000, Resident memory 1 MB, Trigger input TTL. Stimulators: a- General features: Working class I type of equipment B (acoustic and visual) I type of equipment CF (electrical), Safety requirements meets standard IEC 601-1-1 (CEI 62-5), Size 220 mm, 450 mm, 80 mm, Weight approx. 2 Kg, Power supply + 15V (0.7 A) -15V (0.7A) +5V (1A), Power absorption approx. 25 VA, Working conditions +5C/+35C; 30-75%RH, Atm pressure 700/1060 mBar, Storage and conveyance conditions -10/50C; 10-95% RH, Atm pressure 500/1060 mBar, Degree of protection against infiltration of liquids and solids IP20. b- Acoustic stimulator: Transducer Telephonics TDH39, Stimulation modes Tone, Click, Noise, Stimulation rate from 50 ms to 9980 ms (step 10 ms), Tone frequency 250,500,1000,1500,2000,2500,3000,4000,5000,7000 8000 Hz, Stimulus output Right monoaural, Left monoaural, Right monoaural + controlateral noise, Left monoaural + con- trolateral noise, binaural, Acoustic level 0,10,20,30,40,50,60,70,75,80,90 dB SL, Threshold adjustment -4,-2,0, +2, +4, +6 dB, Noise level relative to the stimulus's level -40,-30,-20,-10,0, + 10 dB, Polarity of the stimulus (in Click mode) Condensation, Rarefaction, Alternate, Duration tone mode: from 10 to 150 ms, with 10 ms steps click mode: from 50 to 960 us, 10 us steps. c- Visual stimulator: The stimulator generates two video images (pattern A and B), that are alternatively displayed on screen in response to a trigger command. Generally, each picture is composed of an arbitrary bit map of 512x256 pixels; in detail the current resident software, divides the image in 5 fields: 4 quadrants and a fixation area. Each field can contain one of the following textures: black white spatial frequency horizontal bars 1 spatial frequency horizontal bars 2 spatial frequency vertical bars 1 spatial frequency vertical bars 2 spatial frequency chequer board 1 spatial frequency chequer board 2 and the reverse of all of them. A composite video signal 1 Vpp is transmitted to the monitor with separate positive synchronisms and 8 contrast levels (2,5,10,20,30,50,75,100%) related to the maximum contrast level set on the . The line period is 64 us, the frame frequency 60 Hz. d- Electric stimulator: The electric stimulator is constant current type. Other features are: Amplitude from 0 to 50 mA, with two scales (0-5 and 0-50), Precision +/- 10% full range, Duration limited to 0.5 millisec (see Device Hazard Analysis), Isolation between the applied elements and the ground 1500 V AC (see Device Hazard Analysis), Patient's safety limited energy and stimulation frequency; protection against failure of the current sensor, Light indicators stimulus ON, overload, abnormal functioning.

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and the study information as requested.

The provided document is a Safety and Effectiveness Summary for the Galileo Evoked Potential EEG Biosound Esaote. It describes the device, its intended use, and compares it to predicate devices. However, this document does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria.

This type of submission to a regulatory body (like the FDA for a 510(k)) primarily focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than conducting new performance studies against predefined acceptance criteria. The comparison table (807.92(a)(3)) highlights the technical similarities between the Galileo EP Option and its predicate devices (Esaote Phasis and Nicolet Viking II System).

Therefore, based solely on the provided text, I cannot complete the table or answer most of the questions.

Here's what I can extract and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance CriteriaReported Device Performance
No acceptance criteria or performance study results are provided in this document.N/AN/A (No specific performance metrics are reported against defined criteria.)

2. Sample size used for the test set and the data provenance

  • Sample Size (Test Set): Not applicable. This document does not describe a performance study with a test set.
  • Data Provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. This document does not describe a performance study using expert ground truth.

4. Adjudication method for the test set

  • Not applicable. This document does not describe a performance study with a test set requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This document does not mention an MRMC study. The device is an Evoked Potential Electroencephalograph, which is a diagnostic tool, not an AI-assisted interpretation system for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This document does not describe a standalone algorithm performance study. The device itself is a standalone medical device.

7. The type of ground truth used

  • Not applicable. This document does not describe a performance study that required establishing ground truth. The submission relies on demonstrating substantial equivalence to predicate devices, implying that the established clinical principles and diagnostic interpretations for those predicate devices would apply to this new device.

8. The sample size for the training set

  • Not applicable. This document does not describe any machine learning or AI components that would require a training set.

9. How the ground truth for the training set was established

  • Not applicable. There is no training set described in the document.

In summary: The provided text is a 510(k) summary focused on demonstrating substantial equivalence to predicate devices based on technical specifications and intended use. It does not include information about formal performance studies with acceptance criteria, test sets, or ground truth establishment. Such details would typically be found in more comprehensive testing protocols or clinical study reports, which are not part of this "Safety and Effectiveness Summary."

{0}------------------------------------------------

Safety and Effectiveness Summary Galileo Evoked Potential EEG Biosound Esaote

UNDES AOTTE Biosound, Inc. DUPLICATE 8000 Castleway Drive Indianapolis. IN 46250 317 849.1793 800 428.4374 fax 317 841.8616 http://www.biosound.com/

FEB 26 1997

Safety and Effectiveness Summary

The following safety and effectiveness summary has been prepared pursuant to requirements for 510(k) summaries specified in 21 CFR 807.92(a).

807.92(a)(1)

Submitter Information

Gerald A. Richardson, Official Correspondent

Biosound Esaote

8000 Castleway Drive

Indianapolis, Indiana 46250

Phone: (317) 849-1793

Facsimile: (317) 841-8616

Contact Person: Gerald A. Richardson

Date: February 10, 1997

{1}------------------------------------------------

Safety and Effectiveness Summary Galileo Evoked Potential EEG Biosound Esaote

807.92(a)(2)

Trade Name: Galileo Evoked Potential

Common Name: Evoked Potential Electroencephalograph

Classification Name(s): Electroencephalograph

Classification Number: 84 GWQ

807.92(a)(3)

Predicate Device(s)

Additional Substantial Equivalence information is provided in the following substantial Equivalence Comparision Table.

807.92(a)(5)

Intended Use(s)

The Galileo Electroencephalograph with Evoked Potential option is indicated for use in the detection and display of evoked responses of electrical activity of the brain and the peripheral nervous system.

{2}------------------------------------------------

Substantial Equivalence Comparison Table

ItemGalileo EP OptionEsaote Phasis
K922488
General Structure
Intended useEvoked PotentialsEMG-EP
General system approachComputer based equipment withspecialized hardware/peripheralsSame
User input devicesASCII standard keyboardFunction keyboardSame
User output devicesDigital color monitorStandard Centronics printerDedicated thermal printerSame
CPUMC68000MC68010
Operating channels24/322-4
Signal acquisition12 bit A/D conversion16 bit A/D conversion
Trigger InTTLSame
Trigger OutTTLSame
Accumulator32 bitSame
Preset1-99991-32000
Standard EP
Programs
SomatosensoryYesYes
AuditoryYesYes
VisualYesYes
P300YesYes
Electric Stimulator
OutputControlled current 0-50 mAControlled current 0-100mA
Stimulus indicatorYesYes
Overload indicatorYesYes
Delivered energylimitationYesYes
Acoustic Stimulator
OutputEarphone TDH39Same
ModeClicks, Tone, NoiseSame
PolarityRarefaction, condensation,alternateSame
Level0-130 dB (SPL)0-124 dB (SPL)
Zero level adjustmentYesYes
Visual Stimulators
Output for flashstimulatorYesYes
OutputVideo compositeSame
TypeCheckerboard, bars, stripsSame
ModeReversal, double patternSame

{3}------------------------------------------------

807.92(a)(4)


Device Description

ACQ board
Analog input32 channels
Analog output32 channels
A/D conversion12 bit, 10 $\mu$ sec/point, 4 converter
Accumulator32 bit
ProcessorMotorola MC68000
Resident memory1 MB
Trigger inputTTL

Stimulators

a- General features
Working classI type of equipment B (acoustic and visual)I type of equipment CF (electrical)
Safety requirementsmeets standard IEC 601-1-1 (CEI 62-5)
Size220 mm, 450 mm, 80 mm
Weightapprox. 2 Kg
Power supply+ 15V (0.7 A)-15V (0.7A)+5V (1A)
Power absorptionapprox. 25 VA
Working conditions+5C/+35C; 30-75%RH
4

·

{4}------------------------------------------------

Atm pressure700/1060 mBar
Storage and conveyance conditions-10/50C; 10-95% RH
Atm pressure500/1060 mBar

Degree of protection against infiltration of liquids and solids

IP20
--------

b- Acoustic stimulator

| |

i

TransducerTelephonics TDH39
Stimulation modesTone, Click, Noise
Stimulation ratefrom 50 ms to 9980 ms (step 10 ms)
Tone frequency250,500,1000,1500,2000,2500,3000,4000,5000,70008000 Hz
Stimulus outputRight monoaural, Left monoaural, Right monoaural+ controlateral noise, Left monoaural + con-trolateral noise, binaural
Acoustic level0,10,20,30,40,50,60,70,75,80,90 dB SL
Threshold adjustment-4,-2,0, +2, +4, +6 dB
Noise level relative to the stimulus's level-40,-30,-20,-10,0, + 10 dB
Polarity of the stimulus (in Click mode)Condensation, Rarefaction, Alternate
Durationtone mode: from 10 to 150 ms, with 10 ms stepsclick mode: from 50 to 960 us, 10 us steps

5

{5}------------------------------------------------

c- Visual stimulator

The stimulator generates two video images (pattern A and B), that are alternatively displayed on screen in response to a trigger command.

Generally, each picture is composed of an arbitrary bit map of 512x256 pixels; in detail the current resident software, divides the image in 5 fields: 4 quadrants and a fixation area. Each field can contain one of the following textures:

  • black white spatial frequency horizontal bars 1 spatial frequency horizontal bars 2 spatial frequency vertical bars 1 spatial frequency vertical bars 2 spatial frequency chequer board 1 spatial frequency chequer board 2
    and the reverse of all of them.

A composite video signal 1 Vpp is transmitted to the monitor with separate positive synchronisms and 8 contrast levels (2,5,10,20,30,50,75,100%) related to the maximum contrast level set on the .

The line period is 64 us, the frame frequency 60 Hz.

d- Electric stimulator

The electric stimulator is constant current type. Other features are:

Amplitudefrom 0 to 50 mA, with two scales (0-5 and 0-50)
Precision+/- 10% full range
Durationlimited to 0.5 millisec (see Device Hazard Analysis)
Isolation between the applied elements and the ground1500 V AC (see Device Hazard Analysis)
Patient's safetylimited energy and stimulation frequency; protectionagainst failure of the current sensor
Light indicatorsstimulus ON, overload, abnormal functioning.

§ 882.1870 Evoked response electrical stimulator.

(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).