SIGMA DIAGNOSTIC CX -3 CALIBRATION STANDARD LEVEL 1
Device Facts
| Record ID | K963533 |
|---|---|
| Device Name | SIGMA DIAGNOSTIC CX -3 CALIBRATION STANDARD LEVEL 1 |
| Applicant | Sigma Diagnostics, Inc. |
| Product Code | JIX · Clinical Chemistry |
| Decision Date | Oct 29, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Intended Use
Sigma Diagnostics CX®3 Calibration Standard Level 1 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasma, urine, or cerebrospinal fluid. Calibration Standard Level 1 should be used with CX®3 Calibration Standard Level 2.
Device Story
Calibration reagent for Beckman SYNCHRON CX®3 clinical chemistry analyzer; contains defined concentrations of BUN/urea, calcium, total CO2, creatinine, potassium, sodium, chloride, and glucose; used in clinical laboratory settings by trained personnel; ensures system accuracy for quantitative diagnostic testing; functions as reference standard for instrument calibration; supports clinical decision-making by maintaining analytical precision of patient sample measurements.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Liquid reagent containing specified concentrations of analytes (Calcium, Chloride, Creatinine, Glucose, Potassium, Sodium, Urea, BUN); intended for use with Beckman SYNCHRON CX®3 System; non-reactive components included; no specific materials or software algorithms described.
Indications for Use
Indicated for calibration of the Beckman SYNCHRON CX®3 System for quantitative determination of BUN/urea, calcium, total CO2, creatinine, potassium, sodium, chloride, and glucose in serum, plasma, urine, or cerebrospinal fluid.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Beckman Calibration Standard 1 (Part No. 443360)
Related Devices
- K090283 — CALIBRATOR 1,2, AND 3 MODELS: GEN-CAL1, GEN-CAL2, AND GEN-CAL3 · Genchem, Inc. · Jun 5, 2009
- K992845 — HICHEM CALIBRATOR 1 KIT, HICHEM CALIBRATOR 2 KIT, HICHEM CALIBRATOR 3 KIT, CALIBRATION STANDARDS KIT · Elan Holdings, Inc. · Oct 18, 1999
- K963543 — SIGMA DIAGNOSTICS CX -3 CALIBRATION STANDARD, LEVEL 2 · Sigma Diagnostics, Inc. · Oct 30, 1996
- K981706 — AMM CAL, BIL CAL, CO2 CAL, HDL CAL, IRON/MG CAL, LDL CAL, MC CAL, UPRO CAL, AND ISE CALIBRATOR LEVELS 1 AND 2 · Abbott Laboratories · Jul 9, 1998
- K060120 — TECO CX3 REAGENT SET FOR SYNCHRON CX SYSTEM · Teco Diagnostics · Mar 27, 2006