Intended Use

The DePuy Motech PEAK Fixation System is intended for use in treating fractures of small bones such as the metacarpals, ulna, radius, humerus and metatarsals and in treating fractures of the lateral malleolus, olecranon, and intra-articular distal tibia.

Device Description

The system consists of 5 sizes of channeled reconstruction plates ranging from 26mm to 104mm in length, 13 sizes of 3.5mm cortical screws ranging from 16 to 30mm in length and a steplock washer. The reconstruction plates are 3mm thick and 9mm wide. All plates incorporate a tooth pattern, on one outer surface. The teeth are designed to interlock with the "tabs" on the steplock washer.

AI/ML Overview

This 510(k) summary describes a medical device, the DePuy Motech PEAK Fixation System, and its substantial equivalence to predicate devices, but does not contain any information regarding acceptance criteria or a study proving the device meets specific acceptance criteria in the manner requested.

The provided text focuses on:

  • Device Description and Intended Use: Explaining what the device is, what it's used for, and its components.
  • Basis of Substantial Equivalence: Comparing its materials, design, and intended use to existing devices.
  • Mechanical Testing: A statement that its mechanical properties (rigidity, strength, stiffness in 4-point bending) exceed those of predicate devices.

However, it lacks the specific details required to answer your questions about acceptance criteria, clinical study methodology, expert involvement, and ground truth establishment. This type of information is typically found in detailed study reports, clinical trial summaries, or more extensive sections of a 510(k) submission, which are not present in the provided excerpt.

Therefore, I cannot populate the table or answer the subsequent questions based solely on the input provided. The summary only mentions "mechanical testing shows that..." without describing the acceptance criteria for that testing or the full study details.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

K96 3350

NAME OF FIRM:DePuy Inc.700 Orthopaedic DriveWarsaw, Indiana 46581-0988NOV 14 1996
510(K) CONTACT:Cheryl HastingsManager, Clinical Affairs
TRADE NAME:DePuy Motech PEAK Fixation System
COMMON NAME:Bone Plate
CLASSIFICATION:888.3030 Single/multiple component metallic bone fixationappliance and accessories
DEVICE PRODUCT CODE:Product code: 87 HRS
SUBSTANTIALLYEQUIVALENT DEVICES:> ACE 100° Tubular Plate> Synthes (USA) LC-DCP Plate Set System> Synthes (USA) 1/3 Tubular Stainless Steel Plates

INTENDED USE AND DEVICE DESCRIPTION:

The DePuy Motech PEAK Fixation System is intended for use in treating fractures of small bones such as the metacarpals, ulna, radius, humerus and metatarsals and in treating fractures of the lateral malleolus, olecranon, and intra-articular distal tibia. The system consists of 5 sizes of channeled reconstruction plates ranging from 26mm to 104mm in length, 13 sizes of 3.5mm cortical screws ranging from 16 to 30mm in length and a steplock washer. The reconstruction plates are 3mm thick and 9mm wide. All plates incorporate a tooth pattern, on one outer surface. The teeth are designed to interlock with the "tabs" on the steplock washer.

BASIS OF SUBSTANTIAL EQUIVALENCE:

The DePuy Motech PEAK Fixation System is similar in material, design and intended use to the ACE 100° Tubular Plate and the Synthes LC-DCP Plate Set System. All of these systems consist of CP Titanium bone plates and screws intended to treat fractures of small bones. The ACE plate has a length of 97mm, 8 holes, a width of approximately 9mm and a thickness of approximately 1.3mm and is intended to be used with cancellous screws with diameters of 4.0mm or less. The Synthes system contains plates with lengths from 25mm with 2 to 10 screw holes per plate, and is designed to be used with 3.5mm cortical bone screws. The 3.5mm screws range in length from 10mm to 50mm.

Mechanical testing shows that in 4 point bending the rigidity, strength and stiffness of the DePuy Motech PEAK Fixation System channeled reconstruction plates exceed those of the ACE 100° Tubular Plate and Synthes 1/3 Tubular Stainless Steel Plates.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.