QA-40M DEFIBRILLATOR TESTER

K963190 · Metron U.S., Inc. · DRL · Jul 1, 1997 · Cardiovascular

Device Facts

Record IDK963190
Device NameQA-40M DEFIBRILLATOR TESTER
ApplicantMetron U.S., Inc.
Product CodeDRL · Cardiovascular
Decision DateJul 1, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5325
Device ClassClass 2

Intended Use

The Metron U.S. QA-40M Defibrillator Analyzer is a precision instrument for testing defibrillators, and for ensuring that defibrillators comply with the specified requirements. The QA-40M tester is a device that is connected to the output of a defibrillator and is used to measure the energy delivered by the defibrillator into a standard resistive load and can also provide waveform information. The QA-40M test for ANSI/AAMI DF39 standards. The QA-40M Defibrillator Analyzer's main function is to measure the energy output of a defibrillator. The instrument has a built-in load resistance of 50 ohm, which roughly corresponds to the impedance of the human body. The defibrillator pads are placed on the QA-40M contact plates. Thus, the defibrillator is connected through load resistance. When the defibrillator is discharged, QA-40M will calculate and display the energy delivered. The QA-40M Defibrillator Analyzer performs the following tests and features (1) energy and cardioversion measurement, (2) peak voltage and current reading, (3) storage and playback of output waveform, (4) 12 lead ECG simulation. (5) ECG, performance and arrhythmia simulation, (6) transcutaneous cardiac pacemaker testing (module is optional), (7) integrated pacemaker loads selectable from 50 ohm to 2500 ohm, (8) automatic defibrillator test procedures, (9) large graphic display and (10) RS-232C and Centronic printer interface.

Device Story

QA-40M Defibrillator Analyzer serves as precision test equipment for defibrillators; connects to defibrillator output via 50-ohm resistive load simulating human body impedance. Device captures high-voltage discharge; calculates delivered energy; measures peak voltage/current; records/plays back waveforms. Additional functions include 12-lead ECG simulation, arrhythmia simulation, and optional transcutaneous cardiac pacemaker testing with selectable loads (50-2500 ohm). Operated by biomedical technicians/clinical engineers in hospital/clinical environments. Output displayed on integrated graphic screen; supports RS-232C/Centronic printer interfaces for documentation. Facilitates regulatory compliance testing per ANSI/AAMI DF39 standards; ensures defibrillator reliability for patient safety.

Clinical Evidence

Bench testing only; device performance verified against ANSI/AAMI DF39 standards for defibrillator testing and measurement accuracy.

Technological Characteristics

Precision instrument with 50-ohm fixed load resistance for defibrillator testing; optional module for pacemaker testing (50-2500 ohm). Features include 12-lead ECG simulation, arrhythmia simulation, and waveform storage/playback. Connectivity via RS-232C and Centronic printer ports. Large graphic display for data output. Complies with ANSI/AAMI DF39 standards.

Indications for Use

Indicated for testing and verification of defibrillator performance, including energy delivery, waveform accuracy, and pacemaker function, for use by clinical engineering or biomedical technicians in healthcare settings.

Regulatory Classification

Identification

A defibrillator tester is a device that is connected to the output of a defibrillator and is used to measure the energy delivered by the defibrillator into a standard resistive load. Some testers also provide waveform information.

Related Devices

Submission Summary (Full Text)

{0} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. James D. Quinn Metron U.S., Inc. 1345 Monroe N.W., Suite 244 Grand Rapids, Michigan 49505 JUL 1 1997 Re: K963190 QA-40M Difibrillator Tester Regulatory Class: II (two) Product Code: 74 DRL Dated: April 2, 1997 Received: April 4, 1997 Dear Mr. Quinn: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1} Page 2 - Mr. James D. Quinn This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2} Page 1 of 1 510(k) Number (if known): Device Name: Defibrillator Tester ## Indications for Use: The Metron U.S. QA-40M Defibrillator Analyzer is a precision instrument for testing defibrillators, and for ensuring that defibrillators comply with the specified requirements. The QA-40M tester is a device that is connected to the output of a defibrillator and is used to measure the energy delivered by the defibrillator into a standard resistive load and can also provide waveform information. The QA-40M test for ANSI/AAMI DF39 standards. The QA-40M Defibrillator Analyzer's main function is to measure the energy output of a defibrillator. The instrument has a built-in load resistance of 50 ohm, which roughly corresponds to the impedance of the human body. The defibrillator pads are placed on the QA-40M contact plates. Thus, the defibrillator is connected through load resistance. When the defibrillator is discharged, QA-40M will calculate and display the energy delivered. The QA-40M Defibrillator Analyzer performs the following tests and features (1) energy and cardioversion measurement, (2) peak voltage and current reading, (3) storage and playback of output waveform, (4) 12 lead ECG simulation. (5) ECG, performance and arrhythmia simulation, (6) transcutaneous cardiac pacemaker testing (module is optional), (7) integrated pacemaker loads selectable from 50 ohm to 2500 ohm, (8) automatic defibrillator test procedures, (9) large graphic display and (10) RS-232C and Centronic printer interface. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ Use ☐ (Per 21 CFR 801.109) OR ☐ (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K963190 Over-the-Counter,
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%