K Number
K962455

Validate with FDA (Live)

Date Cleared
1997-01-13

(202 days)

Product Code
Regulation Number
882.1870
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

to perform the measurements needed for electromyography (EMG), nerve conduction velocity (NCV, F wave, and H reflex), and evoked potentials (brainstem, visual, and somatosensory), and repetitive nerve stimulation. The purpose of the proposed device is to allow compatibility with high-impedance electrodes. The proposed device allows electrode inouts to be made closer to the source of the signal for reduced signal noise during procedures requiring high-impedance electrodes.

Device Description

The proposed device consists of the existing four-channel preamplifier and a buffered electrode input box with extension cable. These components provide electrode inputs that are closer to the source of the signal during electromyographic (EMG) testing. The intent of this design is to reduce signal noise during procedures requiring high-impedance electrodes. The proposed device is for use with the Cadwell Sierra (K924723) and 6200A (K931428) EMG instruments.

All device components are reusable and supplied non-sterile. The extension cable with electrode input box is compatible with EtO sterilization guidelines for procedures requiring a sterile field. The input box is available with separate active and reference input connectors or a single phono jack connector.

The attached extension cable connects the input box to the preamplifier by way of a cable adapter. The existing preamplifier will be fitted with three pin DIN connectors to accept the cable adapter.

The extension cable with buffered electrode input box allows electrode inputs to be made closer to the source of the signal for reduced signal noise during procedures requiring high-impedance electrodes. The electrode input box is available with separate active (labeled .) and reference connectors or a single phono jack connector to accommodate recording electrodes with these connector types.

The input box is enclosed in a white polyethylene foam sheath. The sheath houses a circuit board consisting of a buffer circuit, two electrostatic discharge (ESD) networks (one on each side of the buffer circuit), and a separate circuit designed to limit fault currents.

The circuit board is attached to a polyvinyl chloride extension cable terminated in an eight-pin DIN connector. The input box and extension cable are connected to the preamplifier by way of a polyvinyl chloride cable adapter terminated in a three-pin DIN connector. The adapter cable does not qualify as a class II device.

AI/ML Overview

Here's an analysis of the provided text regarding the device's acceptance criteria and studies, structured per your request:

Device: Cadwell Sierra and Cadwell 6200A with the proposed four-channel preamplifier and buffered input box (modification of existing devices).

Purpose of Modification: To allow electrode inputs closer to the signal source for reduced signal noise during procedures requiring high-impedance electrodes.


1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" as a separate, quantified list with pass/fail thresholds. Instead, it compares the technical specifications and safety standards of the proposed device against the predicate device, implying that matching or maintaining these characteristics constitutes acceptance. The "Testing and Validation" section then describes specific tests and their outcomes, which serve as evidence of meeting performance expectations.

For the purpose of this request, I will construct a table based on the provided technical comparison and highlight the directly reported performance from the validation section.

CriteriaAcceptance Criteria (Implied: Match Predicate)Reported Device Performance
Safety ComplianceUL 544, IEC 601-1 Type BF (isolated patient connections)Designed to comply with requirements of UL 544. Classification: isolated patient connections IEC 601-1: Type BF.
Electrode InputsFour buffered electrode inputs with separate active & reference 1.5-mm touch-proof connectors; One remote buffered electrode input for separate active & reference pin jack connectors or single phono jack connector; Preamplifier fitted with 3-pin DIN connector. (This is a description of the new setup, not an acceptance value.)N/A (This describes the feature, not a performance metric for it. The performance relates to noise reduction and signal integrity.)
Isolated Ground Connections2 connection2 connection
Isolation Mode Rejection> 150 dB.> 150 dB.
Common Mode Rejection90 dB90 dB.
Sensitivities2, 5, 10, 20, 50, 100, 200, 500 micro V/div; 1, 2, 5, 10, 20 m V/div.2, 5, 10, 20, 50, 100, 200, 500 micro V/div; 1, 2, 5, 10, 20 m V/div.
Noise (Electrical)2 micro V peak to peak (10 Hz to 10 kHz).The measured value of peak-to-peak noise is less than the allowed value (2 micro V peak to peak (10 Hz to 10 kHz)).
Input Impedance> 1,000 Mohms (common mode)The input impedance is greater than 10 Mohms when the active and reference connectors are subjected to ten strikes of 8-kV ESD each. (This is a specific test result, not a direct match to the >1000 Mohms spec, but likely demonstrating robust performance under stress.)
Notch Filter50 or 60 Hz50 or 60 Hz
Low-cut Filters1- or 2-pole filter. Selectable at 0.04, 0.1, 1, 3, 10, 30, 100, 500 Hz.1- or 2-pole filter. Selectable at 0.04, 0.1, 1, 3, 10, 30, 100, 500 Hz.
High-cut Filters2-pole (12 dB/octave) filter. Selectable at 30, 50, 100, 200, 300, 500 Hz; 1, 1.5, 2, 3, 5, 10, 15 kHz.2-pole (12 dB/octave) filter. Selectable at 30, 50, 100, 200, 300, 500 Hz; 1, 1.5, 2, 3, 5, 10, 15 kHz.
Common Recording Reference Input1 input1 input
Temperature Probe Input20 to 45 °C20 to 45 °C
Gain(Implied: No significant change or distortion)Not affected by passing the signal through the proposed device.
Patient Auxiliary Current(Implied: Below regulatory limits for BF connections)Well below the regulatory limits set forth by IEC 601-1 for BF connections.
Nerve Conduction Waveform Morphology(Implied: No distortion)Does not distort the morphology of the nerve conduction waveform, nor does it significantly affect the onset time, peak time, or amplitude when compared to the signal that does not pass through the proposed device.
EMG Waveform Morphology(Implied: No distortion)Does not distort the morphology of the EMG waveform when compared to the signal that does not pass through the proposed device.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size for clinical tests (Test 6 and Test 7). It mentions "clinical results" but provides no information on the number of subjects, cases, or specific data provenance (country of origin, retrospective/prospective). This information seems to be deferred to "Enclosure 3 of the previous submission dated October 16, 1996."


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish ground truth for the clinical tests (Test 6 and Test 7). It refers to the interpretation of waveform morphology, onset time, peak time, and amplitude, which would typically involve expert analysis in electrophysiology, but no details are provided.


4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test set.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study is mentioned. The device is an electrophysiology component designed to improve signal quality, not an AI diagnostic tool requiring multiple human readers to interpret its output in a comparative effectiveness study. The "clinical results" described are a comparison of signals with and without the proposed device, not a human-AI comparison.


6. Standalone Performance Study

Yes, a standalone performance assessment was conducted for the device in its intended function. The "Testing and Validation" section describes engineering and clinical tests (Tests 1-7) where the device's performance properties (input impedance, gain, noise, patient auxiliary current, waveform morphology) were measured and compared against regulatory limits or control signals (signals not passing through the device). This assessed the algorithm (or in this case, the hardware modification) performance in isolation.


7. Type of Ground Truth Used

The ground truth for the engineering tests (Tests 1-5) appears to be:

  • Regulatory Limits: For patient auxiliary current (IEC 601-1), allowed noise levels.
  • Engineering Specifications: For input impedance, gain (expected to be unchanged).
  • Control Measurements: Direct comparison to signals that do not pass through the proposed device for gain and noise.

For the clinical tests (Tests 6-7), the ground truth was:

  • Comparison to Signals Without the Device: Waveform morphology, onset time, peak time, and amplitude were compared to signals obtained without passing through the proposed device, with the expectation that these characteristics should not be significantly distorted. This implies a baseline reading from the existing system serves as a form of ground truth for assessing non-inferiority or lack of distortion.

8. Sample Size for the Training Set

The document does not refer to a "training set" in the context of machine learning or AI. This device is a hardware modification for an electromyography instrument, not an AI algorithm that undergoes training. The "testing and validation" described pertains to verifying the physical and functional aspects of the hardware.


9. How the Ground Truth for the Training Set Was Established

As there is no "training set" in the AI sense for this device, this question is not applicable.

{0}------------------------------------------------

K968455

IAN | 3 1997

Section 2 - Summary and Certification

December 20, 1996

A. 510(k) Summary of Safety and Effectiveness

The proposed device consists of the existing four-channel preamplifier and a buffered electrode input box with extension cable. These components provide electrode inputs that are closer to the source of the signal during electromyographic (EMG) testing. The intent of this design is to reduce signal noise during procedures requiring high-impedance electrodes. The proposed device is for use with the Cadwell Sierra (K924723) and 6200A (K931428) EMG instruments.

All device components are reusable and supplied non-sterile. The extension cable with electrode input box is compatible with EtO sterilization guidelines for procedures requiring a sterile field. The input box is available with separate active and reference input connectors or a single phono jack connector.

The attached extension cable connects the input box to the preamplifier by way of a cable adapter. The existing preamplifier will be fitted with three pin DIN connectors to accept the cable adapter.

1. Submitter Name and Identification

Cadwell Laboratories, Inc. 909 North Kellogg Street Kennewick, WA 99336 (800) 245-3001 Contact: Chris Bolkan Establishment Registration Number: 3020018

2. Proposed Device Name and Part Number

Trade Name (Proprietary Name): Cadwell Sierra and Cadwell 6200A

Proposed DeviceOrder Number
Four-Channel Preamplifier190155-200
Buffered Input Box with Extension Cable197112-200
Cable Adapter199155-200

Common Name or Usual Name: Electromyography and Evoked Potential Equipment.

Classification: Type II

NameNumber
Electromyograph84GWP
Electromyograph, Diagnostic89IKN

{1}------------------------------------------------

3. Identification of the Substantially Equivalent Device

Reason for Premarket Notification: Modification to an existing device.

Modification of Existing Device: The proposed device is a modification to the existing Cadwell Sierra (K924723) and 6200A (K931428) devices. The Cadwell four-channel preamplifier with buffered input box allows electrode inputs to be made closer to the signal source for reduced noise during procedures requiring high-impedance electrodes. The proposed device complies with the same safety standards as the existing device.

Safety of the Cadwell Sierra (K924723) and 6200A (K931428) Devices

The original device complies with the following safety standards for medical equipment:

  • IEC 601-1 Medical Electrical Equipment. Part 1. General requirements for safety. Type: Class 1 or grounded equipment, continuous operation, with B and BF applied parts.
  • IEC 878 (1988) Graphic symbols for electrical equipment in medical practice.

NFPA 99. Standard for Health Care Facilities.

The original device complies with the following general safety standards for electrical equipment:

  • ANSI/NFPA No. 70 (1990) National electric code.
  • UL 796 Standard for printed wiring boards.
  • UL 94 Standard for tests for flammability of plastic materials for parts in devices and appliances.

Safety of the Cadwell Sierra (K924723) and 6200A (K931428) Four-Channel Preamplifier Device

The original device complies with the following standards:

IEC 601-1 isolated applied part Type BF.

  • American Electroencephalographic Society Guidelines for Clinical Evoked Potential Studies, 1984. Section III Standards for Clinical Evoked Potential Equipment: Minimal Standards, Amplifier Averager. Section IV. Standards for Clinical Evoked Potential Recording: Calibration.

{2}------------------------------------------------

Effectiveness of the Cadwell Sierra (K924723) and 6200A (K931428) Devices

The original device is designed to perform the measurements needed for electromyography (EMG), nerve conduction velocity (NCV, F wave, and H reflex), evoked potentials (brainstern, visual, somatosensory) and repetitive nerve stimulation. The effectiveness of these clinical protocols is described in standard medical school textbooks. Please refer to the following texts for additional information.

  • Aminoff MJ: Electrodiagnosis in Clinical Neurology, Churchill Livingstone Inc., 1980.
  • Chiappa KH: Evoked Potentials in Clinical Medicine, ed 2. Raven Press, 1990.
  • Delisa JA: Manual of Nerve Conduction Velocity and Somatosensory Evoked Potentials, ed 2. Raven Press, 1987.
  • Johnson EW: Practical Electromyography, ed 2, Williams & Wilkins, 1988.
  • Kimura J: Electrodiagnosis in Diseases of Nerve and Muscle: Principles and Practice, ed 2. A Davis Company, 1989.
  • Spehlmann R: Evoked Potential Primer: Visual, Auditory, and Somatosensory Evoked Potentials in Clinical Diagnosis, Butterworth Publishers, 1985.
  • Regan D: Human Brain Electrophysiology: Evoked Potentials and Evoked Magnetic Fields in Science and Medicine. Elsevier Science Publishing Co., Inc., 1989.

4. Description of the Proposed Device

The extension cable with buffered electrode input box allows electrode inputs to be made closer to the source of the signal for reduced signal noise during procedures requiring high-impedance electrodes. The electrode input box is available with separate active (labeled .) and reference connectors or a single phono jack connector to accommodate recording electrodes with these connector types.

The input box is enclosed in a white polyethylene foam sheath. The sheath houses a circuit board consisting of a buffer circuit, two electrostatic discharge (ESD) networks (one on each side of the buffer circuit), and a separate circuit designed to limit fault currents.

{3}------------------------------------------------

The circuit board is attached to a polyvinyl chloride extension cable terminated in an eight-pin DIN connector. The input box and extension cable are connected to the preamplifier by way of a polyvinyl chloride cable adapter terminated in a three-pin DIN connector. The adapter cable does not qualify as a class II device.

5. Statement of Intended Use

The general purpose of the proposed device is identical to the predicate device: "to perform the measurements needed for electromyography (EMG), nerve conduction velocity (NCV, F wave, and H reflex), and evoked potentials (brainstem, visual, and somatosensory), and repetitive nerve stimulation." The purpose of the proposed device is to allow compatibility with high-impedance electrodes. The proposed device allows electrode inouts to be made closer to the source of the signal for reduced signal noise during procedures requiring high-impedance electrodes.

Duration of Use

The four-channel preamplifier with buffered input box is designed for use during the duration of the procedure only. This device is not for chronic use and is labeled accordingly.

Intended Use Environment

Use of the proposed device is to be administered under the direction of a trained physician, surgeon, neurologist, or electrophysiologist in a suitable operating room or clinic.

{4}------------------------------------------------

6. Comparison of Device Technological Characteristics

The proposed device is a modification to the existing four-channel preamplifier used with the Cadwell Sierra (K924723) and 6200A (K931428) devices. Physical modifications to the predicate device are in the following areas:

  • The recording component (electrode input box) from the 1. existing preamplifier is placed at the end of the extension cable. The proposed device provides closer electrode inputs for reduced signal noise during procedures requiring high-impedance electrodes.
    1. The proposed input box and extension cable are compatible with EtO sterilization requirements.
    1. To accommodate the proposed input box and extension cable, the existing preamplifier requires a hardware upgrade from the existing five-pin DIN connector to a three-pin DIN connector. The proposed device also requires a preamplifier cable adapter. The cable adapter does not qualify as a class II device.

The proposed device and the predicate device have nearly identical technical specifications and characteristics. The modification consists of adding a remote electrode input box for use with a single, high-impedance electrode. The input box connects to the existing preamplifier by way of an extension cable and cable adapter. The cable adapter requires a three-pin DIN connector in place of the standard five-pin DIN connector currently on the preamplifier.

{5}------------------------------------------------

CriteriaCadwell Sierra and 6200A withthe existing four-channelpreamplifierCadwell Sierra and 6200A withthe proposed four-channelpreamplifier and buffered inputbox
SafetyDesigned to comply withrequirements of UL 544.Classification: isolated patientconnections IEC 601-1: TypeBF.Designed to comply withrequirements of UL 544.Classification: isolated patientconnections IEC 601-1: TypeBF.
Electrode inputsFour buffered electrode inputswith separate active andreference 1.5-mm touch-proofconnectors or 5-pin DINconnector.Four buffered electrodeinputs with separate activereference 1.5-mm touch-proof connectors.One remote bufferedelectrode input for separateactive and reference pinjack connectors or singlephono jack connector.Preamplifier fitted with 3-pin DIN connector
Isolated groundconnections2 connection2 connection
Isolation moderejection> 150 dB.> 150 dB.
Common moderejection90 dB90 dB.
Sensitivities2, 5, 10, 20, 50, 100, 200, 500micro V/div; 1, 2, 5, 10, 20 mV/div.2, 5, 10, 20, 50, 100, 200, 500micro V/div; 1, 2, 5, 10, 20 mV/div.
Noise2 micro V peak to peak (10 Hz to10 kHz).2 micro V peak to peak (10 Hz to10 kHz).
Input impedance> 1,000 Mohms (commonmode)> 1,000 Mohms (commonmode)
Notch filter50 or 60 Hz50 or 60 Hz
Low-cut filters1 - or 2-pole filter. Selectable at0.04, 0.1, 1, 3, 10, 30, 100, 500Hz.1- or 2-pole filter. Selectable at0.04, 0.1, 1, 3, 10, 30, 100, 500Hz.
High-cut filters2-pole (12 dB/octave) filter.Selectable at 30, 50, 100, 200,300, 500 Hz; 1, 1.5, 2, 3, 5, 10,15 kHz.2-pole (12 dB/octave) filter.Selectable at 30, 50, 100, 200,300, 500 Hz; 1, 1.5, 2, 3, 5, 10,15 kHz.
Common recordingreference input1 input1 input
Temperature probeinput20 to 45 °C20 to 45 °C

:

{6}------------------------------------------------

Testing and Validation

The proposed device underwent engineering and clinical testing to validate that the device functions as a remote buffer preamplifier when used with the appropriate electrodiagnostic device. For validation procedures and results, please see Enclosure 3 of the previous submission dated October 16, 1996.

Conclusion

The results of engineering tests 1 and 2 indicate that the input impedance is greater than 10 Mohms when the active and reference connectors are subjected to ten strikes of 8-kV ESD each.

The following tests were carried out with the proposed device connected to the intended electrodiagnostic instrument. The results from test 3 show that the gain is not affected by passing the signal through the proposed device. The results from test 4 show that the patient auxiliary current is well below the regulatory limits set forth by IEC 601-1 for BF connections. The results from test 5 show that the measured value of peak-to-peak noise is less than the allowed valued.

Clinical results from test 6 show that the proposed device does not distort the morphology of the nerve conduction waveform, nor does it significantly affect the onset time, peak time, or amplitude when compared to the signal that does not pass through the proposed device. Additional clinical results in test 7 indicate that the proposed device does not distort the morphology of the EMG waveform when compared to the signal that does not pass through the proposed device.

§ 882.1870 Evoked response electrical stimulator.

(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).