BARD TIGER GUIDEWIRE MODEL 000540

K962178 · C.R. Bard, Inc. · OCY · Oct 29, 1996 · Gastroenterology, Urology

Device Facts

Record IDK962178
Device NameBARD TIGER GUIDEWIRE MODEL 000540
ApplicantC.R. Bard, Inc.
Product CodeOCY · Gastroenterology, Urology
Decision DateOct 29, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Bard® Tiger™ guidewire is indicated for use with devices having a .035" guidewire compatible lumen. The Tiger guidewire is designed to guide and exchange endoscopic accessories into the G.I. Tract, including, but not limited to, the common bile, cystic, the right, and left hepatic and pancreatic ducts. It is not necessary to withdraw or remove the Tiger guidewire during sphincterotomy.

Device Story

Tiger guidewire facilitates guidance and exchange of endoscopic accessories within G.I. tract; used during procedures involving common bile, cystic, hepatic, and pancreatic ducts. Operated by physicians during endoscopic procedures under visualization or fluoroscopy. Device remains in place during sphincterotomy. Core composed of Bisphenol A Epoxy and fibers; jacket of polyethylene. Nonconductive design aids in procedural safety. Benefits include streamlined accessory exchange and selective cannulation of biliary ductal system.

Clinical Evidence

No clinical data; biocompatibility testing performed to confirm safety of core and jacket materials for mucosal and tissue contact.

Technological Characteristics

Core: Bisphenol A Epoxy and fibers (fiberglass). Jacket: Polyethylene. Nonconductive. Compatible with .035" lumen devices. Mechanical design for endoscopic guidance and exchange.

Indications for Use

Indicated for patients requiring endoscopic accessory guidance and exchange within the G.I. tract, including common bile, cystic, hepatic, and pancreatic ducts, using devices with a .035" lumen.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} OCT 29 1996 K962178 # 6.0 510(k) SUMMARY FOR THE TIGER™ GUIDEWIRE As required under Section 12, part (a)(i)(3A) of the Safe Medical Device Act of 1990, an adequate summary of any information respecting safety and effectiveness follows. ## 6.1 General Information - Name and address of submitter: Bard Interventional Products Division, C.R. Bard, Inc. 129 Concord Road Billerica, MA 01821-7031 - Contact: Beth A. Rochette Regulatory Affairs Manager - Date of Summary: June 6, 1996 - Name of Device Bard® Tiger™ Guidewire - Predicate Device(s): Boston Scientific Zebra™ Exchange Guidewire, Flexmedics FlexFinder Guidewire, and Davol Laparoscopic Suction Irrigation Probe. - Description and Intended Use of Device: The Bard® Tiger™ guidewire is indicated for use with devices having a .035" guidewire compatible lumen. The Tiger guidewire is designed to guide and exchange endoscopic accessories into the G.I. Tract, including, but not limited to, the common bile, cystic, the right, and left hepatic and pancreatic ducts. It is not necessary to withdraw or remove the Tiger guidewire during sphincterotomy. {1} K962178 ρ = 92 # 6.2 Summary of Similarities and Differences The Tiger™ guidewire is substantially equivalent to the currently marketed Boston Scientific Zebra™ exchange guidewire (K931650) and the Flexmedics Corporation FlexFinder guidewire (K923755). The tapered core material is the same as the Davol Laparoscopic Suction Irrigation Probe (K941334), which received concurrence on June 14, 1994. The general design, and functionality of the Tiger guidewire is similar to the Boston Scientific Zebra exchange guidewire and the broader GI indication of the Flexmedics Corporation FlexFinder guidewire. All three of the guidewires have similar principles of operation. The guidewires are used to guide and exchange endoscopic accessory devices through the biopsy channel of an endoscope into the GI tract and to aid in selective cannulation of the biliary ductal system under endoscopic visualization or fluoroscopy. The major differences between the guidewires are: 1. The Tiger guidewire core material is a composite of Bisphenol A Epoxy and fibers, while the Zebra exchange and FlexFinder guidewires are constructed of Nitinol. 2. The Tiger guidewire jacket material is polyethylene, while the Zebra exchange and FlexFinder guidewires are constructed out of the following plastics: Fluropolymer, Urethane, and polyester. The Bisphenol A Epoxy and fibers, commonly known as fiberglass, used in the proposed Tiger guidewire is the same material utilized with the Davol Laparoscopic Suction Irrigation Probe tip. The Davol Laparoscopic Suction Irrigation Probe is a nonconductive fiberglass probe intended for suction and irrigation of the operative site during electrosurgical laparoscopic procedures. The probe tip is advanced to the desired location through a laparoscopic cannula under endoscopic visualization. All three guidewires have a plastic jacket, however, the jacket material of the Tiger guidewire is a slightly different polymer than used in the Zebra exchange and FlexFinder guidewire. Biocompatibility testing has confirmed the material is safe for contacting mucosa and tissue. Both of these materials were chosen for their strength and nonconductive properties. 038
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...