BIOCIRCUITS IOS SERUM PREGNANCY TEST CARTRIDGES/IMMUNOASSAY CONTROLS
K961792 · Biocircuits Corp. · DHA · Aug 21, 1996 · Clinical Chemistry
Device Facts
Record ID
K961792
Device Name
BIOCIRCUITS IOS SERUM PREGNANCY TEST CARTRIDGES/IMMUNOASSAY CONTROLS
Applicant
Biocircuits Corp.
Product Code
DHA · Clinical Chemistry
Decision Date
Aug 21, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The IOS™ Serum Pregnancy Test Cartridge is to be used for the qualitative determination of human chorionic gonadotropin (hCG) in serum for the early detection of pregnancy. It is intended to be used with the IOS™ instrument in clinical laboratories, physicians' office laboratories, and other alternate sites of use close to the point of patient care. The IOS™ Immunoassay Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ immunoassay test cartridges.
Device Story
Device is a two-site simultaneous sandwich immunoassay for serum hCG detection. Input: patient serum sample added to IOS™ cartridge. Operation: sample rehydrates dried reagents; hCG binds to alkaline phosphatase-labeled monoclonal antibody; complex binds to immobilized polyclonal capture antibody on plastic cartridge. Instrument performs automated wash steps and adds buffer to rehydrate substrate. Output: fluorescent signal generated by substrate-enzyme reaction, measured via front-surface fluorometry. Instrument compares signal rate to barcode-stored cut-off value to provide qualitative positive/negative result. Used in clinical labs and physician offices by technologists or office staff. Benefits: rapid, automated, objective interpretation compared to manual visual assays; reduces operator error via automated reagent delivery and standardized interpretation.
Clinical Evidence
Bench testing established 25 mIU/mL analytical sensitivity and confirmed no interference from common substances (LH, FSH, TSH, hemoglobin, bilirubin, cholesterol, triglycerides). Accuracy study compared IOS to ICON II using 142 positive and 149 negative samples, showing 100% and 98% agreement respectively. Clinical study in three physician office labs (n=172) showed 98.7% agreement for positive samples and 100% for negative samples compared to predicate.
Technological Characteristics
Fluorometric enzyme immunoassay; sandwich format. Components: plastic cartridge, immobilized polyclonal capture antibody, alkaline phosphatase-labeled monoclonal antibody, 4-MUP substrate. Automated fluidics for reagent rehydration and washing. Front-surface fluorometry for signal detection. Barcode-based calibration and lot-specific control range management. Standalone instrument platform.
Indications for Use
Indicated for qualitative detection of hCG in serum for early pregnancy diagnosis. Applicable to clinical settings including physician offices and labs. No specific age or gender contraindications listed, though intended for use in patients where pregnancy status is clinically relevant.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
Hybritech ICON II hCG ImmunoConcentration Assay
Related Devices
K963770 — BIOCIRCUITS IOS QUANTITATIVE HCG TEST CARTRIDGES/BIOCIRCUITS IOS IMMUNOASSAY CONTROLS · Biocircuits Corp. · Dec 5, 1996
K133002 — i-STAT Total beta-hCG Test, i-STAT Total beta-hCG Controls, i-STAT Total beta-hCG Calibration Verification Materials · Abbott Point of Care, Inc. · Apr 16, 2015
Submission Summary (Full Text)
{0}
K961792
AUG 21 1996
510(k) Summary of Information Respecting Safety and Effectiveness
A. Submitter:
Biocircuits Corporation
1324 Chesapeake Terrace
Sunnyvale, CA 94089
(408) 745-1961
Contact: Sheila Ramerman
Date Prepared: August 12, 1996
B. Device Names:
Proprietary Name: Biocircuits IOS™ Serum Pregnancy Test Cartridges
Biocircuits IOS™ Immunoassay Controls
(formerly IOS™ Thyroid Controls)
Common Names: Reagents for hCG (human chorionic gonadotropin)
Quality control materials (assayed and unassayed)
Classification Name: human chorionic gonadotropin (hCG) test system
Quality control materials (assayed and unassayed)
C. Legally Marketed Device:
The IOS Serum Pregnancy Test Cartridges are substantially equivalent to the Hybritech ICON II hCG ImmunoConcentration Assay currently manufactured and distributed by Hybritech Incorporated.
D. Device Description:
Serum Pregnancy Test Cartridges: Human chorionic gonadotropin is a glycoprotein that is synthesized by the placenta during pregnancy. hCG appears in both serum and urine relatively soon after implantation of the developing embryo. The presence of hCG, and its rapid rise following conception, is thus the basis of pregnancy testing. hCG testing can aid in the determination of ectopic pregnancy and spontaneous abortion. hCG is also secreted by a wide variety of tumors: gestational trophoblastic tumors, testicular and prostatic tumors, and some breast cancers.
hCG is composed of two noncovalently linked polypeptides: the alpha and beta subunits. The individual subunits lack biological activity, but become active when linked to form the intact hormone. The alpha subunit of hCG is structurally similar to the alpha subunits of LH (luteinizing hormone), FSH (follicle stimulating hormone), and TSH (thyroid stimulating hormone). The beta subunits of each of the hormones are structurally unique and confer the biologic and immunologic specificity to each hormone. The beta subunit of hCG contains a unique chain of 30 carboxy-terminal amino acids that confers its specificity (1). Monoclonal antibodies directed against this portion of the beta subunit permit differentiation between hCG and the other pituitary glycoprotein hormones.
{1}
# Principle of the Test
The IOS™ Serum Pregnancy test is a two-site simultaneous sandwich immunoassay which utilizes two antibodies directed against intact hCG: a polyclonal antibody immobilized on the plastic cartridge surface (capture antibody) and a monoclonal antibody labeled with alkaline phosphatase (detection antibody, 'conjugate'). hCG in the patient sample binds to conjugate in the test cartridge, and this hCG-conjugate complex then binds to the immobilized capture antibody. After a short incubation period, excess hCG-conjugate complex is washed away. Substrate is added, which reacts with the alkaline phosphatase conjugate-hCG-capture antibody 'sandwich' and produces a fluorescent signal. The level of fluorescence is directly proportional to the amount of bound hCG-conjugate, which is directly proportional to the amount of hCG present in the patient sample. All reagents necessary to perform the test are dried in the IOS™ cartridge, and are rehydrated by the addition of patient sample by the operator, or by the addition of buffer by the instrument.
To perform a test, the operator inserts an IOS™ Serum Pregnancy Test Cartridge in the IOS™ instrument. When prompted, the operator adds sample to a sample well. Each cartridge can test two different patient samples, with one patient sample added to the left well and the second to the right well. The operator starts the test sequence when sample addition is complete. Patient sample rehydrates conjugate dried in the sample well; the conjugate binds to hCG present in the patient serum sample. This hCG-conjugate complex then flows into the incubation/reaction chamber where it binds to the immobilized capture antibody. At the end of the incubation time, excess sample-conjugate is washed away by buffer. Buffer is used to rehydrate the dried substrate necessary for signal generation and quantitation of hCG, in a second reaction chamber. Substrate flows into the incubation/reaction chamber where signal is produced as a result of substrate reacting with conjugate bound to captured patient hCG (the 'sandwich'). The fluorescent signal is read as a rate by front-surface fluorometry, compared to the rate for the cut-off value stored in the barcode on the test cartridge, and determined to be higher or lower than the cut-off value.
Immunoassay Controls: The use of materials derived from human blood to monitor quality control of clinical chemistry testing in the clinical laboratory has been widely established over the past several years. The Biocircuits IOS™ Immunoassay Controls are two levels of blood-based material for use with Biocircuits IOS™ Test Cartridges.
To run a control, the operator inserts the Control Cartridge (packaged with the controls) into the IOS™ instrument. The instrument reads the lot number and ranges of acceptable values for the control solutions from the Control Cartridge barcode, and then ejects the Control Cartridge. The operator then inserts a test cartridge and follows the instrument prompts to identify the control level, apply control solutions, and begin the test sequence. The IOS™ instrument performs the required buffer additions to rehydrate assay reagents and perform wash steps as necessary, reads the fluorescence signal generated, and calculates and prints the control result just as it would if the cartridge were used to test a patient sample.
# E. Intended Use:
The IOS™ Serum Pregnancy Test Cartridge is to be used for the qualitative determination of human chorionic gonadotropin (hCG) in serum for the early detection of pregnancy. It is intended to be used with the IOS™ instrument in clinical laboratories, physicians' office laboratories, and other alternate sites of use close to the point of patient care.
The IOS™ Immunoassay Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ immunoassay test cartridges.
{2}
F. Comparison with the Predicate Device:
Table I summarizes the comparative features of the IOS and Hybritech assays.
G. Performance Data:
1. Serum Pregnancy Test Cartridges:
Non-clinical testing performed in the manufacturer's laboratories gave the following results:
**Sensitivity (Detection Limit or Cut-off Value)**
The analytical sensitivity of the IOS™ Serum Pregnancy Test is defined as the detection limit, or cut-off value, of the assay. Sixty replicates of a zero calibrator were run using IOS™ Serum Pregnancy Test Cartridges. The mean value and standard deviation of the signal were calculated. The sensitivity was obtained by interpolating two standard deviations above the mean signal of the zero calibrator. The analytical sensitivity of the IOS™ Serum Pregnancy Test is 25 mIU/mL.
The cut-off value was confirmed by spiking serum samples from normal males with different levels of hCG below, at, and above the cut-off value. Samples containing 0, 15, 20, 25, 30, 35, 40 and 100 mIU/mL hCG were tested in IOS™ Serum Pregnancy Test Cartridges. The IOS™ Serum Pregnancy Test Cartridges gave positive results in 100% of the samples containing 25 mIU/mL hCG; all samples containing >25 mIU/mL hCG also tested 100% positive.
**Accuracy (Correlation)**
Serum samples were tested using the IOS Serum Pregnancy Test and the Hybritech ICON II HCG ImmunoConcentration™ Assay, a qualitative test for serum hCG. Results obtained using the IOS Serum Pregnancy Test agreed with the ICON II Assay in 142 (100%) of the positive samples tested (n = 142). The IOS Serum Pregnancy Test agreed with the ICON II Assay in 146 (98%) of the negative samples tested (n = 149).
**Cross-Reactivity**
Potential cross-reacting substances were added to serum samples containing 0 and 100 mIU/mL hCG at the levels specified below, and tested in the IOS Serum Pregnancy Test. None of the substances interfered with the test results.
- LH at 500 mIU/mL
- FSH at 500 mIU/mL
- TSH at 500 uIU/mL
**Interfering Substances**
Potential interfering substances were added to serum samples containing 0 and 100 mIU/mL hCG at the levels specified below, and tested in the IOS Serum Pregnancy Test. None of the substances interfered with the test results.
- Hemoglobin at 500 mg/dL
- Bilirubin (unconjugated) at 20 mg/dL
- Cholesterol at 750 mg/dL
- Triglycerides at 750 mg/dL
{3}
Clinical testing performed in three typical physicians' office laboratories gave the following results:
Samples were tested by users in three typical physicians' office laboratories. Sites of testing included an OB/GYN practice, an Internal Medicine practice, and a Reproductive Endocrinology/Fertility practice. Users included medical technologists, medical assistants, and front-office personnel. A total of 172 patient samples were tested using the IOS™ Serum Pregnancy Test cartridges in the office laboratories; samples tested included those from pregnant females, non-pregnant females, post-menopausal females, and males. The samples were tested in the manufacturer's laboratory using the ICON II assays. Results obtained using the IOS Serum Pregnancy Test agreed with the ICON II Assay in 81 of 82 (98.7%) of the positive samples tested and in 90 of 90 (100%) of the negative samples tested.
## 2. Immunoassay Controls
The following ranges for the IOS™ Immunoassay Controls were determined in studies in the manufacturer's laboratories. To establish the ranges, the controls were tested in at least 40 cartridges each, over at least 10 days, using several IOS™ instruments. These values only apply to this lot of IOS™ Immunoassay Controls. Different lots of Immunoassay Controls will likely have slightly different ranges. Your laboratory should establish its own range for these controls over time.
| Analyte | Control Level 1 | | Control Level 2 | |
| --- | --- | --- | --- | --- |
| | Mean | Range | Mean | Range |
| T4 (ug/dL) | 8.1 | 6.48-9.72 | 12.9 | 10.32-15.4 |
| T-Uptake (%) | 32.4 | 29.25-35.55 | 39.7 | 36.85-42.55 |
| Serum Preg. Test | Negative | | Positive | |
It is self-evident from the data and information presented here that the Biocircuits IOS™ Serum Pregnancy Test Cartridges are as safe, effective, and perform as well as the Hybritech ICON II hCG Immunoconcentration Assay manufactured and distributed by Hybritech Incorporated.
Attachment: Table I: Assay Comparison
{4}
TABLE 1
Hybritech ICON II vs. Biocircuits IOS™
Assay Comparison
| ATTRIBUTE | Hybritech ICON II hCG | IOS™ Serum Pregnancy Test |
| --- | --- | --- |
| Technology | Immunoconcentration immunoassay | Fluorometric enzyme immunoassay |
| Assay format | Sandwich | Sandwich |
| Enzyme label | Alkaline phosphatase | Alkaline phosphatase |
| Substrate | Bromochloroindolyl phosphate (BCIP) | 4-Methylumbelliferyl phosphate (4-MUP) |
| Reagents | | |
| Immobilization Medium | Paper membrane | Plastic cartridge |
| Dry reagents | Monoclonal antibody only | Monoclonal antibody, polyclonal antibody conjugate, substrate |
| Wet reagents | 3 (for urine) or 4 (for serum), each added by operator as needed | 1 (for all assays), always on board, added by instrument as needed |
| Delivery of wet reagents | Manual, dropwise | Fully automated, precisely measured |
| Stability | Three months at room temperature | Four months at room temperature |
| Storage | Room Temperature (15-30°C) or Refrigerated (2-8°C) | Room Temperature (15-30°C) |
| Sample | | |
| Type | Serum or urine | Serum |
| Volume | 450 uL (serum) | 33 uL |
| | 250 uL (urine) | |
| Preparation | Serum diluted by operator in separate container before addition to icon | None; serum diluted by instrument after addition to cartridge |
| Test Performance | | |
| Hands-on Time | 5+ minutes | Approximately one minute |
| Number of Steps | 6, including serum dilution | One (sample addition to cartridge) |
| Test Result | Qualitative color development | Displayed on screen
Printed alphanumeric hard copy |
| Test Interpretation | By operator
Visual comparison to Positive Reference Zone and to Negative Control Zone | By instrument
Comparison to cut-off value in software |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.