K Number
K961710

Validate with FDA (Live)

Date Cleared
1996-11-01

(183 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AOC Phase II 25 surgical diode laser system has the same general intended use as the predicate diode lasers. This intended use includes incision, excision, coagulation and vaporization applications in the following surgical specialties:

Urology Gynecology Gastroenterology Pulmonology General and Thoracic Surgery Neurosurgery (hemostasis) Ear, Nose, and Throat; Head/Neck Surgery Dermatology/Plastic Surgery

No new intended use was sought in this Premarket Notification.

Device Description

The AOC Medical Systems Model Phase II 25 Fiber-Coupled Diode Laser System delivers up to 25 watts of CW optical power from a compact, air cooled unit. It is a complete, self-contained instrument, including a high efficiency power supply, an adjustable light output with constant current operation and a visible aiming laser. The system couples the laser light into high numerical aperture delivery fibers fitted with standard SMA-905 fiber connectors.

The diode lasers in this unit were fabricated by Applied Optronics Corp. from single quantum well, strained-layer, indium gallium aluminum on gallium arsenide semiconductor material. The diode lasers are sealed in a rugged, factory-aligned, environmentally protective module. High electrical efficiency of the diode lasers eliminate the need for water cooling and assure low maintenance, reliable laser operation for this allsolid-state system. Each individual diode laser emits at a narrow wavelength; however coupling several diodes through a fiber optic bundle results in each diode wavelength contributing to a broader spectrum emitted form the final delivery fiber. The laser energy is centered at 980nm with a total range from approximately 970-1000nm.

The AOC Phase II 25 has a modular design comprised of:

Laser Diode and Cooling Module containing the laser diodes, the heat sink and fans.

Optics Module containing a beamsplitter pick-off for display of the total optical power.

Front panel for simple control of the unit.

Transformer isolated power supplies and control electronics.

Unique attributes of this system include the following:

Multiple, individually fiber-coupled, high power diode lasers mounted on a large thermal footprint to allow sustained high power CW operation without water cooling.

Optical power level continuously adjustable from the front panel, with display of optical output power.

Low power, visible aiming laser for beam targeting prior to activation of the high power lasers. The aiming laser is controlled independently of the high power lasers.

Immediate turn-on capability. No warm-up period. No system alignments or maintenance to be made by the user.

Forced air cooling (no water). Maximum inlet air temperature 30° C.

Individually replaceable, fiber pigtailed diode lasers. Repairs performed at factory.

The microprocessor controlled AOC Phase II 25 diode laser system is designed to allow easy and safe operation of the laser using controls common to other surgical laser systems. The laser system utilizes a menu driven set-up operation for user prompts and program entry. The laser can be set for continuous output or programmed for single or repeat pulses at output powers up to 25 watts. A footswitch or handswitch, activated by the user, always controls laser output. An optional remote control for use in a sterile plastic dressing (e.g., standard OR microscope drape) allows the user to change the power and activate Ready and Standby Modes.

The following pre-packaged, presterilzed disposable optical fibers are available for use with the AOC Phase II 25 laser system:

Fiber ModelType
AOC DCLF-600600um Conical tip
AOC DCLF-10001000um Conical tip
AOC DCLF-600B600um Hemispherical Ball tip
AOC DCLF-1000B1000um Hemispherical Ball tip
AOC DBLF-60S600um Side firing
AOC DBLF-100S1000um Side firing
AOC DBLF-60SF600um Scatter Free Side firing
AOC DBLF-60600um Free Beam
AOC NeosConical (conical shaped)
AOC NeosDelta (wedge shaped) Handpiece, Laparoscopicand Bare Fiber
AOC HSDConical shaped with electrode
AI/ML Overview

This document describes the AOC Phase II 25 Surgical Diode Laser. Based on the provided text, the device is a surgical diode laser system used for incision, excision, coagulation, and vaporization in various surgical specialties.

Here's an analysis of the acceptance criteria and the study information as requested:

Acceptance Criteria and Device Performance

The provided document does not explicitly state formal "acceptance criteria" for the AOC Phase II 25 laser in a quantitative manner. Instead, the submission focuses on demonstrating substantial equivalence to existing predicate devices based on technological characteristics and intended use. The "performance data" section indicates that bench test data was submitted to demonstrate the device meets its rated power and wavelength specifications.

Therefore, the table below reflects what can be inferred as the "performance criteria" based on the comparison to predicate devices and the documented "performance data."

Acceptance Criteria (Inferred)Reported Device Performance
Wavelength: Nominal wavelength matching 980nm predicate devices.980nm (centered at 980nm with a total range from approximately 970-1000nm), matching AOC 25, Cynosure Phototome, and Cytocare DioLase HP-1 lasers.
Output Power: Up to 25 watts.Up to 25 watts of CW optical power. The document states "The output power of the AOC Phase II 25 is about the same as the power of the Diomed 25, Indigo 830, and Sharplan 6020." It acknowledges lower power compared to CytoCare and Cynosure 980nm lasers but states this is not clinically significant for contact fiber applications or can be compensated. Bench test data confirmed the device produces the rated power.
Temporal Output Modes: Ability to operate in Continuous Wave (CW) and electronically gated pulses (single, repeat, countdown, exponential).Continuous, Single Pulse, Countdown Pulse, Repeat Pulse, and Exponential Pulse output. These modes are stated to be found on one or more of the predicate lasers, and its range of timed outputs are within the ranges of cleared products.
Safety Features: Inclusion of standard safety interlocks, aiming beam, foot/hand switch activation, Ready/Standby modes, etc.The AOC Phase II 25 has "the same overall characteristics and features as the predicate diode laser systems," including: adjustable power output, internal monitoring of power output accuracy, prominent power output display, foot switch activation, safety interlocks, a red aiming beam, Ready and Standby Modes, an electrically safe design, visual and audible indicators of operation, and a rugged cabinet. It also includes handswitch activation (like Sharplan 6020) and a Decay (exponential) Pulse mode (like Indigo unit).
Tissue Interaction: Expected to be identical to 980nm predicate lasers."Its laser tissue interactions would be expected to be identical" to other 980nm predicate lasers. The document also notes that "The 980 nm wavelength, like the Nd:YAG wavelength at 1064 nm, is less dependent on variations in tissue pigment and blood content," and absorbed less in pigment/blood rich tissue than 805-830 nm lasers. The 980 nm wavelength also has slightly higher water absorption than 1064 nm. Peer-reviewed clinical articles on tissue interaction at 980 nm were included in the 510(k) submission.

Study Details and Ground Truth Establishment

Based on the provided text, the device's performance was evaluated through bench testing and literature review, rather than a clinical study involving human patients or a complex ground truth assessment.

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Test Set Sample Size: Not applicable in the context of a clinical test set. The submission mentions "Bench test data" which would involve testing the physical device itself. The sample size for these bench tests (e.g., number of units tested, number of measurements) is not specified.
    • Data Provenance: The bench test data would be from the manufacturer (AOC Medical Systems) presumably from their manufacturing site (South Plainfield, NJ, USA). It is prospective in the sense of new device testing, but not a clinical trial.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. No "ground truth" in the sense of expert consensus on patient data was established for the device's performance because no clinical study was conducted. The "ground truth" for the device's technical specifications (power, wavelength) would be derived from calibrated measurement equipment and engineering standards.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. There was no clinical test set requiring adjudication of findings.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This is a surgical laser device; the concept of "human readers" and "AI assistance" in the context of diagnostic image interpretation is not relevant here.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device (surgical laser), not a software algorithm. Its "standalone" performance is its technical specifications (power output, wavelength stability, etc.) as measured during bench testing.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for this device's performance is its adherence to engineering specifications and established physical principles (e.g., measuring power output with a power meter, wavelength with a spectrometer). The submission relied on "Peer reviewed clinical articles on the tissue interaction at 980 nm" to support the expected clinical effects, indicating that existing scientific literature served as a form of "ground truth" for the physiological impact of the laser's wavelength.
  7. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI device that requires a training set.
  8. How the ground truth for the training set was established:

    • Not applicable. No training set was used.

{0}------------------------------------------------

APPLIED OPTRONICS CORP.

K 961710

1 1 1 Corporate Blvd. • Building J • South Plainfield, NJ 07080 TEL (908) 753-6300 · FAX (908) 753-4041

NOV - I 1996

SECTION K

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

Submitter Information 1.

AOC Medical Systems A Division of Applied Optronics Corporation 111 Corporate Blvd. Building J South Plainfield, NJ 07080 Telephone: (908) 753-6300 Fax: (908) 753-4041 Contact Person: Douglass Mead Regulatory Affairs Date Prepared: April 30, 1996

Model AOC Phase II 25 Surgical Diode Laser 2.

Predicate or Legally Marketed Devices: 3.

ManufacturerModelLaser Type
AOC Medical Sys.AOC 25980 nm Diode Laser
CynosurePhototome980 nm Diode Laser
CytocareDioLase HP-1980 nm Diode Laser
DioMedDioMed 25810 nm Diode Laser
Indigo MedicalIDE 830E830 nm Diode Laser
Sharplan Lasers6020810 nm Diode Laser

General System Description 4.

The AOC Medical Systems Model Phase II 25 Fiber-Coupled Diode Laser System delivers up to 25 watts of CW optical power from a compact, air cooled unit. It is a complete, self-contained instrument, including a high efficiency power supply, an adjustable light output with constant current operation and a visible aiming laser. The system couples the laser light into high numerical aperture delivery fibers fitted with standard SMA-905 fiber connectors.

The diode lasers in this unit were fabricated by Applied Optronics Corp. from single quantum well, strained-layer, indium gallium aluminum on gallium arsenide semiconductor material. The diode lasers are sealed in

{1}------------------------------------------------

a rugged, factory-aligned, environmentally protective module. High electrical efficiency of the diode lasers eliminate the need for water cooling and assure low maintenance, reliable laser operation for this allsolid-state system. Each individual diode laser emits at a narrow wavelength; however coupling several diodes through a fiber optic bundle results in each diode wavelength contributing to a broader spectrum emitted form the final delivery fiber. The laser energy is centered at 980nm with a total range from approximately 970-1000nm.

The AOC Phase II 25 has a modular design comprised of:

Laser Diode and Cooling Module containing the laser diodes, the heat sink and fans.

Optics Module containing a beamsplitter pick-off for display of the total optical power.

Front panel for simple control of the unit.

Transformer isolated power supplies and control electronics.

Unique attributes of this system include the following:

Multiple, individually fiber-coupled, high power diode lasers mounted on a large thermal footprint to allow sustained high power CW operation without water cooling.

Optical power level continuously adjustable from the front panel, with display of optical output power.

Low power, visible aiming laser for beam targeting prior to activation of the high power lasers. The aiming laser is controlled independently of the high power lasers.

Immediate turn-on capability. No warm-up period. No system alignments or maintenance to be made by the user.

Forced air cooling (no water). Maximum inlet air temperature 30° C.

Individually replaceable, fiber pigtailed diode lasers. Repairs performed at factory.

The microprocessor controlled AOC Phase II 25 diode laser system is designed to allow easy and safe operation of the laser using controls common to other surgical laser systems. The laser system utilizes a menu driven set-up operation for user prompts and program entry. The laser can be set for continuous output or programmed for single or repeat pulses at output powers up to 25 watts. A footswitch or handswitch, activated by the user, always controls laser output. An optional remote control for use in a sterile plastic dressing (e.g., standard OR microscope drape) allows the user to change the power and activate Ready and Standby Modes.

The following pre-packaged, presterilzed disposable optical fibers are available for use with the AOC Phase II 25 laser system:

{2}------------------------------------------------

Fiber ModelType
AOC DCLF-600600um Conical tip
AOC DCLF-10001000um Conical tip
AOC DCLF-600B600um Hemispherical Ball tip
AOC DCLF-1000B1000um Hemispherical Ball tip
AOC DBLF-60S600um Side firing
AOC DBLF-100S1000um Side firing
AOC DBLF-60SF600um Scatter Free Side firing
AOC DBLF-60600um Free Beam
AOC NeosConical (conical shaped)
AOC NeosDelta (wedge shaped) Handpiece, Laparoscopicand Bare Fiber
AOC HSDConical shaped with electrode

Intended Use: 5.

The AOC Phase II 25 surgical diode laser system has the same general intended use as the predicate diode lasers. This intended use includes incision, excision, coagulation and vaporization applications in the following surgical specialties:

Urology Gynecology Gastroenterology Pulmonology General and Thoracic Surgery Neurosurgery (hemostasis) Ear, Nose, and Throat; Head/Neck Surgery Dermatology/Plastic Surgery

No new intended use was sought in this Premarket Notification.

The wavelength (980 nm) and tissue effects of the AOC Phase II 25 diode laser are closely matched to the predicate lasers at 805-830 and 980 nm. The 980 nm wavelength absorption in tissue is less dependent on variations in tissue pigment or vascularity than at 805 nm and has a slightly higher water absorption than the 1064 nm wavelength. Other predicate 980 nm diode laser systems have been found substantially equivalent to Nd:YAG and 805 nm diode lasers.

{3}------------------------------------------------

Comparison of Technological Characteristics 6.

The AOC Phase II 25 laser output is at the same nominal wavelength, 980 nanometers as the AOC 25, Cynosure Phototome, and Cytocare DioLase HP-1 lasers. Its laser tissue interactions would be expected to be identical. The same individual diode laser components are supplied Optronics for all four laser systems. The AOC Phase II 25 differs in wavelength from the DioMed 25, Indigo 830 and Sharplan 6020 (at 805 -830 nanometers) which has a slightly greater energy absorption in pigment and blood rich tissue. The 980 nm wavelength, like the Nd:YAG wavelength at 1064 nm, is less dependent on variations in tissue pigment and blood content.

The temporal output of all these systems is the same. These lasers are operated most often in the Continuous Wave (steady state) Mode but can be electronically gated to produce time pulses, single or repeated. The AOC Phase II 25 laser has a Continuous, Single Pulse, Countdown Pulse, Repeat Pulse, and Exponetial Pulse output which are found on one or more of the predicate lasers. Its range of timed outputs are within the ranges of the above cleared products.

The output power of the AOC Phase II 25 is about the same as the power of the Diomed 25, Indigo 830, and Sharplan 6020. For applications using contact fibers with sculpted tips which typically operate below 25 watts, the difference in output is not clinically significant. The lower output power of the AOC Phase II 25 in comparison to the CytoCare and Cynosure 980 nm lasers can be compensated for by using smaller diameter fibers for equal power density or exposing target tissue for longer duration (equivalent energy).

The AOC Phase II 25 laser system has the same overall characteristics and features as the predicate diode laser systems. Like these lasers, the AOC Phase II 25 provides for adjustable power output, internal monitoring of power output accuracy, prominent power output display, operation in the Continuous Mode and Single Pulse, and other Modes (with adjustable pulse duration), foot switch activation, safety interlocks, a red aiming beam, Ready and Standby Modes, an electrically safe design, visual and audible indicators of operation, as well as a rugged cabinet designed for the clinical environment.

Like The Sharplan 6020, the AOC Phase II 25 is equipped for handswitch activation of the laser output for convenient surgical control. Also, like the Indigo unit, the AOC Phase II 25 has a Decay (e.g., exponential) Pulse mode.

7. Performance Data

No clinical test data was submitted in support of this submission. Peer reviewed clinical articles on the tissue interaction at 980 nm were included in the 510 (k). Bench test data was submitted demonstrating that the design is able to produce the rated power and wavelength.

Notice: This Summary was prepared in order to comply with CFR 21 807.87. AOC Medical Systems makes no claim that any disease or other condition can be cured and promotes its laser as a surgical tool only for use by qualified physicians. Laser surgery, like any surgery, entails necessary risks, some of which may result in complications, injury or, in the extreme, death.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.