K Number
K961489

Validate with FDA (Live)

Date Cleared
1996-06-11

(54 days)

Product Code
Regulation Number
862.1695
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro quantitative determination of free thyroxine in human serum and plasma.

Device Description

The Elecsys® test principle is based on competition principle. Total duration of assay: 18 minutes (37° C).
• 1st incubation (9 minutes): Sample (15 µL) and a specific anti-T4 antibody labeled with a ruthenium complex (75 µL).
• 2nd incubation (9 minutes): After addition of biotinylated T4 (75 µL) and streptavidin-coated microparticles (35 µL), the still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the solid phase via interaction of biotin and streptavidin.
The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

AI/ML Overview

The provided text describes a 510(k) summary for the "Elecsys® FT4 Assay," a device for the in vitro quantitative determination of free thyroxine in human serum and plasma. The acceptance criteria and the study proving the device meets these criteria are detailed through comparisons to a predicate device, the Enzymun-Test® FT4 (K900883).

Here's the breakdown of the information requested:


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by comparing the performance characteristics of the Elecsys® FT4 to its predicate device, Enzymun-Test® FT4. The goal is to demonstrate substantial equivalence, meaning the new device is as safe and effective as the predicate.

FeaturePredicate Device (Enzymun-Test® FT4) Criteria (Reported Performance)Elecsys® FT4 Reported Performance
Reportable Range0.1 ng/dL - 6.5 ng/dL (1.29 - 83.69 pmol/L)0.023 ng/dL - 7.77 ng/dL (0.3 - 100.0 pmol/L)
Calibration StabilityA calibration is required every 2 weeks.A calibration is recommended every 7 days if kit is not consumed; 4 weeks with same reagent lot if reagent is consumed within 7 days.
PrecisionModified NCCLS (ng/dL):- Low: Within-Run %CV: 0.49, Total %CV: 5.1- Mid: Within-Run %CV: 1.17, Total %CV: 2.9- High: Within-Run %CV: 4.43, Total %CV: 2.0Modified NCCLS (ng/dL):- Sample Control 1: Within-Run %CV: 1.58, Total %CV: 3.54- Sample Control 2: Within-Run %CV: 1.43, Total %CV: 2.72- Unknown Sample: Within-Run %CV: 2.85, Total %CV: 2.85
Sensitivity (Lower Detection Limit)0.1 ng/dL (1.29 pmol/L)0.023 ng/dL (0.3 pmol/L)
Method Comparison (vs. Enzymun-Test® FT4)Least Squares: y = 0.88x + 0.15, r = 0.984, SEE = 0.199, N = 77Least Squares: y = 0.954x + 0.18, r = 0.981, SEE = 1.11, N = 315Passing/Bablok: y = 1.02x + 0.36, r = 0.981, SEE = 1.11, N = 315
Interfering Substances (Bilirubin)No interference at 64.5 mg/dLNo interference at 25 mg/dL
Interfering Substances (Hemoglobin)No interference at 1 g/dLNo interference at 1 g/dL
Interfering Substances (Lipemia)No interference at 1250 mg/dLNo interference at 1500 mg/dL
Interfering Substances (Biotin)No interference at 200 ng/mLNo interference at 30 ng/mL
Specificity (% Cross-reactivity L-T4)100100
Specificity (% Cross-reactivity D-T4)100100
Specificity (% Cross-reactivity L-T3)3.51.53
Specificity (% Cross-reactivity D-T3)2.91.38
Specificity (% Cross-reactivity 3-iodo-L-tyrosine)<0.10.002
Specificity (% Cross-reactivity 3,5-diiodo-L-tyrosine)<0.10.01
Specificity (% Cross-reactivity Tetraiodo-thyroacetic acid)2038.5

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Method Comparison:
    • For the Elecsys® FT4 compared against the Enzymun-Test® FT4: N = 315
    • For the Enzymun-Test® FT4 compared against itself (as a predicate, likely for its own method comparison studies): N = 77
  • Sample Size for Precision: 60 for the Elecsys® FT4 at each control level. For the predicate, the N varied from 117 to 120.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This information is typically found in the full study report, not necessarily in a 510(k) summary. Given the context of in vitro diagnostic assays, the samples would likely be human serum and plasma, but their origin is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not applicable to this type of device and study. This device is an in vitro diagnostic (IVD) assay designed to quantitatively measure a biomarker (free thyroxine). The "ground truth" for such devices is established by reference methods or validated comparative methods (like the predicate device in this case), not by expert consensus on image interpretation or clinical diagnosis. The performance is assessed by analytical accuracy, precision, sensitivity, and specificity against these established methods or known spiked concentrations.


4. Adjudication Method for the Test Set

This is not applicable as the ground truth is established through quantitative analytical measurements against established methods, not through expert review and adjudication.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for imaging devices or AI algorithms where human interpretation is involved. The Elecsys® FT4 Assay is an automated quantitative immunoassay, not a device requiring human interpretation through reading images or cases.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the performance reported for the Elecsys® FT4 is a standalone (algorithm only) performance. This device is an automated immunoassay system that performs the measurement and calculation of free thyroxine levels without human intervention during the assay process. The results are generated directly by the instrument based on its electrochemical detection and built-in calibration curve.


7. The Type of Ground Truth Used

The ground truth for evaluating the Elecsys® FT4 Assay is primarily established through:

  • Reference Method Comparison: Comparison against the predicate device, Enzymun-Test® FT4. This implies that the predicate device's results are considered a valid reference for "ground truth" for demonstrating substantial equivalence.
  • Analytical Standards: Precision and sensitivity are determined using control samples with known concentrations and spiked samples, which serve as the "ground truth" for these performance characteristics.
  • Known Interferents/Analytes: Specificity studies use known concentrations of potentially interfering substances or cross-reacting analytes, where the expected outcome (no interference, or specific cross-reactivity) serves as the ground truth.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for a "training set." For an IVD assay like Elecsys® FT4, the "training" aspect is more related to method development, reagent optimization, and establishing the master calibration curve, which is distinct from the concept of a training set in machine learning. The data provided focuses on the validation of the final assay.


9. How the Ground Truth for the Training Set Was Established

As mentioned above, the concept of a "training set" and its "ground truth" in the machine learning sense is not directly applicable here. For the development and establishment of the assay itself (analogous to 'training' in a broad sense):

  • Master Calibration Curve: The master curve is provided via the reagent bar code and is instrument-specifically generated by 2-point calibration. This curve is developed using a comprehensive set of calibrators with precisely known concentrations of the analyte (FT4). These known concentrations establish the "ground truth" for the dose-response relationship of the assay.
  • Reagent Selection and Optimization: During the development phase, various reagent formulations and conditions would be tested against samples with pre-determined FT4 levels (perhaps measured by highly accurate reference methods or gravimetric preparation) to optimize performance characteristics like sensitivity, specificity, and precision.

{0}------------------------------------------------

K9Cd489

Image /page/0/Picture/1 description: The image contains the word "Diagnostics" in a stylized font. Above the word "Diagnostics" is a black square with some white markings. Below the word "Diagnostics" are the words "a manufacturer".

היישוק

510(k) Summary

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
1.Submittername,address,contactBoehringer Mannheim Corporation2400 Bisso LaneConcord, CA 94524-4117(510) 674-0667; extension 8240Contact Person: Betsy Soares-MaddoxDate Prepared: April 9, 1996
2.Device nameProprietary name: Elecsys® FT4 AssayCommon name: Electrochemiluminescence assay for the determination of free thyroxine (FT4).Classification name: Radioimmunoassay, Free Thyroxine
3.PredicatedeviceWe claim substantial equivalence to the Enzymun-Test® FT4 (K900883)
4.DeviceDescriptionThe Elecsys® test principle is based on competition principle. Total duration of assay: 18 minutes (37° C).• 1st incubation (9 minutes): Sample (15 µL) and a specific anti-T4 antibody labeled with a ruthenium complex (75 µL).• 2nd incubation (9 minutes): After addition of biotinylated T4 (75 µL) and streptavidin-coated microparticles (35 µL), the still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the solid phase via interaction of biotin and streptavidin.

Continued on next page

.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains a black square on the left side and the word "Diagnostics" on the right side. The word is written in a serif font and is slightly italicized. The black square appears to have some vertical lines or patterns within it.

·The reaction mixture is aspirated into the measuring cell where the 4. microparticles are magnetically captured onto the surface of the electrode. Device Description, Unbound substances are then removed with ProCell. Application of a voltage cont. to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame). · Results are determined via a calibration curve which is instrumentspecifically generated by 2-point calibration and a master curve provided via the reagent bar code. Immunoassay for the in vitro quantitative determination of free thyroxine in ર. Intended use human serum and plasma. ર . The Boehringer Mannheim Elecsys® FT4 is substantially equivalent to other Comparison products in commercial distribution intended for similar use. Most notably it to predicate is substantially equivalent to the currently marketed Enzymun-Test® FT4 device (K900883). The following table compares the Elecsys® FT4 with the predicate device, Enzymun-Test® FT4. Specific data on the performance of the test have been incorporated into the draft labeling in attachment 5. Labeling for the predicate device is provided in attachment 6. Similarities: · Intended Use: Immunoassay for the in vitro quantitative determination of free thyroxine ·Sample type: Serum and plasma · Antibody: Polyclonal Sheep anti-FT4 antibodies · Solid phase binding principle: Streptavidin/Biotin · Assay standardization: Equilibrium dialysis Continued on next page

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the word "Diagnostics" in a stylized font. The word is positioned to the right of a dark, rectangular shape. The text "anaları" is present above the word "Diagnostics", but it is less prominent.

6

in

Differences:

Comparison to predicate
device cont.

FeatureElecsys® FT4Enzymun-Test® FT4
Reaction testprincipleElectrochemiluminescenceELISA/1-step sandwich assayusing streptavidin technology
InstrumentrequiredElecsys® 2010ES 300
Assay RangeReportable Range:0.023 ng/dL - 7.77 ng/dL(0.3 - 100.0 pmol/L)Reportable Range:0.1 ng/dL - 6.5 ng/dL(1.29 - 83.69 pmol/L)
CalibrationStabilityA calibration is recommendedevery 7 days if kit is notconsumed; 4 weeks with samereagent lot if reagent isconsumed within 7 days.A calibration is required every2 weeks.

Continued on next page

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Diagnostics" in a serif font. To the left of the word is a black square. Below the word "Diagnostics" are the numbers "2019" and "191".

6.

Comparison to predicate device, (cont.)

Performance Characteristics:

FeatureElecsys® FT4Enzymun-Test® FT4
PrecisionModified NCCLS (pmol/L):Modified NCCLS (ng/dL):
LevelSample Control 1 Control 2Low Mid High
NWithin-Run%CVTotal%CV60 17.53 50.841.58 1.43 2.858.77 17.53 50.843.54 2.72 6.59117 120 1200.49 1.17 4.433.9 2.4 1.20.49 1.17 4.435.1 2.9 2.0
Modified NCCLS (ng/dL):Sample Control 1 Control 2
NWithin-Run%CVTotal%CV60 1.367 3.971.58 1.43 2.850.684 1.367 3.973.54 2.72 2.85
SensitivityLower Detection Limit:0.023 ng/dL(0.3 pmol/L)Lower Detection Limit:0.1 ng/dL(1.29 pmol/L)
MethodComparisonVs Enzymun-Test® FT4Least Squares$y =0.954x + 0.18$$r=0.981$$SEE =1.11$$N=315$Passing/Bablok$y =1.02x + 0.36$$r=0.981$$SEE =1.11$$N=315$Vs Enzymun-Test® FT4Least Squares$y = 0.88x + 0.15$$r=0.984$$SEE = 0.199$$N= 77$

Continued on next page

:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Diagnostics" in a serif font. To the left of the word is a black square with some white markings on it. Below the word "Diagnostics" are the numbers "1 . 11".

6. Comparison

to predicate
device, (cont.)

Performance Characteristics, cont.:

FeatureElecsys® FT4Enzymun-Test® FT4
InterferingsubstancesNo interference at:No interference at:
Bilirubin25 mg/dL64.5 mg/dL
Hemoglobin1 g/dL1 g/dL
Lipemia1500 mg/dL1250 mg/dL
Biotin30 ng/mL200 ng/mL
Specificity% Cross-reactivity% Cross-reactivity
L-T4100100
D-T4100100
L-T31.533.5
D-T31.382.9
3-iodo-L-tyrosine0.002<0.1
3,5-diiodo-L-tyrosine0.01<0.1
Tetraiodo-thyroaceticacid38.520

§ 862.1695 Free thyroxine test system.

(a)
Identification. A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.(b)
Classification. Class II.