K Number
K961400

Validate with FDA (Live)

Date Cleared
1997-01-03

(267 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Scanditronix Medical RFA-300, LDA Utility is used to more effectively perform radiation therapy beam measurements including acceptance testing, trimming information, routine beam checks and quality assurance measurements and on-line data analyses using personal computer (PC) monitor. These uses are similar to those of the predicate identified in section (3) of this Summary.

Device Description

The Scanditronix Medical RFA-300, LDA Utility is an extension to the RFA-300 system used to measure radiation therapy beam intensity. For use, a complete RFA-300 system is needed extended with the following equipment:

  1. Multichannel electrometers emX which provides 12 independent electrometer channels each or a DPD-510 which provides 10 independent electrometer channels.
  2. Linear Detector Array LDA-11 (11 detectors) or LDA-25 (25 detectors).
  3. PC-based software providing control, display, calibration and save functions.
AI/ML Overview

This is a summary of the acceptance criteria and study details for the Scanditronix Medical RFA-300, LDA Utility:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Operational characteristics suitable for intended purpose (effectively performing radiation therapy beam measurements including acceptance testing, trimming information, routine beam checks, quality assurance measurements, and on-line data analyses).Comparison of operational characteristics for the Scanditronix Medical RFA-300, LDA Utility and the predicate product showed similar results that are suitable for their intended purpose.
Minimization of potential electrical hazardsScanditronix Medical adheres to recognized and established industry practice, and all devices are subject to final performance testing. The device is designed for conformance with IEC 601-1 standards for electrical isolation and meets electrical performance standards for CSA and UL certifications.
Meeting accelerator performance standards for measurement and data processing capabilitiesIEC-601C standards for accelerator performance are used as minimum criteria of measurement and data processing capabilities of the Scanditronix Medical RFA-300, LDA. The device was tested and found to fulfill these requirements.
Electromagnetic compatibility (EMC)The Scanditronix Medical RFA-300, LDA Utility has been tested and found to fulfill the requirements concerning electromagnetic compatibility according to the standard IEC 601-1-2 (see appendix XVII).
Safety (implied, as part of overall performance and meeting standards)Testing of operational parameters for the Scanditronix Medical RFA-300, LDA Utility, indicates that the device is safe.
Appropriate verification and dose measurements of radiation beams (implied, as part of overall performance)Testing indicates that the device provides appropriate verification and dose measurements of radiation beams.
Performance equivalent to or better than the legally marketed predicate deviceTesting indicates that the device performs as well as or better than the legally marketed predicate device identified in section (3) of this summary (RFA-300, Wellhöfer WP600, Wellhöfer MDA).

Study Details

  1. Sample Size used for the test set and the data provenance: Not applicable. The device is for radiation therapy beam measurements, not for patient data analysis. The testing involved comparing operational characteristics and adherence to engineering standards.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for this type of device (a measurement tool) is established through engineering standards, comparative performance against predicate devices, and adherence to regulatory compliance. No mention of human expert panel for ground truth assessment in the context of diagnostic interpretation.
  3. Adjudication method for the test set: Not applicable. This document describes testing against engineering standards and comparison with predicate device performance, not a diagnostic adjudication process.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device. It's a measurement tool for radiation therapy.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device itself is standalone in its function for measuring radiation beams. The "human-in-the-loop" involves a technician operating the system and interpreting the measurements as part of a larger radiation therapy quality assurance process. The document primarily focuses on the device's technical specifications and performance against established benchmarks without human involvement in evaluating its output for a diagnostic decision.
  6. The type of ground truth used:
    • Engineering Standards: IEC 601-1, CSA, UL certifications for electrical safety and performance.
    • Accelerator Performance Standards: IEC-601C for measurement and data processing capabilities.
    • Electromagnetic Compatibility Standards: IEC 601-1-2.
    • Comparative Performance: Against identified predicate devices (RFA-300, Wellhöfer WP600, Wellhöfer MDA) regarding operational characteristics.
  7. The sample size for the training set: Not applicable. This device is a measurement instrument, not a machine learning or AI system that requires a "training set" in the conventional sense.
  8. How the ground truth for the training set was established: Not applicable, for the reason stated above.

{0}------------------------------------------------

510(k) Summary

961400 Jan 3, 199-

This summery is submitted in compliance with the FDA interim rule as published 28 April 1992 in 21 CFR 807.92:

(a)(1)Submitted by:Scanditronix Medical ABStålgatan 14,S-754 50 UppsalaSweden
Contact Persons:Lars Göran Larsson+46 18 18 08 52orGörgen Nilsson+46 18 18 08 94
Date of preparation:28 Mar 1996
(2)Trade name of device:Scanditronix Medical RFA-300, LDA Utility
Common name:Radiation field analyzer, LDA
Classification name:(Accessory to)Radionuclide radiation therapysystem, §892,5750; and X-ray radiation therapysystem, §892,5900.
(3)Identification of predicatemarketed device:RFA-300 (Scanditronix Medical AB, or legallyK934303 / S1) , Wellhöfer WP600 (MedicalPhysics instrumentation, K882770) and WellhöferMDA ( Medical Physics instrumentation).
  • (4) Description of the devise:
    The Scanditronix Medical RFA-300, LDA Utility is an extension to the RFA-300 system used to measure radiation therapy beam intensity. For use, a complete RFA-300 system is needed extended with the following equipment:

  • Multichannel electrometers emX which provides 12 independent 1. electrometer channels each or a DPD-510 which provides 10 independent electrometer channels.

    1. Linear Detector Array LDA-11 (11 detectors) or LDA-25 (25 detectors).
    1. PC-based software providing control, display, calibration and save functions.
  • (5) Intended uses:

The Scanditronix Medical RFA-300, LDA Utility is used to more effectively perform radiation therapy beam measurements including acceptance testing, trimming information, routine beam checks and

1(2)

{1}------------------------------------------------

quality assurance measurements and on-line data analyses using personal computer (PC) monitor. These uses are similar to those of the predicate identified in section (3) of this Summary.

  • Technological comparison: (6)
    The Scanditronix Medical RFA-300, LDA Utility, is similar to the predicate marketed device in that radiation beam measurements are made using solid state, single or multi, detectors connected to electrometers which converts data into formats suitable for tabulation, display, or transfer to therapy planning systems. Control of the system is based on the software in a PC which also provides for display, storage, or transfer of data.

  • Non-Clinical tests: (b) (1)
    Comparison of operational characteristics for the Scanditronix Medical RFA-300. LDA Utility and the predicate product show similar results that are suitable for their intended purpose. To minimize potential electrical hazards, Scanditronix Medical adheres to recognized and established industry practice, and all devices are subject to final performance testing. The Scanditronix Medical RFA-300, LDA Utility is designed for conformance with IEC 601-1 standards for electrical isolation, and meets electrical performance standards for CSA and UL certifications. In addition, IEC-601C standards for accelerator performance are used as minimum criteria of measurement and data processing capabilities of the Scanditronix Medical RFA-300, LDA. The Scanditronix Medical RFA-300, LDA Utility has been tested and found to fulfil the requirements concerning electromagnetic compatibility according to the standard IEC 601-1-2 (see appendix XVII).

  • Clinical tests: (2)
    Due to the fact that the system is not directly involved with patients, no clinical testing was performed.

  • (3) Test conclusions:
    Testing of operational parameters for the Scanditronix Medical RFA-300, LDA Utility, indicate that the device is safe, that it provides appropriate verification and dose measurements of radiation beams, and that it performs as well as or better than the legally marketed predicate device identified in section (3) of this summary.

304
2(2)

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.